title: Isolation amplifiers
heading: Breaking the ohmic path between patient and mains-powered equipment, the isolation barrier's capacitance and the leakage it allows, isolation-mode rejection alongside common-mode rejection, transformer, optical and capacitive isolators, and the isolated power supplies that feed them
eyebrow: S15 · Lesson 16 · Isolation amplifiers
xp: 50
---

::: hook
A monitor's electrodes connect the patient, through a few kilohms, to circuitry powered from the wall. If that circuitry's ground drifts, or a fault puts line voltage on it, current flows through the patient. An isolation amplifier puts a barrier in the signal path that passes the ECG but almost no current: kilovolts can stand across it while a few microamperes leak through its capacitance. Every patient-connected input in a hospital crosses one, and its capacitance, voltage rating and rejection set how safe and how quiet the measurement is.
:::

::: play
- The curve is the current that leaks through an isolation barrier of capacitance C when the voltage across it is the line voltage, against frequency.
- The dashed line is 10 µA, the limit for patient leakage through a cardiac-connected input under normal conditions.
- At 60 Hz, a barrier must stay below about 220 pF at 120 V; fast transients and the higher harmonics of a switching supply push leakage up in proportion to frequency.
:::

```sim plot
view: graph
x: f = 10..100000 log
y: I = 0.01..10000 log
slider: C = 1..1000 = 100 log
slider: V = 120..240 = 120
plot: 2*pi*f*C*1e-12*V*1e6 | label: leakage, µA
plot: 10 + 0*f | dashed | label: 10 µA limit
```

::: predict
## What isolation capacitance keeps the 60 Hz leakage from a 120 V line below 10 µA?

- [ ] About 2 pF
- [x] About 220 pF
- [ ] About 22 nF
- [ ] About 2 µF

::: reveal
The barrier must present at least 120 V/10 µA = 12 MΩ at 60 Hz, so $C = 1/(2\pi \times 60 \times 12\text{ M}\Omega)$ = 221 pF, Webster's Example 14.8. Practical isolators have a few picofarads, but the isolated power supply that feeds the patient side adds its own interwinding capacitance, and the total is what counts.
:::
:::

::: nameit
## The isolation barrier

An **isolation amplifier** is an instrumentation amplifier followed by a unity-gain stage that carries the signal across a barrier with no ohmic connection: separate supplies and separate grounds on each side. The barrier is modeled by its isolation capacitance and resistance (typically a few picofarads and more than 10¹⁰ Ω). It must withstand an **isolation voltage**, between the input common and the output common, of 1 to 10 kV without breakdown, and pass leakage of only microamperes at line frequency: $I = 2\pi fC_\text{iso}V$.

## Three kinds of rejection

The output of an isolation amplifier with gain G is
$$v_o = G\left(v_\text{SIG} + \frac{v_\text{CM}}{\text{CMRR}} + \frac{v_\text{ISO}}{\text{IMRR}}\right)$$
The input stage rejects common-mode voltage between the electrodes and the floating input common, as any instrumentation amplifier does (above 100 dB). The barrier rejects the isolation-mode voltage between the two commons, often the full line voltage, with its **isolation-mode rejection ratio**, typically 120 to 160 dB at 60 Hz and falling with frequency.

## How the signal crosses

- **Transformer** isolators modulate a carrier (FM or pulse width) through a small transformer, with signal bandwidths up to tens of kilohertz; a second winding at about 25 kHz, rectified and filtered, powers the input side.
- **Optical** isolators drive an LED whose light reaches a photodiode on the output side; a matched photodiode on the input side, in a feedback loop, cancels the LED's nonlinearity. They pass DC without a modulator.
- **Capacitive** isolators encode the signal digitally or as a frequency and pass it across a pair of small ceramic capacitors, with isolation up to 8 kV and bandwidths to 70 kHz.

Modern designs often digitize on the patient side and isolate the digital stream instead, so the analog signal never crosses the barrier.

## Isolated power

The patient side needs power, from a battery or an isolated DC–DC converter. The converter's transformer has interwinding capacitance of tens of picofarads, often more than the signal isolator's, so it usually dominates the leakage budget. Standards such as IEC 60601-1 set patient leakage limits: 10 µA under normal conditions and 50 µA with a single fault for cardiac-connected (CF) inputs.
:::

::: example
## A transformer-isolated ECG amplifier with gain 1000 has a CMRR of 100 dB and an IMRR of 120 dB. Its input sees 1 mV of ECG, 10 mV of common-mode voltage, and the full 120 V line between its commons. What's the output, and what fraction of it is error?

::: solution
1. **Common-mode term.** Referred to the input, with 100 dB = 10⁵:
   $$\begin{aligned} e_\text{CM} &= \frac{10\ \text{mV}}{10^5} \\ &= 0.1\ \mu\text{V} \end{aligned}$$
2. **Isolation-mode term.** With 120 dB = 10⁶:
   $$\begin{aligned} e_\text{ISO} &= \frac{120\ \text{V}}{10^6} \\ &= 120\ \mu\text{V} \end{aligned}$$
3. **Output.**
   $$\begin{aligned} v_o &= 1000 \times (1\ \text{mV} + 0.1\ \mu\text{V} \\ &\quad + 120\ \mu\text{V}) \\ &= 1.12\ \text{V} \end{aligned}$$
4. **Fraction that is error.** The error terms add to 120.1 µV, nearly all of it line voltage leaking across the isolation barrier. That is 12% of the 1 mV signal.
5. **A better barrier.** A 140 dB IMRR would cut it to 1.2%. The isolation-mode voltage is so large that the IMRR sets the floor, and the CMRR hardly matters.

**Answer:** the output is 1.12 V. The error is 12% of the signal, nearly all of it line voltage leaking across the barrier.
:::
:::

::: example
## An isolated DC–DC converter with 20 pF of interwinding capacitance feeds the patient side of an amplifier whose signal isolator adds 2 pF. What leakage flows at 230 V, 50 Hz, and how much margin is left under a 10 µA limit?

::: solution
1. **Barrier capacitance.** The two capacitances are in parallel across the barrier, 22 pF in all.
2. **Leakage at 230 V, 50 Hz.**
   $$\begin{aligned} I &= 2\pi \times 50 \times 22\ \text{pF} \times 230\ \text{V} \\ &= 1.6\ \mu\text{A} \end{aligned}$$
   a sixth of the 10 µA limit.
3. **The converter's share.** It contributes 20 of the 22 pF, 91% of the leakage.
4. **At 60 Hz and 120 V.** The same barrier leaks 1.0 µA.
5. **What the margin is for.** It allows several such inputs to connect to one patient.

**Answer:** 1.6 µA leaks at 230 V, 50 Hz, a sixth of the 10 µA limit, which leaves room for several such inputs on one patient.
:::
:::

::: example
## An isolation amplifier's IMRR is 140 dB at 60 Hz and falls by 20 dB per decade above. A switching supply puts a 10 V, 60 kHz ripple between the commons. What input-referred error does that cause?

::: solution
1. **IMRR at 60 kHz.** That's three decades above 60 Hz, so the IMRR has fallen to
   $$140 - 3 \times 20 = 80\ \text{dB}$$
2. **Input-referred error.** 80 dB is a ratio of 10⁴:
   $$\begin{aligned} e &= \frac{10\ \text{V}}{10^4} \\ &= 1\ \text{mV} \end{aligned}$$
   as large as the ECG itself, although at a frequency the ECG filter removes.
3. **Fast transients.** Isolation-mode transients, such as an electrosurgical unit's RF or a defibrillator's pulse, couple the same way, which is why isolated inputs also need filtering and transient protection.

**Answer:** the ripple appears as 1 mV at the input, as large as the ECG itself, though at a frequency the ECG filter removes.
:::
:::

::: check
kind: parts
given: V = 120 | 230; f = 50 | 60; I = 10 | 50

## A barrier must limit leakage from a {V} V, {f} Hz line to {I} µA.

::: part
answer: V/(I*1e-6)/1e6
unit: MΩ
tolerance: 0.5%

### What impedance must it present?

::: misses
- [any] Voltage over current.
:::
:::

::: part
answer: 1/(2*pi*f*V/(I*1e-6))*1e12
unit: pF
tolerance: 0.5%

### What is the largest capacitance it can have?

::: misses
- [1/(f*V/(I*1e-6))*1e12] Use the angular frequency, 2πf.
- [any] $C = I/(2\pi fV)$.
:::
:::
:::

::: check
kind: parts
given: G = 100 | 1000; s = 0.5 | 1; c = 5 | 50; C = 90 | 110; v = 120 | 230; M = 110 | 140

## An isolation amplifier with gain {G} has CMRR {C} dB and IMRR {M} dB; its input sees {s} mV of signal, {c} mV of common-mode voltage and {v} V between its commons.

::: part
answer: v/10^(M/20)*1e6
unit: µV
tolerance: 0.5%

### What input-referred error does the isolation-mode voltage cause?

::: misses
- [v/10^(M/10)*1e6] Decibels for voltage ratios use 20 log.
- [any] $v_\text{ISO}/\text{IMRR}$.
:::
:::

::: part
answer: G*(s*1e-3 + c*1e-3/10^(C/20) + v/10^(M/20))
unit: V
tolerance: 0.5%

### What is the total output?

::: misses
- [G*s*1e-3] Add the two error terms.
- [any] $G(v_\text{SIG} + v_\text{CM}/\text{CMRR} + v_\text{ISO}/\text{IMRR})$.
:::
:::
:::

::: check
kind: parts
given: a = 2 | 10 | 40; b = 1 | 5; V = 120 | 230; f = 50 | 60; n = 1 | 4

## An isolated converter with {a} pF of interwinding capacitance and a signal isolator with {b} pF feed one input; {n} such inputs connect to a patient, with {V} V at {f} Hz across the barriers.

::: part
answer: 2*pi*f*(a + b)*1e-12*V*1e6
unit: µA
tolerance: 0.5%

### How much leaks through one input's barrier?

::: misses
- [2*pi*f*a*1e-12*V*1e6] The isolator's capacitance is in parallel with the converter's.
- [any] $2\pi fCV$ with the capacitances added.
:::
:::

::: part
answer: n*2*pi*f*(a + b)*1e-12*V*1e6
unit: µA
tolerance: 0.5%

### How much in total through all {n}?

::: misses
- [any] The barriers are in parallel.
:::
:::
:::

::: check
kind: parts
given: M = 120 | 140 | 160; F = 1000 | 30000 | 100000; v = 1 | 10

## An isolator's IMRR is {M} dB at 60 Hz and falls by 20 dB per decade above. A {v} V disturbance at {F} Hz appears between its commons.

::: part
answer: M - 20*log(F/60)
unit: dB
tolerance: 0.5%

### What is its IMRR at {F} Hz?

::: misses
- [any] Subtract 20 dB per decade above 60 Hz.
:::
:::

::: part
answer: v/10^((M - 20*log(F/60))/20)*1e6
unit: µV
tolerance: 0.5%

### What input-referred error results?

::: misses
- [v/10^(M/20)*1e6] Use the IMRR at the disturbance's frequency.
- [any] Disturbance over the IMRR there.
:::
:::
:::

::: check
kind: parts
given: V = 2 | 5; R = 10 | 50; t = 5 | 10

## An ECG input is protected from a {V} kV defibrillator pulse lasting {t} ms by a {R} kΩ series resistor ahead of clamp diodes that hold the amplifier's input near 0 V.

::: part
answer: V*1e3/(R*1e3)*1e3
unit: mA
tolerance: 0.5%

### What current flows through the resistor?

::: misses
- [any] Nearly the whole pulse is across the resistor.
:::
:::

::: part
answer: (V*1e3)^2/(R*1e3)*t*1e-3
unit: J
tolerance: 0.5%

### How much energy must the resistor absorb?

::: misses
- [any] $V^2t/R$.
:::
:::
:::

::: check
kind: choice

## Why do many modern monitors digitize the signal on the patient side and isolate the digital data instead of the analog signal?

- [ ] Digital isolators need no power on the patient side
- [x] A digital stream crosses the barrier without any error from the barrier's nonlinearity, drift or limited IMRR at the signal's frequencies, and digital isolators are small and cheap
- [ ] It removes the need for an isolated power supply
- [ ] Analog signals can't cross a capacitive barrier

::: misses
- [1] The converter on the patient side still needs isolated power.
- [3] The patient side still needs power from across the barrier or a battery.
- [4] Capacitive analog isolators exist; the advantage is accuracy.
:::
:::

::: check
kind: derivation
vars: G = 10..10000, s = 0.0001..0.005, c = 0.001..0.1, C = 1000..10000000, v = 1..300, M = 100000..1000000000

## An isolation amplifier has gain G, CMRR C and IMRR M (as ratios); its input sees signal s, common-mode voltage c and isolation-mode voltage v.

::: step
answer: G*(s + c/C + v/M)

### What is its output?
:::

::: step
answer: (c/C + v/M)/s

### What fraction of the signal is error?
:::

::: right
Right. The isolation-mode voltage can be ten thousand times the common-mode voltage, so IMRR usually sets the floor.
:::
:::

::: explain
## Why does an isolation amplifier make a patient safer even though its barrier still has capacitance?

::: model
Without isolation, the electrodes connect the patient through the amplifier's input circuit to the equipment's ground and supplies, so a ground difference or a fault drives current straight through the patient, limited only by a few kilohms. The isolation barrier replaces that path with a tiny capacitance whose impedance at line frequency is tens of megohms, so even the full line voltage across it pushes only microamperes, below the level that can disturb the heart. The barrier still passes the signal by modulating a carrier, light or a digital code across it, so the measurement works while the dangerous low-frequency current path is broken.
:::
:::

::: practice
::: check
given: V = 120 | 230; f = 50 | 60; I = 10 | 20
answer: 1/(2*pi*f*V/(I*1e-6))*1e12
unit: pF
tolerance: 0.5%

## Largest barrier capacitance: {V} V, {f} Hz, {I} µA?

::: misses
- [any] $I/(2\pi fV)$.
:::
:::

::: check
given: v = 120 | 230; M = 120 | 150
answer: v/10^(M/20)*1e6
unit: µV
tolerance: 0.5%

## Input-referred error: {v} V at IMRR {M} dB?

::: misses
- [any] $v/10^{M/20}$.
:::
:::

::: check
given: a = 5 | 30; V = 120 | 230; f = 50 | 60
answer: 2*pi*f*a*1e-12*V*1e6
unit: µA
tolerance: 0.5%

## Leakage: {a} pF at {V} V, {f} Hz?

::: misses
- [any] $2\pi fCV$.
:::
:::
:::