C11 · Lesson 18 · Low-level measurements
Microvolts, picoamps and gigaohms, where the wiring itself becomes the signal
Measure a 1 mΩ current shunt carrying 100 mA and the signal is 100 µV. Measure a photodiode's dark current or a gate's leakage and it's a few picoamps. At these levels the measuring setup produces errors as large as the signal: every junction of two metals is a thermocouple, every bit of insulation leaks, every cable has capacitance that takes seconds to charge through a gigaohm, and the resistors themselves hiss. Lesson 16 dealt with noise; this lesson deals with the systematic errors that averaging can't remove, and the techniques that do: reversing the source, guarding, and choosing materials and settling times on purpose.
- The dashed line is a thermal EMF in the measurement leads, starting at an offset and drifting by µV between successive readings. A single reading includes all of it.
- Current reversal takes readings with the excitation flipped and halves their difference, which removes a constant offset. But the offset drifts between the two readings, which leaves an error of half the drift per reading (the solid line).
- The delta method combines three alternating readings, , which also removes linear drift: its error, the line along zero, stays there whatever the drift.
Predict
A 1 mΩ shunt carries 100 mA. Its leads join copper through solder, which makes about 3 µV per °C of temperature difference, and the two joints differ by 2 °C. Roughly what error does that cause?
6%. The signal is = 100 µV, and the thermocouple formed by the two solder joints adds about = 6 µV. A good DMM resolves well under a microvolt, so the error is entirely in the wiring. Reversing the current flips the signal but not the thermal EMF, so the difference of two readings cancels it.
Name it
Thermal EMFs
Wherever two different metals meet at different temperatures, a Seebeck voltage appears. Typical coefficients against copper:
| Junction with copper | About |
|---|---|
| copper (clean, crimped) | ≤ 0.2 µV/°C |
| gold, silver | 0.3 µV/°C |
| tin–lead solder | 1 to 3 µV/°C |
| Kovar (IC leads, glass seals) | about 40 µV/°C |
| oxidized copper | about 1000 µV/°C |
Keep every connection in the signal path copper to copper, keep them at the same temperature, and let the setup settle thermally after handling it.
Reversal and the delta method
Thermal EMFs don't care which way the excitation current flows; the signal does. With readings and at and ,
cancels any constant offset. If the offset drifts linearly in time and readings are evenly spaced, alternating and taking
cancels the drift as well (with and at ). Nanovoltmeters and resistance bridges run this continuously.
Leakage and guarding
To measure picoamps, the current has to go only where you want it. Insulation of resistance with voltage across it leaks : 10 V across 100 GΩ is 100 pA. A guard is a conductor surrounding the high-impedance node, driven by a buffer to the same voltage as the node. The insulation between node and guard then has almost no voltage across it, so almost nothing leaks; whatever leaks from the guard to the outside comes from the buffer, not from the measurement. Guards appear as triaxial cables, as copper rings around sensitive pins on circuit boards, and as driven shields.
Guarding also hides the cable's capacitance: with no voltage change across it, it draws no charging current, which speeds settling.
Settling with high source resistance
A source of resistance driving a cable and input of capacitance settles as . Reaching a fraction of the final value takes : with 1 GΩ and 200 pF, 1.4 s to settle to 0.1%. Reading too soon looks like a small, repeatable error.
Noise floors
A resistor's Johnson noise, , sets the floor: 1 MΩ in a 10 Hz bandwidth gives 0.41 µV rms at room temperature. Low-level voltage measurements need low source resistance; low-level current measurements need high feedback resistance, since the current noise of a resistor, , falls as it grows.
Worked example
A 1 mΩ shunt carries ±100 mA in turn. The readings are 106 µV and −94 µV. What is the shunt's resistance, and what was the thermal offset?
Show the solution
- Shunt voltage. It is half the difference of the two readings:
- Resistance.
- Thermal offset. The offset is the average of the two readings, the thermocouple's contribution:
- A single reading. A single reading at +100 mA would have reported 1.06 mΩ, 6% high.
Answer the shunt is 1.000 mΩ, and the thermal offset was 6 µV.
Worked example
A picoammeter measures a sensor's 50 pA through a coaxial cable whose insulation between center and shield is 100 GΩ. The center sits at 10 V. What does leakage add? And if the shield is replaced by a guard held within 1 mV of the center?
Show the solution
- Unguarded. The full 10 V sits across the insulation: twice the signal.
- Guarded. Only 1 mV sits across it: a 0.02% error.
- The guard's own leakage. The guard itself still leaks 100 pA to the outer shield, but the guard's buffer supplies that current, not the sensor, so the meter never sees it.
Answer unguarded, leakage adds 100 pA, twice the signal; guarded, it adds 0.01 pA, a 0.02% error.
Worked example
A pH electrode has about 1 GΩ of source resistance and drives 200 pF of cable and meter input. How long must the meter wait to settle within 0.1% after the electrode goes in? And if a guard reduces the effective capacitance to 2 pF?
Show the solution
- Time constant.
- Settling to 0.1%.
- With the guard. The time constant drops to 2 ms and the wait to 14 ms.
- Where this matters. Meters for high-impedance sensors use guarded inputs or put the buffer at the electrode for this reason; the same applies to measuring a capacitor's leakage or a high-value resistor.
Answer about 1.38 s, and 14 ms with the guard.
Check
A four-wire measurement reads 112 µV with the current forward and −96 µV with it reversed. What is the true signal?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A connection pair with 1 µV/°C sits across a 0.5 °C temperature difference, in the path of a 1000 µV signal. What error does it cause, in percent?
Check
A high-impedance node at 10 V is insulated from its shield by 10 GΩ.
(a)
What current leaks if the shield is at 0 V?
Write units the way you normally would: 1.5 pA and 0.0015 nA are the same answer.
(b)
What leaks if the shield becomes a guard within 5 mV of the node?
Check
A 1 GΩ source drives 50 pF of cable and input. How long until the reading is within 0.001% of its final value?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
What is the rms Johnson noise of a 10 kΩ resistor at 300 K in a 1 Hz bandwidth, in nV?
Write units the way you normally would: 1.5 nV, 1500 pV and 0.0015 µV are the same answer.
Check
A guard around a picoamp input works because:
Check
Readings are taken at equal intervals with the excitation at , , . The signal is and the offset drifts as , with readings at = 0, 1, 2.
Step 1
What is , at = 0 with ?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is ?
Step 3
Form with .
Check
Simulate nanovolt-style measurements. A signal of V = 50 µV is read with the excitation alternating +, −, +, −, … at equal intervals, 20 readings in all (k = 0 to 19, the sign (-1)**k), while the thermal offset is 5 + d*k µV with d = 0.5 µV per reading, plus a curvature term 0.01*k**2. Compute two_point, the average of the 10 pair estimates , and delta, the average of the 18 three-point estimates for k = 0 to 17 (sign + when reading k is positive). Then set err2 and err3, their errors from V.
Python 3.14 with numpy and matplotlib, running in your browser. Tab indents; press Esc and then Tab to move on.
Explain why
A technician measures a 10 GΩ resistor by applying 100 V and reading the current with a picoammeter through a long, unguarded coaxial cable. The reading creeps upward for half a minute and settles about 15% high. Give two likely causes and how to fix each.
The creep is settling: 10 GΩ charging the cable's and meter's capacitance has a time constant of seconds, and dielectric absorption in the cable's insulation adds a slower tail, so the technician should wait several time constants or use a guarded or shorter cable to reduce the capacitance. The 15% excess is most likely leakage: with the cable's center at 100 V and its shield at ground, the insulation's resistance carries current in parallel with the resistor under test, and the meter can't tell them apart. Driving the cable's shield as a guard at the same potential as the center (or arranging the circuit so the high-impedance side sits at the meter's virtual ground with no voltage across the cable) removes that path, and clean, dry insulation such as PTFE helps as well.
Practice
Keep practicing
8 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.