title: "Problem set 1: measurement systems, sensors and optics"
heading: "Problem set 1: sensitivity, nonlinearity and resolution, first- and second-order sensor dynamics, strain gauges and bridges, capacitive and piezoelectric sensors, and phase-sensitive and optical detection"
eyebrow: S15 · Problem set 1
type: problems
---

::: hook
These problems cover lessons 1 to 5: transducer outputs and nonlinearity, input-referred resolution and combined errors, step, ramp and sinusoid responses of first-order probes, catheter amplitude ratios and overshoot, gauge resistance changes and bridge outputs, loaded potentiometers, capacitive and piezoelectric sensor signals, phase-sensitive demodulation, fibers and photodiodes.
:::

::: pool
pick: 12

::: check
kind: parts
given: s = 5 | 20; V = 3 | 6; P = 80 | 250; a = 0.001 | 0.004; X = 200 | 300

## A transducer of {s} µV/V/mmHg at {V} V excitation; separately, an output $sx + ax^2$ with a = {a} µV/mmHg² over 0 to {X} mmHg at the same s.

::: part
answer: s*V*P/1000
unit: mV
tolerance: 0.5%

### Output at {P} mmHg?

::: misses
- [any] $sVP$.
:::
:::

::: part
answer: 100*a*X^2/4/(s*X + a*X^2)
tolerance: 0.5%

### Nonlinearity against the endpoint line, in percent?

::: misses
- [any] $aX^2/4$ over full scale.
:::
:::
:::

::: check
kind: parts
given: R = 3 | 10; N = 12 | 16; G = 200 | 2000; a = 0.5 | 1.5; b = 0.4 | 2; c = 1 | 3

## A gain of {G} into an {N}-bit converter spanning {R} V; independent errors {a}, {b} and {c} mmHg.

::: part
answer: R/(2^N*G)*1e6
unit: µV
tolerance: 0.5%

### Input-referred step?

::: misses
- [any] $R/(2^NG)$.
:::
:::

::: part
answer: sqrt(a^2 + b^2 + c^2)
tolerance: 0.5%

### Combined error in mmHg?

::: misses
- [any] Root sum of squares.
:::
:::
:::

::: check
kind: parts
given: T = 1.5 | 3; u = 8 | 20; s = 15 | 40; e = 0.05 | 0.2

## A probe with τ = {u} s meets a {T} °C step.

::: part
answer: T*exp(-s/u)
unit: °C
tolerance: 0.5%

### Remaining error after {s} s?

::: misses
- [any] $\Delta Te^{-t/\tau}$.
:::
:::

::: part
answer: u*ln(T/e)
unit: s
tolerance: 0.5%

### Time to come within {e} °C?

::: misses
- [any] $\tau\ln(\Delta T/\varepsilon)$.
:::
:::
:::

::: check
kind: parts
given: r = 0.5 | 2; u = 10 | 40; f = 0.05 | 0.5

## A first-order probe with τ = {u} s.

::: part
answer: r*u/60
unit: °C
tolerance: 0.5%

### Lag behind a ramp of {r} °C/min?

::: misses
- [any] Rate × τ.
:::
:::

::: part
answer: 1/sqrt(1 + (2*pi*f*u)^2)
tolerance: 0.5%

### Amplitude ratio at {f} Hz?

::: misses
- [any] $1/\sqrt{1 + (2\pi f\tau)^2}$.
:::
:::
:::

::: check
kind: parts
given: n = 20 | 40; z = 0.15 | 0.35; f = 8 | 15

## A catheter system with $f_n$ = {n} Hz and ζ = {z}.

::: part
answer: 1/sqrt((1 - (f/n)^2)^2 + (2*z*f/n)^2)
tolerance: 0.5%

### Amplitude ratio at {f} Hz?

::: misses
- [any] Second-order magnitude.
:::
:::

::: part
answer: 100*exp(-pi*z/sqrt(1 - z^2))
tolerance: 0.5%

### Step overshoot in percent?

::: misses
- [any] $100e^{-\pi\zeta/\sqrt{1 - \zeta^2}}$.
:::
:::
:::

::: check
kind: parts
given: G = 2 | 2.1; e = 300 | 1200; R = 120 | 350; V = 2.5 | 10

## A {R} Ω gauge with G = {G} at {e} microstrain, in a bridge excited at {V} V.

::: part
answer: G*e*1e-6*R
unit: Ω
tolerance: 0.5%

### Resistance change?

::: misses
- [any] $G\varepsilon R$.
:::
:::

::: part
answer: V*G*e*1e-6/4*1000
unit: mV
tolerance: 0.5%

### Approximate quarter-bridge output?

::: misses
- [any] $VG\varepsilon/4$.
:::
:::

::: part
answer: V*G*e*1e-6*1000
unit: mV
tolerance: 0.5%

### Output if four such gauges form a full bridge?

::: misses
- [any] $VG\varepsilon$.
:::
:::
:::

::: check
kind: parts
given: x = 0.03 | 0.08; V = 3 | 6

## A quarter bridge at {V} V with ΔR/R = {x} in its active arm.

::: part
answer: V*x/(2*(2 + x))*1000
unit: mV
tolerance: 0.05%

### Exact output?

::: misses
- [any] $Vx/(2(2 + x))$.
:::
:::

::: part
answer: 100*(x/4 - x/(2*(2 + x)))/(x/(2*(2 + x)))
tolerance: 0.5%

### Percent error of the linear approximation?

::: misses
- [any] Compare $Vx/4$ with the exact output.
:::
:::
:::

::: check
kind: parts
given: a = 15 | 40; G = 2 | 2.1; T = 2 | 8; x = 0.3 | 0.5; p = 2 | 10; L = 20 | 50

## A foil gauge (G = {G}) with a {a} ppm/°C temperature coefficient; a {p} kΩ pot at {x} of its travel with a {L} kΩ load.

::: part
answer: a*T/G
tolerance: 0.5%

### Apparent strain for a {T} °C change, in microstrain?

::: misses
- [any] $\alpha\Delta T/G$.
:::
:::

::: part
answer: x/(1 + p/L*x*(1 - x))
tolerance: 0.05%

### Wiper output as a fraction of the excitation?

::: misses
- [any] $x/(1 + (R_p/R_L)x(1 - x))$.
:::
:::
:::

::: check
kind: parts
given: A = 0.5 | 2; d = 0.1 | 0.4; r = 2 | 5; x = 1 | 4

## A capacitive sensor: {A} cm² plates {d} mm apart, relative permittivity {r}.

::: part
answer: 8.854e-12*r*A*1e-4/(d*1e-3)*1e12
unit: pF
tolerance: 0.5%

### Capacitance?

::: misses
- [any] $\varepsilon_0\varepsilon_rA/d$.
:::
:::

::: part
answer: 8.854e-12*r*A*1e-4/(d*1e-3)*1e12*x/(d*1000)*1000
unit: fF
tolerance: 0.5%

### Change for a {x} µm gap reduction?

::: misses
- [any] $C\Delta d/d$.
:::
:::
:::

::: check
kind: parts
given: k = 150 | 450; F = 0.5 | 3; C = 1 | 4; R = 0.5 | 5

## A piezoelectric sensor with d = {k} pC/N and {C} nF, read across {R} GΩ, under a {F} N step.

::: part
answer: k*F/(C*1000)
unit: V
tolerance: 0.5%

### Initial voltage?

::: misses
- [any] $dF/C$.
:::
:::

::: part
answer: 1/(2*pi*R*1e9*C*1e-9)
unit: Hz
tolerance: 0.5%

### Low-frequency corner?

::: misses
- [any] $1/(2\pi RC)$.
:::
:::

::: part
answer: k*F/(C*1000)*exp(-1/(R*C))
unit: V
tolerance: 0.5%

### Voltage after 1 s?

::: misses
- [any] $Ve^{-t/RC}$.
:::
:::
:::

::: check
kind: parts
given: k = 100 | 400; F = 0.3 | 2; f = 50 | 500; c = 0.002 | 0.006; d = 4 | 12

## A charge amplifier with $C_f$ = {f} pF reads a sensor with d = {k} pC/N; a transducer whose sensitivity falls by a fraction {c} per °C.

::: part
answer: k*F/f
unit: V
tolerance: 0.5%

### Output size for {F} N?

::: misses
- [any] $dF/C_f$.
:::
:::

::: part
answer: 100*c*d
tolerance: 0.5%

### Percent gain error after {d} °C?

::: misses
- [any] Fraction per degree times degrees.
:::
:::
:::

::: check
kind: parts
given: H = 60 | 150; k = 8 | 12; n = 30 | 80; z = 0.3 | 0.6

## A heart rate of {H} per minute whose first {k} harmonics must be reproduced by a catheter with $f_n$ = {n} Hz and ζ = {z}.

::: part
answer: k*H/60
unit: Hz
tolerance: 0.5%

### Highest harmonic needed?

::: misses
- [any] k times the heart rate in hertz.
:::
:::

::: part
answer: 100*(1/sqrt((1 - (k*H/60/n)^2)^2 + (2*z*k*H/60/n)^2) - 1)
tolerance: 0.5%

### Percent error in that harmonic's amplitude?

::: misses
- [any] Amplitude ratio minus 1.
:::
:::
:::

::: check
kind: parts
given: A = 30 | 60 | 120; f = 15 | 30 | 50; q = 20 | 40

## An LVDT's 4 kHz carrier reaches a phase-sensitive demodulator with an in-phase amplitude of {A} mV at {f}° to the reference, plus {q} mV of quadrature pickup from cable capacitance.

::: part
answer: 2/pi*A*cos(f*pi/180)
unit: mV
tolerance: 0.5%

### Average output from the in-phase signal?

::: misses
- [any] $(2/\pi)A\cos\varphi$.
:::
:::

::: part
answer: 2/pi*A*cos(f*pi/180)
unit: mV
tolerance: 0.5%

### Average output with the quadrature pickup added?

::: misses
- [any] A component at 90° to the reference averages to zero; the output doesn't change.
:::
:::

::: part
answer: 1/(2*pi*10*220e-9)/1000
unit: kΩ
tolerance: 0.5%

### Resistor for a 10 Hz RC output filter with 220 nF?

::: misses
- [any] $R = 1/(2\pi f_cC)$.
:::
:::
:::

::: check
kind: parts
given: a = 1.49 | 1.5; b = 1.40 | 1.42; e = 0.7 | 0.85; l = 660 | 880; P = 5 | 20

## A fiber with core index {a} and cladding {b} carries {P} µW at {l} nm to a photodiode with quantum efficiency {e}.

::: part
answer: sqrt(a^2 - b^2)
tolerance: 0.5%

### The fiber's numerical aperture?

::: misses
- [any] $\sqrt{n_1^2 - n_2^2}$.
:::
:::

::: part
answer: asin(sqrt(a^2 - b^2))*180/pi
unit: °
tolerance: 0.5%

### Acceptance half-angle in air?

::: misses
- [any] arcsin of the NA.
:::
:::

::: part
answer: e*(l/1000)/1.23984*P
unit: µA
tolerance: 0.5%

### Photocurrent?

::: misses
- [any] $\eta\lambda[\mu\text{m}]/1.24$ A/W times the power.
:::
:::
:::

::: check
kind: parts
given: R = 120 | 350; L = 0.5 | 2; G = 2 | 2.1; T = 10 | 25; e = 200 | 800

## A {R} Ω gauge (G = {G}) measures {e} microstrain through two-wire leads totaling {L} Ω of copper (0.0039/°C), which warm by {T} °C during the test.

::: part
answer: e*R/(R + L)
unit: microstrain
tolerance: 0.5%

### What strain does the bridge indicate, before the leads warm?

::: misses
- [any] The leads dilute the gauge's change by $R/(R + R_L)$.
:::
:::

::: part
answer: L*0.0039*T/(R*G)*1e6
unit: microstrain
tolerance: 0.5%

### How much apparent strain does the warming add?

::: misses
- [any] $R_L\alpha\Delta T/(RG)$.
:::
:::
:::
:::