title: "Problem set 6: pressure, flow and plethysmography"
heading: "Problem set 6: catheter natural frequency and damping, bubbles and flush tests, leveling, cuff methods and mean pressure, Fick and indicator dilution, thermodilution, electromagnetic, transit-time and Doppler flowmeters, and plethysmography"
eyebrow: S15 · Problem set 6
type: problems
---

::: hook
These problems cover lessons 23, 24 and 26 to 28: catheter dynamics from dimensions, the effect of a bubble, damping from a flush test, hydrostatic errors, mean pressure and oscillometric ratios, deflation errors, Fick and Stewart–Hamilton outputs with their tails, thermodilution, stroke volume, flow from magnetic induction, transit times and Doppler shifts, and volumes from venous occlusion and impedance plethysmography.
:::

::: pool
pick: 12

::: check
kind: parts
given: r = 0.4 | 0.7; L = 0.8 | 1.5; c = 0.008 | 0.03

## A saline catheter of radius {r} mm and length {L} m on a transducer of compliance {c} mm³ per 100 mmHg.

::: part
answer: r*1e-3/2*sqrt(1/((c*1e-9/(100*133.322))*pi*1000*L))
unit: Hz
tolerance: 0.5%

### Natural frequency?

::: misses
- [any] $\frac{r}{2}\sqrt{E_d/\pi\rho L}$.
:::
:::

::: part
answer: 4*1e-3/(r*1e-3)^3*sqrt(L*(c*1e-9/(100*133.322))/(pi*1000))
tolerance: 0.5%

### Damping ratio (η = 0.001 Pa·s)?

::: misses
- [any] $\frac{4\eta}{r^3}\sqrt{L/\pi\rho E_d}$.
:::
:::
:::

::: check
kind: parts
given: f = 100 | 200; t = 5 | 10; b = 0.5 | 2; z = 0.03 | 0.06

## A catheter system with $f_n$ = {f} Hz, ζ = {z} and transducer compliance {t} × 10⁻¹⁶ m³/Pa gains a {b} mm³ bubble.

::: part
answer: f*sqrt((t*1e-16)/(t*1e-16 + b*1e-9/101325))
unit: Hz
tolerance: 0.5%

### New natural frequency?

::: misses
- [any] Scales as the square root of stiffness.
:::
:::

::: part
answer: z*sqrt((t*1e-16 + b*1e-9/101325)/(t*1e-16))
tolerance: 0.5%

### New damping ratio?

::: misses
- [any] Scales inversely with the square root of stiffness.
:::
:::
:::

::: check
kind: parts
given: d = 0.15 | 0.4; T = 30 | 60; h = 8 | 15

## A flush test gives a peak ratio of {d} with peaks {T} ms apart; the transducer sits {h} cm above the heart (saline column).

::: part
answer: -ln(d)/sqrt(4*pi^2 + ln(d)^2)
tolerance: 0.5%

### Damping ratio?

::: misses
- [any] $-\ln d/\sqrt{4\pi^2 + \ln^2d}$.
:::
:::

::: part
answer: 1000/T/sqrt(1 - ln(d)^2/(4*pi^2 + ln(d)^2))
unit: Hz
tolerance: 0.5%

### Natural frequency?

::: misses
- [any] $f_d/\sqrt{1 - \zeta^2}$.
:::
:::

::: part
answer: 1000*9.81*h/100/133.322
unit: mmHg
tolerance: 0.5%

### Size of the leveling error?

::: misses
- [any] $\rho gh$.
:::
:::
:::

::: check
kind: parts
given: s = 115 | 150; d = 70 | 90; r = 2 | 4; H = 60 | 90

## A pressure of {s}/{d} mmHg measured with a cuff deflating at {r} mmHg/s, heart rate {H} per minute.

::: part
answer: d + (s - d)/3
unit: mmHg
tolerance: 0.5%

### Mean arterial pressure?

::: misses
- [any] DBP + (SBP − DBP)/3.
:::
:::

::: part
answer: r*60/H
unit: mmHg
tolerance: 0.5%

### Deflation between beats?

::: misses
- [any] $r \times 60/H$.
:::
:::
:::

::: check
kind: parts
given: M = 85 | 105; w = 20 | 40; k = 0.5 | 0.6; q = 0.75 | 0.85

## An oscillometric envelope $e^{-((P - M)/w)^2}$ with M = {M} mmHg and w = {w} mmHg.

::: part
answer: M + w*sqrt(-ln(k))
unit: mmHg
tolerance: 0.5%

### Systolic estimate at ratio {k}?

::: misses
- [any] $M + w\sqrt{-\ln k}$.
:::
:::

::: part
answer: M - w*sqrt(-ln(q))
unit: mmHg
tolerance: 0.5%

### Diastolic estimate at ratio {q}?

::: misses
- [any] $M - w\sqrt{-\ln q}$.
:::
:::
:::

::: check
kind: parts
given: v = 220 | 320; d = 35 | 60; H = 65 | 95; B = 1.7 | 2.1

## Oxygen uptake {v} mL/min with an arteriovenous difference of {d} mL/L; heart rate {H}; body surface area {B} m².

::: part
answer: v/d
unit: L/min
tolerance: 0.5%

### Cardiac output?

::: misses
- [any] Fick.
:::
:::

::: part
answer: v/d*1000/H
unit: mL
tolerance: 0.5%

### Stroke volume?

::: misses
- [any] CO/HR.
:::
:::

::: part
answer: v/d/B
unit: L/min/m²
tolerance: 0.5%

### Cardiac index in L/min/m²?

::: misses
- [any] CO/BSA.
:::
:::
:::

::: check
kind: parts
given: m = 3 | 8; A = 35 | 70; c = 0.4 | 1; u = 3 | 5

## A {m} mg dye bolus; the first pass accumulates {A} mg·s/L before recirculation, where the concentration is {c} mg/L falling with τ = {u} s.

::: part
answer: m/(A + c*u)*60
unit: L/min
tolerance: 0.5%

### Flow with the tail included?

::: misses
- [any] Area + $c_1\tau$.
:::
:::

::: part
answer: 100*((m/A) - m/(A + c*u))/(m/(A + c*u))
tolerance: 0.5%

### Percent error from ignoring the tail?

::: misses
- [any] Compare the two flows.
:::
:::
:::

::: check
kind: parts
given: V = 5 | 10; T = 0 | 22; A = 2.5 | 6

## {V} mL of injectate at {T} °C into blood at 37 °C; curve area {A} °C·s; K = 1.08.

::: part
answer: V*1e-3*(37 - T)*1.08/A*60
unit: L/min
tolerance: 0.5%

### Cardiac output?

::: misses
- [any] Thermodilution.
:::
:::

::: part
answer: V*1e-3*(37 - T)*1.08*60/5
unit: °C·s
tolerance: 0.5%

### Curve area, in °C·s, that the same injectate would give at 5 L/min?

::: misses
- [any] Solve for the area.
:::
:::
:::

::: check
kind: parts
given: B = 0.02 | 0.05; d = 4 | 12; v = 0.2 | 0.8

## An electromagnetic probe applies {B} T across a {d} mm vessel with mean velocity {v} m/s.

::: part
answer: B*d*1e-3*v*1e6
unit: µV
tolerance: 0.5%

### Voltage?

::: misses
- [any] $Bdv$.
:::
:::

::: part
answer: v*pi*(d*1e-3)^2/4*60000
unit: L/min
tolerance: 0.5%

### Volume flow?

::: misses
- [any] $\bar{v}\pi d^2/4$.
:::
:::
:::

::: check
kind: parts
given: L = 8 | 15; v = 0.3 | 0.9; t = 30 | 45

## A transit-time path of {L} mm at {t}° to blood at {v} m/s.

::: part
answer: 2*L*1e-3*v*cos(t*pi/180)/1540^2*1e9
unit: ns
tolerance: 0.5%

### Time difference?

::: misses
- [any] $2Lv\cos\theta/c^2$.
:::
:::

::: part
answer: L*1e-3/1540*1e6
unit: µs
tolerance: 0.5%

### Travel time with no flow?

::: misses
- [any] L/c.
:::
:::
:::

::: check
kind: parts
given: f = 3 | 7; v = 0.3 | 1.2; t = 45 | 60

## A {f} MHz Doppler at {t}° to blood moving at {v} m/s.

::: part
answer: 2*f*1e6*v*cos(t*pi/180)/1540
unit: Hz
tolerance: 0.5%

### Shift?

::: misses
- [any] $2f_0v\cos\theta/c$.
:::
:::

::: part
answer: 100*(cos((t - 5)*pi/180)/cos(t*pi/180) - 1)
tolerance: 0.5%

### Percent velocity error if the true angle is 5° smaller than assumed?

::: misses
- [any] Ratio of cosines.
:::
:::
:::

::: check
kind: parts
given: f = 2 | 5; d = 5 | 12; t = 30 | 60

## A {f} MHz pulsed Doppler at a depth of {d} cm and {t}°.

::: part
answer: 1540/(2*d*1e-2)
unit: Hz
tolerance: 0.5%

### Highest pulse repetition frequency?

::: misses
- [any] c/2d.
:::
:::

::: part
answer: 1540^2/(8*f*1e6*d*1e-2*cos(t*pi/180))
unit: m/s
tolerance: 0.5%

### Highest unaliased velocity?

::: misses
- [any] $c^2/(8f_0d\cos\theta)$.
:::
:::
:::

::: check
kind: parts
given: r = 1.4 | 1.6; L = 8 | 12; Z = 25 | 40; d = 0.05 | 0.12; I = 1 | 2

## A four-electrode impedance plethysmograph drives {I} mA at 100 kHz through a {L} cm limb segment of {Z} Ω. Each pulse lowers the impedance by {d} Ω; blood resistivity is {r} Ω·m.

::: part
answer: I*Z
unit: mV
tolerance: 0.5%

### Steady voltage across the sensing electrodes?

::: misses
- [any] IZ.
:::
:::

::: part
answer: I*d
unit: mV
tolerance: 0.5%

### Pulsatile signal?

::: misses
- [any] I ΔZ.
:::
:::

::: part
answer: r*(L/100)^2*d/Z^2*1e6
unit: mL
tolerance: 0.5%

### Volume added by each pulse?

::: misses
- [any] $\rho_bL^2|\Delta Z|/Z^2$.
:::
:::
:::

::: check
kind: parts
given: s = 0.3 | 0.6 | 0.9; v = 900 | 1200; t = 2 | 6

## In venous-occlusion plethysmography a {v} mL calf segment swells at {s} mL/s after the cuff inflates. After release, the extra volume drains with a time constant of {t} s.

::: part
answer: s*60/(v/100)
tolerance: 0.5%

### Arterial inflow, in mL/min per 100 mL of tissue?

::: misses
- [any] Slope in mL/min divided by the number of 100 mL portions.
:::
:::

::: part
answer: 3*t
unit: s
tolerance: 0.5%

### About how long does 95% of the extra volume take to drain?

::: misses
- [any] Three time constants.
:::
:::
:::

::: check
kind: parts
given: h = 3 | 5 | 10; P = 5 | 8

## A transducer sits {h} cm below the right atrium, connected by saline (1000 kg/m³).

::: part
answer: 1000*9.81*h*1e-2/133.322
unit: mmHg
tolerance: 0.5%

### What offset does that add?

::: misses
- [any] $\rho gh$ in mmHg.
:::
:::

::: part
answer: 1000*9.81*h*1e-2/133.322/P*100
unit: %
tolerance: 0.5%

### What percent error is that on a central venous pressure of {P} mmHg?

::: misses
- [any] Offset over the venous pressure.
:::
:::
:::
:::