title: Clinical laboratory instruments
heading: Spectrophotometry with Beer's law and the absorbance that minimizes error, calibration against standards, automated analyzers, chromatography with internal standards, electrophoresis, and the hematology analyzer's impedance counting, red-cell indices and coincidence
eyebrow: S15 · Lesson 35 · Clinical laboratory instruments
xp: 50
---

::: hook
Most of the numbers in a patient's chart never touch the patient's body directly. A tube of blood goes to the laboratory, where analyzers run hundreds of tests an hour: glucose and cholesterol by the color a reagent turns, drug levels by how long a compound takes to leave a column, and blood counts by pushing cells one at a time through a hole 80 µm across and watching the resistance jump. Each is an electrical measurement of something chemical or cellular. How does a spectrophotometer turn transmitted light into a concentration, and why does it work best when the sample absorbs about two-thirds of it?
:::

::: play
- A spectrophotometer reads transmittance T and reports concentration through the absorbance $A = -\log_{10}T$, which is proportional to concentration.
- Every photometer has some fixed error in reading T. The curve is the resulting relative error in concentration, in percent, against the sample's absorbance, for the transmittance error (in percent of full scale) set by the slider.
- The error is smallest at A = 0.434 (T = 37%) and grows quickly at both ends. That's why labs dilute samples that read above A ≈ 1 and choose reactions that give A well above 0.1.
:::

```sim plot
view: graph
x: A = 0.05..2
y: e = 0..12
slider: dT = 0.1..1 = 0.5
plot: 0.4343*dT*10^A/A | samples: 400 | label: concentration error
marker: A = 0.434
marker: A = 0.11
marker: A = 1
readout: error at the optimum, % = 0.4343*dT*10^0.434343/0.434343 | digits: 3
readout: error at an absorbance of 1, % = 0.4343*dT*10 | digits: 3
```

::: predict
## A sample transmits 10% of the light. A second sample of the same substance, at twice the concentration in the same cuvette, transmits what fraction?

- [ ] 5%
- [x] 1%
- [ ] 20%
- [ ] 0.1%

::: reveal
Beer's law makes absorbance proportional to concentration: 10% transmittance is A = 1, so twice the concentration gives A = 2, which is 1% transmittance. Each layer of absorber removes the same fraction of what reaches it, so transmittance falls exponentially with concentration and path length.
:::
:::

::: nameit
## Beer's law

Light of power $P_0$ passing a path L through a solution of concentration C leaves with

$$P = P_0\,10^{-aLC}, \qquad A = \log_{10}\frac{P_0}{P} = aLC$$

where a is the absorptivity of the substance at that wavelength. Transmittance is reported as a percentage, so $A = 2 - \log_{10}(\%T)$. A spectrophotometer has a source, a wavelength selector (filter or monochromator) set to the absorption peak of the reaction's colored product, a cuvette, a photodetector and readout.

## Calibration

With a and L fixed, concentration follows from a standard of known concentration $C_s$: $C_u = C_s A_u/A_s$. That one-point method assumes Beer's law holds across the range. Solvent absorption, stray light and cuvette reflections bend the line, so labs run several standards spanning the range and read unknowns from the calibration curve.

## Where the photometer is most accurate

A fixed error ΔT in reading transmittance gives a concentration error

$$\frac{\Delta C}{C} = \frac{0.4343\,\Delta T}{T\,A}$$

which is smallest at $A = 1/\ln 10 = 0.434$. Webster's working range is about 0.11 to 1.0; outside it, errors grow quickly.

## Automated analyzers, chromatography and electrophoresis

Chemistry analyzers automate the same photometry: they pipette sample and reagent into cuvettes on a rotor or track, time the reaction and read absorbance, often at two wavelengths to cancel turbidity, at hundreds to thousands of tests an hour.

A **chromatograph** separates a mixture as a carrier gas or liquid sweeps it along a column, each compound held back by its affinity for the stationary phase. Compounds are identified by retention time and quantified by peak area. A known amount of an **internal standard** added to every sample corrects for losses: the unknown's concentration is the standard's times the ratio of peak areas, divided by a response factor from calibration. Two peaks are separated when their **resolution** $R_s = 2(t_2 - t_1)/(w_1 + w_2)$ reaches about 1.5.

**Electrophoresis** separates charged molecules, such as serum proteins, by moving them through a gel or paper in an electric field E. Each travels a distance $d = \mu E t$ set by its mobility μ. Heating from the current raises mobility, so the voltage or current is held constant.

## The hematology analyzer

Blood is diluted in a conducting solution and drawn through an aperture between two electrodes, about 50 to 100 µm wide. The aperture's resistance is $R = \rho L/A$, and each cell displaces conducting fluid, raising the resistance by about $\Delta R \approx \rho v/A^2$ for a cell of volume v (about 1.5 times that for a sphere, from how the current bends around it). With a constant current, each cell makes a voltage pulse whose count gives the cell count and whose height gives the cell's volume. Counts are corrected for **coincidence**, two cells in the aperture at once, which statistics predicts from the concentration and sensing volume. Red cells, white cells and platelets are counted at different dilutions; white cells are also classified by light scatter in a flow cell (flow cytometry).

From the red-cell count (millions per µL), hematocrit (percent) and hemoglobin (g/dL), the **red-cell indices** follow:

$$\text{MCV} = \frac{10\,\text{HCT}}{\text{RBC}}\ \text{fL}, \qquad \text{MCH} = \frac{10\,\text{Hb}}{\text{RBC}}\ \text{pg}, \qquad \text{MCHC} = \frac{100\,\text{Hb}}{\text{HCT}}\ \%$$

Normal ranges are about 82 to 98 fL, 27 to 31 pg and 32 to 36%.
:::

::: example
## A standard of 10 mg/dL reads 40% transmittance and a patient sample 55%. What's the patient's concentration, and is the measurement in a good range?

::: solution
1. **Absorbances.**
   $$\begin{aligned} A_s &= 2 - \log 40 \\ &= 0.398 \\ A_u &= 2 - \log 55 \\ &= 0.260 \end{aligned}$$
2. **Concentration.**
   $$\begin{aligned} C_u &= 10 \times \frac{A_u}{A_s} \\ &= 6.52\ \text{mg/dL} \end{aligned}$$
3. **Is the range good?** Both absorbances are inside the 0.11 to 1.0 range, near the 0.434 optimum, so a 0.5% transmittance error costs only about 1.4% (standard) to 1.5% (sample) in concentration.
4. **Linearity.** The one-point calculation is only as good as Beer's law over this range; a lab would confirm linearity with standards above and below.

**Answer:** the patient's concentration is 6.52 mg/dL, measured in a good range: both absorbances sit near the 0.434 optimum.
:::
:::

::: example
## A hematology aperture is 80 µm across and 100 µm long, filled with diluent of resistivity 1.5 Ω·m, and driven by 1 mA. How big is the pulse from a 90 fL red cell?

::: solution
1. **Aperture area.**
   $$\begin{aligned} A &= \pi(40\ \mu\text{m})^2 \\ &= 5.03 \times 10^{-9}\ \text{m}^2 \end{aligned}$$
2. **Aperture resistance.**
   $$\begin{aligned} R &= \frac{1.5 \times 100 \times 10^{-6}}{5.03 \times 10^{-9}}\ \Omega \\ &= 29.8\ \text{k}\Omega \end{aligned}$$
   so 1 mA puts 29.8 V across it.
3. **Change from one cell.** A 90 fL cell (90 µm³) raises the resistance by about
   $$\begin{aligned} \Delta R &= \frac{\rho v}{A^2} \\ &= \frac{1.5 \times 9 \times 10^{-17}}{(5.03 \times 10^{-9})^2}\ \Omega \\ &= 5.3\ \Omega \end{aligned}$$
4. **Pulse size.** That's a 5.3 mV pulse on 29.8 V, a change of 1.8 parts in 10,000: the fraction of the aperture's volume the cell fills.
5. **What the analyzer does with it.** It counts thousands of such pulses a second, and a histogram of their heights gives the volume distribution, from which the MCV and the spread of sizes (RDW) are computed.

**Answer:** a 90 fL red cell gives a pulse of about 5.3 mV on the 29.8 V across the aperture.
:::
:::

::: example
## Blood with 5 million red cells per microliter is diluted 1:50,000 and drawn through an 80 µm, 100 µm long aperture. How often is a second cell in the aperture when one is counted?

::: solution
1. **Diluted concentration.** Diluting 5 million cells per microliter 1:50,000 leaves 100 cells per µL.
2. **Aperture volume.**
   $$\begin{aligned} V &= \pi(40\ \mu\text{m})^2 \times 100\ \mu\text{m} \\ &= 5.03 \times 10^{-4}\ \mu\text{L} \end{aligned}$$
3. **Mean number inside.**
   $$\begin{aligned} \lambda &= 100 \times 5.03 \times 10^{-4} \\ &= 0.050\ \text{cells} \end{aligned}$$
4. **Chance of a second cell.** By Poisson statistics, the chance that another cell is present along with the one being counted is
   $$\begin{aligned} P &= 1 - e^{-\lambda} \\ &= 4.9\% \end{aligned}$$
5. **Consequence.** Such coincidences merge two pulses into one, undercounting and inflating apparent volume, so analyzers apply a coincidence correction that grows with concentration.

**Answer:** a second cell is in the aperture 4.9% of the time a cell is counted.
:::
:::

::: check
given: T = 8 | 25 | 50 | 70
answer: 2 - log(T)
tolerance: 0.5%

## A sample reads {T}% transmittance. What's its absorbance?

::: right
Right. $A = 2 - \log_{10}(\%T)$ = {answer}.
:::

::: misses
- [-log(T)] Write T as a fraction, or use $2 - \log(\%T)$.
- [2 - ln(T)] Absorbance uses the base-10 logarithm.
- [any] Not yet. $A = \log_{10}(100/\%T)$.
:::
:::

::: check
given: C = 5 | 8 | 20; s = 20 | 35; u = 15 | 45 | 60
answer: C*(2 - log(u))/(2 - log(s))
tolerance: 0.5%

## A standard of {C} mg/dL reads {s}% transmittance; the unknown reads {u}%. What's the unknown's concentration, in mg/dL?

::: right
Right. $C_u = C_s A_u/A_s$ = {answer} mg/dL.
:::

::: misses
- [C*u/s] Concentration is proportional to absorbance, not to transmittance.
- [C*s/u] Convert both readings to absorbance first.
- [any] Not yet. Convert both readings to absorbance, then scale the standard.
:::
:::

::: check
given: A = 0.2 | 0.434 | 0.8 | 1.5; d = 0.2 | 0.5 | 1
answer: 0.4343*d*10^A/A
unit: %
tolerance: 1%

## A photometer reads transmittance to within ±{d}% (of full scale). What's the relative concentration error for a sample with A = {A}?

::: right
Right. $0.4343\,\Delta T/(TA)$ = {answer}%.
:::

::: misses
- [d/A] Divide by the transmittance too: $T = 10^{-A}$.
- [any] Not yet. Differentiate $A = -\log_{10}T$ and divide by A.
:::
:::

::: check
given: H = 36 | 42 | 48; R = 3.8 | 4.6 | 5.4
answer: 10*H/R
unit: fL
tolerance: 0.5%

## A blood count gives hematocrit {H}% and red cells {R} million per µL. What's the MCV, in fL?

::: right
Right. MCV = 10 HCT/RBC = {answer}: the packed volume shared among the cells.
:::

::: misses
- [H/R] With HCT in percent and the count in millions per µL, the factor is 10.
- [any] Not yet. Divide the volume fraction by the number of cells per volume.
:::
:::

::: check
given: b = 11 | 13.5 | 16; H = 34 | 40 | 46
answer: 100*b/H
unit: %
tolerance: 0.5%

## Hemoglobin is {b} g/dL and hematocrit {H}%. What's the MCHC?

::: right
Right. MCHC = 100 Hb/HCT = {answer}%: the hemoglobin concentration inside the cells.
:::

::: misses
- [b/H] Multiply by 100 for a percentage.
- [any] Not yet. Divide hemoglobin by the packed-cell fraction.
:::
:::

::: check
given: r = 1.2 | 1.6; d = 50 | 80 | 100; L = 60 | 100
answer: r*L*1e-6/(pi*(d*1e-6/2)^2)/1000
unit: kΩ
tolerance: 0.5%

## An impedance-counting aperture is {d} µm in diameter and {L} µm long, filled with diluent of resistivity {r} Ω·m. What's its resistance?

::: right
Right. $R = \rho L/(\pi d^2/4)$ = {answer} kΩ.
:::

::: misses
- [r*L*1e-6/(pi*(d*1e-6)^2)/1000] Use the radius, d/2, for the area.
- [any] Not yet. $R = \rho L/A$ with the aperture's cross-section.
:::
:::

::: check
given: u = 2 | 5 | 10; E = 5 | 10 | 20; t = 20 | 40 | 60
answer: u*1e-5*E*t*60
unit: cm
tolerance: 0.5%

## A protein with mobility {u} × 10⁻⁵ cm²/(V·s) migrates in a field of {E} V/cm. How far does it move in {t} minutes?

::: right
Right. $d = \mu E t$ = {answer} cm.
:::

::: misses
- [u*1e-5*E*t] Convert minutes to seconds.
- [any] Not yet. Distance is mobility times field times time.
:::
:::

::: check
kind: parts
given: N = 4 | 5 | 6; D = 10000 | 50000; d = 50 | 100; L = 60 | 100

## Blood with {N} million cells per µL is diluted 1:{D} and drawn through a {d} µm diameter aperture {L} µm long.

::: part
answer: N*1e6/D*pi*(d/2*1e-4)^2*(L*1e-4)*1e3
tolerance: 0.5%

### How many cells does the aperture hold on average?

::: misses
- [N*1e6/D*pi*(d*1e-4)^2*(L*1e-4)*1e3] Use the radius.
- [any] Concentration times aperture volume (1 cm³ = 1000 µL).
:::
:::

::: part
answer: 1 - exp(-N*1e6/D*pi*(d/2*1e-4)^2*(L*1e-4)*1e3)
tolerance: 0.5%

### What is the chance another cell is present when one passes?

::: misses
- [any] $1 - e^{-\lambda}$.
:::
:::
:::

::: check
kind: derivation
vars: H = 20..60, R = 2..7, b = 5..20

## A blood count gives hematocrit H%, red cells R million per µL and hemoglobin b g/dL.

::: step
answer: 10*H/R

### What is the mean cell volume, in femtoliters?
:::

::: step
answer: 10*b/R

### What is the mean cell hemoglobin, in picograms?
:::

::: step
answer: 100*b/H

### What is the mean cell hemoglobin concentration, in g/dL?
:::

::: right
Right. The three indices are ratios of the three measured quantities, so an error in any one shows up in two of them.
:::
:::

::: explain
## Why does a spectrophotometer measure concentration least accurately when the sample absorbs almost nothing or almost everything?

::: model
Concentration is proportional to absorbance, the logarithm of the transmitted fraction, while the photometer's errors are roughly fixed amounts of transmittance. When the sample absorbs almost nothing, T is near 1 and the absorbance is a tiny number computed from the small difference between two nearly equal readings, so a fixed error in T is a large fraction of it. When the sample absorbs almost everything, T is tiny, and a fixed error in T is a large fraction of T itself, which the logarithm turns into a large error in absorbance. In between the two effects balance: the relative error $0.4343\,\Delta T/(TA)$ is smallest at A = 0.434, where 37% of the light gets through. Labs dilute strong samples and choose reactions and path lengths that put absorbances in the middle of the range.
:::
:::

::: practice

::: check
given: T = 2..95
answer: 2 - log(T)
tolerance: 0.5%

## Absorbance at {T}% transmittance?

::: misses
- [any] $2 - \log_{10}(\%T)$.
:::
:::

::: check
given: H = 30..55; R = 3..6.5
answer: 10*H/R
unit: fL
tolerance: 0.5%

## MCV in fL for HCT {H}% and RBC {R} million/µL?

::: misses
- [any] 10 HCT/RBC.
:::
:::

::: check
given: C = 1..50; s = 10..80; u = 10..80
answer: C*(2 - log(u))/(2 - log(s))
tolerance: 0.5%

## Standard {C} mg/dL at {s}%T, unknown at {u}%T. Concentration in mg/dL?

::: misses
- [any] $C_s A_u/A_s$.
:::
:::
:::