title: "Problem set 2: sensors, dynamics and temperature"
heading: "Problem set 2: calibration lines and nonlinearity, loading and drift, first- and second-order responses, bridges and their compensation, capacitive and piezoelectric sensors, demodulation and lock-in noise, photodiodes, thermistors, RTDs, thermocouples and radiation thermometry"
eyebrow: S15 · Problem set 2
type: problems
---

::: hook
These problems cover lessons 1 to 8: characterizing a sensor statically and dynamically, conditioning its signal in bridges, charge amplifiers and demodulators, and measuring temperature five different ways. Several take one sensor from its calibration through its dynamics to the circuit that reads it.
:::

::: pool
pick: 12

::: check
kind: parts
given: a = 0.3 | 0.6; b = 2.7 | 2.95; c = 5.0 | 5.3

## A transducer reads {a}, {b} and {c} mV at 0, 50 and 100 mmHg.

::: part
answer: (c - a)/100
unit: mV
tolerance: 0.5%

### What sensitivity, in mV per mmHg, does the least-squares line give?

::: misses
- [c/100] The slope comes from the spread of the readings, not the last one alone.
- [any] For equally spaced inputs the slope is (c − a)/100.
:::
:::

::: part
answer: abs(2*b - a - c)/(3*(c - a))*100
unit: %
tolerance: 0.5%

### What nonlinearity, as a percent of span, does the line leave?

::: misses
- [any] The largest deviation from the fitted line over the span.
:::
:::
:::

::: check
kind: parts
given: s = 5 | 50; Z = 300 | 1000 | 3000; k = 0.001 | 0.005; T = 10 | 25

## A sensor of {s} µV/unit and output impedance {Z} Ω drives a 1 MΩ input; its sensitivity drifts by {k} per °C, and the room warms {T} °C.

::: part
answer: s*1e6/(1e6 + Z)
unit: µV
tolerance: 0.5%

### What sensitivity per unit does the amplifier see?

::: misses
- [any] The loading divider keeps $Z_i/(Z_s + Z_i)$.
:::
:::

::: part
answer: k*T*100
unit: %
tolerance: 0.5%

### By what percent does the drift change every reading?

::: misses
- [any] Drift per degree times degrees.
:::
:::
:::

::: check
kind: parts
given: u = 5 | 15 | 40; r = 0.2 | 1; T = 1 | 3

## A temperature probe has a {u} s time constant.

::: part
answer: r*u/60
unit: °C
tolerance: 0.5%

### How far behind does it lag a temperature rising {r} °C per minute?

::: misses
- [r*u] Convert the rate to degrees per second.
- [any] Rate times time constant.
:::
:::

::: part
answer: u*ln(T/0.05)
unit: s
tolerance: 0.5%

### How long does it take to come within 0.05 °C after a {T} °C step?

::: misses
- [any] $\tau\ln(T/0.05)$.
:::
:::
:::

::: check
kind: parts
given: d = 0.2 | 0.4 | 0.6; P = 20 | 40 | 60; H = 70 | 120

## A flush test's peaks shrink to {d} of the one before, {P} ms apart; the patient's heart rate is {H} per minute.

::: part
answer: -ln(d)/sqrt(4*pi^2 + ln(d)^2)
tolerance: 0.5%

### What is the damping ratio?

::: misses
- [-ln(d)/sqrt(pi^2 + ln(d)^2)] For peaks a full period apart, the denominator has 4π².
- [any] $-\ln d/\sqrt{4\pi^2 + \ln^2 d}$.
:::
:::

::: part
answer: 1/sqrt((1 - (10*H/60/(1000/P/sqrt(1 - ln(d)^2/(4*pi^2 + ln(d)^2))))^2)^2 + (2*(-ln(d)/sqrt(4*pi^2 + ln(d)^2))*10*H/60/(1000/P/sqrt(1 - ln(d)^2/(4*pi^2 + ln(d)^2))))^2)
tolerance: 0.5%

### By what factor does the system scale the pressure wave's tenth harmonic?

::: misses
- [any] Find $f_n = f_d/\sqrt{1 - \zeta^2}$, then the second-order amplitude ratio at 10H/60 Hz.
:::
:::
:::

::: check
kind: parts
given: G = 2 | 2.1 | 100; e = 200 | 1000; V = 2.5 | 10

## Four gauges (G = {G}) at ±{e} microstrain form a full bridge excited at {V} V; separately, one such gauge forms a quarter bridge.

::: part
answer: V*G*e*1e-6*1e3
unit: mV
tolerance: 0.5%

### What does the full bridge output?

::: misses
- [V*G*e*1e-6/4*1e3] That's a quarter bridge; four active arms give four times as much.
- [any] $VG\epsilon$.
:::
:::

::: part
answer: G*e*1e-6/2*100
unit: %
tolerance: 0.5%

### By what percent does the quarter bridge's linear formula overstate its output?

::: misses
- [any] About x/2, with x = Gε.
:::
:::
:::

::: check
kind: parts
given: G = 2 | 2.1; e = 100 | 500; a = 10 | 50; T = 5 | 30; V = 5 | 10

## An active gauge (G = {G}) at {e} microstrain and an identical unstrained dummy gauge, both with a temperature coefficient of {a} ppm/°C, form a half bridge excited at {V} V; both warm by {T} °C.

::: part
answer: V*G*e*1e-6/4*1e3
unit: mV
tolerance: 1%

### About what does the bridge output?

::: misses
- [V*(G*e*1e-6 + a*1e-6*T)/4*1e3] The dummy's identical temperature change cancels the active gauge's.
- [any] The strain alone, as for a quarter bridge.
:::
:::

::: part
answer: a*T/G
unit: microstrain
tolerance: 0.5%

### What apparent strain would a single uncompensated gauge have shown?

::: misses
- [any] $\alpha\Delta T/G$.
:::
:::
:::

::: check
kind: parts
given: A = 1 | 2; d = 25 | 100; x = 0.5 | 2; r = 1 | 3

## A capacitive pressure sensor's diaphragm, {A} cm² with relative permittivity {r} in the gap, sits {d} µm from a fixed plate and deflects {x} µm toward it.

::: part
answer: 8.854e-12*r*A*1e-4/(d*1e-6)*1e12
unit: pF
tolerance: 0.5%

### What is its capacitance at rest?

::: misses
- [any] $\epsilon_0\epsilon_rA/d$.
:::
:::

::: part
answer: 8.854e-12*r*A*1e-4/((d - x)*1e-6)*1e12 - 8.854e-12*r*A*1e-4/(d*1e-6)*1e12
unit: pF
tolerance: 0.5%

### By how much does the capacitance rise?

::: misses
- [any] $C(d - x) - C(d)$.
:::
:::
:::

::: check
kind: parts
given: k = 2.3 | 100 | 300; F = 1 | 20; C = 10 | 100 | 1000; R = 1 | 100

## A piezoelectric sensor (d = {k} pC/N) under {F} N feeds a charge amplifier with $C_f$ = {C} pF and $R_f$ = {R} GΩ.

::: part
answer: k*F/C
unit: V
tolerance: 0.5%

### What step output does it give?

::: misses
- [any] $dF/C_f$.
:::
:::

::: part
answer: 1/(2*pi*R*1e9*C*1e-12)
unit: Hz
tolerance: 0.5%

### What is its low-frequency corner?

::: misses
- [any] $1/(2\pi R_fC_f)$.
:::
:::
:::

::: check
kind: parts
given: A = 20 | 100; p = 30 | 45 | 80; f = 1 | 10; C = 10 | 100

## An LVDT's carrier reaches a phase-sensitive demodulator at {A} mV and {p}° from the reference; the output filter's corner must be {f} Hz with a {C} nF capacitor.

::: part
answer: 2/pi*A*cos(p*pi/180)
unit: mV
tolerance: 0.5%

### What average output does the demodulator give?

::: misses
- [2/pi*A] Only the in-phase component survives.
- [any] $(2/\pi)A\cos\varphi$.
:::
:::

::: part
answer: 1/(2*pi*f*C*1e-9)/1e3
unit: kΩ
tolerance: 0.5%

### What filter resistor does that corner need?

::: misses
- [any] $R = 1/(2\pi f_cC)$.
:::
:::
:::

::: check
kind: parts
given: n = 0.5 | 5 | 50; t = 0.3 | 3; s = 0.2 | 2

## A lock-in amplifier with a single-pole output filter of time constant {t} s sees noise of {n} nV/√Hz and an in-phase signal of {s} µV.

::: part
answer: n*sqrt(1/(4*t))
unit: nV
tolerance: 0.5%

### What noise remains at the output?

::: misses
- [any] $n\sqrt{1/(4\tau)}$.
:::
:::

::: part
answer: s*1000/(n*sqrt(1/(4*t)))
tolerance: 0.5%

### What signal-to-noise ratio (amplitude) does that give?

::: misses
- [s/(n*sqrt(1/(4*t)))] Convert microvolts to nanovolts.
- [any] Signal over output noise.
:::
:::

::: part
answer: 3*t
unit: s
tolerance: 0.5%

### About how long must it wait after a change to settle within 5%?

::: misses
- [any] Three time constants.
:::
:::
:::

::: check
kind: parts
given: e = 0.6 | 0.85; P = 1 | 50; l = 660 | 940; B = 100 | 10000

## A photodiode with quantum efficiency {e} receives {P} µW at {l} nm, over a {B} Hz bandwidth.

::: part
answer: e*l/1239.84*P
unit: µA
tolerance: 0.5%

### What photocurrent does it produce?

::: misses
- [any] Responsivity $\eta\lambda/1239.84$ A/W times power.
:::
:::

::: part
answer: sqrt(2*1.602176634e-19*e*l/1239.84*P*1e-6*B)*1e12
unit: pA
tolerance: 0.5%

### How much shot noise does it carry?

::: misses
- [any] $\sqrt{2qIB}$.
:::
:::
:::

::: check
kind: parts
given: R = 5 | 10 | 30; B = 3400 | 3900; T = 35 | 42

## A thermistor is {R} kΩ at 25 °C with B = {B} K.

::: part
answer: R*exp(B*(1/(T + 273.15) - 1/298.15))
unit: kΩ
tolerance: 0.5%

### What is it at {T} °C?

::: misses
- [any] $R_{25}e^{B(1/T - 1/298.15)}$.
:::
:::

::: part
answer: -B/(T + 273.15)^2*100
unit: %
tolerance: 0.5%

### What is its temperature coefficient there, per °C?

::: misses
- [B/(T + 273.15)^2*100] The coefficient is negative.
- [any] $-B/T^2$.
:::
:::
:::

::: check
kind: parts
given: R = 100 | 1000; T = 37 | 100; I = 1 | 5; d = 2 | 10

## A platinum RTD with $R_0$ = {R} Ω (α = 0.00385/°C) at {T} °C carries {I} mA, and its dissipation constant is {d} mW/°C.

::: part
answer: R*(1 + 0.00385*T)
unit: Ω
tolerance: 0.5%

### What is its resistance?

::: misses
- [any] $R_0(1 + \alpha T)$.
:::
:::

::: part
answer: (I*1e-3)^2*R*(1 + 0.00385*T)*1e3/d
unit: °C
tolerance: 0.5%

### How much does self-heating raise its reading?

::: misses
- [any] $I^2R/\delta$.
:::
:::
:::

::: check
kind: parts
given: S = 40.6 | 52; V = 0.5 | 1.2; c = 20 | 25

## A thermocouple with a Seebeck coefficient of {S} µV/°C reads {V} mV, with its reference junction at {c} °C.

::: part
answer: V*1000/S
unit: °C
tolerance: 0.5%

### What temperature difference does that show?

::: misses
- [any] Voltage over Seebeck coefficient.
:::
:::

::: part
answer: c + V*1000/S
unit: °C
tolerance: 0.5%

### What is the measuring junction's temperature?

::: misses
- [V*1000/S] Add the reference junction's temperature.
- [any] Cold-junction compensation adds it back.
:::
:::
:::

::: check
kind: parts
given: e = 0.95 | 0.98; T = 36 | 38; a = 20 | 25; A = 1 | 3

## An infrared thermometer views {A} cm² of skin (emissivity {e}) at {T} °C, in a {a} °C room.

::: part
answer: e*5.670374419e-8*A*1e-4*((T + 273.15)^4 - (a + 273.15)^4)*1e3
unit: mW
tolerance: 0.5%

### What net power does the skin radiate?

::: misses
- [e*5.670374419e-8*A*1e-4*(T + 273.15)^4*1e3] It also absorbs radiation from the room.
- [any] $\epsilon\sigma A(T^4 - T_a^4)$.
:::
:::

::: part
answer: 4*(T + 273.15)^3/((T + 273.15)^4 - (a + 273.15)^4)*100
unit: %
tolerance: 0.5%

### By what percent does that power change per °C of skin temperature?

::: misses
- [4/(T + 273.15)*100] That ignores the room's radiation; the net power changes faster.
- [any] $4T^3/(T^4 - T_a^4)$.
:::
:::
:::
:::