title: Electrical safety of medical equipment
heading: What current does to the body, macroshock and microshock thresholds, skin and body resistance, leakage current through stray capacitance, why the ground wire matters, ground-potential differences and catheters, isolated power systems and line isolation monitors, and patient isolation
eyebrow: S15 · Lesson 46 · Electrical safety of medical equipment
xp: 50
---

::: hook
In a hospital, patients are connected to electrical equipment through low-resistance paths that ordinary appliances never provide: gel electrodes that bypass dry skin, and catheters filled with conductive saline that lead straight to the heart. A current too small to feel through the hands can trigger fibrillation when it enters the heart directly. Electrical safety in medicine is therefore an engineering discipline with its own numbers: how much current is dangerous by which path, how much leaks through every power supply's stray capacitance, and what grounding, isolation and monitoring keep it below the limits.
:::

::: play
- The current through a person who touches 120 V, 60 Hz, against their body resistance, compared with rough thresholds for current entering through the skin: perception (about 0.5 mA), let-go (about 10 mA) and ventricular fibrillation (about 100 mA).
- Dry skin keeps an accidental touch near the perception level; wet or broken skin, or an electrode with gel, lets the same voltage drive a dangerous current.
- For current applied directly to the heart, the danger level drops by about a factor of a thousand.
:::

```sim plot
view: graph
x: R = 0.3..300 log
y: I = 0.01..1000 log
slider: V = 12..240 step 6 = 120
plot: V/R | samples: 300 | label: body current, mA (R in kΩ)
marker: y = 0.5 | label: perception
marker: y = 10 | label: let-go
marker: y = 100 | label: fibrillation (skin contact)
readout: current through 1 kΩ, mA = V/1 | digits: 4
```

::: predict
## A device's power supply has 2500 pF of stray capacitance between the 120 V, 60 Hz line and its metal case. How much current flows through someone touching the case if its ground wire is broken?

- [ ] None: capacitors block current
- [ ] About 0.1 µA
- [x] About 110 µA
- [ ] About 50 mA

::: reveal
At 60 Hz, 2500 pF has a reactance of 1/(2π × 60 × 2.5 nF) = 1.06 MΩ, so it passes 120 V/1.06 MΩ = 113 µA, more than the usual 100 µA limit for chassis leakage. Through intact skin that's below perception; through a catheter to the heart it could be lethal. With the ground wire intact, nearly all of it flows to ground instead.
:::
:::

::: nameit
## Physiological effects

For 60 Hz current entering and leaving through the skin of the limbs (**macroshock**), typical thresholds are: perception about 0.5 mA; let-go (the largest current at which a person can still release a grip) about 6 to 10 mA; ventricular fibrillation roughly 75 to 400 mA; burns and sustained contraction at amperes. Thresholds rise at frequencies above about 1 kHz, which is why electrosurgery can pass large currents at several hundred kilohertz. When current enters the heart directly through a catheter or pacing lead (**microshock**), fibrillation can occur at tens to hundreds of microamperes, so patient connections that reach the heart are limited to about **10 µA**.

## Resistance of the body

Dry, intact skin is the main barrier, from tens to hundreds of kilohms; wet or broken skin, or gel electrodes, drop it to around a kilohm. The body's interior is a few hundred ohms. Current is set by $I = V/R$ along whatever path exists.

## Leakage current and grounding

Every line-powered device leaks current to its case through stray and filter capacitance: $I = V2\pi fC$. The protective ground wire carries it away; the current divides between the ground path and a person touching the case in inverse proportion to their resistances, so an intact ground (a fraction of an ohm) takes nearly all of it. A broken ground sends it all through the person. Limits for chassis leakage are about 100 µA in normal conditions.

## Ground-potential differences

Two grounded devices can sit at slightly different potentials because ground currents flow through the wiring's resistance. A patient connected to both, through a saline catheter of a few hundred ohms, carries $\Delta V/R$: a few millivolts can exceed 10 µA. **Equipotential grounding**, bonding all equipment near the patient to one point, prevents this.

## Isolation

**Isolated power systems** in operating rooms supply equipment from an isolation transformer, so a single fault to ground doesn't complete a circuit; a **line isolation monitor** continuously measures the current that would flow through a fault, the hazard current $V2\pi fC$ of the system's capacitance to ground, and alarms at 5 mA. **Isolated patient inputs** (lesson 15's front ends behind isolation barriers) keep patient leads floating so they can't become a return path.
:::

::: example
## What current does a 120 V touch drive through dry skin (100 kΩ) and through a wet or gelled path (1 kΩ)?

::: solution
1. **Through dry skin.**
   $$\begin{aligned} I &= \frac{120\ \text{V}}{100\ \text{k}\Omega} \\ &= 1.2\ \text{mA} \end{aligned}$$
   just above perception.
2. **Through a wet or gelled path.**
   $$\begin{aligned} I &= \frac{120\ \text{V}}{1\ \text{k}\Omega} \\ &= 120\ \text{mA} \end{aligned}$$
   in the fibrillation range for a hand-to-hand path.
3. **What decides the outcome.** The voltage is the same; the path decides the outcome.

**Answer:** the touch drives 1.2 mA through dry skin and 120 mA through the wet path.
:::
:::

::: example
## Two outlets near a patient differ by 20 mV. A saline catheter and the patient's body form a 500 Ω path between equipment on each. What flows, and what potential difference would keep it under 10 µA? What hazard current does a line isolation monitor report for 30 nF of system capacitance at 120 V, 60 Hz?

::: solution
1. **Current through the catheter path.**
   $$\begin{aligned} I &= \frac{20\ \text{mV}}{500\ \Omega} \\ &= 40\ \mu\text{A} \end{aligned}$$
   four times the microshock limit.
2. **Allowed difference.** Staying under 10 µA requires less than
   $$\begin{aligned} V &= 10\ \mu\text{A} \times 500\ \Omega \\ &= 5\ \text{mV} \end{aligned}$$
   between grounds.
3. **Line isolation monitor.** It reports
   $$\begin{aligned} I &= 120 \times 2\pi \times 60 \times 30\ \text{nF} \\ &= 1.36\ \text{mA} \end{aligned}$$
   under its 5 mA alarm limit.

**Answer:** 40 µA flows, the grounds must differ by less than 5 mV to keep it under 10 µA, and the monitor reports 1.36 mA.
:::
:::

::: example
## A hot wire touches a grounded metal case. The fault path has 0.2 Ω of wiring and 0.1 Ω in the ground conductor. What fault current flows, what voltage does the case reach before the breaker opens, and what flows through a person (1 kΩ) touching it?

::: solution
1. **Fault current.** On a 120 V circuit:
   $$\begin{aligned} I &= \frac{120\ \text{V}}{0.3\ \Omega} \\ &= 400\ \text{A} \end{aligned}$$
   far above a 20 A breaker's rating, so it trips in a few milliseconds.
2. **Case voltage.** Until it does, the case sits above ground at
   $$\begin{aligned} V &= 400 \times 0.1 \\ &= 40\ \text{V} \end{aligned}$$
3. **Current through a person.** 40 V drives 40 mA through 1 kΩ: dangerous, but for milliseconds.
4. **Without the ground conductor.** The case would sit at 120 V indefinitely and the breaker would never trip; the ground wire's job is to turn a shock hazard into a short circuit.

**Answer:** 400 A flows, the case reaches 40 V until the breaker opens, and 40 mA flows through the person for those few milliseconds.
:::
:::

::: check
given: V = 24 | 120 | 230; R = 1 | 10 | 100
answer: V/R
unit: mA
tolerance: 0.5%

## A person touches {V} V through a total path resistance of {R} kΩ. What current flows?

::: right
Right. V/R = {answer} mA.
:::

::: misses
- [any] Not yet. Ohm's law.
:::
:::

::: check
given: V = 120 | 230; f = 50 | 60; C = 500 | 2500 | 10000
answer: V*2*pi*f*C*1e-12*1e6
unit: µA
tolerance: 0.5%

## A device has {C} pF between its {V} V, {f} Hz supply and its case. What leakage current flows through that capacitance?

::: right
Right. $V2\pi fC$ = {answer} µA.
:::

::: misses
- [V*f*C*1e-12*1e6] Use ω = 2πf.
- [any] Not yet. Voltage over the capacitive reactance.
:::
:::

::: check
given: d = 5 | 20 | 50; R = 300 | 500 | 1000
answer: d*1e-3/R*1e6
unit: µA
tolerance: 0.5%

## Two pieces of equipment connected to a patient differ in ground potential by {d} mV, and the path between them through a catheter is {R} Ω. What current flows?

::: right
Right. ΔV/R = {answer} µA.
:::

::: misses
- [any] Not yet. Ohm's law along the catheter path.
:::
:::

::: check
given: I = 100 | 300 | 500; g = 0.1 | 0.5 | 2; p = 1 | 10
answer: I*g/(g + p*1000)
unit: µA
tolerance: 0.5%

## {I} µA of leakage reaches a case connected to ground through {g} Ω. A person touching it offers a {p} kΩ path to ground. How much flows through the person?

::: right
Right. $I R_g/(R_g + R_p)$ = {answer} µA.
:::

::: misses
- [I] That's with the ground wire broken.
- [any] Not yet. Current divides in inverse proportion to resistance.
:::
:::

::: check
given: V = 120 | 230; f = 50 | 60; C = 10 | 30 | 80
answer: V*2*pi*f*C*1e-9*1e3
unit: mA
tolerance: 0.5%

## An isolated power system at {V} V, {f} Hz has {C} nF of capacitance to ground. What hazard current does its line isolation monitor report?

::: right
Right. $V2\pi fC$ = {answer} mA.
:::

::: misses
- [any] Not yet. The current a fault to ground would draw through the system's capacitance.
:::
:::

::: check
given: L = 10 | 20; R = 300 | 500 | 1000
answer: L*1e-6*R*1000
unit: mV
tolerance: 0.5%

## A catheter path has {R} Ω, and the microshock limit is {L} µA. What's the largest ground-potential difference allowed between the two connected devices?

::: right
Right. $I_\text{limit}R$ = {answer} mV.
:::

::: misses
- [any] Not yet. Ohm's law, solved for voltage.
:::
:::

::: check
kind: parts
given: V = 120 | 230; w = 0.1 | 0.3; g = 0.05 | 0.2 | 0.5; R = 1 | 5

## A hot-to-case fault occurs on a {V} V circuit with {w} Ω of wiring and {g} Ω in the ground conductor; a person of {R} kΩ touches the case.

::: part
answer: V/(w + g)
unit: A
tolerance: 0.5%

### What fault current flows?

::: misses
- [any] The supply voltage over the loop's resistance.
:::
:::

::: part
answer: V/(w + g)*g/(R*1e3)*1e3
unit: mA
tolerance: 0.5%

### What current flows through the person until the breaker trips?

::: misses
- [V/(R*1e3)*1e3] The ground conductor holds the case far below the full supply voltage.
- [any] The case voltage, fault current times ground resistance, over the person's resistance.
:::
:::
:::

::: check
kind: derivation
vars: V = 100..250, f = 50..60, C = 10..5000, g = 0.01..1, R = 500..100000

## A device has C pF between its V-volt, f-hertz supply and its case, which is grounded through g ohms; a person of R ohms touches the case.

::: step
answer: 2*pi*f*C*1e-12*V*1e6

### What leakage current flows, in microamperes?
:::

::: step
answer: 2*pi*f*C*1e-12*V*1e6*g/(g + R)

### How much of it passes through the person?
:::

::: right
Right. An intact ground diverts almost all leakage; a broken one sends it all through whoever touches the case.
:::
:::

::: explain
## Why is a current that's harmless through the hands dangerous through a catheter?

::: model
When current enters and leaves through the skin of the limbs, it spreads through the whole torso, and only a small fraction passes through the heart, so a large total current is needed before the current density in heart muscle reaches the level that can trigger fibrillation. A saline-filled catheter or pacing lead delivers current directly to a small area of the heart, so all of it is concentrated there, and the current density can reach that level with a thousand times less total current. That's why the limits for connections that reach the heart are measured in microamperes, while the ordinary shock limits are in milliamperes.
:::
:::

::: practice

::: check
given: V = 5 | 400; R = 0.5 | 500
answer: V/R
unit: mA
tolerance: 0.5%

## Body current: {V} V, {R} kΩ?

::: misses
- [any] V/R.
:::
:::

::: check
given: V = 100 | 250; f = 50 | 60; C = 100 | 20000
answer: V*2*pi*f*C*1e-12*1e6
unit: µA
tolerance: 0.5%

## Leakage: {V} V, {f} Hz, {C} pF?

::: misses
- [any] $V2\pi fC$.
:::
:::

::: check
given: d = 1 | 100; R = 100 | 2000
answer: d*1e-3/R*1e6
unit: µA
tolerance: 0.5%

## Microshock: {d} mV over {R} Ω?

::: misses
- [any] ΔV/R.
:::
:::
:::