title: "Problem set 8: respiration, oximetry, chemical sensors, the laboratory and X-ray imaging"
heading: "Problem set 8: BTPS corrections, helium dilution and body plethysmography, FEV1 and alveolar ventilation, ratio of ratios and oxygen content, respiratory mechanics, Nernstian and amperometric sensors, spectrophotometry and blood counts, image quality, attenuation, Hounsfield units and photon noise"
eyebrow: S15 · Problem set 8
type: problems
---

::: hook
These problems cover lessons 31 to 38: gas-volume corrections, lung volumes by dilution and Boyle's law, forced expiration and ventilation, end-tidal CO₂, compliance, resistance and forced oscillation, oximeter ratios and calibrations, oxygen content, pH and ion-selective electrodes, Henderson–Hasselbalch, Faraday currents and enzyme saturation, absorbance and red-cell indices, photon-limited detection and MTF, and X-ray attenuation, contrast, CT numbers and dose.
:::

::: pool
pick: 12

::: check
kind: parts
given: T = 21 | 25; W = 18.7 | 23.8; P = 720 | 760; V = 3 | 5

## A spirometer at {T} °C (water vapor {W} mmHg), barometric pressure {P} mmHg, records {V} L.

::: part
answer: 310/(273.15 + T)*(P - W)/(P - 47)
tolerance: 0.5%

### BTPS factor?

::: misses
- [any] Temperature ratio times dry-gas pressure ratio.
:::
:::

::: part
answer: V*310/(273.15 + T)*(P - W)/(P - 47)
unit: L
tolerance: 0.5%

### Volume in the lungs?

::: misses
- [any] Multiply by the factor.
:::
:::
:::

::: check
kind: parts
given: V = 4 | 8; a = 9 | 12; b = 5.5 | 7; P = 960 | 1000; v = 25 | 50; p = 9 | 14

## Helium in a {V} L spirometer falls from {a}% to {b}%; a body box sees {v} mL of compression and {p} cmH₂O rise from {P} cmH₂O.

::: part
answer: V*(a - b)/b
unit: L
tolerance: 0.5%

### FRC by helium dilution?

::: misses
- [any] $V_1(C_1 - C_2)/C_2$.
:::
:::

::: part
answer: P*v/p
unit: mL
tolerance: 0.5%

### Volume by plethysmography?

::: misses
- [any] $P\Delta V/\Delta P$.
:::
:::
:::

::: check
kind: parts
given: F = 3 | 5; u = 0.35 | 1.1; V = 450 | 650; D = 130 | 170; f = 12 | 18

## A forced expiration with FVC = {F} L and τ = {u} s; quiet breathing at {V} mL, {D} mL dead space, {f} per minute.

::: part
answer: F*(1 - exp(-1/u))
unit: L
tolerance: 0.5%

### FEV1?

::: misses
- [any] $\text{FVC}(1 - e^{-1/\tau})$.
:::
:::

::: part
answer: (V - D)*f/1000
unit: L/min
tolerance: 0.5%

### Alveolar ventilation?

::: misses
- [any] $(V_T - V_D)f$.
:::
:::

::: part
answer: V*f/1000
unit: L/min
tolerance: 0.5%

### Minute ventilation?

::: misses
- [any] $V_Tf$.
:::
:::
:::

::: check
kind: parts
given: c = 4.5 | 6; P = 520 | 650 | 760

## End-tidal gas with {c}% CO₂ (of the dry gas) at a barometric pressure of {P} mmHg.

::: part
answer: c/100*(P - 47)
unit: mmHg
tolerance: 0.5%

### End-tidal CO₂ partial pressure?

::: misses
- [any] $F(P_B - 47)$.
:::
:::

::: part
answer: 100*40/(P - 47)
tolerance: 0.5%

### What CO₂ percentage would correspond to 40 mmHg at this pressure?

::: misses
- [any] 40 mmHg over the dry-gas pressure.
:::
:::
:::

::: check
kind: parts
given: a = 0.015 | 0.03; b = 0.9 | 1.3; c = 0.025 | 0.04; d = 1 | 1.4

## Red AC/DC = {a}/{b}; infrared AC/DC = {c}/{d}.

::: part
answer: (a/b)/(c/d)
tolerance: 0.5%

### Ratio of ratios?

::: misses
- [any] $(AC_r/DC_r)/(AC_{ir}/DC_{ir})$.
:::
:::

::: part
answer: 110 - 25*(a/b)/(c/d)
tolerance: 0.5%

### Displayed SpO₂ with the 110 − 25R calibration, in percent?

::: misses
- [any] 110 − 25R.
:::
:::
:::

::: check
kind: parts
given: S = 75 | 92; H = 10 | 15; s = 0.85 | 0.97; P = 60 | 95

## Saturation {S}% in the Beer–Lambert model (coefficients 3227/320 at 660 nm, 693/1214 at 940 nm); blood with {H} g/dL Hb at saturation {s} and $P_aO_2$ = {P} mmHg.

::: part
answer: (3227*(1 - S/100) + 320*S/100)/(693*(1 - S/100) + 1214*S/100)
tolerance: 0.5%

### Theoretical R?

::: misses
- [any] Coefficient-weighted ratio.
:::
:::

::: part
answer: 1.34*H*s + 0.003*P
unit: mL/dL
tolerance: 0.5%

### Oxygen content?

::: misses
- [any] $1.34[\text{Hb}]S + 0.003P$.
:::
:::
:::

::: check
kind: parts
given: T = 25 | 37; d = 0.2 | 0.5; b = 15 | 28; p = 30 | 55

## A pH electrode at {T} °C; blood with {b} mM bicarbonate and $P_{CO_2}$ = {p} mmHg.

::: part
answer: 8.314462618*(T + 273.15)/96485.33212*ln(10)*d*1000
unit: mV
tolerance: 0.5%

### Voltage change for {d} pH units?

::: misses
- [any] $2.303(RT/F)\Delta\text{pH}$.
:::
:::

::: part
answer: 6.1 + log(b/(0.0307*p))
tolerance: 0.1%

### Blood pH?

::: misses
- [any] Henderson–Hasselbalch.
:::
:::
:::

::: check
kind: parts
given: a = 3 | 5; K = 0.0005 | 0.005; n = 130 | 145; f = 1 | 5; e = 2 | 4

## A K⁺ electrode with $K_{K,Na}$ = {K} in {a} mM K⁺ and {n} mM Na⁺; an amperometric sensor reacting {f} pmol/s with {e} electrons each.

::: part
answer: 100*K*n/a
tolerance: 0.5%

### Percent overreading?

::: misses
- [any] $K_{ij}a_j/a_i$.
:::
:::

::: part
answer: e*96485.33212*f*1e-12*1e9
unit: nA
tolerance: 0.5%

### Sensor current?

::: misses
- [any] $nF \times$ rate.
:::
:::
:::

::: check
kind: parts
given: m = 60 | 150; K = 12 | 30; G = 5 | 12

## A glucose sensor with $i_\max$ = {m} nA and $K_m$ = {K} mM at {G} mM.

::: part
answer: m*G/(K + G)
unit: nA
tolerance: 0.5%

### Current?

::: misses
- [any] Michaelis–Menten.
:::
:::

::: part
answer: G*K/(K + G)
unit: mM
tolerance: 0.5%

### What glucose does a meter using the initial linear slope report?

::: misses
- [any] Current divided by $i_\max/K_m$.
:::
:::

::: part
answer: 100*G/(K + G)
tolerance: 0.5%

### Percent low reading of a linear calibration?

::: misses
- [any] $G/(K_m + G)$.
:::
:::
:::

::: check
kind: parts
given: m = 0.16 | 0.22; x = 12 | 30; b = 0.45 | 0.7; t = 1 | 4

## Soft tissue with μ = {m} cm⁻¹, {x} cm thick; a bone of μ = {b} cm⁻¹ replacing {t} cm of it.

::: part
answer: exp(-m*x)
tolerance: 0.5%

### Transmitted fraction through the tissue?

::: misses
- [any] $e^{-\mu x}$.
:::
:::

::: part
answer: ln(2)/m
unit: cm
tolerance: 0.5%

### Half-value layer of the tissue?

::: misses
- [any] ln 2/μ.
:::
:::

::: part
answer: 1 - exp(-(b - m)*t)
tolerance: 0.5%

### Bone contrast?

::: misses
- [any] $1 - e^{-(\mu_b - \mu_s)t}$.
:::
:::
:::

::: check
kind: parts
given: m = 0.15 | 0.26 | 0.45; w = 0.19 | 0.2; e = 0.004 | 0.02

## A voxel with μ = {m} cm⁻¹ against water's {w}; a target relative noise of {e}.

::: part
answer: 1000*(m - w)/w
tolerance: 0.5%

### CT number?

::: misses
- [any] $1000(\mu - \mu_w)/\mu_w$.
:::
:::

::: part
answer: 1/e^2
tolerance: 0.5%

### Detected photons needed?

::: misses
- [any] $1/\epsilon^2$.
:::
:::
:::

::: check
kind: parts
given: k = 80 | 120; N = 256 | 1024; r = 0.01 | 0.05

## A {k} kV X-ray tube; a CT detector with {N} bins; a projection that transmits {r} of the beam.

::: part
answer: 1.23984/k
unit: nm
tolerance: 0.5%

### Shortest wavelength?

::: misses
- [any] hc/E = 1.24 nm·keV/E.
:::
:::

::: part
answer: pi/2*N
tolerance: 0.5%

### Views needed?

::: misses
- [any] About (π/2)N.
:::
:::

::: part
answer: ln(1/r)
tolerance: 0.5%

### Projection value ∫μ dl?

::: misses
- [any] $\ln(I_0/I)$.
:::
:::
:::

::: check
kind: parts
given: a = 0.15 | 0.2; b = 0.15 | 0.25; R = 2 | 4 | 8; I = 0.01 | 0.02

## A patient's lung compliance is {a} L/cmH₂O and chest-wall compliance {b} L/cmH₂O, with airway resistance {R} cmH₂O·s/L and inertance {I} cmH₂O·s²/L.

::: part
answer: a*b/(a + b)
tolerance: 0.5%

### Total compliance, in L/cmH₂O?

::: misses
- [any] Series combination.
:::
:::

::: part
answer: R*a*b/(a + b)
unit: s
tolerance: 0.5%

### Time constant of passive exhalation?

::: misses
- [any] R times the total compliance.
:::
:::

::: part
answer: 1/(2*pi*sqrt(I*a*b/(a + b)))
unit: Hz
tolerance: 0.5%

### Resonant frequency in a forced-oscillation test?

::: misses
- [any] $1/(2\pi\sqrt{IC})$ with the total compliance.
:::
:::
:::

::: check
kind: parts
given: C = 100 | 200; s = 30 | 45; u = 50 | 65; H = 38 | 44; R = 4.2 | 5

## A cholesterol standard of {C} mg/dL reads {s}% transmittance and a patient's serum {u}%. The same patient's blood count shows hematocrit {H}% and {R} million red cells per µL.

::: part
answer: C*(2 - log(u))/(2 - log(s))
tolerance: 0.5%

### The patient's cholesterol, in mg/dL?

::: misses
- [any] $C_sA_u/A_s$ with $A = 2 - \log\%T$.
:::
:::

::: part
answer: 10*H/R
unit: fL
tolerance: 0.5%

### MCV, in fL?

::: misses
- [any] 10 HCT/RBC.
:::
:::

::: part
answer: 100*0.4343*0.005*10^(2 - log(u))/(2 - log(u))
unit: %
tolerance: 1%

### Relative concentration error from a 0.5% transmittance error at the patient's reading?

::: misses
- [any] $0.4343\,\Delta T/(TA)$.
:::
:::
:::

::: check
kind: parts
given: N = 1000 | 4000 | 16000; C = 0.02 | 0.05; J = 4 | 5

## An X-ray detector pixel averages {N} photons. A detail has contrast {C}, and detection needs an SNR of J = {J}.

::: part
answer: C*sqrt(N)
tolerance: 0.5%

### The detail's SNR at this exposure?

::: misses
- [any] $C\sqrt N$.
:::
:::

::: part
answer: (J/C)^2
tolerance: 0.5%

### Photons per pixel needed to detect it?

::: misses
- [any] $(J/C)^2$.
:::
:::

::: part
answer: (J/C)^2/N
tolerance: 0.5%

### By what factor must the dose rise?

::: misses
- [any] The ratio of photons needed to photons available.
:::
:::
:::
:::