title: Cardiac output by indicator dilution
heading: The Fick principle with oxygen, the Stewart–Hamilton method for a dye bolus, correcting the curve for recirculation, thermodilution with a cold injectate and its correction constant, and stroke volume and cardiac index
eyebrow: S15 · Lesson 26 · Cardiac output by indicator dilution
xp: 50
---

::: hook
How much blood does the heart pump each minute? No flowmeter can be clamped around a beating heart in a patient, but conservation of mass gives a way around that. Add a known amount of something to the blood, measure how diluted it becomes downstream, and the flow follows. Adolf Fick did it in 1870 with the oxygen the lungs add; Stewart and Hamilton did it with a dye; today's pulmonary artery catheters do it with a few milliliters of cold saline and a thermistor (lesson 7) near the catheter's tip.
:::

::: play
- A dye bolus of 5 mg is injected upstream, and its concentration downstream follows a smooth curve whose area is m/F: the faster the flow, the more diluted the dye, and the smaller the area.
- Here the curve is modeled as $c = Kt^2e^{-t/\tau}$, with K chosen so the area matches m/F.
- Real curves show a second bump as dye recirculates; the first pass is extrapolated by fitting an exponential to its downslope.
:::

```sim plot
view: graph
x: t = 0..30
y: c = 0..15
slider: F = 2..10 step 0.1 = 5
slider: u = 1..4 step 0.1 = 2
plot: 5/(2*F/60*u^3)*t^2*exp(-t/u) | samples: 400 | label: dye concentration, mg/L
readout: area under the curve, mg·s/L = 5/(F/60) | digits: 4
readout: peak concentration, mg/L = 5/(2*F/60*u^3)*(2*u)^2*exp(-2) | digits: 4
```

::: predict
## A patient consumes 250 mL of oxygen per minute. Arterial blood carries 200 mL of oxygen per liter and mixed venous blood 150 mL/L. What's the cardiac output?

- [ ] 1.25 L/min
- [x] 5 L/min
- [ ] 50 L/min
- [ ] 0.2 L/min

::: reveal
Each liter of blood passing through the lungs picks up 200 − 150 = 50 mL of oxygen, and 250 mL is taken up each minute, so 250/50 = 5 L of blood must pass per minute. That's the Fick principle: uptake equals flow times the concentration difference.
:::
:::

::: nameit
## The Fick principle

If a substance is added to (or removed from) the blood at a known rate $\dot{m}$ and its concentrations upstream and downstream are known, conservation of mass gives

$$F = \frac{\dot{m}}{C_\text{out} - C_\text{in}}$$

For the heart and lungs: cardiac output = oxygen uptake/(arterial − mixed venous oxygen content).

## Indicator dilution

Inject an amount m as a bolus and record the downstream concentration c(t). All of it passes the sampling point, carried by flow F, so $m = F\int c\,dt$:

$$F = \frac{m}{\int_0^\infty c\,dt}$$

This is the **Stewart–Hamilton** method. Recirculation adds a second rise before the first pass has ended, so the downslope is fitted with an exponential, $c_1e^{-(t - t_1)/\tau}$, whose remaining area is $c_1\tau$.

## Thermodilution

A cold injectate of volume $V_i$ at temperature $T_i$, injected into blood at $T_b$, cools the blood downstream; a thermistor records the temperature drop $\Delta T_b(t)$. A heat balance gives

$$\text{CO} = \frac{V_i(T_b - T_i)K}{\int\Delta T_b\,dt}$$

where K corrects for the different densities and specific heats of injectate and blood (about 1.08 for dextrose solution) and for catheter losses. Cold saline is harmless and dissipates, so measurements can be repeated freely.

## Derived quantities

Stroke volume = cardiac output/heart rate. **Cardiac index** = cardiac output/body surface area, normally about 2.5 to 4 L/min/m².
:::

::: example
## 5 mg of dye gives a first-pass curve with an area of 60 mg·s/L. What's the cardiac output? And what if the record was cut at a point where the concentration was 0.8 mg/L, falling with a 4 s time constant, after 40 mg·s/L had accumulated?

::: solution
1. **Full curve.**
   $$\begin{aligned} F &= \frac{5\ \text{mg}}{60\ \text{mg·s/L}} \\ &= 0.0833\ \text{L/s} \\ &= 5.0\ \text{L/min} \end{aligned}$$
2. **Tail of the cut record.** The tail adds
   $$0.8 \times 4 = 3.2\ \text{mg·s/L}$$
   so the area is 43.2 mg·s/L.
3. **Flow with the tail.**
   $$\begin{aligned} F &= \frac{5}{43.2}\ \text{L/s} \\ &= 6.9\ \text{L/min} \end{aligned}$$
4. **Ignoring the tail.** Using only the 40 mg·s/L would overstate the flow by 8%.

**Answer:** the cardiac output is 5.0 L/min; for the cut record, adding the tail gives 6.9 L/min, and ignoring it would overstate the flow by 8%.
:::
:::

::: example
## 10 mL of injectate at 0 °C enters blood at 37 °C, and the temperature curve's area is 4.4 °C·s. With K = 1.08, find the cardiac output, and the stroke volume and cardiac index at 70 beats per minute and 1.9 m².

::: solution
1. **Cardiac output.**
   $$\begin{aligned} \text{CO} &= \frac{0.010 \times 37 \times 1.08}{4.4}\ \text{L/s} \\ &= 0.0908\ \text{L/s} \\ &= 5.45\ \text{L/min} \end{aligned}$$
2. **Stroke volume.** At 70 beats per minute:
   $$\begin{aligned} \text{SV} &= \frac{5450}{70}\ \text{mL} \\ &= 78\ \text{mL} \end{aligned}$$
3. **Cardiac index.**
   $$\begin{aligned} \text{CI} &= \frac{5.45}{1.9} \\ &= 2.87\ \text{L/min/m}^2 \end{aligned}$$

**Answer:** the cardiac output is 5.45 L/min, the stroke volume 78 mL, and the cardiac index 2.87 L/min/m².
:::
:::

::: example
## A dye curve gives a cardiac output of 5 L/min and a mean transit time of 12 s between the injection site and the sampling site. How much blood lies between them?

::: solution
1. **Volume from flow and transit time.** Indicator dilution measures volumes as well as flows: the volume between injection and sampling is flow times mean transit time.
   $$\begin{aligned} V &= Q\bar t \\ &= \frac{5}{60}\ \text{L/s} \times 12\ \text{s} \\ &= 1.0\ \text{L} \end{aligned}$$
2. **Mean transit time.** It is the first moment of the corrected curve:
   $$\bar t = \frac{\int tc\,dt}{\int c\,dt}$$
3. **What the volume is.** Between a right-heart injection and an arterial sample this is the central blood volume, which rises in heart failure as blood pools in the lungs.

**Answer:** 1.0 L of blood lies between the injection and sampling sites.
:::
:::

::: check
given: v = 200 | 250 | 350; a = 180 | 200; d = 30 | 50 | 70
answer: v/d
unit: L/min
tolerance: 0.5%

## A patient takes up {v} mL of oxygen per minute. Arterial blood carries {a} mL/L and mixed venous blood {d} mL/L less. What's the cardiac output?

::: right
Right. Uptake over the content difference: {answer} L/min.
:::

::: misses
- [v/a] Divide by the arterial–venous difference, not the arterial content.
- [any] Not yet. The Fick principle.
:::
:::

::: check
given: m = 2 | 5 | 10; A = 30 | 60 | 120
answer: m/A*60
unit: L/min
tolerance: 0.5%

## A {m} mg dye bolus gives a first-pass concentration curve whose area is {A} mg·s/L. What's the flow?

::: right
Right. $m/\int c\,dt$ = {answer} L/min.
:::

::: misses
- [m/A] That's in liters per second.
- [any] Not yet. Stewart–Hamilton.
:::
:::

::: check
given: A = 30 | 45; c = 0.5 | 1.2; u = 3 | 6; m = 2 | 5
answer: m/(A + c*u)*60
unit: L/min
tolerance: 0.5%

## A {m} mg bolus's curve accumulates {A} mg·s/L before recirculation, at which point the concentration is {c} mg/L and falling exponentially with a {u} s time constant. What's the flow?

::: right
Right. Add the tail $c_1\tau$, then $m/\text{area}$: {answer} L/min.
:::

::: misses
- [m/A*60] Include the exponential tail's area.
- [any] Not yet. Total area = measured part + $c_1\tau$.
:::
:::

::: check
given: V = 5 | 10; T = 0 | 20; A = 2 | 4 | 8
answer: V*1e-3*(37 - T)*1.08/A*60
unit: L/min
tolerance: 0.5%

## {V} mL of injectate at {T} °C enters blood at 37 °C; the temperature curve's area is {A} °C·s and K = 1.08. What's the cardiac output?

::: right
Right. $V_i(T_b - T_i)K/\int\Delta T_b\,dt$ = {answer} L/min.
:::

::: misses
- [V*1e-3*(37 - T)/A*60] Include the correction constant K.
- [any] Not yet. Heat balance.
:::
:::

::: check
given: C = 4 | 5.5 | 7; H = 60 | 80 | 110
answer: C*1000/H
unit: mL
tolerance: 0.5%

## A cardiac output of {C} L/min at {H} beats per minute means what stroke volume?

::: right
Right. CO/HR = {answer} mL.
:::

::: misses
- [any] Not yet. Output per beat.
:::
:::

::: check
given: C = 4 | 5.5 | 7; B = 1.6 | 1.9 | 2.2
answer: C/B
unit: L/min/m²
tolerance: 0.5%

## What's the cardiac index, in L/min/m², for an output of {C} L/min and a body surface area of {B} m²?

::: right
Right. {answer} L/min/m².
:::

::: misses
- [any] Not yet. Output over body surface area.
:::
:::

::: check
kind: parts
given: m = 2.5 | 5 | 10; A = 30 | 60 | 120; t = 8 | 15

## A {m} mg dye bolus gives a curve with an area of {A} mg·s/L and a mean transit time of {t} s.

::: part
answer: m/A*60
unit: L/min
tolerance: 0.5%

### What is the cardiac output?

::: misses
- [m/A] That's in liters per second.
- [any] $Q = m/\int c\,dt$.
:::
:::

::: part
answer: m/A*t
unit: L
tolerance: 0.5%

### What volume lies between injection and sampling?

::: misses
- [m/A*60*t] Use the flow in liters per second.
- [any] Flow times mean transit time.
:::
:::
:::

::: check
kind: derivation
vars: m = 1..20, A = 10..200, H = 40..180

## A dye bolus of m mg gives a first-pass curve of area A mg·s/L, at a heart rate of H beats per minute.

::: step
answer: m/A*60

### What is the cardiac output, in L/min?
:::

::: step
answer: m/A*60/H*1000

### What is the stroke volume, in milliliters?
:::

::: right
Right. The Stewart–Hamilton principle: the indicator is conserved, so its total divided by the curve's area is the flow.
:::
:::

::: explain
## Why must the recirculation part of a dye-dilution curve be removed, and why does thermodilution suffer less from it?

::: model
The Stewart–Hamilton formula counts every bit of indicator passing the sampling point exactly once. Blood carrying dye travels around the circulation and returns within a quarter of a minute or so, so the downstream concentration rises again before the first pass has fully cleared. Counting that returning dye would add area and understate the flow, so the first pass's downslope is extrapolated as an exponential and the recirculation is discarded. With thermodilution the cold injectate warms toward body temperature as it spreads through tissues and mixes with the large volume of blood, so little of the temperature deficit survives a full circuit, and the indicator vanishes by itself, which also allows repeated measurements.
:::
:::

::: practice

::: check
given: v = 150 | 500; d = 25 | 80
answer: v/d
unit: L/min
tolerance: 0.5%

## Fick: {v} mL/min uptake, {d} mL/L difference?

::: misses
- [any] Uptake over difference.
:::
:::

::: check
given: m = 1 | 20; A = 10 | 300
answer: m/A*60
unit: L/min
tolerance: 0.5%

## Stewart–Hamilton: {m} mg, area {A} mg·s/L?

::: misses
- [any] $m/\int c\,dt$.
:::
:::

::: check
given: V = 3 | 10; T = 0 | 25; A = 1 | 10
answer: V*1e-3*(37 - T)*1.08/A*60
unit: L/min
tolerance: 0.5%

## Thermodilution: {V} mL at {T} °C, area {A} °C·s, K = 1.08?

::: misses
- [any] $V_i(T_b - T_i)K/\text{area}$.
:::
:::
:::