title: Final exam
heading: "Final exam: biomedical instrumentation"
eyebrow: S15 · Final exam
type: problems
---

::: hook
Twelve problems from the whole course: sensors and their dynamics, temperature measurement, membrane potentials and conduction, the ECG, electrodes and biopotential amplifiers, interference, signal processing, blood pressure and flow, respiration, oximetry and chemical sensors, the clinical laboratory, image quality and medical imaging, therapeutic devices, electrical safety, diagnostic accuracy and device power budgets. Give yourself three hours and work without the lessons open. "Another set" deals a different exam.
:::

::: pool
pick: 12

::: check
kind: parts
given: s = 5 | 25; V = 2.5 | 10; P = 60 | 240; S = 15 | 30; a = 0.002 | 0.008; X = 150 | 300

## A transducer of {s} µV/V/mmHg at {V} V; another with output $Sx + ax^2$ µV (S = {S}, a = {a}) over 0 to {X} mmHg.

::: part
answer: s*V*P/1000
unit: mV
tolerance: 0.5%

### First transducer's output at {P} mmHg?

::: misses
- [any] $sVP$.
:::
:::

::: part
answer: 100*a*X^2/4/(S*X + a*X^2)
tolerance: 0.5%

### Second one's nonlinearity in percent of full scale?

::: misses
- [any] $aX^2/4$ over full scale.
:::
:::
:::

::: check
kind: parts
given: T = 2 | 4; u = 5 | 15; s = 10 | 25; n = 20 | 60; z = 0.1 | 0.3; f = 8 | 16

## A probe with τ = {u} s meets a {T} °C step; a catheter has $f_n$ = {n} Hz and ζ = {z}.

::: part
answer: T*exp(-s/u)
unit: °C
tolerance: 0.5%

### Probe error after {s} s?

::: misses
- [any] $\Delta Te^{-t/\tau}$.
:::
:::

::: part
answer: 1/sqrt((1 - (f/n)^2)^2 + (2*z*f/n)^2)
tolerance: 0.5%

### Catheter amplitude ratio at {f} Hz?

::: misses
- [any] Second-order magnitude.
:::
:::

::: part
answer: 100*exp(-pi*z/sqrt(1 - z^2))
tolerance: 0.5%

### Catheter step overshoot, percent?

::: misses
- [any] $100e^{-\pi\zeta/\sqrt{1 - \zeta^2}}$.
:::
:::
:::

::: check
kind: parts
given: G = 2 | 2.1; e = 250 | 1000; V = 3 | 10; k = 1 | 4; c = 100 | 400; F = 0.5 | 2; f = 100 | 1000

## Gauges (G = {G}) at ±{e} microstrain in a bridge at {V} V with {k} active arms; a piezoelectric sensor (d = {c} pC/N) into a charge amplifier with $C_f$ = {f} pF.

::: part
answer: V*G*e*1e-6*k/4*1000
unit: mV
tolerance: 0.5%

### Bridge output?

::: misses
- [any] $V(k/4)G\varepsilon$.
:::
:::

::: part
answer: c*F/f
unit: V
tolerance: 0.5%

### Charge amplifier output for {F} N?

::: misses
- [any] $dF/C_f$.
:::
:::
:::

::: check
kind: parts
given: R = 3 | 30; B = 3300 | 4300; T = 33 | 41; I = 50 | 150; d = 1 | 2

## A thermistor of {R} kΩ at 25 °C, B = {B} K, at {T} °C carrying {I} µA with δ = {d} mW/°C.

::: part
answer: R*exp(B*(1/(T + 273.15) - 1/298.15))
unit: kΩ
tolerance: 0.5%

### Resistance?

::: misses
- [any] B equation.
:::
:::

::: part
answer: -100*B/(T + 273.15)^2
unit: %/°C
tolerance: 0.5%

### Coefficient in %/°C?

::: misses
- [any] $-B/T^2$.
:::
:::

::: part
answer: (I*1e-6)^2*R*exp(B*(1/(T + 273.15) - 1/298.15))*1e3/(d*1e-3)
unit: °C
tolerance: 0.5%

### Self-heating?

::: misses
- [any] $I^2R/\delta$.
:::
:::
:::

::: check
kind: parts
given: i = 130 | 150; o = 4 | 8; r = 0.03 | 0.1; d = 2 | 10; M = 8000 | 20000; R = 70 | 150

## K⁺ at {i} mM inside and {o} mM outside, Na⁺ at 12 and 145 mM, $P_{Na}/P_K$ = {r}, 37 °C; an axon {d} µm across with $R_m$ = {M} Ω·cm² and $R_i$ = {R} Ω·cm.

::: part
answer: 26.727*ln(o/i)
unit: mV
tolerance: 0.5%

### Potassium Nernst potential?

::: misses
- [any] $(RT/F)\ln([K]_o/[K]_i)$.
:::
:::

::: part
answer: 26.727*ln((o + 145*r)/(i + 12*r))
unit: mV
tolerance: 0.5%

### GHK potential?

::: misses
- [any] Permeability-weighted.
:::
:::

::: part
answer: 10*sqrt(M*d*1e-4/(4*R))
unit: mm
tolerance: 0.5%

### Length constant?

::: misses
- [any] $\sqrt{R_md/4R_i}$.
:::
:::
:::

::: check
kind: parts
given: a = 0.3 | 0.8; b = 0.7 | 1.5; D = 15 | 30; q = 0.35 | 0.42

## Lead I = {a} mV, lead II = {b} mV; R waves {D} mm apart at 25 mm/s; QT = {q} s.

::: part
answer: b - a
unit: mV
tolerance: 0.5%

### Lead III?

::: misses
- [any] II − I.
:::
:::

::: part
answer: atan((2*b - a)/(sqrt(3)*a))*180/pi
unit: °
tolerance: 0.5%

### Axis?

::: misses
- [any] $\tan^{-1}[(2\,\text{II} - \text{I})/(\sqrt{3}\,\text{I})]$.
:::
:::

::: part
answer: q/sqrt(D/25)
unit: s
tolerance: 0.5%

### Bazett QTc?

::: misses
- [any] $QT/\sqrt{RR}$.
:::
:::
:::

::: check
kind: parts
given: s = 0.5 | 2; R = 50 | 200; C = 0.05 | 0.3; f = 10 | 40; F = 25 | 50; r = 3 | 8

## An electrode with $R_s$ = {s} kΩ, $R_d$ = {R} kΩ, $C_d$ = {C} µF; an instrumentation amplifier with $R_1$ = {r} kΩ and $R_2$ = {F} kΩ.

::: part
answer: sqrt((s + R/(1 + (2*pi*f*R*1e3*C*1e-6)^2))^2 + (R*2*pi*f*R*1e3*C*1e-6/(1 + (2*pi*f*R*1e3*C*1e-6)^2))^2)
unit: kΩ
tolerance: 0.5%

### Electrode impedance at {f} Hz?

::: misses
- [any] $|R_s + R_d \| C_d|$.
:::
:::

::: part
answer: 1 + 2*F/r
tolerance: 0.5%

### Amplifier first-stage gain?

::: misses
- [any] $1 + 2R_2/R_1$.
:::
:::
:::

::: check
kind: parts
given: v = 5 | 20; d = 10 | 40; Z = 100 | 500; o = 2 | 6; i = 0.1 | 0.3; F = 4 | 8; a = 25 | 50

## {v} mV of common mode, {d} kΩ electrode mismatch, {Z} MΩ common-mode input impedance; a driven right leg with $R_o$ = {o} MΩ, $R_f$ = {F} MΩ, $R_a$ = {a} kΩ and {i} µA of displacement current.

::: part
answer: v*1e-3*d*1e3/(Z*1e6)*1e6
unit: µV
tolerance: 0.5%

### Mismatch error?

::: misses
- [any] $v_{cm}\Delta Z/Z_{cm}$.
:::
:::

::: part
answer: o*1e6*i*1e-6/(1 + 2*F*1e6/(a*1e3))*1e3
unit: mV
tolerance: 0.5%

### Driven-right-leg common-mode voltage?

::: misses
- [any] $R_oi_d/(1 + 2R_f/R_a)$.
:::
:::
:::

::: check
kind: parts
given: s = 0.5 | 2; n = 10 | 25; t = 3 | 6; c = 0.05 | 0.5; f = 1 | 2; Q = 20 | 40

## An evoked response of {s} µV in {n} µV of background; a high-pass at {c} Hz; a 60 Hz notch with Q = {Q}.

::: part
answer: (t*n/s)^2
tolerance: 0.5%

### Sweeps for an SNR of {t}?

::: misses
- [any] $(\text{target}\times n/s)^2$.
:::
:::

::: part
answer: atan(c/f)*180/pi
unit: °
tolerance: 0.5%

### High-pass phase lead at {f} Hz?

::: misses
- [any] $\tan^{-1}(f_c/f)$.
:::
:::

::: part
answer: 60/Q
unit: Hz
tolerance: 0.5%

### Notch width?

::: misses
- [any] $f_0/Q$.
:::
:::
:::

::: check
kind: parts
given: r = 0.4 | 0.6; L = 0.8 | 1.5; c = 0.01 | 0.04; S = 110 | 150; D = 70 | 90; h = 5 | 12

## A saline catheter (radius {r} mm, {L} m) on a transducer of {c} mm³ per 100 mmHg; a pressure of {S}/{D} mmHg; a transducer {h} cm below the heart.

::: part
answer: r*1e-3/2*sqrt(1/((c*1e-9/(100*133.322))*pi*1000*L))
unit: Hz
tolerance: 0.5%

### Natural frequency?

::: misses
- [any] $\frac{r}{2}\sqrt{E_d/\pi\rho L}$.
:::
:::

::: part
answer: D + (S - D)/3
unit: mmHg
tolerance: 0.5%

### Mean arterial pressure?

::: misses
- [any] DBP + (SBP − DBP)/3.
:::
:::

::: part
answer: 1000*9.81*h/100/133.322
unit: mmHg
tolerance: 0.5%

### Leveling error?

::: misses
- [any] $\rho gh$.
:::
:::
:::

::: check
kind: parts
given: v = 220 | 300; d = 40 | 60; V = 5 | 10; T = 0 | 20; A = 3 | 6; f = 2 | 5; u = 0.3 | 1; t = 30 | 60

## Fick: {v} mL/min uptake and {d} mL/L difference; thermodilution: {V} mL at {T} °C, area {A} °C·s, K = 1.08; Doppler: {f} MHz at {t}° on {u} m/s.

::: part
answer: v/d
unit: L/min
tolerance: 0.5%

### Fick output?

::: misses
- [any] Uptake over difference.
:::
:::

::: part
answer: V*1e-3*(37 - T)*1.08/A*60
unit: L/min
tolerance: 0.5%

### Thermodilution output?

::: misses
- [any] $V_i(T_b - T_i)K/\text{area}$.
:::
:::

::: part
answer: 2*f*1e6*u*cos(t*pi/180)/1540
unit: Hz
tolerance: 0.5%

### Doppler shift?

::: misses
- [any] $2f_0v\cos\theta/c$.
:::
:::
:::

::: check
kind: parts
given: V = 3 | 6; a = 9 | 12; b = 5 | 7; u = 0.4 | 1; R = 0.5 | 1; H = 10 | 15; s = 0.9 | 0.98

## Helium dilution from {a}% to {b}% in a {V} L spirometer; a forced expiration with τ = {u} s; an oximeter reading R = {R}; blood with {H} g/dL Hb at saturation {s}, $P_aO_2$ 90 mmHg.

::: part
answer: V*(a - b)/b
unit: L
tolerance: 0.5%

### FRC?

::: misses
- [any] $V_1(C_1 - C_2)/C_2$.
:::
:::

::: part
answer: 1 - exp(-1/u)
tolerance: 0.5%

### FEV1/FVC?

::: misses
- [any] $1 - e^{-1/\tau}$.
:::
:::

::: part
answer: 110 - 25*R
tolerance: 0.5%

### Calibrated SpO₂ in percent?

::: misses
- [any] 110 − 25R.
:::
:::

::: part
answer: 1.34*H*s + 0.003*90
unit: mL/dL
tolerance: 0.5%

### Oxygen content?

::: misses
- [any] $1.34[\text{Hb}]S + 0.003P$.
:::
:::
:::

::: check
kind: parts
given: T = 25 | 37; p = 0.2 | 0.4; b = 18 | 28; c = 35 | 50; m = 80 | 150; K = 15 | 30; G = 5 | 12

## A pH electrode at {T} °C; blood with {b} mM bicarbonate at $P_{CO_2}$ = {c} mmHg; a glucose sensor with $i_\max$ = {m} nA and $K_m$ = {K} mM at {G} mM.

::: part
answer: 8.314462618*(T + 273.15)/96485.33212*ln(10)*p*1000
unit: mV
tolerance: 0.5%

### Voltage change for {p} pH units?

::: misses
- [any] $2.303(RT/F)\Delta\text{pH}$.
:::
:::

::: part
answer: 6.1 + log(b/(0.0307*c))
tolerance: 0.1%

### Blood pH?

::: misses
- [any] Henderson–Hasselbalch.
:::
:::

::: part
answer: m*G/(K + G)
unit: nA
tolerance: 0.5%

### Glucose current?

::: misses
- [any] Michaelis–Menten.
:::
:::
:::

::: check
kind: parts
given: m = 0.16 | 0.22; x = 10 | 25; w = 0.19 | 0.2; M = 0.24 | 0.4; a = 1.38 | 1.63; b = 1.9 | 7.8; f = 3 | 7; d = 5 | 12

## Tissue with μ = {m} cm⁻¹, {x} cm thick; a voxel of μ = {M} against water's {w}; an ultrasound boundary from {a} to {b} MRayl; a {f} MHz echo from {d} cm.

::: part
answer: exp(-m*x)
tolerance: 0.5%

### X-ray transmission?

::: misses
- [any] $e^{-\mu x}$.
:::
:::

::: part
answer: 1000*(M - w)/w
tolerance: 0.5%

### CT number?

::: misses
- [any] $1000(\mu - \mu_w)/\mu_w$.
:::
:::

::: part
answer: ((b - a)/(b + a))^2
tolerance: 0.5%

### Reflected intensity fraction?

::: misses
- [any] $((Z_2 - Z_1)/(Z_2 + Z_1))^2$.
:::
:::

::: part
answer: 2*0.5*f*d
unit: dB
tolerance: 0.5%

### Round-trip loss at 0.5 dB/(cm·MHz)?

::: misses
- [any] $2\alpha fd$.
:::
:::
:::

::: check
kind: parts
given: B = 1.5 | 3; G = 10 | 25; W = 40 | 100; A = 200 | 400; t = 60 | 150; R = 500 | 3000; E = 15 | 90

## A {B} T scanner with a {G} mT/m readout gradient and {W} kHz bandwidth; {A} MBq of ¹⁸F; a spin echo (TR {R} ms, TE {E} ms) for T1 = 1000 ms, T2 = 100 ms.

::: part
answer: 42.577*B
unit: MHz
tolerance: 0.5%

### Larmor frequency?

::: misses
- [any] $\bar{\gamma}B_0$.
:::
:::

::: part
answer: W*1e3/(42.577e6*G*1e-3)*100
unit: cm
tolerance: 0.5%

### Field of view?

::: misses
- [any] $\text{BW}/(\bar{\gamma}G)$.
:::
:::

::: part
answer: A*2^(-t/109.8)
unit: MBq
tolerance: 0.5%

### Activity after {t} min?

::: misses
- [any] $A_02^{-t/T_{1/2}}$.
:::
:::

::: part
answer: (1 - exp(-R/1000))*exp(-E/100)
tolerance: 0.5%

### Relative signal?

::: misses
- [any] Spin-echo equation.
:::
:::
:::

::: check
kind: parts
given: I = 0.5 | 1.5; c = 0.3 | 0.6; t = 0.2 | 0.5; C = 30 | 100; V = 2 | 5; R = 50 | 80; T = 4 | 8

## Tissue with rheobase {I} mA and chronaxie {c} ms; a defibrillator: {C} µF at {V} kV into {R} Ω, truncated after {T} ms.

::: part
answer: I*(1 + c/t)
unit: mA
tolerance: 0.5%

### Threshold for {t} ms pulses?

::: misses
- [any] $I_{rh}(1 + c/t)$.
:::
:::

::: part
answer: 0.5*C*1e-6*(V*1000)^2
unit: J
tolerance: 0.5%

### Stored energy?

::: misses
- [any] $\frac{1}{2}CV^2$.
:::
:::

::: part
answer: 1 - exp(-2*T*1e-3/(R*C*1e-6))
tolerance: 0.5%

### Fraction delivered?

::: misses
- [any] $1 - e^{-2t/RC}$.
:::
:::
:::

::: check
kind: parts
given: V = 120 | 230; C = 1000 | 3000; d = 10 | 30; r = 400 | 800; a = 3 | 10; s = 3 | 10; D = 0.005 | 0.02; K = 200 | 500

## A {V} V, 60 Hz device with {C} pF to its case; {d} mV between grounds across a {r} Ω catheter path; a wearable ({a} mA active, {s} µA asleep, duty {D}) on {K} mAh.

::: part
answer: V*2*pi*60*C*1e-12*1e6
unit: µA
tolerance: 0.5%

### Leakage current?

::: misses
- [any] $V2\pi fC$.
:::
:::

::: part
answer: d*1e-3/r*1e6
unit: µA
tolerance: 0.5%

### Catheter current?

::: misses
- [any] ΔV/R.
:::
:::

::: part
answer: K/((a*1e-3*D + s*1e-6*(1 - D))*1e3)/24
tolerance: 0.5%

### Wearable life in days?

::: misses
- [any] Capacity over average current.
:::
:::
:::

::: check
kind: parts
given: V = 0.45 | 0.55; Q = 0.5 | 0.75; R = 6 | 12; C = 0.04 | 0.06; E = 5 | 10

## A volume-controlled breath delivers {V} L at {Q} L/s with PEEP {E} cmH₂O to lungs with R = {R} cmH₂O·s/L and C = {C} L/cmH₂O.

::: part
answer: E + R*Q + V/C
tolerance: 0.5%

### Peak airway pressure, in cmH₂O?

::: misses
- [any] PEEP + RQ + V/C.
:::
:::

::: part
answer: E + V/C
tolerance: 0.5%

### Plateau pressure, in cmH₂O?

::: misses
- [any] PEEP + V/C, with no flow.
:::
:::

::: part
answer: 3*R*C
unit: s
tolerance: 0.5%

### Time for 95% of passive exhalation?

::: misses
- [any] 3RC.
:::
:::
:::

::: check
kind: parts
given: s = 0.85 | 0.95; c = 0.9 | 0.98; p = 0.02 | 0.15

## A rhythm test with sensitivity {s} and specificity {c} is used where the prevalence is {p}.

::: part
answer: 100*s*p/(s*p + (1 - c)*(1 - p))
unit: %
tolerance: 0.5%

### Positive predictive value?

::: misses
- [any] Bayes' rule.
:::
:::

::: part
answer: 100*c*(1 - p)/(c*(1 - p) + (1 - s)*p)
unit: %
tolerance: 0.1%

### Negative predictive value?

::: misses
- [any] True negatives over all negatives.
:::
:::
:::

::: check
kind: parts
given: r = 1.2 | 1.6; d = 60 | 90; L = 80 | 120; v = 80 | 95

## A hematology aperture is {d} µm in diameter and {L} µm long, filled with diluent of resistivity {r} Ω·m and driven by 1 mA. Red cells of {v} fL pass through it.

::: part
answer: r*L*1e-6/(pi*(d*1e-6/2)^2)/1000
unit: kΩ
tolerance: 0.5%

### Aperture resistance?

::: misses
- [any] $\rho L/A$.
:::
:::

::: part
answer: r*v*1e-18/(pi*(d*1e-6/2)^2)^2
unit: Ω
tolerance: 0.5%

### Resistance change for one cell, from $\rho v/A^2$?

::: misses
- [any] $\rho v/A^2$ with v in cubic meters.
:::
:::

::: part
answer: r*v*1e-18/(pi*(d*1e-6/2)^2)^2
unit: mV
tolerance: 0.5%

### Pulse height at 1 mA?

::: misses
- [any] 1 mA times the resistance change.
:::
:::
:::
:::