title: Direct blood-pressure measurement
heading: Measuring arterial pressure through a fluid-filled catheter, the catheter and transducer as a second-order system with natural frequency and damping set by their dimensions, why a small air bubble ruins the response, the fast-flush test, and hydrostatic offsets
eyebrow: S15 · Lesson 23 · Direct blood-pressure measurement
xp: 50
---

::: hook
In intensive care and in the catheter lab, arterial pressure is measured beat by beat through a thin plastic tube filled with saline, connected to a disposable silicon bridge transducer (lesson 3). The column of liquid has mass, the transducer's diaphragm and the tubing have compliance, and the narrow tube has viscous friction: a mass, a spring and a damper. Whether the trace on the monitor is the patient's real pressure waveform or a ringing caricature of it depends on the numbers lesson 2 introduced, and this lesson shows where those numbers come from.
:::

::: play
- A catheter of radius 0.5 mm and length 1 m, filled with saline, connected to a transducer whose compliance is 0.01 mm³ per 100 mmHg. A small air bubble of volume $V_b$ adds compliance $V_b/P_\text{atm}$.
- The amplitude ratio of the whole system against frequency. Without the bubble, $f_n$ is 163 Hz and the response is flat across the pressure waveform's harmonics.
- A 1 mm³ bubble drops $f_n$ to 43 Hz: the harmonics near 20 Hz get exaggerated, though the damping rises too.
:::

```sim plot
view: graph
x: f = 1..300 log
y: A = 0..4
slider: b = 0..5 step 0.1 = 0
slider: r = 0.3..1 step 0.05 = 0.5
plot: 1/sqrt((1 - (f/(r*1e-3/2*sqrt(1/((7.5006e-16 + b*1e-9/101325)*pi*1000))))^2)^2 + (2*(4*1e-3/(r*1e-3)^3*sqrt((7.5006e-16 + b*1e-9/101325)/(pi*1000)))*f/(r*1e-3/2*sqrt(1/((7.5006e-16 + b*1e-9/101325)*pi*1000))))^2) | samples: 600 | label: amplitude ratio
readout: natural frequency, Hz = r*1e-3/2*sqrt(1/((7.5006e-16 + b*1e-9/101325)*pi*1000)) | digits: 4
readout: damping ratio = 4*1e-3/(r*1e-3)^3*sqrt((7.5006e-16 + b*1e-9/101325)/(pi*1000)) | digits: 3
```

::: predict
## A 1 mm³ air bubble gets into a catheter system whose transducer has a compliance of 0.01 mm³ per 100 mmHg. What happens to its natural frequency?

- [ ] Nothing: the bubble is tiny
- [ ] It rises
- [x] It falls by about a factor of 4
- [ ] It falls to zero

::: reveal
Air at atmospheric pressure compresses by $V_b/P$ per pascal: 1 mm³/101 kPa = 9.9 × 10⁻¹⁵ m³/Pa, thirteen times the transducer's 7.5 × 10⁻¹⁶. Natural frequency goes as the square root of stiffness, so it falls by √14.2 ≈ 3.8, from 163 to 43 Hz.
:::
:::

::: nameit
## The catheter as a second-order system

Model the liquid in a catheter of radius r and length L as a slug of mass $\rho L\pi r^2$, the transducer's diaphragm as a spring of volume stiffness $E_d = \Delta P/\Delta V$, and Poiseuille friction in the tube as a damper. Then

$$f_n = \frac{r}{2}\sqrt{\frac{E_d}{\pi\rho L}}, \qquad \zeta = \frac{4\eta}{r^3}\sqrt{\frac{L}{\pi\rho E_d}}$$

with ρ and η the liquid's density and viscosity (about 1000 kg/m³ and 0.001 Pa·s for saline). Wider, shorter catheters and stiffer transducers raise $f_n$; damping falls steeply with radius.

## Bubbles

Compliances add, and a gas bubble is very compliant: isothermally, $\Delta V/\Delta P = V_b/P$. Even 1 mm³ dominates a modern transducer. A bubble lowers $f_n$ (as $\sqrt{E_d}$) and raises ζ (as $1/\sqrt{E_d}$); the frequency effect usually matters more, because the waveform's harmonics then fall near the resonance.

## The fast-flush test

Opening the flush valve briefly and releasing it applies a pressure step, and the system rings. If successive oscillation peaks shrink by a ratio d, then

$$\zeta = \frac{-\ln d}{\sqrt{4\pi^2 + (\ln d)^2}}, \qquad f_n = \frac{f_d}{\sqrt{1 - \zeta^2}}$$

where $f_d$ is the ringing frequency, one over the time between peaks.

## Units and leveling

1 mmHg = 133.32 Pa. The transducer reads pressure relative to its own height, so it's leveled with the heart: a height error h adds $\rho gh$, about 0.74 mmHg per centimeter of saline.
:::

::: example
## Find $f_n$ and ζ for the sim's catheter without a bubble, and with the 1 mm³ bubble.

::: solution
1. **Transducer compliance.**
   $$\begin{aligned} C_t &= \frac{0.01\ \text{mm}^3}{100 \times 133.32\ \text{Pa}} \\ &= 7.50 \times 10^{-16}\ \text{m}^3/\text{Pa} \end{aligned}$$
   so $E_d$ = 1.33 × 10¹⁵ Pa/m³.
2. **Natural frequency without a bubble.**
   $$\begin{aligned} f_n &= \frac{0.5\ \text{mm}}{2}\sqrt{\frac{1.33 \times 10^{15}}{\pi \times 1000 \times 1}} \\ &= 163\ \text{Hz} \end{aligned}$$
3. **Damping without a bubble.**
   $$\begin{aligned} \zeta &= \frac{4 \times 0.001}{(0.5\ \text{mm})^3} \\ &\quad \times \sqrt{\frac{1}{\pi \times 1000 \times 1.33 \times 10^{15}}} \\ &= 0.016 \end{aligned}$$
4. **With the bubble.** The compliance becomes 1.06 × 10⁻¹⁴ m³/Pa, which gives
   $$\begin{aligned} f_n &= 43\ \text{Hz} \\ \zeta &= 0.059 \end{aligned}$$

**Answer:** without the bubble, $f_n$ = 163 Hz and ζ = 0.016; with the 1 mm³ bubble, $f_n$ = 43 Hz and ζ = 0.059.
:::
:::

::: example
## A flush test shows successive peaks shrinking to 0.3 of the previous one, 50 ms apart. Find ζ and $f_n$. And what error does a transducer 10 cm below the heart give?

::: solution
1. **Log of the decay.**
   $$\ln 0.3 = -1.204$$
2. **Damping ratio.**
   $$\begin{aligned} \zeta &= \frac{1.204}{\sqrt{39.48 + 1.45}} \\ &= 0.188 \end{aligned}$$
3. **Natural frequency.** Peaks 50 ms apart ring at 20 Hz:
   $$\begin{aligned} f_n &= \frac{20\ \text{Hz}}{\sqrt{1 - 0.035}} \\ &= 20.4\ \text{Hz} \end{aligned}$$
   an underdamped, low-frequency system that will overstate systolic pressure.
4. **Level error.** A 10 cm level error adds a saline column of
   $$\begin{aligned} \Delta P &= \frac{1000 \times 9.81 \times 0.1}{133.32} \\ &= 7.4\ \text{mmHg} \end{aligned}$$
   (7.8 mmHg if the column were blood).

**Answer:** ζ = 0.188 and $f_n$ = 20.4 Hz, and a transducer 10 cm below the heart adds 7.4 mmHg to the reading.
:::
:::

::: example
## A transducer is placed 5 cm below a patient's right atrium. What error does that cause in a central venous pressure of 6 mmHg, and in an arterial pressure of 100 mmHg?

::: solution
1. **Saline column.**
   $$\begin{aligned} \rho gh &= 1000 \times 9.81 \times 0.05 \\ &= 490\ \text{Pa} \\ &= 3.7\ \text{mmHg} \end{aligned}$$
   added to every reading.
2. **Arterial pressure.** On 100 mmHg that's 3.7%.
3. **Central venous pressure.** On 6 mmHg it's 61%: the reading would say 9.7 mmHg and suggest fluid overload where there is none.
4. **The remedy.** That's why the transducer is leveled to the phlebostatic axis, at the fourth intercostal space halfway through the chest, and re-zeroed whenever the patient's bed moves.

**Answer:** the transducer adds 3.7 mmHg to every reading: 3.7% of an arterial pressure of 100 mmHg, but 61% of a central venous pressure of 6 mmHg, which would read 9.7 mmHg.
:::
:::

::: check
given: r = 0.3 | 0.5 | 0.8; L = 0.5 | 1 | 2; c = 0.005 | 0.01 | 0.05
answer: r*1e-3/2*sqrt(1/((c*1e-9/(100*133.322))*pi*1000*L))
unit: Hz
tolerance: 0.5%

## A saline-filled catheter of radius {r} mm and length {L} m feeds a transducer whose compliance is {c} mm³ per 100 mmHg. What's the natural frequency?

::: right
Right. $\frac{r}{2}\sqrt{E_d/(\pi\rho L)}$ = {answer} Hz.
:::

::: misses
- [any] Not yet. Convert the compliance to m³/Pa, invert it for $E_d$, then use $f_n = \frac{r}{2}\sqrt{E_d/\pi\rho L}$.
:::
:::

::: check
given: r = 0.3 | 0.5 | 0.8; L = 0.5 | 1 | 2; c = 0.005 | 0.01 | 0.05
answer: 4*1e-3/(r*1e-3)^3*sqrt(L*(c*1e-9/(100*133.322))/(pi*1000))
tolerance: 0.5%

## For that catheter (radius {r} mm, length {L} m, transducer compliance {c} mm³ per 100 mmHg, η = 0.001 Pa·s), what's the damping ratio?

::: right
Right. $\frac{4\eta}{r^3}\sqrt{L/(\pi\rho E_d)}$ = {answer}.
:::

::: misses
- [any] Not yet. $\zeta = (4\eta/r^3)\sqrt{L/(\pi\rho E_d)}$.
:::
:::

::: check
given: f = 80 | 150 | 250; t = 7.5 | 15; b = 0.2 | 1 | 3
answer: f*sqrt((t*1e-16)/(t*1e-16 + b*1e-9/101325))
unit: Hz
tolerance: 0.5%

## A catheter system has $f_n$ = {f} Hz with a transducer compliance of {t} × 10⁻¹⁶ m³/Pa. A {b} mm³ air bubble enters. What's the new natural frequency?

::: right
Right. $f_n$ scales as the square root of stiffness: {answer} Hz.
:::

::: misses
- [f*(t*1e-16)/(t*1e-16 + b*1e-9/101325)] Natural frequency goes as the square root of the stiffness ratio.
- [any] Not yet. Add the bubble's compliance $V_b/P_\text{atm}$.
:::
:::

::: check
given: d = 0.1 | 0.25 | 0.5
answer: -ln(d)/sqrt(4*pi^2 + ln(d)^2)
tolerance: 0.5%

## In a fast-flush test, each oscillation peak is {d} of the one before. What's the damping ratio?

::: right
Right. $-\ln d/\sqrt{4\pi^2 + (\ln d)^2}$ = {answer}.
:::

::: misses
- [-ln(d)/sqrt(pi^2 + ln(d)^2)] Successive peaks are a full period apart: use 4π².
- [any] Not yet. Logarithmic decrement over one period.
:::
:::

::: check
given: h = 5 | 10 | 20; p = 1000 | 1060
answer: p*9.81*h/100/133.322
unit: mmHg
tolerance: 0.5%

## A pressure transducer sits {h} cm below the heart, connected by a liquid column of density {p} kg/m³. How much does it overread?

::: right
Right. $\rho gh$ = {answer} mmHg.
:::

::: misses
- [p*9.81*h/100] That's in pascals; divide by 133.32.
- [any] Not yet. Hydrostatic pressure, then convert to mmHg.
:::
:::

::: check
given: P = 80 | 120 | 200
answer: P*133.322/1000
unit: kPa
tolerance: 0.5%

## What's {P} mmHg in kilopascals?

::: right
Right. {answer} kPa.
:::

::: misses
- [any] Not yet. 1 mmHg = 133.32 Pa.
:::
:::

::: check
kind: parts
given: H = 60 | 90 | 150; f = 15 | 25 | 40; z = 0.1 | 0.2 | 0.6

## A pressure wave with heart rate {H} per minute is recorded through a catheter system with $f_n$ = {f} Hz and ζ = {z}.

::: part
answer: 10*H/60
unit: Hz
tolerance: 0.5%

### What is the frequency of its tenth harmonic?

::: misses
- [any] Ten times the heart rate in hertz.
:::
:::

::: part
answer: 1/sqrt((1 - (10*H/60/f)^2)^2 + (2*z*10*H/60/f)^2)
tolerance: 0.5%

### By what factor does the system scale that harmonic's amplitude?

::: misses
- [any] $1/\sqrt{(1 - r^2)^2 + (2\zeta r)^2}$ with $r = f/f_n$.
:::
:::
:::

::: check
kind: derivation
vars: p = 990..1100, h = -0.3..0.3, P = 2..150

## A transducer sits a height h (m) below the reference level, connected by liquid of density ρ (kg/m³), while measuring a pressure P (mmHg).

::: step
answer: p*9.81*h

### What offset does the liquid column add, in pascals?
:::

::: step
answer: p*9.81*h/133.322/P*100

### What percent error is that?
:::

::: right
Right. The same offset that hardly matters on arterial pressure can swamp a venous one.
:::
:::

::: explain
## Why does a tiny air bubble in a pressure line distort the arterial waveform so much?

::: model
The catheter system resonates where the mass of the liquid column bounces on the elasticity of everything that can change volume: the transducer's diaphragm, the tubing walls and any gas. A modern transducer is extremely stiff, so even a cubic millimeter of air, which compresses easily, can be ten times more compliant than everything else combined. The added compliance lowers the natural frequency in proportion to the square root of the stiffness, moving the resonance down into the range of the pressure waveform's harmonics, where they're amplified and the trace overshoots and rings. That's why lines are carefully flushed of bubbles and checked with a flush test.
:::
:::

::: practice

::: check
given: r = 0.2 | 1.2; L = 0.2 | 3; c = 0.002 | 0.1
answer: r*1e-3/2*sqrt(1/((c*1e-9/(100*133.322))*pi*1000*L))
unit: Hz
tolerance: 0.5%

## $f_n$: r = {r} mm, L = {L} m, compliance {c} mm³/100 mmHg?

::: misses
- [any] $\frac{r}{2}\sqrt{E_d/\pi\rho L}$.
:::
:::

::: check
given: d = 0.02 | 0.1 | 0.2 | 0.3 | 0.5 | 0.8
answer: -ln(d)/sqrt(4*pi^2 + ln(d)^2)
tolerance: 0.5%

## ζ from a peak ratio of {d}?

::: misses
- [any] $-\ln d/\sqrt{4\pi^2 + \ln^2d}$.
:::
:::

::: check
given: h = 1 | 40; p = 1000 | 1060
answer: p*9.81*h/100/133.322
unit: mmHg
tolerance: 0.5%

## Level error: {h} cm, ρ = {p}?

::: misses
- [any] $\rho gh$ in mmHg.
:::
:::
:::