C11 · Lesson 8 · Oscilloscopes

Bandwidth, rise time, sampling and triggering

An oscilloscope draws a voltage against time, so it's tempting to believe what it draws. But a scope is itself a low-pass filter followed by an ADC: it rounds off edges faster than its own rise time, shrinks signals near its bandwidth, and with the wrong settings can show an aliased waveform that isn't there at all. Choosing a scope and setting it up well comes down to a few numbers: bandwidth and rise time, sample rate and memory depth, and how the trigger decides which part of a repeating signal to show.

  • A step with rise time tr (10 to 90%) enters a scope whose bandwidth gives it its own rise time ts=0.35/BW. The dashed curve is the true signal; the solid one is what the scope displays.
  • With ts much shorter than tr, the display is faithful. Make them equal: the displayed edge is 41% slower than the real one.
  • The readout gives the displayed rise time, close to tr2+ts2. That's why a scope should be three to five times faster than the edges it measures.

Predict

A 350 MHz scope (rise time 1 ns) displays an edge that truly rises in 1 ns. What rise time does it show?

Name it

Bandwidth and rise time

A scope's bandwidth is the frequency where a sine reads 3 dB low (70.7% of its true amplitude). For the usual response shape, its rise time is

ts≈0.35BW

and the displayed rise time of an edge is about tr2+ts2 (add the probe's too, in quadrature). The rule of thumb: bandwidth three to five times the highest frequency of interest. A sine near the bandwidth is understated: with a single-pole response, a sine at f reads 1/1+(f/BW)2 of its amplitude.

Sampling and memory

Digital scopes sample at a rate fs, at least 2.5 times the bandwidth to avoid aliasing (C9.21), and store samples in memory: the window it can capture at full rate is memory depth over sample rate. A long window with shallow memory forces a lower sample rate, and aliasing can appear.

Vertical resolution

Most scopes use 8-bit converters: 256 levels across the screen's full height (usually 8 divisions), so at 100 mV/div one level is 3.1 mV. Averaging repetitive signals, or high-resolution modes, gains bits.

Triggering

The trigger picks the moment that's placed at the reference point on the screen, so a repeating signal stays still: an edge crossing a level, with a slope, and optionally a holdoff time during which new triggers are ignored. Input coupling is DC (everything), AC (blocks DC, so a small ripple on a large DC level can be magnified), and the input impedance is usually 1 MΩ with ~15 pF, or 50 Ω for fast signals on matched cables.

Worked example

A scope with 1.5 ns rise time shows an edge rising in 1.8 ns. How fast is the edge really?

Show the solution
  1. Subtract in quadrature. tr=1.82−1.52ns=1.0ns
  2. What that means. The scope accounts for most of what's displayed; subtracting in quadrature recovers the signal's own rise time, though the result gets uncertain when the two are close.

Answer the edge really rises in about 1.0 ns.

Worked example

A scope samples at 2 GS/s with 1 M points of memory. How long a window can it capture at full rate?

Show the solution
  1. Window at full rate. T=1062×109s=0.5ms
  2. A longer window. To view 10 ms, it must drop to 100 MS/s, so a 20 MHz bandwidth limit (or less) is needed to avoid aliasing at that setting.

Answer 0.5 ms at the full 2 GS/s.

Worked example

A 100 MHz oscilloscope with a Gaussian response measures a 50 MHz sine. How low does its amplitude read?

Show the solution
  1. The Gaussian response. It is −3 dB at its bandwidth: |H|=e−ln⁡2(f/BW)2
  2. At half the bandwidth. (f/BW)2=0.25 and ln⁡2=0.347, so |H|=e−0.347×0.25=0.917 That's 8.3% low.
  3. Rule of thumb. That's why the rule of thumb is a scope with three to five times the bandwidth of the signal for amplitude measurements.

Answer it reads 8.3% low.

Check

What is the rise time of a 350 MHz scope?

Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.

Check

An edge with a 2 ns rise time is viewed on a scope with a 0.35 ns rise time. About what rise time is displayed?

Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.

Check

A scope with a 0.7 ns rise time displays an edge rising in 3 ns. What is the edge's own rise time?

Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.

Check

A single-pole 100 MHz scope measures a 100 MHz sine. What percentage of its true amplitude does it show?

Check

A scope with 10 Mpts of memory samples at 0.5 GS/s. How long a window can it record?

Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.

Check

A scope with an 8-bit converter is set to 50 mV/div across 8 divisions. How big is one quantization step?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

A 5 V supply's 20 mV switching ripple needs to be seen in detail. Which input setting helps most?

Check

A scope with 0.35 ns rise time shows an edge of 3 ns.

(a)

How fast is the edge really?

Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.

(b)

What is the scope's bandwidth?

Check

A 100 MHz scope with a Gaussian response measures a sine at 0.3 times its bandwidth.

(a)

What fraction of the amplitude does it show?

(b)

How low is that reading?

Explain why

Why should a scope's bandwidth be several times higher than the frequency of the signal you care about?