C1 · Lesson 8 · Meters and loading

Why connecting a meter changes what it measures

A voltmeter has to draw a little current to measure anything, and an ammeter has a little resistance of its own. Connect one to a circuit and the circuit is no longer quite the one you meant to measure. With a 10 MΩ multimeter on a divider of 10 kΩ resistors you'll never notice. On a divider of megohm resistors, the reading can be off by several percent, and an engineer who doesn't expect it will chase a fault that isn't there. Thevenin's theorem (lesson 15) explains exactly how much, and the same thinking shows how meters themselves are built.

  • A divider of two equal resistors from 10 V, with a voltmeter of input resistance RM across its output. The meter reads the loaded voltage.
  • With RM = 10 MΩ (a typical multimeter) and 100 kΩ resistors, the reading is half a percent low. Make the resistors 1 MΩ.
  • How large must RM be, compared with the divider's resistors, for an error under 0.1%?

Predict

A 10 MΩ multimeter measures the midpoint of a divider made of two 1 MΩ resistors from 10 V. What does it read?

Name it

Voltmeter loading

A voltmeter is a resistance RM placed across the two points it measures. Replace the circuit, seen from those points, by its Thevenin equivalent Vth behind Rth, and the meter reads

Vread=VthRMRth+RM

The relative error is −RthRth+RM≈−RthRM: keep RM at least a hundred to a thousand times Rth. Handheld multimeters have about 10 MΩ; an oscilloscope input 1 MΩ, or 10 MΩ through a ×10 probe.

Ammeter burden

An ammeter is a small resistance RA (the burden) put in series with the branch. With the rest of the circuit as Vth behind Rth, it reads VthRth+RA instead of VthRth. A multimeter's current ranges can have ohms of burden, which matters in low-voltage circuits.

Building meters

A moving-coil meter needs a full-scale current Ifs through its coil of resistance Rc.

  • Voltmeter for full scale V: a series multiplier Rs=VIfs−Rc. Its sensitivity, 1/Ifs in Ω/V, is the input resistance per volt of range.
  • Ammeter for full scale I: a parallel shunt that carries the rest, Rsh=RcIfsI−Ifs.

Digital meters work the same way inside: a voltage range is a resistor divider, and a current range is a known shunt whose voltage is measured.

Small resistances

Measuring a fraction of an ohm with two leads adds the leads' resistance to the reading. A four-wire (Kelvin) measurement sends current through one pair of leads and senses voltage with another pair that carries almost no current, so the lead resistance drops out.

Worked example

A 10 MΩ meter measures the midpoint of a 100 kΩ/100 kΩ divider from 10 V, and then of a 1 MΩ/1 MΩ divider. What does it read each time?

Show the solution
  1. Thevenin equivalents. The two dividers are 5 V behind 50 kΩ and 5 V behind 500 kΩ.
  2. Reading on the 100 kΩ divider. v=5×1010.05=4.975V 0.5% low.
  3. Reading on the 1 MΩ divider. v=5×1010.5=4.76V 4.8% low.

Answer the meter reads 4.975 V on the 100 kΩ divider and 4.76 V on the 1 MΩ one.

Worked example

A meter movement needs 1 mA for full scale and has a 50 Ω coil. Turn it into a 10 V voltmeter and a 1 A ammeter.

Show the solution
  1. Voltmeter. Add a resistor in series with the coil: Rs=1010−3−50=9950Ω This gives an input resistance of 10 kΩ (1000 Ω/V).
  2. Ammeter. Add a shunt in parallel with the coil: Rsh=50×10−31−10−3=0.05005Ω It carries 999 mA while the coil carries 1 mA.

Answer 9950 Ω in series makes the 10 V voltmeter, and a 0.05005 Ω shunt in parallel makes the 1 A ammeter.

Worked example

An analog meter rated 20 kΩ/V is set to its 10 V range and measures the midpoint of a 100 kΩ/100 kΩ divider from 10 V. What does it read?

Show the solution
  1. Meter resistance. On the 10 V range, Rm=20kΩ/V×10V=200kΩ
  2. Thevenin equivalent of the divider. The midpoint is 5 V behind 100‖100=50kΩ
  3. Reading. v=5×20050+200=4.0V 20% low.
  4. Compare with a digital meter. A 10 MΩ digital meter would read 4.98 V.
  5. Why manuals named the meter. Old service manuals listed voltages "measured with a 20 kΩ/V meter" for exactly this reason: the reading depends on the meter as much as the circuit.

Answer the meter reads 4.0 V, 20% low.

Check

A point has a Thevenin equivalent of 5 V behind 50 kΩ. A meter with 10 MΩ input resistance measures it. What does it read?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

Check

A divider of two 100 kΩ resistors is measured at its midpoint by a 1 MΩ meter. By what percentage does the reading fall short?

Check

A 1 V source drives 10 Ω. An ammeter with 10 Ω of burden is put in series. What does it read?

Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.

Check

A movement needs 0.05 mA for full scale and has a 50 Ω coil. What shunt turns it into a 5 mA ammeter?

Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.

Check

A movement needs 1 mA for full scale and has a 100 Ω coil. What series resistor makes a 0.5 V voltmeter?

Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.

Check

A 0.1 Ω resistor is measured with two leads of 0.02 Ω each. By what percentage does the reading come out high?

Check

An analog meter rated 20 kΩ/V is on its 2.5 V range, measuring the midpoint of a 100 kΩ / 100 kΩ divider from 2.5 V.

(a)

What's the meter's resistance?

Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.

(b)

What does it read?

Check

A point is V behind R (kΩ), measured by a meter of M (kΩ).

Step 1

What does the meter read?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why does an oscilloscope probe often have a ×10 setting, when it makes signals ten times smaller on the screen?