title: Biopotential electrodes
heading: How electrodes turn ionic current into electronic current, half-cell potentials and the offsets they create, polarizable and nonpolarizable electrodes, the double-layer equivalent circuit and its impedance against frequency, skin, motion artifact, and microelectrodes
eyebrow: S15 · Lesson 14 · Biopotential electrodes
xp: 50
---

::: hook
In the body, current is carried by ions; in the amplifier, by electrons. Every biopotential measurement depends on the junction where one becomes the other, and that junction is a small electrochemical cell with its own voltage, its own impedance and its own noise. A good electrode, like the silver/silver-chloride snap on an ECG pad, keeps its voltage steady and its impedance low. A poor one adds hundreds of millivolts of drifting offset and turns every movement of the patient into a signal larger than the heartbeat.
:::

::: play
- The magnitude of an electrode's impedance against frequency, from the model $R_s + R_d \| C_d$: a series resistance (electrolyte and gel) plus the charge-transfer resistance in parallel with the double-layer capacitance.
- At low frequencies the impedance is $R_s + R_d$; above the corner $1/(2\pi R_dC_d)$ the capacitor shorts out $R_d$ and only $R_s$ remains.
- Biopotentials sit at the low end, where the impedance is largest, so amplifier input impedance must be far larger still.
:::

```sim plot
view: graph
x: f = 0.01..10000 log
y: Z = 0.1..1000 log
slider: s = 0.2..5 step 0.1 = 1
slider: R = 10..1000 = 100 log
slider: C = 0.01..10 = 0.1 log
plot: sqrt((s + R/(1 + (2*pi*f*R*1e3*C*1e-6)^2))^2 + (R*2*pi*f*R*1e3*C*1e-6/(1 + (2*pi*f*R*1e3*C*1e-6)^2))^2) | samples: 500 | label: |Z|, kΩ
readout: corner frequency, Hz = 1/(2*pi*R*1e3*C*1e-6) | digits: 3
readout: |Z| at 10 Hz, kΩ = sqrt((s + R/(1 + (2*pi*10*R*1e3*C*1e-6)^2))^2 + (R*2*pi*10*R*1e3*C*1e-6/(1 + (2*pi*10*R*1e3*C*1e-6)^2))^2) | digits: 4
```

::: predict
## Two Ag/AgCl electrodes on the chest show a DC offset of 300 mV between them, the most a standard ECG input must tolerate. The amplifier's first stage runs from ±5 V supplies. What's the largest first-stage gain that avoids saturating on the offset alone?

- [ ] 1000
- [ ] 300
- [x] About 16
- [ ] 1

::: reveal
300 mV times a gain of 16.7 reaches 5 V, so the first stage can only have a gain of about 16 (less with real output swing). The rest of the gain must come after a high-pass filter removes the DC offset, which is why biopotential amplifiers are built in stages.
:::
:::

::: nameit
## The electrode–electrolyte interface

At a metal in contact with an electrolyte, oxidation and reduction reactions exchange charge, and ions arrange themselves into a charged **double layer** a few nanometers thick. At equilibrium the interface sits at its **half-cell potential**, measured against the standard hydrogen electrode: about +0.223 V for Ag/AgCl, +0.80 V for silver, −0.76 V for zinc. Two identical electrodes cancel, but differences in skin, gel and contact leave offsets of tens to hundreds of millivolts.

## Polarizable and nonpolarizable electrodes

A **perfectly polarizable** electrode (platinum, roughly) passes no real charge across the interface; it behaves like a capacitor, and current shifts its voltage (the overpotential). A **nonpolarizable** electrode passes current freely without changing its potential. Ag/AgCl comes close: the silver chloride layer exchanges chloride ions with the gel, giving a stable, low-noise, low-drift contact, which makes it the standard for ECG.

## The equivalent circuit

$$Z(f) = R_s + \frac{R_d}{1 + j2\pi fR_dC_d}$$

with the half-cell potential in series. $R_s$ is the electrolyte and gel; $R_d$ and $C_d$ model charge transfer and the double layer. The stratum corneum of dry skin adds a similar, larger RC section, which abrasion and gel reduce.

## Motion artifact

Moving the electrode disturbs the double layer and the skin's own potential, producing low-frequency artifacts of millivolts or more. Ag/AgCl with gel, firm attachment and good skin preparation reduce them; a high-pass filter removes what's slower than the ECG.

## Microelectrodes

Glass micropipettes and fine metal tips for recording inside single cells have megohm resistances. Together with stray capacitance of picofarads they form a low-pass filter, $f_c = 1/(2\pi RC)$, that can blur fast action potentials unless the amplifier cancels the capacitance.
:::

::: example
## An electrode has $R_s$ = 1 kΩ, $R_d$ = 100 kΩ and $C_d$ = 0.1 µF. Find its impedance at 0.1 Hz, 10 Hz and 1 kHz, and its corner frequency.

::: solution
1. **Corner frequency.**
   $$\begin{aligned} f_c &= \frac{1}{2\pi R_d C_d} \\ &= \frac{1}{2\pi \times 10^5 \times 10^{-7}}\ \text{Hz} \\ &= 15.9\ \text{Hz} \end{aligned}$$
2. **Impedance at each frequency.**
   $$\begin{aligned} |Z(0.1\ \text{Hz})| &\approx 101\ \text{k}\Omega \\ |Z(10\ \text{Hz})| &= 85.5\ \text{k}\Omega \\ |Z(1\ \text{kHz})| &= 1.89\ \text{k}\Omega \end{aligned}$$
   approaching $R_s$ at 1 kHz.
3. **Where the ECG sits.** The ECG band sits mostly below the corner, at the high-impedance end.

**Answer:** the corner is at 15.9 Hz, and |Z| is about 101 kΩ at 0.1 Hz, 85.5 kΩ at 10 Hz and 1.89 kΩ at 1 kHz.
:::
:::

::: example
## A 100 kΩ electrode drives an amplifier input of 10 MΩ. What fraction of the signal is lost? And what bandwidth does a 10 MΩ microelectrode with 10 pF of stray capacitance allow?

::: solution
1. **Divider loss.**
   $$\begin{aligned} \text{loss} &= \frac{100\ \text{k}}{100\ \text{k} + 10\ \text{M}} \\ &= 0.99\% \end{aligned}$$
   at low frequencies, and much more if the two electrodes differ, as lesson 15 shows.
2. **Microelectrode bandwidth.**
   $$\begin{aligned} f_c &= \frac{1}{2\pi \times 10^7 \times 10^{-11}}\ \text{Hz} \\ &= 1.59\ \text{kHz} \end{aligned}$$
   low enough to round off a 1 ms action potential.

**Answer:** 0.99% of the signal is lost, and the microelectrode's bandwidth is 1.59 kHz.
:::
:::

::: example
## A pair of electrodes presents 10 kΩ of resistance at body temperature. How much thermal noise does that add over the 0.05 to 150 Hz diagnostic band, compared with a 1 mV ECG?

::: solution
1. **Noise power.** For 10 kΩ at 310 K over the 150 Hz band:
   $$\begin{aligned} v_n^2 &= 4kTRB \\ &= 4 \times 1.38 \times 10^{-23} \times 310 \\ &\quad \times 10^4 \times 150\ \text{V}^2 \end{aligned}$$
2. **Noise voltage.**
   $$v_n = 0.16\ \mu\text{V rms}$$
   nearly 80 dB below the 1 mV ECG.
3. **What limits surface ECG.** Thermal noise is rarely the limit; motion artifact and fluctuations of the half-cell potential, often tens of microvolts, are.
4. **EEG.** At 10 to 100 µV, EEG is small enough that the same 0.16 µV starts to matter, and electrodes for small signals are prepared to bring their impedance down.

**Answer:** the electrodes add 0.16 µV RMS of thermal noise, nearly 80 dB below a 1 mV ECG.
:::
:::

::: check
given: s = 0.5 | 2; R = 50 | 200; C = 0.05 | 0.5; f = 10 | 30 | 100
answer: sqrt((s + R/(1 + (2*pi*f*R*1e3*C*1e-6)^2))^2 + (R*2*pi*f*R*1e3*C*1e-6/(1 + (2*pi*f*R*1e3*C*1e-6)^2))^2)
unit: kΩ
tolerance: 0.5%

## An electrode has $R_s$ = {s} kΩ, $R_d$ = {R} kΩ and $C_d$ = {C} µF. What's the magnitude of its impedance at {f} Hz?

::: right
Right. $|R_s + R_d/(1 + j\omega R_dC_d)|$ = {answer} kΩ.
:::

::: misses
- [s + R] That's the DC value; at {f} Hz the capacitor carries part of the current.
- [any] Not yet. Combine $R_d$ and $C_d$ in parallel, then add $R_s$.
:::
:::

::: check
given: R = 20 | 100 | 500; C = 0.02 | 0.1 | 1
answer: 1/(2*pi*R*1e3*C*1e-6)
unit: Hz
tolerance: 0.5%

## Where's the corner frequency of an electrode with $R_d$ = {R} kΩ and $C_d$ = {C} µF?

::: right
Right. $1/(2\pi R_dC_d)$ = {answer} Hz.
:::

::: misses
- [1/(R*1e3*C*1e-6)] That's in radians per second.
- [any] Not yet. The RC corner.
:::
:::

::: check
given: o = 100 | 300 | 500; V = 2.5 | 5 | 12
answer: V/(o*1e-3)
tolerance: 0.5%

## Electrodes present a {o} mV DC offset to an amplifier stage whose output can swing ±{V} V. What's the largest DC-coupled gain it can use?

::: right
Right. $V_\text{max}/V_\text{offset}$ = {answer}.
:::

::: misses
- [V/o] Convert millivolts to volts.
- [any] Not yet. The offset times the gain must stay within the swing.
:::
:::

::: check
given: a = 0.223 | 0.80; b = -0.76 | -0.44 | 0.34
answer: a - b
unit: V
tolerance: 0.5%

## Two electrodes with half-cell potentials {a} V and {b} V share an electrolyte. What voltage appears between them?

::: right
Right. The difference, {answer} V.
:::

::: misses
- [a + b] Subtract the half-cell potentials.
- [any] Not yet. Each is measured against the same hydrogen reference.
:::
:::

::: check
given: R = 2 | 10 | 50; C = 2 | 10 | 30
answer: 1/(2*pi*R*1e6*C*1e-12)
unit: Hz
tolerance: 0.5%

## A microelectrode of {R} MΩ sees {C} pF of stray capacitance. Where's the resulting low-pass corner?

::: right
Right. $1/(2\pi RC)$ = {answer} Hz.
:::

::: misses
- [any] Not yet. RC low-pass corner.
:::
:::

::: check
given: Z = 20 | 100 | 500; R = 1 | 10 | 100
answer: 100*Z/(Z + R*1000)
tolerance: 0.5%

## A {Z} kΩ electrode drives an amplifier with {R} MΩ input resistance. What percent of the signal is lost?

::: right
Right. $Z_e/(Z_e + Z_\text{in})$ = {answer}%.
:::

::: misses
- [any] Not yet. The voltage divider formed with the input.
:::
:::

::: check
kind: parts
given: R = 2 | 20 | 200; B = 40 | 150 | 1000; s = 10 | 1000

## An electrode pair presents {R} kΩ at 37 °C, recorded over {B} Hz of bandwidth, for a {s} µV signal.

::: part
answer: sqrt(4*1.380649e-23*310.15*R*1e3*B)*1e6
unit: µV
tolerance: 0.5%

### How much thermal noise does it add?

::: misses
- [sqrt(1.380649e-23*310.15*R*1e3*B)*1e6] Thermal noise is $\sqrt{4kTRB}$.
- [any] $\sqrt{4kTRB}$.
:::
:::

::: part
answer: 20*log(s/(sqrt(4*1.380649e-23*310.15*R*1e3*B)*1e6))
unit: dB
tolerance: 0.5%

### What signal-to-noise ratio does that leave, considering only this noise?

::: misses
- [any] $20\log(v_s/v_n)$.
:::
:::
:::

::: check
kind: derivation
vars: s = 0.1..10, R = 1..1000, C = 0.001..10

## An electrode's model has series resistance $R_s$, and a double layer of resistance $R_d$ in parallel with capacitance $C_d$ (kΩ and µF).

::: step
answer: s + R

### What impedance does it present at DC, in kΩ?
:::

::: step
answer: s

### And at very high frequencies?
:::

::: step
answer: 1/(2*pi*R*1e3*C*1e-6)

### Where is the corner between them, in hertz?
:::

::: right
Right. A recording amplifier's input impedance must stay well above the DC value, the largest the electrode presents.
:::
:::

::: explain
## Why are Ag/AgCl electrodes the standard for recording the ECG?

::: model
An electrode must pass small currents between ionic tissue and electronic wires without its own potential changing, because any change in its half-cell potential looks like signal. Silver coated with silver chloride in a chloride gel is nearly nonpolarizable: the reaction Ag + Cl⁻ ⇌ AgCl + e⁻ runs freely in both directions, so charge crosses the interface easily and the potential stays fixed by the chloride concentration. That gives a stable, reproducible half-cell potential, low impedance and low noise, and much smaller motion artifacts than polarizable metals like stainless steel or platinum, whose double layer behaves like a capacitor whose charge shifts whenever the electrode moves.
:::
:::

::: practice

::: check
given: R = 5 | 1000; C = 0.01 | 5
answer: 1/(2*pi*R*1e3*C*1e-6)
unit: Hz
tolerance: 0.5%

## Corner: $R_d$ = {R} kΩ, $C_d$ = {C} µF?

::: misses
- [any] $1/(2\pi R_dC_d)$.
:::
:::

::: check
given: o = 10 | 800; V = 1 | 15
answer: V/(o*1e-3)
tolerance: 0.5%

## Max gain: {o} mV offset, ±{V} V swing?

::: misses
- [any] Swing over offset.
:::
:::

::: check
given: Z = 1 | 2000; R = 0.5 | 1000
answer: 100*Z/(Z + R*1000)
tolerance: 0.5%

## Loss: {Z} kΩ electrode, {R} MΩ input?

::: misses
- [any] $Z_e/(Z_e + Z_\text{in})$.
:::
:::
:::