title: Indirect blood pressure and heart sounds
heading: Cuff measurements by auscultation and Korotkoff sounds, the oscillometric method with its peak at mean arterial pressure and fixed-ratio estimates of systolic and diastolic pressure, mean pressure from systolic and diastolic, deflation rate and cuff size errors, and the phonocardiogram
eyebrow: S15 · Lesson 24 · Indirect blood pressure and heart sounds
xp: 50
---

::: hook
Most blood pressures are measured with a cuff, without breaking the skin. A clinician listening with a stethoscope hears tapping sounds start as the cuff pressure falls below systolic and fade near diastolic. An automatic monitor hears nothing at all: it watches the small pressure oscillations that each heartbeat makes in the cuff, finds where they're largest, and estimates systolic and diastolic pressures from ratios calibrated against direct measurements. Both methods are simple to use and subtle in their errors.
:::

::: play
- The envelope of cuff oscillation amplitude against cuff pressure, modeled as a bell centered at the mean arterial pressure with width w.
- An oscillometric monitor finds the peak (MAP), then reports systolic where the amplitude has fallen to 55% of the peak on the high-pressure side, and diastolic where it's 85% on the low side.
- The estimates depend on the envelope's shape, which varies between people: the fixed ratios are empirical.
:::

```sim plot
view: graph
x: P = 40..180
y: A = 0..1.05
slider: M = 70..120 step 1 = 93
slider: w = 15..50 step 1 = 30
plot: exp(-((P - M)/w)^2) | samples: 400 | label: oscillation amplitude (relative)
marker: x = M + w*sqrt(-ln(0.55)) | label: systolic estimate
marker: x = M - w*sqrt(-ln(0.85)) | label: diastolic estimate
readout: systolic estimate, mmHg = M + w*sqrt(-ln(0.55)) | digits: 4
readout: diastolic estimate, mmHg = M - w*sqrt(-ln(0.85)) | digits: 4
```

::: predict
## A cuff deflates at 3 mmHg per second and the heart rate is 75 per minute. How precisely can the auscultatory method pin down systolic pressure?

- [ ] To within 0.1 mmHg
- [x] To within about 2.4 mmHg
- [ ] To within 3 mmHg per minute
- [ ] To within 75 mmHg

::: reveal
A Korotkoff sound can only occur once per beat, every 0.8 s, and in that time the cuff falls 3 × 0.8 = 2.4 mmHg. The first sound marks systolic pressure only to within one beat's worth of deflation, which is why guidelines call for 2 to 3 mmHg per beat.
:::
:::

::: nameit
## Auscultation

The cuff is inflated above systolic pressure, closing the brachial artery, then deflated steadily. When cuff pressure falls just below systolic, blood spurts through during each peak, and the turbulence makes the tapping **Korotkoff sounds** heard below the cuff. They muffle and disappear as cuff pressure reaches diastolic, when the artery stays open throughout the cycle. Each reading is uncertain by the pressure drop between beats.

## Oscillometry

The arterial pulse makes the cuff pressure oscillate slightly. The oscillation amplitude is largest when the cuff pressure equals the **mean arterial pressure**, where the arterial wall is unloaded and most compliant. Monitors filter the oscillations out of the cuff-pressure record (a band-pass of roughly 0.5 to 10 Hz), find the peak, and estimate systolic and diastolic pressures where the envelope falls to fixed fractions of the peak, typically around 0.5 to 0.6 on the high side and 0.7 to 0.85 on the low side, chosen empirically.

## Mean arterial pressure

The arterial waveform spends longer near diastolic than near systolic, so the mean is closer to diastolic:

$$\text{MAP} \approx \text{DBP} + \frac{\text{SBP} - \text{DBP}}{3}$$

The difference SBP − DBP is the **pulse pressure**.

## Sources of error

A cuff that's too narrow for the arm reads high, because cuff pressure isn't fully transmitted to the artery; the inflatable bladder should be about 40% of the arm's circumference wide. Deflating too fast, a cuff above heart level, movement and irregular rhythms all add error.

## Heart sounds

The closing of the heart's valves makes the first and second heart sounds, S1 and S2, mostly between about 20 and 200 Hz; murmurs from turbulent flow reach higher. A **phonocardiogram** records them with a chest microphone, an accelerometer or a piezoelectric sensor (lesson 4), filtered into bands to separate sounds from murmurs.
:::

::: example
## A pressure is 120/80 mmHg. What's its mean? And for an oscillometric envelope centered at 93 mmHg with w = 30 mmHg, what do the 0.55 and 0.85 ratios report?

::: solution
1. **Mean pressure.**
   $$\begin{aligned} \text{MAP} &= 80 + \frac{40}{3} \\ &= 93.3\ \text{mmHg} \end{aligned}$$
2. **Systolic estimate.** At the 0.55 ratio:
   $$\begin{aligned} P_s &= 93 + 30\sqrt{-\ln 0.55} \\ &= 116.2\ \text{mmHg} \end{aligned}$$
3. **Diastolic estimate.** At the 0.85 ratio:
   $$\begin{aligned} P_d &= 93 - 30\sqrt{-\ln 0.85} \\ &= 80.9\ \text{mmHg} \end{aligned}$$
4. **A limitation.** A wider envelope from a stiffer artery would move both estimates, an error the fixed ratios can't see.

**Answer:** the mean is 93.3 mmHg, and the envelope reports a systolic pressure of 116.2 mmHg and a diastolic of 80.9 mmHg.
:::
:::

::: example
## What bladder width suits an arm of 32 cm circumference? And how long after deflation starts at 180 mmHg (3 mmHg/s) should the first sound come for a systolic of 120?

::: solution
1. **Bladder width.**
   $$\begin{aligned} w &= 0.4 \times 32 \\ &= 12.8\ \text{cm} \end{aligned}$$
2. **Time to reach 120 mmHg.**
   $$\begin{aligned} t &= \frac{180 - 120}{3} \\ &= 20\ \text{s} \end{aligned}$$

**Answer:** a 12.8 cm bladder suits the arm, and the first sound should come 20 s after deflation starts.
:::
:::

::: example
## An artery has a wall with Young's modulus 0.5 MPa, 0.5 mm thick, and an inner diameter of 10 mm, filled with blood of density 1060 kg/m³. What pulse-wave velocity does the Moens–Korteweg equation give, how long does the pulse take to travel 0.8 m, and how much shorter is that if a pressure rise stiffens the wall by 20%?

::: solution
1. **Pulse-wave velocity.**
   $$\begin{aligned} c &= \sqrt{\frac{Eh}{\rho d}} \\ &= \sqrt{\frac{0.5 \times 10^6 \times 0.5 \times 10^{-3}}{1060 \times 0.01}} \\ &= 4.86\ \text{m/s} \end{aligned}$$
2. **Transit time over 0.8 m.**
   $$\begin{aligned} t &= \frac{0.8}{4.86}\ \text{s} \\ &= 165\ \text{ms} \end{aligned}$$
3. **Stiffer wall.** A 20% stiffer wall speeds the wave by $\sqrt{1.2}$:
   $$\begin{aligned} c' &= 4.86 \times \sqrt{1.2} \\ &= 5.32\ \text{m/s} \end{aligned}$$
   cutting the transit time to 150 ms, 14 ms shorter.
4. **Why it matters.** Because arterial stiffness rises with pressure, pulse transit time, measured from the ECG's R wave to the pulse at a finger, tracks blood pressure without a cuff, once it's calibrated for each person.

**Answer:** the pulse travels at 4.86 m/s and takes 165 ms to cover 0.8 m; a 20% stiffer wall makes that 14 ms shorter, 150 ms.
:::
:::

::: check
given: s = 110 | 130 | 160; d = 70 | 85 | 95
answer: d + (s - d)/3
unit: mmHg
tolerance: 0.5%

## Systolic pressure is {s} mmHg and diastolic {d} mmHg. What's the mean arterial pressure?

::: right
Right. DBP + (SBP − DBP)/3 = {answer} mmHg.
:::

::: misses
- [(s + d)/2] The pressure spends longer near diastolic, so the mean sits a third of the way up.
- [any] Not yet. DBP plus a third of the pulse pressure.
:::
:::

::: check
given: m = 85 | 95 | 110; d = 65 | 75
answer: 3*m - 2*d
unit: mmHg
tolerance: 0.5%

## An oscillometric monitor finds a mean pressure of {m} mmHg, and the diastolic is {d} mmHg. What systolic pressure is consistent with the one-third rule?

::: right
Right. SBP = 3 MAP − 2 DBP = {answer} mmHg.
:::

::: misses
- [2*m - d] That would make the mean the midpoint; use the one-third rule.
- [any] Not yet. Solve MAP = DBP + (SBP − DBP)/3.
:::
:::

::: check
given: M = 85 | 100; w = 20 | 35; k = 0.5 | 0.55 | 0.6
answer: M + w*sqrt(-ln(k))
unit: mmHg
tolerance: 0.5%

## An oscillometric envelope follows $e^{-((P - \text{MAP})/w)^2}$ with MAP = {M} mmHg and w = {w} mmHg. Where on the high-pressure side does it fall to {k} of its peak?

::: right
Right. MAP + $w\sqrt{-\ln k}$ = {answer} mmHg.
:::

::: misses
- [M + w*(1 - k)] The envelope is a bell curve, not a straight line.
- [any] Not yet. Solve $e^{-x^2} = k$.
:::
:::

::: check
given: r = 2 | 3 | 5; H = 50 | 75 | 100
answer: r*60/H
unit: mmHg
tolerance: 0.5%

## A cuff deflates at {r} mmHg/s with a heart rate of {H} per minute. How far does it fall between beats?

::: right
Right. Rate × beat interval = {answer} mmHg.
:::

::: misses
- [r*H/60] The beat interval is 60/H seconds.
- [any] Not yet. Deflation rate times time per beat.
:::
:::

::: check
given: c = 22 | 32 | 42
answer: 0.4*c
unit: cm
tolerance: 0.5%

## What bladder width suits an arm with a {c} cm circumference?

::: right
Right. 40% of the circumference: {answer} cm.
:::

::: misses
- [any] Not yet. About 40% of the arm's circumference.
:::
:::

::: check
given: P = 160 | 180 | 200; s = 110 | 130; r = 2 | 3
answer: (P - s)/r
unit: s
tolerance: 0.5%

## A cuff starts deflating from {P} mmHg at {r} mmHg/s. When does it reach a systolic pressure of {s} mmHg?

::: right
Right. $(P_0 - \text{SBP})/r$ = {answer} s.
:::

::: misses
- [any] Not yet. Pressure drop over rate.
:::
:::

::: check
kind: parts
given: E = 0.3 | 0.8 | 2; h = 0.4 | 1; d = 4 | 20; L = 0.5 | 1

## An artery's wall has Young's modulus {E} MPa and thickness {h} mm, with an inner diameter of {d} mm; blood density is 1060 kg/m³.

::: part
answer: sqrt(E*1e6*h*1e-3/(1060*d*1e-3))
unit: m/s
tolerance: 0.5%

### What pulse-wave velocity does the Moens–Korteweg equation give?

::: misses
- [sqrt(E*1e6*h*1e-3/(1060*d*1e-3*2))] This form of the equation uses the inner diameter, not twice it.
- [any] $c = \sqrt{Eh/(\rho d)}$.
:::
:::

::: part
answer: L/sqrt(E*1e6*h*1e-3/(1060*d*1e-3))*1e3
unit: ms
tolerance: 0.5%

### How long does the pulse take to travel {L} m?

::: misses
- [any] Distance over velocity.
:::
:::
:::

::: check
kind: derivation
vars: s = 90..200, d = 50..110

## A pressure reads s over d mmHg.

::: step
answer: d + (s - d)/3

### What does the usual formula give for the mean arterial pressure?
:::

::: step
answer: s - d

### What is the pulse pressure?
:::

::: right
Right. The one-third rule assumes the waveform spends most of each beat near diastolic pressure; at high heart rates the true mean sits higher.
:::
:::

::: explain
## Why are oscillations in cuff pressure largest at the mean arterial pressure?

::: model
The cuff's pressure oscillations come from the artery under it changing volume with each pulse. How much the volume changes depends on the arterial wall's compliance, which depends on the transmural pressure, the arterial pressure minus the cuff pressure. The wall is most compliant when the transmural pressure is near zero: neither stretched taut nor collapsed. As the arterial pressure swings through the cardiac cycle, the volume swing is largest when the cuff pressure sits at the center of that swing, the mean arterial pressure, so the oscillations peak there. Systolic and diastolic pressures don't mark any such feature, which is why they're estimated from empirical ratios.
:::
:::

::: practice

::: check
given: s = 90 | 220; d = 50 | 120
answer: d + (s - d)/3
unit: mmHg
tolerance: 0.5%

## MAP: {s}/{d} mmHg?

::: misses
- [any] DBP + (SBP − DBP)/3.
:::
:::

::: check
given: M = 60 | 130; w = 10 | 50; k = 0.4 | 0.9
answer: M - w*sqrt(-ln(k))
unit: mmHg
tolerance: 0.5%

## Low-side ratio point: MAP {M}, w = {w}, k = {k}?

::: misses
- [any] MAP − $w\sqrt{-\ln k}$.
:::
:::

::: check
given: r = 1 | 8; H = 40 | 150
answer: r*60/H
unit: mmHg
tolerance: 0.5%

## Drop per beat: {r} mmHg/s, {H} per minute?

::: misses
- [any] $r \times 60/H$.
:::
:::
:::