title: "Problem set 11: biomedical instrumentation across the course"
heading: "Problem set 11: problems that chain sensors to amplifiers, biopotentials to front ends, pressure and flow to their transducers, optics and chemistry to readings, imaging to its limits, and therapy and safety to their budgets"
eyebrow: S15 · Problem set 11
type: problems
---

::: hook
These problems draw on the whole course, and most chain several ideas: a transducer's output carried through an amplifier and converter, an ECG's common-mode problem traced from the power line to the input, a catheter's bubble turned into a pressure error, a Doppler shift turned into a flow, a pacemaker's pulse turned into a battery life. Work them as a review before the final.
:::

::: pool
pick: 12

::: check
kind: parts
given: s = 5 | 10; V = 5 | 6; P = 100 | 150; G = 200 | 500; N = 12 | 16; R = 5 | 10

## A {s} µV/V/mmHg transducer at {V} V feeds a gain of {G} and an {N}-bit converter spanning {R} V.

::: part
answer: s*V*P*G/1e6
unit: V
tolerance: 0.5%

### Amplifier output at {P} mmHg?

::: misses
- [any] $sVPG$.
:::
:::

::: part
answer: R/2^N/(s*1e-6*V*G)
tolerance: 0.5%

### Resolution in mmHg?

::: misses
- [any] One converter step referred back through gain and sensitivity.
:::
:::
:::

::: check
kind: parts
given: G = 2 | 100; e = 300 | 800; V = 3 | 5; g = 50 | 500; N = 12 | 16

## A full bridge of gauges (G = {G}) at ±{e} microstrain and {V} V, amplified {g} times into an {N}-bit converter spanning ±5 V.

::: part
answer: V*G*e*1e-6*g
unit: V
tolerance: 0.5%

### Amplified output?

::: misses
- [any] $VG\varepsilon \times$ gain.
:::
:::

::: part
answer: V*G*e*1e-6*g/(10/2^N)
tolerance: 0.5%

### How many converter steps is that?

::: misses
- [any] Output over the step size.
:::
:::
:::

::: check
kind: parts
given: R = 8 | 12; B = 3800 | 4100; T = 36 | 39; d = 0.05 | 0.1

## A {R} kΩ (25 °C) thermistor with B = {B} K near {T} °C, read to {d} °C.

::: part
answer: R*exp(B*(1/(T + 273.15) - 1/298.15))*B/(T + 273.15)^2*d*1000
unit: Ω
tolerance: 0.5%

### Resistance change to resolve?

::: misses
- [any] $|\alpha|R\Delta T$.
:::
:::

::: part
answer: 100*B/(T + 273.15)^2*d
tolerance: 0.5%

### As a percent of its resistance?

::: misses
- [any] $|\alpha|\Delta T$.
:::
:::
:::

::: check
kind: parts
given: V = 120 | 230; C = 2 | 4; Z = 30 | 60; d = 10 | 30; M = 50 | 200; c = 90 | 110

## A {V} V, 60 Hz line couples {C} pF to a patient grounded through {Z} kΩ; electrodes differ by {d} kΩ; the amplifier has {M} MΩ common-mode input impedance and {c} dB CMRR.

::: part
answer: V*2*pi*60*C*1e-12*Z*1e3*1e3
unit: mV
tolerance: 0.5%

### Common-mode voltage?

::: misses
- [any] $i_dZ_G$.
:::
:::

::: part
answer: (V*2*pi*60*C*1e-12*Z*1e3)*(d*1e3/(M*1e6) + 1/10^(c/20))*1e6
unit: µV
tolerance: 0.5%

### Total input-referred 60 Hz error?

::: misses
- [any] $v_{cm}(\Delta Z/Z_{cm} + 1/\text{CMRR})$.
:::
:::
:::

::: check
kind: parts
given: a = 0.5 | 0.9; b = 0.9 | 1.4; D = 18 | 25; o = 200 | 300; V = 4 | 5; T = 1000 | 2000

## Lead I = {a} mV, lead II = {b} mV; R waves {D} mm apart at 25 mm/s; electrode offsets up to {o} mV into a ±{V} V first stage; total gain {T}.

::: part
answer: atan((2*b - a)/(sqrt(3)*a))*180/pi
unit: °
tolerance: 0.5%

### Axis?

::: misses
- [any] $\tan^{-1}[(2\,\text{II} - \text{I})/(\sqrt{3}\,\text{I})]$.
:::
:::

::: part
answer: 1500/D
tolerance: 0.5%

### Rate per minute?

::: misses
- [any] 1500/D.
:::
:::

::: part
answer: T/(V/(o*1e-3))
tolerance: 0.5%

### Gain needed after the high-pass?

::: misses
- [any] Total over the first-stage limit.
:::
:::
:::

::: check
kind: parts
given: r = 0.4 | 0.6; L = 1 | 1.5; c = 0.01 | 0.03; b = 0.5 | 2

## A catheter (radius {r} mm, {L} m, saline) on a transducer of {c} mm³ per 100 mmHg compliance; a {b} mm³ bubble enters.

::: part
answer: r*1e-3/2*sqrt(1/((c*1e-9/(100*133.322))*pi*1000*L))
unit: Hz
tolerance: 0.5%

### Natural frequency without the bubble?

::: misses
- [any] $\frac{r}{2}\sqrt{E_d/\pi\rho L}$.
:::
:::

::: part
answer: r*1e-3/2*sqrt(1/((c*1e-9/(100*133.322) + b*1e-9/101325)*pi*1000*L))
unit: Hz
tolerance: 0.5%

### With the bubble?

::: misses
- [any] Add $V_b/P_\text{atm}$ to the compliance.
:::
:::
:::

::: check
kind: parts
given: v = 230 | 280; d = 40 | 55; H = 65 | 85; f = 3 | 5; u = 0.4 | 0.8; t = 45 | 60

## Oxygen uptake {v} mL/min, arteriovenous difference {d} mL/L, heart rate {H}; a {f} MHz Doppler at {t}° on blood at {u} m/s.

::: part
answer: v/d*1000/H
unit: mL
tolerance: 0.5%

### Stroke volume?

::: misses
- [any] Fick output over heart rate.
:::
:::

::: part
answer: 2*f*1e6*u*cos(t*pi/180)/1540
unit: Hz
tolerance: 0.5%

### Doppler shift?

::: misses
- [any] $2f_0v\cos\theta/c$.
:::
:::
:::

::: check
kind: parts
given: m = 4 | 8; A = 40 | 60; c = 0.5 | 1; u = 3 | 5; B = 0.02 | 0.04; D = 20 | 25

## A {m} mg dye bolus: {A} mg·s/L before recirculation, then {c} mg/L falling with τ = {u} s; all of that flow leaves through an aorta {D} mm across, under an electromagnetic probe at {B} T.

::: part
answer: m/(A + c*u)*60
unit: L/min
tolerance: 0.5%

### Flow?

::: misses
- [any] Area including the tail.
:::
:::

::: part
answer: m/(A + c*u)/1000/(pi*(D*1e-3)^2/4)
unit: m/s
tolerance: 0.5%

### Mean velocity in the aorta?

::: misses
- [any] Flow over cross-section.
:::
:::

::: part
answer: B*D*1e-3*m/(A + c*u)/1000/(pi*(D*1e-3)^2/4)*1e6
unit: µV
tolerance: 0.5%

### Probe voltage?

::: misses
- [any] $Bdv$.
:::
:::
:::

::: check
kind: parts
given: a = 0.02 | 0.03; b = 1 | 1.2; c = 0.03 | 0.045; d = 1 | 1.4; H = 12 | 15

## Red AC/DC = {a}/{b}, infrared AC/DC = {c}/{d}; {H} g/dL of hemoglobin and a $P_aO_2$ of 80 mmHg.

::: part
answer: 110 - 25*(a/b)/(c/d)
tolerance: 0.5%

### SpO₂ from the 110 − 25R calibration, in percent?

::: misses
- [any] Find R, then the line.
:::
:::

::: part
answer: 1.34*H*(110 - 25*(a/b)/(c/d))/100 + 0.003*80
unit: mL/dL
tolerance: 0.5%

### Oxygen content at that saturation?

::: misses
- [any] $1.34[\text{Hb}]S + 0.003P$.
:::
:::
:::

::: check
kind: parts
given: m = 0.18 | 0.21; x = 15 | 25; e = 0.005 | 0.01; f = 3 | 5; d = 8 | 14

## X-rays through {x} cm of tissue (μ = {m} cm⁻¹) at a target noise of {e}; {f} MHz ultrasound to {d} cm at 0.5 dB/(cm·MHz).

::: part
answer: 1/e^2/exp(-m*x)
tolerance: 0.5%

### Incident photons per measurement needed?

::: misses
- [any] Detected photons over the transmitted fraction.
:::
:::

::: part
answer: 2*0.5*f*d
unit: dB
tolerance: 0.5%

### Ultrasound round-trip loss?

::: misses
- [any] $2\alpha fd$.
:::
:::
:::

::: check
kind: parts
given: B = 1.5 | 3; G = 10 | 20; W = 1 | 2; R = 500 | 2500; E = 20 | 80

## A {B} T scanner with a {G} mT/m slice gradient and a {W} kHz RF pulse; a spin echo at TR = {R} ms, TE = {E} ms for T1 = 900 ms, T2 = 90 ms.

::: part
answer: 42.577*B
unit: MHz
tolerance: 0.5%

### RF frequency?

::: misses
- [any] $\bar{\gamma}B_0$.
:::
:::

::: part
answer: W*1e3/(42.577e6*G*1e-3)*1000
unit: mm
tolerance: 0.5%

### Slice thickness?

::: misses
- [any] $\Delta f/(\bar{\gamma}G)$.
:::
:::

::: part
answer: (1 - exp(-R/900))*exp(-E/90)
tolerance: 0.5%

### Relative signal?

::: misses
- [any] $(1 - e^{-TR/T_1})e^{-TE/T_2}$.
:::
:::
:::

::: check
kind: parts
given: V = 2 | 3.5; t = 0.3 | 0.5; R = 500 | 700; H = 60 | 80; i = 5 | 8; C = 1 | 1.5

## A pacemaker: {V} V, {t} ms pulses into {R} Ω at {H} per minute, {i} µA of circuit current, a {C} Ah cell.

::: part
answer: V^2*t*1e-3/R*1e6
unit: µJ
tolerance: 0.5%

### Pulse energy?

::: misses
- [any] $V^2t/R$.
:::
:::

::: part
answer: C/((V/R*t*1e-3*H/60 + i*1e-6)*24*365.25)
tolerance: 0.5%

### Battery life in years?

::: misses
- [any] Capacity over average current.
:::
:::
:::

::: check
kind: parts
given: V = 5 | 10; T = 0 | 15; A = 3 | 6; H = 70 | 100; B = 1.7 | 2.1

## Thermodilution: {V} mL of injectate at {T} °C, a curve area of {A} °C·s and K = 1.08, at {H} beats per minute, in a patient of {B} m² body surface area.

::: part
answer: V*(37 - T)*1.08/A*60/1000
unit: L/min
tolerance: 0.5%

### What is the cardiac output?

::: misses
- [any] $V_i(T_b - T_i)K/\int\Delta T\,dt$, converted to L/min.
:::
:::

::: part
answer: V*(37 - T)*1.08/A*60/1000/B
unit: L/min/m^2
tolerance: 0.5%

### What is the cardiac index?

::: misses
- [any] Output over body surface area.
:::
:::

::: part
answer: V*(37 - T)*1.08/A*60/H
unit: mL
tolerance: 0.5%

### What is the stroke volume?

::: misses
- [any] Output over heart rate.
:::
:::
:::

::: check
kind: parts
given: G = 500 | 1000; s = 1 | 2; c = 10 | 50; C = 100 | 120; v = 120 | 230; M = 120 | 150

## An isolated ECG amplifier (gain {G}, CMRR {C} dB, IMRR {M} dB) sees {s} mV of signal, {c} mV of common mode and {v} V across its barrier.

::: part
answer: (c*1e-3/10^(C/20) + v/10^(M/20))/(s*1e-3)*100
unit: %
tolerance: 0.5%

### What fraction of the signal is error?

::: misses
- [any] Sum both input-referred errors, over the signal.
:::
:::

::: part
answer: G*(s*1e-3 + c*1e-3/10^(C/20) + v/10^(M/20))
unit: V
tolerance: 0.5%

### What is the output?

::: misses
- [any] Gain times signal plus both errors.
:::
:::
:::

::: check
kind: parts
given: P = 0.1 | 0.5 | 1; D = 0.6 | 0.85; b = 12 | 16; a = 0.008 | 0.012; c = 0.015 | 0.02

## An oximeter's infrared perfusion index is {P}% with the DC at {D} of a {b}-bit range; its red and infrared AC/DC ratios are {a} and {c}, calibrated as 110 − 25R.

::: part
answer: P/100*D*2^b
unit: counts
tolerance: 0.5%

### How many counts does the infrared pulse span?

::: misses
- [any] AC/DC × DC counts.
:::
:::

::: part
answer: 110 - 25*a/c
unit: %
tolerance: 0.5%

### What saturation does it display?

::: misses
- [110 - 25*c/a] R is red over infrared.
- [any] 110 − 25R.
:::
:::
:::
:::