title: Low-level measurements
heading: Microvolts, picoamps and gigaohms, where the wiring itself becomes the signal
eyebrow: C11 · Lesson 18 · Low-level measurements
xp: 40
---

::: hook
Measure a 1 mΩ current shunt carrying 100 mA and the signal is 100 µV. Measure a photodiode's dark current or a gate's leakage and it's a few picoamps. At these levels the measuring setup produces errors as large as the signal: every junction of two metals is a thermocouple, every bit of insulation leaks, every cable has capacitance that takes seconds to charge through a gigaohm, and the resistors themselves hiss. Lesson 16 dealt with noise; this lesson deals with the systematic errors that averaging can't remove, and the techniques that do: reversing the source, guarding, and choosing materials and settling times on purpose.
:::

::: play
- The dashed line is a thermal EMF in the measurement leads, starting at an offset and drifting by $d$ µV between successive readings. A single reading includes all of it.
- **Current reversal** takes readings with the excitation flipped and halves their difference, which removes a constant offset. But the offset drifts between the two readings, which leaves an error of half the drift per reading (the solid line).
- The **delta method** combines three alternating readings, $(V_1 - 2V_2 + V_3)/4$, which also removes linear drift: its error, the line along zero, stays there whatever the drift.
:::

```sim plot
view: graph
x: t = 0..10
y: e = -10..30
slider: d = 0..5 step 0.1 = 2
slider: o = 0..20 step 1 = 6
plot: o + d*t | dashed | label: thermal EMF, µV
plot: -d/2 + 0*t | label: error left by two-point reversal, µV
plot: 0*t | color: ch3 | label: error left by the delta method, µV
readout: two-point reversal error in µV = -d/2
```

::: predict
## A 1 mΩ shunt carries 100 mA. Its leads join copper through solder, which makes about 3 µV per °C of temperature difference, and the two joints differ by 2 °C. Roughly what error does that cause?

- [ ] 0.06%
- [ ] 0.6%
- [x] 6%
- [ ] 60%

::: reveal
6%. The signal is $10^{-3} \times 0.1$ = 100 µV, and the thermocouple formed by the two solder joints adds about $3 \times 2$ = 6 µV. A good DMM resolves well under a microvolt, so the error is entirely in the wiring. Reversing the current flips the signal but not the thermal EMF, so the difference of two readings cancels it.
:::
:::

::: nameit
## Thermal EMFs

Wherever two different metals meet at different temperatures, a Seebeck voltage appears. Typical coefficients against copper:

| Junction with copper | About |
|---|---|
| copper (clean, crimped) | ≤ 0.2 µV/°C |
| gold, silver | 0.3 µV/°C |
| tin–lead solder | 1 to 3 µV/°C |
| Kovar (IC leads, glass seals) | about 40 µV/°C |
| oxidized copper | about 1000 µV/°C |

Keep every connection in the signal path copper to copper, keep them at the same temperature, and let the setup settle thermally after handling it.

## Reversal and the delta method

Thermal EMFs don't care which way the excitation current flows; the signal does. With readings $V_+$ and $V_-$ at $+I$ and $-I$,

$$V = \frac{V_+ - V_-}{2}$$

cancels any constant offset. If the offset drifts linearly in time and readings are evenly spaced, alternating $+I, -I, +I$ and taking

$$V = \frac{V_1 - 2V_2 + V_3}{4}$$

cancels the drift as well (with $V_1$ and $V_3$ at $+I$). Nanovoltmeters and resistance bridges run this continuously.

## Leakage and guarding

To measure picoamps, the current has to go only where you want it. Insulation of resistance $R_{ins}$ with voltage $V$ across it leaks $V/R_{ins}$: 10 V across 100 GΩ is 100 pA. A **guard** is a conductor surrounding the high-impedance node, driven by a buffer to the same voltage as the node. The insulation between node and guard then has almost no voltage across it, so almost nothing leaks; whatever leaks from the guard to the outside comes from the buffer, not from the measurement. Guards appear as triaxial cables, as copper rings around sensitive pins on circuit boards, and as driven shields.

Guarding also hides the cable's capacitance: with no voltage change across it, it draws no charging current, which speeds settling.

## Settling with high source resistance

A source of resistance $R_s$ driving a cable and input of capacitance $C$ settles as $e^{-t/R_sC}$. Reaching a fraction $\varepsilon$ of the final value takes $R_sC\ln(1/\varepsilon)$: with 1 GΩ and 200 pF, 1.4 s to settle to 0.1%. Reading too soon looks like a small, repeatable error.

## Noise floors

A resistor's Johnson noise, $\sqrt{4kTRB}$, sets the floor: 1 MΩ in a 10 Hz bandwidth gives 0.41 µV rms at room temperature. Low-level voltage measurements need low source resistance; low-level current measurements need high feedback resistance, since the current noise of a resistor, $\sqrt{4kTB/R}$, falls as it grows.
:::

::: example
## A 1 mΩ shunt carries ±100 mA in turn. The readings are 106 µV and −94 µV. What is the shunt's resistance, and what was the thermal offset?

::: solution
1. **Shunt voltage.** It is half the difference of the two readings:
   $$\begin{aligned} V &= \frac{106 - (-94)}{2}\ \mu\text{V} \\ &= 100\ \mu\text{V} \end{aligned}$$
2. **Resistance.**
   $$\begin{aligned} R &= \frac{100\ \mu\text{V}}{100\ \text{mA}} \\ &= 1.000\ \text{m}\Omega \end{aligned}$$
3. **Thermal offset.** The offset is the average of the two readings, the thermocouple's contribution:
   $$\frac{106 + (-94)}{2}\ \mu\text{V} = 6\ \mu\text{V}$$
4. **A single reading.** A single reading at +100 mA would have reported 1.06 mΩ, 6% high.

**Answer:** the shunt is 1.000 mΩ, and the thermal offset was 6 µV.
:::
:::

::: example
## A picoammeter measures a sensor's 50 pA through a coaxial cable whose insulation between center and shield is 100 GΩ. The center sits at 10 V. What does leakage add? And if the shield is replaced by a guard held within 1 mV of the center?

::: solution
1. **Unguarded.** The full 10 V sits across the insulation:
   $$\frac{10\ \text{V}}{10^{11}\ \Omega} = 100\ \text{pA}$$
   twice the signal.
2. **Guarded.** Only 1 mV sits across it:
   $$\frac{10^{-3}\ \text{V}}{10^{11}\ \Omega} = 0.01\ \text{pA}$$
   a 0.02% error.
3. **The guard's own leakage.** The guard itself still leaks 100 pA to the outer shield, but the guard's buffer supplies that current, not the sensor, so the meter never sees it.

**Answer:** unguarded, leakage adds 100 pA, twice the signal; guarded, it adds 0.01 pA, a 0.02% error.
:::
:::

::: example
## A pH electrode has about 1 GΩ of source resistance and drives 200 pF of cable and meter input. How long must the meter wait to settle within 0.1% after the electrode goes in? And if a guard reduces the effective capacitance to 2 pF?

::: solution
1. **Time constant.**
   $$\begin{aligned} \tau &= R_sC \\ &= 10^9 \times 200 \times 10^{-12}\ \text{s} \\ &= 0.2\ \text{s} \end{aligned}$$
2. **Settling to 0.1%.**
   $$\begin{aligned} t &= \tau\ln(1/\varepsilon) \\ &= 0.2\ \text{s} \times \ln(1000) \\ &= 0.2 \times 6.91\ \text{s} \\ &= 1.38\ \text{s} \end{aligned}$$
3. **With the guard.** The time constant drops to 2 ms and the wait to 14 ms.
4. **Where this matters.** Meters for high-impedance sensors use guarded inputs or put the buffer at the electrode for this reason; the same applies to measuring a capacitor's leakage or a high-value resistor.

**Answer:** about 1.38 s, and 14 ms with the guard.
:::
:::

::: check
given: Vp = 104 | 112 | 155; Vm = -96 | -88 | -145
answer: (Vp - Vm)/2
unit: µV
tolerance: 0.5%

## A four-wire measurement reads {Vp} µV with the current forward and {Vm} µV with it reversed. What is the true signal?

::: right
Right. $(V_+ - V_-)/2$ = {answer}: the offset doesn't reverse, so it drops out.
:::

::: misses
- [(Vp + Vm)/2] That's the offset. The signal is half the difference.
- [Vp - Vm] Divide by two: the difference counts the signal twice.
- [Vp] A single reading includes the thermal offset.
- [any] Not yet. $(V_+ - V_-)/2$.
:::
:::

::: check
given: s = 0.5 | 1 | 3 | 40; dT = 0.5 | 2 | 5; V = 50 | 200 | 1000
answer: s*dT/V*100
tolerance: 0.5%

## A connection pair with {s} µV/°C sits across a {dT} °C temperature difference, in the path of a {V} µV signal. What error does it cause, in percent?

::: right
Right. The offset is $s\Delta T$, and $100 s\Delta T/V$ = {answer}%.
:::

::: misses
- [s*dT/V] Give a percentage.
- [s/V*100] Multiply by the temperature difference.
- [any] Not yet. Offset $= s\Delta T$; divide by the signal.
:::
:::

::: check
kind: parts
given: V = 5 | 10 | 50; R = 10 | 100 | 1000; dv = 0.1 | 1 | 5

## A high-impedance node at {V} V is insulated from its shield by {R} GΩ.

::: part
answer: V/(R*1e9)*1e12
unit: pA
tolerance: 0.5%

### What current leaks if the shield is at 0 V?

::: misses
- [V/R] Put the resistance in ohms: {R} GΩ is {R} × 10⁹ Ω.
- [any] Not yet. $V/R_{ins}$, in picoamps.
:::
:::

::: part
answer: dv*1e-3/(R*1e9)*1e12
unit: pA
tolerance: 0.5%

### What leaks if the shield becomes a guard within {dv} mV of the node?

::: misses
- [V/(R*1e9)*1e12] With the guard, only the small difference drives leakage.
- [dv/(R*1e9)*1e12] Put the guard's offset in volts: {dv} mV.
- [any] Not yet. $\Delta V/R_{ins}$.
:::
:::

::: right
Right. Guarding divides the leakage by the ratio of the node's voltage to the guard's error.
:::
:::

::: check
given: R = 0.1 | 1 | 10; C = 50 | 200 | 1000; e = 0.1 | 0.01 | 0.001
answer: R*1e9*C*1e-12*ln(100/e)
unit: s
tolerance: 0.5%

## A {R} GΩ source drives {C} pF of cable and input. How long until the reading is within {e}% of its final value?

::: right
Right. $R_sC\ln(1/\varepsilon)$ with $\varepsilon$ = {e}/100: {answer}.
:::

::: misses
- [R*1e9*C*1e-12*ln(1/e)] Convert the percentage to a fraction first.
- [R*1e9*C*1e-12] One time constant only gets within 37%.
- [R*1e9*C*1e-12*log(100/e)] Use the natural logarithm.
- [any] Not yet. $t = R_sC\ln(1/\varepsilon)$.
:::
:::

::: check
given: R = 1 | 10 | 100; B = 1 | 10 | 100; T = 77 | 300
answer: sqrt(4*1.380649e-23*T*R*1e3*B)*1e9
unit: nV
tolerance: 0.5%

## What is the rms Johnson noise of a {R} kΩ resistor at {T} K in a {B} Hz bandwidth, in nV?

::: right
Right. $\sqrt{4kTRB}$ = {answer}.
:::

::: misses
- [sqrt(4*1.380649e-23*T*R*B)*1e9] Put the resistance in ohms.
- [sqrt(1.380649e-23*T*R*1e3*B)*1e9] Johnson noise is $\sqrt{4kTRB}$.
- [any] Not yet. $\sqrt{4kTRB}$.
:::
:::

::: check
kind: choice

## A guard around a picoamp input works because:

- [x] It holds the insulation around the input at nearly zero voltage, so almost no current leaks through it
- [ ] It's grounded, so leakage currents flow to ground instead of the input
- [ ] It shields the input from electric fields, which induce currents
- [ ] It raises the insulation resistance by adding another layer

::: right
Right. A guard is driven to the input's own voltage. Leakage needs a voltage across the insulation; with the guard there's almost none, and the guard's own leakage to the outside is supplied by its buffer, not by the source being measured.
:::

::: misses
- [2] A grounded shield leaves the full input voltage across the insulation. A guard follows the input.
- [3] Shielding helps with interference, but guarding is about leakage and charging current.
- [4] The insulation is the same. What changes is the voltage across it.
:::
:::

::: check
kind: derivation
vars: V = 1e-6..1e-3, a = -1e-5..1e-5, b = -1e-6..1e-6

## Readings are taken at equal intervals with the excitation at $+I$, $-I$, $+I$. The signal is $\pm V$ and the offset drifts as $a + bt$, with readings at $t$ = 0, 1, 2.

::: step
answer: V + a

### What is $V_1$, at $t$ = 0 with $+I$?
:::

::: step
answer: -V + a + b

### What is $V_2$?
:::

::: step
answer: ((V + a) - 2*(-V + a + b) + (V + a + 2*b))/4

### Form $(V_1 - 2V_2 + V_3)/4$ with $V_3 = V + a + 2b$.
:::

::: right
Right. It simplifies to exactly $V$: both $a$ and $b$ cancel.
:::
:::

::: check
kind: python
given: V = 10 | 50 | 100; d = 0.5 | 2 | 5

## Simulate nanovolt-style measurements. A signal of `V` = {V} µV is read with the excitation alternating +, −, +, −, … at equal intervals, 20 readings in all (`k` = 0 to 19, the sign `(-1)**k`), while the thermal offset is `5 + d*k` µV with `d` = {d} µV per reading, plus a curvature term `0.01*k**2`. Compute `two_point`, the average of the 10 pair estimates $(V_{2m} - V_{2m+1})/2$, and `delta`, the average of the 18 three-point estimates $\pm(V_k - 2V_{k+1} + V_{k+2})/4$ for `k` = 0 to 17 (sign + when reading `k` is positive). Then set `err2` and `err3`, their errors from `V`.

```python starter
two_point = delta = V
err2 = err3 = 0.0
print(err2, err3)
```

```python solution
r = [V * (-1) ** k + 5 + d * k + 0.01 * k ** 2 for k in range(20)]
two_point = sum((r[2 * m] - r[2 * m + 1]) / 2 for m in range(10)) / 10
delta = sum((-1) ** k * (r[k] - 2 * r[k + 1] + r[k + 2]) / 4 for k in range(18)) / 18
err2, err3 = two_point - V, delta - V
print(err2, err3)
```

```python tests
r = [V * (-1) ** k + 5 + d * k + 0.01 * k ** 2 for k in range(20)]
t2 = sum((r[2 * m] - r[2 * m + 1]) / 2 for m in range(10)) / 10
t3 = sum((-1) ** k * (r[k] - 2 * r[k + 1] + r[k + 2]) / 4 for k in range(18)) / 18
assert abs(two_point - t2) < 1e-9 and abs(delta - t3) < 1e-9, "Compute both averages from the 20 readings as described."
assert abs(err2 - (t2 - V)) < 1e-9 and abs(err3 - (t3 - V)) < 1e-9, "The errors are the estimates minus V."
assert abs(err3) < 0.01 and abs(err2) > 0.2, "The three-point estimate should leave only the small curvature error, while pairs leave half the drift per reading."
```

::: right
Right. Pairs leave an error of half the drift per reading, because the offset changes between the two halves of each pair; three-point estimates remove linear drift completely and leave only the tiny curvature term.
:::
:::

::: explain
## A technician measures a 10 GΩ resistor by applying 100 V and reading the current with a picoammeter through a long, unguarded coaxial cable. The reading creeps upward for half a minute and settles about 15% high. Give two likely causes and how to fix each.

::: model
The creep is settling: 10 GΩ charging the cable's and meter's capacitance has a time constant of seconds, and dielectric absorption in the cable's insulation adds a slower tail, so the technician should wait several time constants or use a guarded or shorter cable to reduce the capacitance. The 15% excess is most likely leakage: with the cable's center at 100 V and its shield at ground, the insulation's resistance carries current in parallel with the resistor under test, and the meter can't tell them apart. Driving the cable's shield as a guard at the same potential as the center (or arranging the circuit so the high-impedance side sits at the meter's virtual ground with no voltage across the cable) removes that path, and clean, dry insulation such as PTFE helps as well.
:::
:::

::: practice

::: check
given: Vp = 220 | 510; Vm = -180 | -490
answer: (Vp - Vm)/2
unit: µV
tolerance: 0.5%

## Forward and reversed readings are {Vp} µV and {Vm} µV. What is the signal?

::: misses
- [(Vp + Vm)/2] That's the offset.
- [any] $(V_+ - V_-)/2$.
:::
:::

::: check
given: V = 1 | 10; R = 10 | 1000
answer: V/(R*1e9)*1e12
unit: pA
tolerance: 0.5%

## How much current leaks through {R} GΩ of insulation with {V} V across it?

::: misses
- [V/R] Convert gigaohms to ohms.
- [any] $I = V/R$.
:::
:::
:::