C11 · Lesson 10 · Function generators and power supplies
Sources that behave the way their front panels say, mostly
Set a function generator to 1 V peak-to-peak, connect it to a scope, and the scope shows 2 V. Nothing is broken: the generator has a 50 Ω output and assumes it's driving a 50 Ω load, so its display shows what a matched load would see. Bench power supplies have their own surprises: set to 12 V with a 0.5 A limit, a supply becomes a 0.5 A current source the moment the load wants more, and its voltage drops to whatever that current gives. Knowing how a source is built tells you what it will actually deliver.
- A generator set to "1 V amplitude into 50 Ω" is really a 2 V source behind 50 Ω. The load is adjustable; the meter shows the amplitude the load actually gets.
- At = 50 Ω the load sees the 1 V the panel promises. At high impedance (like a scope's 1 MΩ input) it sees 2 V.
- Generators often have a "High Z" setting that just doubles the displayed number; the output itself is the same.
Predict
A generator is set to 1 Vpp (its default 50 Ω load setting) and connected straight to a scope with a 1 MΩ input. What does the scope show?
2 Vpp. Internally it's a 2 Vpp source behind 50 Ω, set up so that a 50 Ω load gets half, the 1 Vpp displayed. A 1 MΩ load takes almost none of the drop, so it sees nearly the full 2 Vpp.
Name it
Function generators
A generator is a voltage source (the open-circuit EMF) behind a 50 Ω output resistance. Its displayed amplitude assumes a 50 Ω load, so the EMF is twice the display, and a load gets
Amplitudes appear as peak-to-peak, RMS, or dBm. dBm is power into 50 Ω: mW, , and for a sine ; 0 dBm is 0.632 Vpp and 10 dBm is 2 Vpp. Use 50 Ω cable and terminations at high frequencies so the generator sees the load it expects.
Power supplies
A lab supply has two settings, a voltage and a current limit , and works in whichever mode the load requires:
- constant voltage when the load draws less than : the output is ;
- constant current when the load would draw more: the output is , and the voltage falls to .
The crossover is at . Set the limit just above what the circuit should draw, so a fault produces a current-limited supply instead of smoke.
Remote sensing uses a second pair of wires to measure the voltage at the load, so the supply corrects for the drop in its output leads.
Worked example
A generator set to 2 Vpp (50 Ω mode) drives a 100 Ω load. What does the load get?
Show the solution
- The EMF. In 50 Ω mode the display shows half the EMF, so the EMF is 4 Vpp.
- Divide into 100 Ω.
Answer the load gets 2.67 Vpp, neither the displayed 2 V nor the open-circuit 4 V.
Worked example
A supply is set to 12 V and 0.5 A. What does it deliver into 100 Ω, and into 10 Ω?
Show the solution
- Crossover.
- Into 100 Ω. It stays in constant voltage:
- Into 10 Ω. It limits the current at 0.5 A:
Answer into 100 Ω it delivers 12 V and 0.12 A; into 10 Ω it limits at 0.5 A, 5 V.
Worked example
A generator in 50 Ω mode is set to 1 Vpp and connected to a scope's 1 MΩ input. What does the scope show?
Show the solution
- Open-circuit output. The generator's display assumes a 50 Ω load, so its open-circuit output is 2 Vpp.
- Into 1 MΩ. Almost none of it drops across the source resistance:
- The fix. Either set the generator to high-impedance mode or terminate the scope input in 50 Ω.
Answer the scope shows 2.0 Vpp.
Check
A generator is set to 1 V amplitude for a 50 Ω load. What amplitude does a 25 Ω load actually get?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A generator outputs a 0 dBm sine into 50 Ω. What is its peak-to-peak voltage?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A supply is set to 12 V with a 0.1 A limit, driving a 100 Ω load.
(a)
At what load resistance does it switch between modes?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
What voltage does the 100 Ω load get?
Check
A supply delivers 3 A through leads of 0.02 Ω each. How much voltage is lost in the leads, which remote sensing would correct?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A sine of 2 Vpp is delivered into 50 Ω. What power is that in dBm?
Check
A circuit should draw 80 mA from a 5 V supply. Where should the current limit be set?
Check
A generator in 50 Ω mode is set to 2 Vpp and drives 25 Ω.
(a)
What peak-to-peak voltage does the load get?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What is the generator's open-circuit voltage?
Check
A supply is set to 12 V with a 0.1 A current limit and drives 500 Ω.
(a)
At what load resistance does it cross from voltage to current limiting?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
What current flows into 500 Ω?
Explain why
Why does a function generator's display show half its open-circuit voltage?
At high frequencies the generator drives 50 Ω cables that should be terminated in 50 Ω to avoid reflections, so its designers assume that load. A 50 Ω load and the generator's internal 50 Ω form a divider that halves the internal voltage, and the display shows that halved value, what a properly terminated load actually receives. Connected to a high-impedance input instead, almost nothing drops across the internal resistor, and the load sees the full internal voltage, twice what's displayed.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.