title: Electromagnetic and ultrasonic flowmeters
heading: The electromagnetic flowmeter and Faraday's law in flowing blood, volume flow from mean velocity, transit-time ultrasonic meters and their nanosecond differences, the Doppler shift from moving blood cells, and the depth–velocity limit of pulsed Doppler
eyebrow: S15 · Lesson 27 · Electromagnetic and ultrasonic flowmeters
xp: 50
---

::: hook
Indicator dilution gives an average over many heartbeats. To see flow beat by beat in a single vessel, the measurement has to respond to the moving blood itself. Blood conducts, so moving it through a magnetic field generates a voltage, B5's motional EMF in a liquid. Blood also carries red cells that scatter ultrasound, so sound sent into a vessel comes back Doppler-shifted by their motion, and that shift falls conveniently in the audio range. Doppler ultrasound is how clinicians hear flow in a fetal heart or find a narrowed carotid artery.
:::

::: play
- The Doppler shift $f_d = 2f_0v\cos\theta/c$ from blood moving at speed v, insonated at frequency $f_0$ at angle θ to the flow, with c = 1540 m/s in tissue.
- The shift falls to zero as the beam turns perpendicular to the vessel, and small angle errors near 90° cause large velocity errors, so scanners work at 60° or less.
- Pulsed Doppler samples the shift at the pulse repetition frequency, so shifts above half of it alias; deeper vessels force slower pulsing.
:::

```sim plot
view: graph
x: t = 0..90
y: f = 0..10
slider: v = 0.1..2 step 0.05 = 0.5
slider: F = 2..10 step 0.5 = 5
plot: 2*F*1e6*v*cos(t*pi/180)/1540/1000 | samples: 300 | label: Doppler shift, kHz
readout: shift at 60°, kHz = 2*F*1e6*v*0.5/1540/1000 | digits: 4
readout: velocity error from a 5° angle error at 60°, % = 100*(cos(55*pi/180)/cos(60*pi/180) - 1) | digits: 3
```

::: predict
## 5 MHz ultrasound meets blood moving at 0.5 m/s, at 60° to the flow. What's the Doppler shift?

- [ ] 5 MHz
- [ ] About 16 kHz
- [x] About 1.6 kHz
- [ ] About 0.3 Hz

::: reveal
$f_d = 2 \times 5\,\text{MHz} \times 0.5 \times \cos 60°/1540$ = 1.62 kHz. The factor of 2 comes from the double trip: the moving cells receive a shifted frequency and re-emit it shifted again. A shift of a few kilohertz is easy to hear, which is why Doppler instruments have loudspeakers.
:::
:::

::: nameit
## Electromagnetic flowmeters

A conducting liquid moving at velocity v across a magnetic field B between electrodes a distance d apart generates

$$e = Bdv$$

For axially symmetric flow in a round vessel the voltage depends on the mean velocity, whatever the profile, so volume flow follows as $Q = \bar{v}\pi d^2/4$. The signal is microvolts, so the field is driven by AC or switched square waves and the electrodes are read synchronously, which rejects electrode offsets and drift. Perivascular probes require the vessel to be exposed, so the method is used in surgery and research.

## Transit-time flowmeters

Ultrasound sent along a path of length L at angle θ to the flow travels faster downstream than upstream. The difference in travel times is

$$\Delta t = \frac{2Lv\cos\theta}{c^2}$$

a few nanoseconds, measured by timing or phase comparison. Clamped around a vessel, transit-time probes measure volume flow in bypass grafts during surgery.

## Doppler flowmeters

Cells moving at v scatter ultrasound of frequency $f_0$ back with a shift

$$f_d = \frac{2f_0v\cos\theta}{c}$$

so $v = f_dc/(2f_0\cos\theta)$. **Continuous-wave** Doppler uses separate transmit and receive crystals and sees all depths at once. **Pulsed** Doppler gates the echo to a chosen depth d, which requires waiting $2d/c$ between pulses: the pulse rate can't exceed $c/2d$, and shifts above half the pulse rate alias, so

$$v_\max = \frac{c^2}{8f_0d\cos\theta}$$

Deeper or faster flow needs a lower $f_0$.
:::

::: example
## A perivascular probe applies 0.03 T across a 10 mm artery with a mean velocity of 0.3 m/s. What voltage and flow result? And what's the transit-time difference along a 10 mm path at 45° with v = 0.5 m/s?

::: solution
1. **Induced voltage.**
   $$\begin{aligned} e &= 0.03 \times 0.01 \times 0.3\ \text{V} \\ &= 90\ \mu\text{V} \end{aligned}$$
2. **Volume flow.**
   $$\begin{aligned} Q &= 0.3 \times \frac{\pi \times (0.01)^2}{4} \\ &= 2.36 \times 10^{-5}\ \text{m}^3/\text{s} \\ &= 1.41\ \text{L/min} \end{aligned}$$
3. **Transit-time difference.**
   $$\begin{aligned} \Delta t &= \frac{2 \times 0.01 \times 0.5 \times 0.707}{1540^2}\ \text{s} \\ &= 3.0\ \text{ns} \end{aligned}$$

**Answer:** the probe gives 90 µV for a flow of 1.41 L/min, and the transit-time difference is 3.0 ns.
:::
:::

::: example
## What's the fastest flow a 5 MHz pulsed Doppler can measure at 8 cm depth and 60°?

::: solution
1. **Highest pulse rate.**
   $$\begin{aligned} \text{PRF} &= \frac{1540}{2 \times 0.08}\ \text{Hz} \\ &= 9.6\ \text{kHz} \end{aligned}$$
   so shifts up to 4.8 kHz can be measured.
2. **Fastest flow.**
   $$\begin{aligned} v_\max &= \frac{1540^2}{8 \times 5\ \text{MHz} \times 0.08 \times 0.5} \\ &= 1.48\ \text{m/s} \end{aligned}$$
3. **What it means.** A stenotic jet at 3 m/s would alias; dropping to 2.5 MHz doubles the limit.

**Answer:** the fastest flow it can measure is 1.48 m/s.
:::
:::

::: example
## An 8 MHz continuous-wave Doppler probe at 45° watches a 6 mm carotid artery whose peak velocity is 1 m/s, with a parabolic profile. What range of Doppler shifts does it hear, and what's the volume flow?

::: solution
1. **Largest shift.** The fastest blood, at the center, shifts by
   $$\begin{aligned} f_d &= \frac{2f_0v\cos\theta}{c} \\ &= \frac{2 \times 8 \times 10^6 \times 1 \times 0.707}{1540}\ \text{Hz} \\ &= 7.35\ \text{kHz} \end{aligned}$$
2. **Spread of shifts.** A parabolic profile has every velocity from 0 to the peak in equal amounts of cross-section, so the spectrum is flat from 0 to 7.35 kHz, with its mean at half the maximum, 3.67 kHz.
3. **Volume flow.** The mean velocity, 0.5 m/s, times the area:
   $$\begin{aligned} Q &= 0.5\ \text{m/s} \times \pi \times 3^2\ \text{mm}^2 \\ &= 14.1\ \text{mL/s} \\ &= 850\ \text{mL/min} \end{aligned}$$
4. **Why it's audible.** The shifts fall in the audible range, which is why Doppler units have loudspeakers.

**Answer:** it hears shifts spread evenly from 0 to 7.35 kHz, with a mean of 3.67 kHz, and the volume flow is 14.1 mL/s, 850 mL/min.
:::
:::

::: check
given: B = 0.01 | 0.03 | 0.05; d = 3 | 8 | 15; v = 0.1 | 0.4 | 1
answer: B*d*1e-3*v*1e6
unit: µV
tolerance: 0.5%

## An electromagnetic flowmeter applies {B} T across a {d} mm vessel where blood moves at {v} m/s. What voltage appears?

::: right
Right. $Bdv$ = {answer} µV.
:::

::: misses
- [any] Not yet. Motional EMF.
:::
:::

::: check
given: d = 3 | 8 | 15; v = 0.1 | 0.4 | 1
answer: v*pi*(d*1e-3)^2/4*60000
unit: L/min
tolerance: 0.5%

## Blood moves at a mean {v} m/s in a vessel {d} mm across. What's the volume flow?

::: right
Right. $\bar{v}\pi d^2/4$ = {answer} L/min.
:::

::: misses
- [v*pi*(d*1e-3)^2*60000] Use the radius, or d²/4.
- [any] Not yet. Mean velocity times cross-section.
:::
:::

::: check
given: f = 2 | 5 | 10; v = 0.2 | 0.6 | 1.5; t = 30 | 45 | 60
answer: 2*f*1e6*v*cos(t*pi/180)/1540
unit: Hz
tolerance: 0.5%

## {f} MHz ultrasound meets blood at {v} m/s, at {t}° to the flow. What's the Doppler shift?

::: right
Right. $2f_0v\cos\theta/c$ = {answer} Hz.
:::

::: misses
- [f*1e6*v*cos(t*pi/180)/1540] The echo is shifted twice: include the factor of 2.
- [any] Not yet. The Doppler equation with c = 1540 m/s.
:::
:::

::: check
given: s = 500 | 2000 | 6000; f = 2 | 5 | 8; t = 30 | 60
answer: s*1540/(2*f*1e6*cos(t*pi/180))
unit: m/s
tolerance: 0.5%

## A {f} MHz Doppler at {t}° measures a shift of {s} Hz. What's the blood velocity?

::: right
Right. $f_dc/(2f_0\cos\theta)$ = {answer} m/s.
:::

::: misses
- [s*1540/(2*f*1e6)] Divide by cos θ to get the speed along the vessel.
- [any] Not yet. Invert the Doppler equation.
:::
:::

::: check
given: L = 5 | 10 | 20; v = 0.2 | 0.5 | 1; t = 30 | 45
answer: 2*L*1e-3*v*cos(t*pi/180)/1540^2*1e9
unit: ns
tolerance: 0.5%

## A transit-time probe's path is {L} mm long at {t}° to blood moving at {v} m/s. What's the upstream–downstream time difference?

::: right
Right. $2Lv\cos\theta/c^2$ = {answer} ns.
:::

::: misses
- [any] Not yet. Difference of L/(c − v cos θ) and L/(c + v cos θ).
:::
:::

::: check
given: f = 2 | 3.5 | 5; d = 4 | 8 | 15; t = 0 | 60
answer: 1540^2/(8*f*1e6*d*1e-2*cos(t*pi/180))
unit: m/s
tolerance: 0.5%

## A {f} MHz pulsed Doppler looks at a vessel {d} cm deep, at {t}°. What's the fastest velocity it can measure without aliasing?

::: right
Right. $c^2/(8f_0d\cos\theta)$ = {answer} m/s.
:::

::: misses
- [1540^2/(4*f*1e6*d*1e-2*cos(t*pi/180))] The pulse rate is c/2d, and only half of it is usable: the denominator has an 8.
- [any] Not yet. Combine the depth limit on the pulse rate with Nyquist.
:::
:::

::: check
kind: parts
given: f = 2 | 5 | 8; v = 0.4 | 1.2; t = 30 | 60; d = 3 | 8

## A {f} MHz Doppler at {t}° watches a {d} mm vessel with a parabolic profile peaking at {v} m/s.

::: part
answer: 2*f*1e6*v*cos(t*pi/180)/1540
unit: Hz
tolerance: 0.5%

### What is the highest Doppler shift it measures?

::: misses
- [f*1e6*v*cos(t*pi/180)/1540] The shift is doubled, out and back.
- [any] $2f_0v\cos\theta/c$.
:::
:::

::: part
answer: v/2*pi*(d/2)^2*60
unit: mL/min
tolerance: 0.5%

### What is the volume flow?

::: misses
- [v*pi*(d/2)^2*60] The mean velocity of a parabolic profile is half the peak.
- [any] Mean velocity times area.
:::
:::
:::

::: check
kind: derivation
vars: B = 0.01..0.1, d = 0.001..0.03, v = 0.01..2

## An electromagnetic flowmeter applies B tesla across a vessel of diameter d (m) with mean velocity v (m/s).

::: step
answer: B*d*v

### What voltage appears across the vessel?
:::

::: step
answer: pi*d^2/4*v

### What is the volume flow?
:::

::: step
answer: 4*B/(pi*d)

### What voltage does each unit of volume flow give?
:::

::: right
Right. For a given field the sensitivity to flow rises as the vessel narrows.
:::
:::

::: explain
## Why do pulsed Doppler instruments face a trade-off between depth and maximum velocity?

::: model
A pulsed Doppler instrument measures the shift by comparing successive echoes from the same depth, so it samples the Doppler signal once per transmitted pulse. To know which depth an echo came from, it must wait for the echo from the chosen depth to return before sending the next pulse, which takes 2d/c, so deeper vessels force a lower pulse rate. By the sampling theorem, a Doppler shift above half the pulse rate aliases and appears as a lower or reversed velocity. A deep vessel therefore limits the highest shift that can be measured, and the only ways out are a lower transmit frequency, which lowers the shift for a given velocity, a smaller cosine of the angle, or continuous-wave Doppler, which gives up depth information entirely.
:::
:::

::: practice

::: check
given: f = 1 | 15; v = 0.05 | 4; t = 0 | 70
answer: 2*f*1e6*v*cos(t*pi/180)/1540
unit: Hz
tolerance: 0.5%

## Doppler shift: {f} MHz, {v} m/s, {t}°?

::: misses
- [any] $2f_0v\cos\theta/c$.
:::
:::

::: check
given: B = 0.005 | 0.1; d = 1 | 25; v = 0.05 | 2
answer: B*d*1e-3*v*1e6
unit: µV
tolerance: 0.5%

## EM flowmeter: {B} T, {d} mm, {v} m/s?

::: misses
- [any] $Bdv$.
:::
:::

::: check
given: f = 1 | 10; d = 1 | 20; t = 0 | 60
answer: 1540^2/(8*f*1e6*d*1e-2*cos(t*pi/180))
unit: m/s
tolerance: 0.5%

## Aliasing limit: {f} MHz, {d} cm, {t}°?

::: misses
- [any] $c^2/(8f_0d\cos\theta)$.
:::
:::
:::