title: Multimeters in practice
heading: Loading, burden, four-wire resistance and true RMS
eyebrow: C11 · Lesson 4 · Multimeters in practice
xp: 40
---

::: hook
C1 showed that a voltmeter loads the circuit it measures. A bench meter hides several more traps. Its current ranges drop a voltage that can upset the circuit (the **burden**). Its leads add a fraction of an ohm that swamps the reading of a small resistor unless you use four wires. And on AC, a cheap meter that assumes a sine wave can be wrong by 10% or more on a square wave or a motor drive's chopped output. Knowing what the meter actually does turns its number from a guess into a measurement.
:::

::: play
- A 1 A current passes through a small resistor $R_x$ via two leads of resistance $R_\text{lead}$ each. A two-wire meter measures the voltage at the source end, leads included; a four-wire (Kelvin) meter senses right at the resistor.
- With 0.1 Ω under test and 0.2 Ω leads, the two-wire reading suggests 0.5 Ω: five times too high. The four-wire reading is exact, because its sense wires carry no current.
- Shrink the lead resistance: the two readings converge. Four-wire measurement matters whenever the resistor is comparable to the leads.
:::
:::

```sim circuit
I1 0 a 1            @ 0,0 up
RL1 a b 0.2         @ 0,1 right
RX b c 0.1          @ 2,1 down
RL2 c 0 0.2         @ 2,0 left
slider: RL1 0.001..1 log
slider: RL2 0.001..1 log
slider: RX 0.01..10 log
meter: v(a), v(b, c)
```

::: predict
## An average-responding meter, calibrated to read correctly on sine waves, measures a ±1 V square wave. What does it display?

- [ ] 0.71 V
- [ ] 1.00 V
- [x] 1.11 V
- [ ] 1.41 V

::: reveal
1.11 V. It rectifies and averages (giving 1 V for the square wave), then multiplies by 1.11, the ratio of RMS to rectified average for a sine. The square wave's true RMS is 1.00 V, so the reading is 11% high. A true-RMS meter would read 1.00 V.
:::
:::

::: nameit
## Voltage: loading

A meter of input resistance $R_m$ (typically 10 MΩ) across a source of Thévenin resistance $R_s$ reads $R_m/(R_s + R_m)$ of the true voltage. Below about 1% error, $R_s$ must be under $R_m/100$.

## Current: burden

An ammeter is a small shunt resistor; the drop across it, the **burden voltage**, $IR_\text{shunt}$, is subtracted from the circuit. A meter's 200 mA range might use 1 Ω, dropping 0.1 V at 100 mA, enough to matter in a 1 V circuit.

## Resistance: two wires or four

A two-wire measurement passes current through the leads and includes their resistance, $R_x + 2R_\text{lead}$. A **four-wire (Kelvin)** measurement forces current through one pair and senses voltage with another pair that carries no current, so it reads $R_x$ alone. It's standard below about 10 Ω and for sense resistors and windings.

## AC: what "RMS" means

The RMS value is $\sqrt{\overline{v^2}}$, the DC voltage that heats a resistor equally. An **average-responding** meter instead measures the rectified average and multiplies by 1.11 ($\pi/2\sqrt2$, right only for sines). A **true-RMS** meter computes $\sqrt{\overline{v^2}}$, up to a stated **crest factor** (peak over RMS). Useful RMS values:

| Waveform, amplitude A | RMS |
|---|---|
| sine | $A/\sqrt2$ |
| square (±A) | A |
| triangle | $A/\sqrt3$ |
| DC $V_0$ plus a sine | $\sqrt{V_0^2 + A^2/2}$ |
| pulses of height A, duty D | $A\sqrt D$ (crest factor $1/\sqrt D$) |
:::

::: example
## A 0.1 Ω current-sense resistor is measured two-wire with 0.2 Ω leads, then four-wire. What does each read?

::: solution
1. **Two-wire.** The reading includes both leads:
   $$\begin{aligned} R &= (0.1 + 2 \times 0.2)\ \Omega \\ &= 0.5\ \Omega \end{aligned}$$
   five times the true value.
2. **Four-wire.** It reads the true 0.1 Ω.
3. **In the sim.** The sim shows the same at 1 A: 0.5 V against 0.1 V.

**Answer:** two-wire reads 0.5 Ω, five times the true value, and four-wire reads 0.1 Ω.
:::
:::

::: example
## An average-responding meter measures a triangle wave of 1 V amplitude. What does it read, and what is the true RMS?

::: solution
1. **Rectified average.** The triangle's rectified average is 0.5 V.
2. **What the meter shows.** It scales the average by the sine's form factor, 1.11:
   $$1.11 \times 0.5\ \text{V} = 0.555\ \text{V}$$
3. **True RMS.**
   $$\frac{1\ \text{V}}{\sqrt3} = 0.577\ \text{V}$$
4. **The error.** The meter reads 3.8% low.

**Answer:** the meter reads 0.555 V, and the true RMS is 0.577 V, so the meter is 3.8% low.
:::
:::

::: example
## A 5 V supply drives 50 Ω, and a meter on its 200 mA range, with a 1 Ω shunt, measures the current. What does it read, and what is its burden voltage?

::: solution
1. **Without the meter.**
   $$\frac{5\ \text{V}}{50\ \Omega} = 100\ \text{mA}$$
2. **With the meter.** The shunt adds 1 Ω:
   $$\frac{5\ \text{V}}{51\ \Omega} = 98.0\ \text{mA}$$
   The meter's own resistance has changed the circuit by 2.0%.
3. **Burden voltage.** The drop across the shunt:
   $$98\ \text{mA} \times 1\ \Omega = 98\ \text{mV}$$
4. **In low-voltage circuits.** Measure the drop across a known resistor already in the circuit instead.

**Answer:** it reads 98.0 mA, 2.0% below the 100 mA that flows without it, and its burden voltage is 98 mV.
:::
:::

::: check
given: Rs = 10 | 100 | 1000; Rm = 1 | 10
answer: 100*Rs*1e3/(Rs*1e3 + Rm*1e6)
unit: %
tolerance: 0.5%

## A {Rm} MΩ meter measures a source with {Rs} kΩ of Thévenin resistance. By what percentage does it read low?

::: right
Right. $R_s/(R_s + R_m)$ = {answer}%.
:::

::: misses
- [any] Not yet. The divider formed by $R_s$ and $R_m$.
:::
:::

::: check
given: I = 20 | 100 | 400; R = 0.1 | 1 | 10
answer: I*R
unit: mV
tolerance: 0.5%

## An ammeter's range uses a {R} Ω shunt. What burden voltage does it drop at {I} mA?

::: right
Right. $IR$ = {answer} mV.
:::

::: misses
- [any] Not yet. Current times shunt resistance.
:::
:::

::: check
given: Rx = 0.05 | 0.5 | 2; Rl = 0.05 | 0.15
answer: Rx + 2*Rl
unit: Ω
tolerance: 0.5%

## A {Rx} Ω resistor is measured two-wire through leads of {Rl} Ω each. What does the meter read?

::: right
Right. $R_x + 2R_\text{lead}$ = {answer} Ω.
:::

::: misses
- [Rx + Rl] Both leads are in the current path.
- [Rx] That's what a four-wire measurement reads.
- [any] Not yet. Add both leads.
:::
:::

::: check
kind: parts
given: A = 1 | 5 | 10

## An average-responding meter (calibrated for sines) measures waves of amplitude {A} V.

::: part
answer: 1.1107*A
unit: V
tolerance: 0.5%

### What does it show for a ±{A} V square wave?

::: misses
- [A] That's the true RMS; this meter shows 1.11 × the rectified average.
- [any] Not yet. Rectified average × 1.11.
:::
:::

::: part
answer: 1.1107*A/2
unit: V
tolerance: 0.5%

### What does it show for a triangle wave of amplitude {A} V?

::: misses
- [A/sqrt(3)] That's the true RMS; the meter's reading differs.
- [any] Not yet. The triangle's rectified average is A/2.
:::
:::
:::

::: check
given: V0 = 1 | 2 | 5; A = 1 | 3 | 4
answer: sqrt(V0^2 + A^2/2)
unit: V
tolerance: 0.5%

## A signal is {V0} V DC plus a sine of amplitude {A} V. What does a true-RMS meter on DC+AC coupling read?

::: right
Right. $\sqrt{V_0^2 + A^2/2}$ = {answer} V.
:::

::: misses
- [V0 + A/sqrt(2)] RMS values combine in quadrature, not by adding.
- [any] Not yet. $\sqrt{V_0^2 + A^2/2}$.
:::
:::

::: check
given: D = 0.01 | 0.05 | 0.1 | 0.25
answer: 1/sqrt(D)
tolerance: 0.5%

## What is the crest factor of a pulse train with duty cycle {D}?

::: right
Right. Peak A over RMS $A\sqrt D$: {answer}.
:::

::: misses
- [1/D] RMS goes as $\sqrt D$, so the crest factor is $1/\sqrt D$.
- [any] Not yet. Peak over RMS.
:::
:::

::: check
kind: choice

## You need to measure the 0.2 Ω winding resistance of a motor to 1%. What method?

- [ ] A two-wire measurement on the meter's lowest range
- [x] A four-wire (Kelvin) measurement
- [ ] Measure with AC instead of DC
- [ ] Short the leads and subtract nothing

::: misses
- [1] The leads alone add a few tenths of an ohm, more than the winding.
- [3] AC adds the winding's inductance to the problem.
- [4] Lead resistance must be eliminated or subtracted; four-wire eliminates it.
:::
:::

::: check
kind: parts
given: R = 0.05 | 0.1 | 1; L = 0.1 | 0.25

## A {R} Ω resistor is measured with leads of {L} Ω each.

::: part
answer: R + 2*L
unit: Ω
tolerance: 0.5%

### What does a two-wire measurement read?

::: misses
- [R + L] Both leads are in series with it.
- [any] R plus both leads.
:::
:::

::: part
answer: 2*L/R*100
unit: %
tolerance: 0.5%

### By what percentage is that high?

::: misses
- [any] The lead resistance over the true value.
:::
:::
:::

::: check
kind: parts
given: V = 3.3 | 5 | 12; R = 20 | 100; Rs = 1 | 2 | 5

## A {V} V source drives {R} Ω, and an ammeter with {Rs} Ω of shunt resistance is inserted.

::: part
answer: V/(R + Rs)*1000
unit: mA
tolerance: 0.5%

### What does it read?

::: misses
- [V/R*1000] The shunt adds to the circuit's resistance.
- [any] $V/(R + R_s)$.
:::
:::

::: part
answer: -Rs/(R + Rs)*100
unit: %
tolerance: 0.5%

### What is its error relative to the true current?

::: misses
- [any] $-R_s/(R + R_s)$.
:::
:::
:::

::: explain
## Why does a four-wire measurement eliminate lead resistance?

::: model
In a two-wire measurement the same leads carry the test current and sense the voltage, so the voltage drop across the leads, current times lead resistance, is added to the drop across the resistor. A four-wire meter uses one pair of leads to force the current and a separate pair, connected right at the resistor's terminals, to sense the voltage. The sense leads go into a high-impedance input and carry essentially no current, so there's no drop across them, and the voltage they report is the resistor's alone.
:::
:::

::: practice

::: check
given: Rs = 1 | 10 | 50 | 100 | 500 | 1000; Rm = 10
answer: 100*Rs*1e3/(Rs*1e3 + Rm*1e6)
unit: %
tolerance: 0.5%

## Loading error with {Rs} kΩ source and {Rm} MΩ meter?

::: misses
- [any] $R_s/(R_s + R_m)$.
:::
:::

::: check
given: A = 0.5 | 1 | 2 | 3 | 5 | 6 | 10
answer: A/sqrt(3)
unit: V
tolerance: 0.5%

## True RMS of a triangle of amplitude {A} V?

::: misses
- [any] $A/\sqrt3$.
:::
:::

::: check
given: Rx = 0.1 | 0.2 | 0.5 | 1 | 2 | 10; Rl = 0.1
answer: Rx + 2*Rl
unit: Ω
tolerance: 0.5%

## Two-wire reading of {Rx} Ω with {Rl} Ω leads?

::: misses
- [any] $R_x + 2R_\text{lead}$.
:::
:::
:::