title: Biotelemetry and implant links
heading: Sending physiological signals by radio, sampling and bit rates for telemetered biopotentials, the frequency bands for medical telemetry and implants, how tissue absorbs radio waves, the energy cost per bit, and powering implants through the skin by inductive coupling
eyebrow: S15 · Lesson 49 · Biotelemetry
xp: 50
---

::: hook
Webster calls radiotelemetry the best isolation there is: a battery-powered transmitter on the patient has no connection at all to the mains. Hospital cardiac monitors have long used it, and today every implant that talks to a programmer or a phone, and every one recharged through the skin, depends on a wireless link through tissue that absorbs radio waves and limits how much power may be deposited in it. The design questions are how many bits must be sent, at what frequency, with what energy, and how much power can be pushed in.
:::

::: play
- Two coils couple a fraction k of their magnetic flux, and each has quality factor Q. The curve is the best power-transfer efficiency the link can reach with an optimal load, $\eta = k^2Q^2/(1 + \sqrt{1 + k^2Q^2})^2$, against k.
- Implants sit centimeters from their external coils, so k is a few percent. High-Q coils rescue the link: what matters is the product kQ, not k alone.
- Efficiency rises steeply once kQ passes about 1.
:::

```sim plot
view: graph
x: k = 0.001..0.5 log
y: e = 0..1
slider: Q = 10..300 = 100 log
plot: k^2*Q^2/(1 + sqrt(1 + k^2*Q^2))^2 | label: best efficiency
```

::: predict
## An implant's coil couples k = 0.05 to an external coil; both have Q = 100. What's the best power-transfer efficiency?

- [ ] About 0.25%, since only 5% of the flux couples
- [ ] About 25%
- [x] About 67%
- [ ] About 99%

::: reveal
With $k^2Q_1Q_2$ = 0.0025 × 10⁴ = 25, the optimum efficiency is 25/(1 + √26)² = 0.67. Resonating both coils lets energy slosh between them many times per cycle, so a weak coupling transfers power efficiently as long as the coils lose little per cycle. That is why implant chargers use resonant, high-Q coils.
:::
:::

::: nameit
## Telemetry systems

A telemetry transmitter amplifies the signal, digitizes it and modulates a carrier; the receiver recovers the data. Webster's rule is to sample at least five times the highest frequency of interest, though twice suffices in theory, so the bit rate is

$$\text{bit rate} = \text{channels} \times 5f_\text{max} \times \text{bits per sample}$$

Telemetry is immune to ground loops but not to interference at its own frequency, nor to dropouts when the antennas' orientations null the link, so good systems report signal loss.

## Frequency bands

Hospital patient telemetry in the US uses the Wireless Medical Telemetry Service bands (608 to 614, 1395 to 1400 and 1427 to 1432 MHz) to avoid interference from other users. Implants use the Medical Implant Communication Service band, 402 to 405 MHz, with channels up to 300 kHz wide and transmitted power limited to 25 µW EIRP, and wearables commonly use Bluetooth Low Energy at 2.4 GHz.

## Radio through tissue

Tissue is a lossy dielectric: muscle at 403 MHz has a relative permittivity of about 57 and a conductivity of about 0.8 S/m. A plane wave's field decays as $e^{-\alpha x}$ with
$$\alpha = \omega\sqrt{\frac{\mu\epsilon}{2}}\sqrt{\sqrt{1 + \left(\frac{\sigma}{\omega\epsilon}\right)^2} - 1}$$
about 19 Np/m for muscle at 403 MHz: 1.65 dB per centimeter. Higher frequencies lose more, which is one reason implants use the 400 MHz band rather than 2.4 GHz. The absorbed power is also limited by the SAR rules of lesson 48.

## Energy per bit

A radio's energy cost is roughly constant per bit sent, from nanojoules for low-power links to tens of nanojoules with overhead, so average power is the bit rate times the energy per bit, and sending features or detected events rather than raw samples saves energy in proportion.

## Inductive power

Implants that need more power than a primary battery can supply for their lifetime are powered or recharged by a coil outside the body coupled to one inside. For coupling coefficient k and quality factors $Q_1$ and $Q_2$, the best efficiency, with the load chosen optimally, is
$$\eta_\max = \frac{k^2Q_1Q_2}{\left(1 + \sqrt{1 + k^2Q_1Q_2}\right)^2}$$
Coupling falls steeply with distance and misalignment, so links are designed for the worst position. The implant can send data back by changing its coil's load, which the external reader sees as a change in its own coil's current (load modulation), with no transmitter of its own.
:::

::: example
## A three-lead ECG telemeter keeps components up to 150 Hz, samples at five times that with 12 bits, and transmits at 5 nJ per bit. What bit rate and average radio power does it need?

::: solution
1. **Sample rate.** Five times 150 Hz is 750 samples a second per lead.
2. **Bit rate.**
   $$\begin{aligned} R &= 3 \times 750 \times 12 \\ &= 27\ \text{kb/s} \end{aligned}$$
3. **Radio power.** At 5 nJ per bit:
   $$P = 135\ \mu\text{W}$$
   That's 3.24 mWh a day, 0.9 mAh from a 3.6 V cell.
4. **Sending less.** Sending only detected beats and their intervals, a few hundred bits a minute, would cut that by a factor of a thousand.

**Answer:** the telemeter needs 27 kb/s, and its radio averages 135 µW.
:::
:::

::: example
## An implant needs 10 mW delivered to its electronics. Its coil couples k = 0.02 to the external coil, with Q = 100 outside and 50 inside. What's the best efficiency, and how much power must the external unit put into its coil?

::: solution
1. **Coupling figure.**
   $$\begin{aligned} k^2Q_1Q_2 &= 0.0004 \times 5000 \\ &= 2 \end{aligned}$$
2. **Best efficiency.**
   $$\begin{aligned} \eta_\max &= \frac{2}{(1 + \sqrt3)^2} \\ &= 0.27 \end{aligned}$$
3. **External power.**
   $$\begin{aligned} P &= \frac{10\ \text{mW}}{0.27} \\ &= 37\ \text{mW} \end{aligned}$$
   and the remaining 27 mW heats the coils and tissue.
4. **Closer coils.** If k doubles, $k^2Q_1Q_2$ rises to 8 and the efficiency to 0.50, so the external coil needs 20 mW instead of 37.

**Answer:** the best efficiency is 0.27, so the external unit must put 37 mW into its coil.
:::
:::

::: example
## How much does muscle attenuate a 403 MHz signal from an implant 3 cm deep, and how does that compare with 2.45 GHz, where muscle has a relative permittivity of about 53 and a conductivity of about 1.7 S/m?

::: solution
1. **At 403 MHz.** With α ≈ 19 Np/m, 3 cm costs
   $$\begin{aligned} L &= 8.686 \times 19 \times 0.03 \\ &= 5.0\ \text{dB} \end{aligned}$$
2. **At 2.45 GHz.** The same formula gives α ≈ 45 Np/m, 3.9 dB/cm, so the path costs 11.7 dB, more than twice as much.
3. **Near-field coupling.** The shorter wavelength also makes the antenna's near-field coupling to tissue worse.
4. **Which band.** The lower band wins for deep implants, though the higher one offers more bandwidth for shallow wearables.

**Answer:** 3 cm of muscle costs 5.0 dB at 403 MHz and 11.7 dB at 2.45 GHz, more than twice as much.
:::
:::

::: check
kind: parts
given: c = 1 | 3 | 12; f = 40 | 150 | 500; b = 10 | 12 | 16; e = 1 | 10 | 50

## A telemeter sends {c} channels with content up to {f} Hz, sampled at five times that with {b} bits, at {e} nJ per bit.

::: part
answer: c*5*f*b/1e3
unit: kb/s
tolerance: 0.5%

### What bit rate does it need?

::: misses
- [c*2*f*b/1e3] Webster's rule samples at five times the highest frequency.
- [any] Channels × 5f × bits.
:::
:::

::: part
answer: c*5*f*b*e*1e-9*1e6
unit: µW
tolerance: 0.5%

### What average radio power does that take?

::: misses
- [any] Bit rate times energy per bit.
:::
:::
:::

::: check
kind: parts
given: k = 0.01 | 0.03 | 0.1; a = 50 | 200; b = 20 | 80; P = 5 | 50

## An implant's coil couples k = {k} to an external coil; their quality factors are {a} and {b}, and the implant needs {P} mW.

::: part
answer: k^2*a*b/(1 + sqrt(1 + k^2*a*b))^2
tolerance: 0.5%

### What is the best efficiency the link can reach?

::: misses
- [k^2] High-Q coils transfer far more than the coupled fraction of flux.
- [any] $k^2Q_1Q_2/(1 + \sqrt{1 + k^2Q_1Q_2})^2$.
:::
:::

::: part
answer: P/(k^2*a*b/(1 + sqrt(1 + k^2*a*b))^2)
unit: mW
tolerance: 0.5%

### How much power must the external coil deliver?

::: misses
- [any] Needed power over efficiency.
:::
:::
:::

::: check
kind: parts
given: a = 19 | 34 | 60; d = 1 | 3 | 6

## Tissue attenuates a radio wave's field by {a} Np/m.

::: part
answer: 8.686*a*d/100
unit: dB
tolerance: 0.5%

### How many decibels does a path of {d} cm cost?

::: misses
- [4.343*a*d/100] A neper of field attenuation is 8.686 dB.
- [any] $8.686\alpha d$.
:::
:::

::: part
answer: 100/a
unit: cm
tolerance: 0.5%

### At what depth has the field fallen to 1/e?

::: misses
- [any] $1/\alpha$.
:::
:::
:::

::: check
kind: parts
given: e = 40 | 57 | 70; s = 0.5 | 0.8 | 1.5; f = 100 | 403 | 2450

## Tissue has relative permittivity {e} and conductivity {s} S/m, at {f} MHz.

::: part
answer: s/(2*pi*f*1e6*e*8.854e-12)
tolerance: 0.5%

### What is its loss tangent, σ/ωε?

::: misses
- [any] σ/(2πf ε₀εᵣ).
:::
:::

::: part
answer: 2*pi*f*1e6*sqrt(4*pi*1e-7*e*8.854e-12/2)*sqrt(sqrt(1 + (s/(2*pi*f*1e6*e*8.854e-12))^2) - 1)
unit: Np/m
tolerance: 0.5%

### What is its field attenuation constant?

::: misses
- [any] $\alpha = \omega\sqrt{\mu\epsilon/2}\sqrt{\sqrt{1 + (\sigma/\omega\epsilon)^2} - 1}$.
:::
:::
:::

::: check
kind: parts
given: B = 25 | 100 | 300; r = 50 | 250

## An implant's link has a {B} kHz channel and needs {r} kb/s.

::: part
answer: r/B
unit: b/s/Hz
tolerance: 0.5%

### What spectral efficiency does it need?

::: misses
- [any] Bit rate over bandwidth.
:::
:::
:::

::: check
kind: choice

## Why do implant data links favor the 400 MHz band over 2.4 GHz?

- [ ] 2.4 GHz is reserved for hospitals
- [x] Tissue absorbs radio waves more strongly as frequency rises, so a lower frequency loses less on the way out of the body, while still allowing small antennas and adequate bandwidth
- [ ] Lower frequencies carry more data
- [ ] 400 MHz waves don't heat tissue at all

::: misses
- [1] 2.4 GHz is an unlicensed band used widely.
- [3] Higher bands generally offer more bandwidth.
- [4] They do, but less per unit of power at depth.
:::
:::

::: check
kind: derivation
vars: k = 0.001..0.3, a = 10..300, b = 10..300

## Two resonant coils with coupling k and quality factors $Q_1$ and $Q_2$ transfer power to an optimal load.

::: step
answer: k^2*a*b

### What figure of merit sets the link's efficiency?
:::

::: step
answer: k^2*a*b/(1 + sqrt(1 + k^2*a*b))^2

### What is the best efficiency?
:::

::: right
Right. For a weak link it's about $k^2Q_1Q_2/4$; for a strong one it approaches 1 − 2/(k√(Q₁Q₂)).
:::
:::

::: explain
## How can two coils that share only a few percent of their magnetic flux transfer most of the power between them?

::: model
In a resonant link, each coil is tuned with a capacitor to the operating frequency, so energy oscillates between its magnetic and electric fields many times, Q times per radian, before it is lost. The receiving coil captures only a small fraction k of the transmitting coil's flux on each cycle, but because neither coil loses much energy per cycle, the exchange repeats until the receiving side's load has taken most of it. What matters is how fast energy crosses compared with how fast it is lost, which is the product kQ: once it exceeds about one, the efficiency is high even though k itself is small.
:::
:::

::: practice
::: check
given: c = 2 | 8; f = 100 | 250; b = 12 | 14
answer: c*5*f*b/1e3
unit: kb/s
tolerance: 0.5%

## Bit rate: {c} channels to {f} Hz, {b} bits, 5× sampling?

::: misses
- [any] c × 5f × b.
:::
:::

::: check
given: k = 0.02 | 0.08; a = 80 | 150; b = 40 | 100
answer: k^2*a*b/(1 + sqrt(1 + k^2*a*b))^2
tolerance: 0.5%

## Link efficiency: k = {k}, Q = {a} and {b}?

::: misses
- [any] $k^2Q_1Q_2/(1 + \sqrt{1 + k^2Q_1Q_2})^2$.
:::
:::

::: check
given: a = 20 | 50; d = 2 | 5
answer: 8.686*a*d/100
unit: dB
tolerance: 0.5%

## Attenuation: {a} Np/m over {d} cm?

::: misses
- [any] $8.686\alpha d$.
:::
:::
:::