title: Plethysmography and impedance measurements
heading: Venous-occlusion plethysmography for limb blood flow, impedance plethysmography from a cylinder model, four-electrode measurement at 100 kHz, impedance cardiography and apnea monitoring, and hot-film velocity sensors
eyebrow: S15 · Lesson 28 · Plethysmography and impedance measurements
xp: 50
---

::: hook
The flowmeters of lesson 27 need a probe on or in the vessel. Often all a clinician can reach is the outside of a limb or the chest, yet they still want to know whether blood is getting into the calf, whether a vein is blocked by a clot, or whether a sleeping infant is still breathing. Each pulse of blood swells a limb by a fraction of a milliliter and each breath changes the chest's electrical impedance by a fraction of an ohm. How do instruments turn those small changes into volumes and flows?
:::

::: play
- Venous-occlusion plethysmography of the calf. At t = 0 a thigh cuff is inflated to about 50 mmHg, which stops venous outflow but not arterial inflow, so the calf swells. At 20 s the cuff is released and the extra blood drains.
- The initial slope of the volume curve is the arterial inflow. As venous pressure rises toward the cuff pressure, the swelling levels off.
- After release a healthy calf empties with a time constant of a couple of seconds. Raise the outflow time constant to see the slow emptying that signals a deep-vein thrombosis.
:::

```sim plot
view: graph
x: t = 0..45 s
y: V = 0..30
slider: F = 0.1..1 = 0.5
slider: tau = 1..15 = 2 s
plot: step(20 - t)*12*F*(1 - exp(-t/(12))) + step(t - 20)*12*F*(1 - exp(-20/12))*exp(-(t - 20)/tau) | samples: 900 | label: added volume, mL
plot: F*t + 0*tau | dashed | color: ch3 | label: initial inflow slope
readout: arterial inflow, mL/min = 60*F | digits: 3
readout: per 100 mL of a 1000 mL calf, mL/min = 6*F | digits: 3
```

::: predict
## An impedance plethysmograph drives a constant 1 mA at 100 kHz through a limb segment of 30 Ω. Each heartbeat lowers the impedance by 0.1 Ω. How big is the signal to be measured?

- [ ] 30 mV
- [x] 0.1 mV on top of 30 mV
- [ ] 0.1 V
- [ ] 3 µV

::: reveal
The segment's voltage is 1 mA × 30 Ω = 30 mV, and the pulsatile change is 1 mA × 0.1 Ω = 0.1 mV, a third of a percent. The instrument has to subtract the steady 30 mV and amplify the remainder, much as a strain-gauge bridge does in lesson 3.
:::
:::

::: nameit
## Venous-occlusion plethysmography

A **plethysmograph** measures a change in volume. A rigid chamber around a limb, filled with water or air, measures the limb's swelling directly. A cuff upstream inflated to about 50 mmHg blocks venous return without stopping arterial inflow, so the initial rate of swelling is the arterial flow, $F = dV/dt$, usually reported per 100 mL of tissue. When the cuff is released, the limb drains; slow drainage suggests a venous obstruction.

## Impedance plethysmography

Model a limb segment of length L as a cylinder of resistance Z. Each pulse adds a volume ΔV of blood of resistivity $\rho_b$ in parallel, lowering the impedance by ΔZ. Following Swanson's derivation in Webster, with ΔZ ≪ Z,

$$\Delta V = \rho_b\,\frac{L^2\,|\Delta Z|}{Z^2}$$

It assumes the arteries expand uniformly, blood's resistivity stays constant (about 1.5 Ω·m) and the current runs parallel to the vessels, which is reasonable for most limb segments.

Measurements use **four electrodes**: two outer ones drive a constant current and two inner ones sense the voltage, so the large and variable skin–electrode impedances don't add to the reading. The current is about 1 mA at around 100 kHz, where electrode impedance is low and the current is far above any stimulation threshold, which makes a milliampere safe. A balancing circuit removes the steady voltage so the small pulsatile part can be amplified.

## Applications

- **Venous thrombosis:** impedance replaces the chamber in venous-occlusion studies.
- **Apnea monitoring:** electrodes across the chest track breathing, since inhaled air raises the chest's impedance; infant monitors alarm when breaths stop, rejecting the heartbeat's smaller ripple by pattern recognition.
- **Impedance cardiography:** band electrodes around the neck and lower chest give a beat-by-beat estimate of stroke volume, by Kubicek's formula $SV = \rho_b (L/Z_0)^2\,T\,(dZ/dt)_\text{max}$ with T the ejection time. It works fairly well in healthy subjects and less well in very sick patients.
- **Photoplethysmography**, the optical version, tracks blood volume by absorbed light and underlies pulse oximetry (lesson 33).

## Thermal velocity sensors

A heated film or bead in the bloodstream loses heat faster when the blood moves faster. Held at a constant temperature rise ΔT above the blood, its power follows King's law, $P = (a + b\sqrt{v})\Delta T$, so measuring P gives the velocity v after calibration. These sensors are small and fast enough to follow velocity within a beat, but they need a separate temperature sensor and can't tell direction.
:::

::: example
## A 10 cm limb segment measures 30 Ω, and each pulse lowers it by 0.1 Ω. Taking $\rho_b$ = 1.5 Ω·m, how much blood does each pulse add?

::: solution
1. **Blood volume per pulse.**
   $$\begin{aligned} \Delta V &= \frac{1.5 \times (0.1)^2 \times 0.1}{30^2}\ \text{m}^3 \\ &= 1.67 \times 10^{-6}\ \text{m}^3 \\ &= 1.67\ \text{mL} \end{aligned}$$
2. **Sanity check.** That's a volume change of a few tenths of a percent of the segment, consistent with the 0.33% impedance change.
3. **The formula's assumptions.** If the blood's resistivity drops with flow, as it does when red cells align in the stream, the computed volume is off by the same fraction.

**Answer:** each pulse adds 1.67 mL of blood.
:::
:::

::: example
## Band electrodes 30 cm apart on a patient's chest read $Z_0$ = 25 Ω, a peak dZ/dt of 1.2 Ω/s and an ejection time of 0.3 s. Estimate the stroke volume with $\rho_b$ = 1.35 Ω·m.

::: solution
1. **Stroke volume.**
   $$\begin{aligned} SV &= 1.35 \times (0.30/25)^2 \\ &\quad \times 0.3 \times 1.2 \\ &= 1.35 \times 1.44 \times 10^{-4} \times 0.36 \\ &= 7.0 \times 10^{-5}\ \text{m}^3 \\ &= 70\ \text{mL} \end{aligned}$$
   a typical resting stroke volume.
2. **Cardiac output.** At 70 beats a minute that's 4.9 L/min, which can be compared against the indicator-dilution methods of lesson 26.

**Answer:** the stroke volume is about 70 mL, a cardiac output of 4.9 L/min at 70 beats a minute.
:::
:::

::: example
## A mercury strain gauge around a calf of 35 cm circumference stretches by 0.35 mm in the first 10 s of venous occlusion. Taking a 10 cm segment as a cylinder, what's the arterial inflow per 100 mL of tissue per minute?

::: solution
1. **Fractional volume change.** For a cylinder, volume goes as circumference squared:
   $$\begin{aligned} \frac{\Delta V}{V} &= \frac{2\Delta C}{C} \\ &= \frac{2 \times 0.35}{350} \\ &= 0.2\% \end{aligned}$$
2. **Segment volume.**
   $$\begin{aligned} V &= \frac{C^2L}{4\pi} \\ &= 975\ \text{mL} \end{aligned}$$
3. **Inflow.** The segment gains 1.95 mL in 10 s, 11.7 mL/min.
4. **Per 100 mL of tissue.** That's 1.2 mL per 100 mL per minute, a normal resting calf flow.
5. **Why the gauge lasted.** It needs no electrodes and reads the whole limb's swelling, which is why it remained the standard for venous-occlusion studies.

**Answer:** the arterial inflow is 1.2 mL per 100 mL of tissue per minute, a normal resting calf flow.
:::
:::

::: check
given: r = 1.3 | 1.5 | 1.7; L = 5 | 10 | 15; Z = 20 | 30 | 50; d = 0.05 | 0.1 | 0.2
answer: r*(L/100)^2*d/Z^2*1e6
unit: mL
tolerance: 0.5%

## A limb segment {L} cm long measures {Z} Ω, and a pulse lowers it by {d} Ω. With blood resistivity {r} Ω·m, how much volume does the pulse add?

::: right
Right. $\rho_b L^2|\Delta Z|/Z^2$ = {answer} mL.
:::

::: misses
- [r*(L/100)*d/Z^2*1e6] L appears squared.
- [r*(L/100)^2*d/Z*1e6] Z appears squared.
- [any] Not yet. $\Delta V = \rho_b L^2|\Delta Z|/Z^2$.
:::
:::

::: check
given: s = 0.2 | 0.5 | 0.8; v = 800 | 1000 | 1500
answer: s*60/(v/100)
tolerance: 0.5%

## After a venous-occlusion cuff inflates, a {v} mL calf segment swells at {s} mL/s. What's the blood flow in mL per minute per 100 mL of tissue?

::: right
Right. {s} × 60 mL/min spread over {v}/100 hundreds of milliliters: {answer}.
:::

::: misses
- [s*60] Divide by the number of 100 mL portions of tissue.
- [s/(v/100)] Convert seconds to minutes.
- [any] Not yet. Convert the slope to mL/min and normalize to 100 mL.
:::
:::

::: check
given: I = 0.5 | 1 | 2; Z = 20 | 30 | 40; d = 0.05 | 0.1 | 0.2
answer: I*d
unit: mV
tolerance: 0.5%

## A four-electrode plethysmograph drives {I} mA through a {Z} Ω segment. What's the pulsatile signal for a {d} Ω change?

::: right
Right. I ΔZ = {answer} mV, riding on {I} × {Z} mV.
:::

::: misses
- [I*Z] That's the steady voltage. The pulsatile part is I ΔZ.
- [any] Not yet. The signal is the current times the change in impedance.
:::
:::

::: check
given: r = 1.3 | 1.5; L = 25 | 30 | 35; Z = 20 | 25 | 30; T = 0.25 | 0.3; g = 0.8 | 1.2 | 1.6
answer: r*(L/100/Z)^2*T*g*1e6
unit: mL
tolerance: 0.5%

## Impedance cardiography: bands {L} cm apart, $Z_0$ = {Z} Ω, $(dZ/dt)_\text{max}$ = {g} Ω/s, ejection time {T} s, $\rho_b$ = {r} Ω·m. Estimate the stroke volume.

::: right
Right. $\rho_b(L/Z_0)^2 T (dZ/dt)_\text{max}$ = {answer} mL.
:::

::: misses
- [r*(L/100/Z)*T*g*1e6] Square $L/Z_0$.
- [any] Not yet. Use Kubicek's formula.
:::
:::

::: check
given: a = 1 | 2 | 3; b = 4 | 6 | 8; D = 4 | 6; P = 0.05 | 0.075 | 0.1
answer: ((P/D - a/1000)/(b/1000))^2
unit: m/s
tolerance: 0.5%

## A hot-film sensor held {D} K above blood draws {P} W. Its calibration is $P = (a + b\sqrt v)\Delta T$ with a = {a} mW/K and b = {b} mW/(K·(m/s)^½). What's the blood velocity?

::: right
Right. $\sqrt v = (P/\Delta T - a)/b$, so v = {answer} m/s.
:::

::: misses
- [(P/D - a/1000)/(b/1000)] That's $\sqrt v$. Square it.
- [any] Not yet. Solve King's law for $\sqrt v$, then square.
:::
:::

::: check
kind: parts
given: C = 25 | 35 | 45; c = 0.1 | 0.3 | 0.6; t = 5 | 10

## A strain gauge around a limb of {C} cm circumference stretches by {c} mm in {t} s of venous occlusion.

::: part
answer: 2*c/(C*10)*100
unit: %
tolerance: 0.5%

### By what percent does the limb's volume grow?

::: misses
- [c/(C*10)*100] Volume goes as circumference squared: double the fractional change.
- [any] $\Delta V/V = 2\Delta C/C$.
:::
:::

::: part
answer: 2*c/(C*10)*100*60/t
unit: mL/min
tolerance: 0.5%

### What inflow is that, per 100 mL of tissue?

::: misses
- [any] Percent per second times 60.
:::
:::
:::

::: check
kind: derivation
vars: r = 1..3, L = 0.05..0.3, Z = 10..100, d = 0.01..1

## A limb segment of length L (m) is modeled as a cylinder of blood-like conductor with resistivity ρ (Ω·m) and impedance Z.

::: step
answer: r*L^2/Z

### What volume does the impedance imply, in cubic meters?
:::

::: step
answer: r*L^2*d/Z^2

### What volume change does an impedance drop of ΔZ mean?
:::

::: right
Right. The same cylinder model gives the volume and its change, which is why impedance plethysmography needs only Z, ΔZ and the electrode spacing.
:::
:::

::: explain
## Why does impedance plethysmography use four electrodes and a 100 kHz current?

::: model
The quantity of interest is a tiny change, a fraction of an ohm, in the impedance of the tissue between the electrodes. With only two electrodes, the measured impedance would include the skin–electrode impedances, which are hundreds to thousands of ohms, drift with sweat and pressure, and change with movement: they would swamp the signal. With four, the outer pair carries a known current and the inner pair senses voltage through a high-impedance amplifier that draws almost no current, so the inner electrodes' own impedances drop nothing and only the tissue's impedance appears. A high frequency helps twice: electrode impedance, which is largely capacitive, falls with frequency, and nerve and muscle can't respond to 100 kHz, so a current of a milliampere, large enough for a clean signal, is safe and unfelt.
:::
:::

::: practice

::: check
given: r = 1.2..1.8; L = 3..20; Z = 10..60; d = 0.02..0.3
answer: r*(L/100)^2*d/Z^2*1e6
unit: mL
tolerance: 0.5%

## ΔV for L = {L} cm, Z = {Z} Ω, ΔZ = {d} Ω, $\rho_b$ = {r}?

::: misses
- [any] $\rho_b L^2|\Delta Z|/Z^2$.
:::
:::

::: check
given: s = 0.1..1.5; v = 500..2000
answer: s*60/(v/100)
tolerance: 0.5%

## Flow per 100 mL per minute: {s} mL/s into {v} mL?

::: misses
- [any] Slope × 60 ÷ (volume/100).
:::
:::
:::