title: Respiratory measurements
heading: Lung volumes and capacities, spirometry and the forced expiration with FEV1 over FVC, correcting gas volumes to body conditions, pneumotachographs, measuring the volume a spirometer can't reach by helium dilution and body plethysmography, and ventilation and end-tidal carbon dioxide
eyebrow: S15 · Lesson 31 · Respiratory measurements
xp: 50
---

::: hook
Breathing is easy to see and surprisingly hard to measure well. The volume a patient blows out depends on the temperature and humidity of the gas, so the same breath measures differently in the lungs and in the instrument. Some air can never be blown out at all, so measuring total lung capacity takes a gas-mixing trick or a sealed box and Boyle's law. And the single most used lung test, the forced expiration, reduces to a ratio of two volumes that says whether the airways are narrowed.
:::

::: play
- A simple model of a forced expiration: the exhaled volume rises as $V(t) = \text{FVC}(1 - e^{-t/\tau})$, where the time constant τ is airway resistance times lung compliance.
- **FEV1**, the volume out in the first second, over the forced vital capacity: $1 - e^{-1/\tau}$. Narrowed airways (asthma, COPD) raise τ and lower the ratio, below 0.7 by the usual criterion.
- Restrictive disease shrinks FVC but leaves the ratio normal or high.
:::

```sim plot
view: graph
x: t = 0..6
y: V = 0..6
slider: F = 1..6 step 0.1 = 4.5
slider: u = 0.2..3 step 0.05 = 0.4
plot: F*(1 - exp(-t/u)) | samples: 300 | label: exhaled volume, L
marker: x = 1 | label: 1 s
readout: FEV1, L = F*(1 - exp(-1/u)) | digits: 3
readout: FEV1/FVC = 1 - exp(-1/u) | digits: 3
readout: peak flow, L/min = F/u*60 | digits: 4
```

::: predict
## A spirometer at 23 °C records 4.0 L exhaled. Corrected to body temperature (37 °C) and saturation with water vapor, is the true lung volume larger or smaller?

- [ ] Smaller, by about 8%
- [x] Larger, by about 8%
- [ ] The same
- [ ] Larger, by about 50%

::: reveal
In the lungs the gas was warmer and fully saturated, so it occupied more volume than it does after cooling in the instrument. The BTPS factor is (310/296) × (760 − 21.1)/(760 − 47) = 1.085 at 23 °C and sea level, so the 4.0 L breath was 4.34 L in the lungs.
:::
:::

::: nameit
## Volumes and capacities

A quiet breath moves the **tidal volume** (about 0.5 L). Beyond it lie the inspiratory and expiratory reserve volumes, and after a full exhalation the **residual volume** remains. **Vital capacity** is the most that can be exhaled after a full inhalation; **total lung capacity** adds the residual volume; **functional residual capacity** (FRC) is what's left at the end of a quiet breath, the expiratory reserve plus the residual volume. Spirometry can't measure the residual volume, so it can't give FRC or TLC.

## Correcting gas volumes

Instruments measure gas at ambient temperature and pressure, saturated (ATPS). Body conditions (BTPS) are 37 °C and saturated, with water vapor at 47 mmHg:

$$V_{BTPS} = V_{ATPS}\,\frac{310}{273 + T}\,\frac{P_B - P_{H_2O}(T)}{P_B - 47}$$

## Spirometers and pneumotachographs

Older spirometers collect gas in a bell; modern ones measure flow and integrate it. A **pneumotachograph** passes the flow through a small, linear resistance (a fine screen or capillary bundle) and measures the pressure drop, $Q = \Delta P/R$, with a differential pressure sensor (lessons 3 and 4). In a forced expiration the key results are FVC, FEV1, their ratio and the peak flow.

## Volumes a spirometer can't reach

- **Helium dilution:** a spirometer of volume $V_1$ holding helium at fraction $C_1$ is connected at FRC; after mixing, $C_1V_1 = C_2(V_1 + \text{FRC})$, so $\text{FRC} = V_1(C_1 - C_2)/C_2$.
- **Body plethysmography:** in a sealed box, the subject pants against a closed shutter; compressing the lung gas by ΔV raises its pressure by ΔP, and Boyle's law gives $V = P\Delta V/\Delta P$, with P the absolute pressure of the dry alveolar gas.

## Ventilation and CO₂

Minute ventilation is tidal volume × rate; **alveolar ventilation** subtracts the anatomical dead space (about 150 mL) from each breath. A **capnograph** measures exhaled CO₂ by infrared absorption; end-tidal CO₂ of about 5% corresponds to $0.05(P_B - 47)$ ≈ 36 mmHg, close to arterial.
:::

::: example
## Helium at 10% in a 5 L spirometer falls to 6.5% after the subject breathes in from FRC. What's the FRC? And what does a body box give for ΔV = 30 mL and ΔP = 10 cmH₂O at P = 970 cmH₂O?

::: solution
1. **Helium dilution.**
   $$\begin{aligned} \text{FRC} &= 5 \times \frac{10 - 6.5}{6.5} \\ &= 2.69\ \text{L} \end{aligned}$$
2. **Body box.**
   $$\begin{aligned} V &= \frac{970 \times 30}{10} \\ &= 2910\ \text{mL} \end{aligned}$$
3. **Why they differ.** The body box reads a little larger, as is typical, because it also counts gas trapped behind closed airways.

**Answer:** helium dilution gives an FRC of 2.69 L, and the body box 2910 mL.
:::
:::

::: example
## In the sim's model with FVC = 4.5 L and τ = 0.4 s, find FEV1, the ratio and the peak flow. What's the alveolar ventilation at 500 mL tidal volume, 150 mL dead space and 14 breaths per minute?

::: solution
1. **FEV1.**
   $$\begin{aligned} \text{FEV}_1 &= 4.5\left(1 - e^{-2.5}\right) \\ &= 4.13\ \text{L} \end{aligned}$$
   a ratio of 0.918.
2. **Peak flow.**
   $$\begin{aligned} \dot V_\max &= \frac{4.5}{0.4} \\ &= 11.25\ \text{L/s} \\ &= 675\ \text{L/min} \end{aligned}$$
3. **Alveolar ventilation.**
   $$\begin{aligned} \dot V_A &= (500 - 150) \times 14 \\ &= 4.9\ \text{L/min} \end{aligned}$$
   out of a minute ventilation of 7 L/min.

**Answer:** FEV1 is 4.13 L, a ratio of 0.918, the peak flow is 11.25 L/s (675 L/min), and the alveolar ventilation is 4.9 L/min.
:::
:::

::: example
## Arterial blood has a $P_{CO_2}$ of 40 mmHg, and mixed expired gas 28 mmHg. What fraction of each 500 mL breath is dead space?

::: solution
1. **Bohr's equation.** It treats expired gas as alveolar gas diluted by dead-space gas with no CO₂:
   $$\begin{aligned} \frac{V_D}{V_T} &= \frac{P_aCO_2 - P_ECO_2}{P_aCO_2} \\ &= \frac{40 - 28}{40} \\ &= 0.3 \end{aligned}$$
2. **Dead space per breath.** That's 150 mL of each 500 mL breath, which only moves in and out of the airways.
3. **Clinical use.** A rising ratio, as in a pulmonary embolism that cuts perfusion to ventilated lung, shows up as a widening gap between arterial and expired CO₂.

**Answer:** 0.3 of each breath, 150 mL, is dead space.
:::
:::

::: check
given: T = 20 | 23 | 26; W = 17.5 | 21.1 | 25.2; P = 700 | 760
answer: 310/(273.15 + T)*(P - W)/(P - 47)
tolerance: 0.5%

## A spirometer at {T} °C (water vapor pressure {W} mmHg) and barometric pressure {P} mmHg measures a volume. By what factor must it be multiplied to get the BTPS volume?

::: right
Right. $\frac{310}{273 + T}\frac{P_B - P_{H_2O}}{P_B - 47}$ = {answer}.
:::

::: misses
- [310/(273.15 + T)] Include the water-vapor correction too.
- [any] Not yet. Temperature ratio times dry-gas pressure ratio.
:::
:::

::: check
given: V = 3 | 5 | 8; a = 8 | 10; b = 5 | 6.5
answer: V*(a - b)/b
unit: L
tolerance: 0.5%

## A {V} L spirometer starts with {a}% helium; after the subject joins at FRC and the gas mixes, it reads {b}%. What's the FRC?

::: right
Right. $V_1(C_1 - C_2)/C_2$ = {answer} L.
:::

::: misses
- [V*(a - b)/a] Divide by the final concentration.
- [any] Not yet. Helium is conserved: $C_1V_1 = C_2(V_1 + \text{FRC})$.
:::
:::

::: check
given: P = 950 | 1000; v = 20 | 40 | 60; p = 8 | 15
answer: P*v/p
unit: mL
tolerance: 0.5%

## In a body box, panting against a closed shutter compresses the lung gas by {v} mL and raises its pressure by {p} cmH₂O, from an absolute dry-gas pressure of {P} cmH₂O. What's the lung volume?

::: right
Right. $P\Delta V/\Delta P$ = {answer} mL.
:::

::: misses
- [any] Not yet. Boyle's law for small changes.
:::
:::

::: check
given: u = 0.3 | 0.6 | 1.2
answer: 1 - exp(-1/u)
tolerance: 0.5%

## In the single-exponential model of a forced expiration with time constant {u} s, what's FEV1/FVC?

::: right
Right. $1 - e^{-1/\tau}$ = {answer}.
:::

::: misses
- [exp(-1/u)] That's the fraction still in the lungs after one second.
- [any] Not yet. The fraction exhaled in the first second.
:::
:::

::: check
given: V = 400 | 500 | 700; D = 120 | 150 | 180; f = 10 | 14 | 20
answer: (V - D)*f/1000
unit: L/min
tolerance: 0.5%

## A patient breathes {V} mL per breath at {f} breaths per minute with {D} mL of dead space. What's the alveolar ventilation?

::: right
Right. $(V_T - V_D)f$ = {answer} L/min.
:::

::: misses
- [V*f/1000] That's the minute ventilation; subtract the dead space from each breath.
- [any] Not yet. Only gas beyond the dead space reaches the alveoli.
:::
:::

::: check
given: c = 4 | 5 | 6; P = 700 | 760
answer: c/100*(P - 47)
unit: mmHg
tolerance: 0.5%

## End-tidal gas contains {c}% CO₂ (of the dry gas) at a barometric pressure of {P} mmHg. What's its CO₂ partial pressure in the saturated alveolar gas?

::: right
Right. $F_{CO_2}(P_B - 47)$ = {answer} mmHg.
:::

::: misses
- [c/100*P] Subtract the water vapor pressure first.
- [any] Not yet. Fraction of the dry-gas pressure.
:::
:::

::: check
kind: parts
given: a = 35 | 40 | 50; e = 22 | 28; V = 400 | 600

## Arterial $P_{CO_2}$ is {a} mmHg and mixed expired $P_{CO_2}$ {e} mmHg, with a tidal volume of {V} mL.

::: part
answer: (a - e)/a
tolerance: 0.5%

### What fraction of each breath is dead space?

::: misses
- [e/a] That's the fraction that reached the alveoli.
- [any] $(P_aCO_2 - P_ECO_2)/P_aCO_2$.
:::
:::

::: part
answer: (a - e)/a*V
unit: mL
tolerance: 0.5%

### How many milliliters is that?

::: misses
- [any] The fraction times the tidal volume.
:::
:::
:::

::: check
kind: derivation
vars: V = 200..1000, f = 6..40, D = 100..200

## A patient breathes V mL per breath at f breaths per minute, with D mL of dead space.

::: step
answer: V*f

### What is the minute ventilation, in mL/min?
:::

::: step
answer: (V - D)*f

### What is the alveolar ventilation?
:::

::: right
Right. Shallow, fast breathing wastes more of each breath on dead space, so the same minute ventilation can leave the alveoli short.
:::
:::

::: explain
## Why can't a spirometer measure total lung capacity, and how do dilution and plethysmography get around that?

::: model
A spirometer measures only gas that moves in or out of the mouth, and even after the hardest possible exhalation the residual volume stays in the lungs, so every spirometer volume is a difference from some unknown starting amount. Helium dilution adds a known amount of an insoluble gas and measures how much it's diluted once it has mixed into the lungs, which reveals the total gas volume it spread into. Body plethysmography uses Boyle's law instead: compressing the gas in the chest slightly and measuring the pressure change gives the gas volume directly, including gas in regions that don't communicate well with the mouth, which dilution can miss.
:::
:::

::: practice

::: check
given: V = 2 | 10; a = 5 | 15; b = 2 | 4.5
answer: V*(a - b)/b
unit: L
tolerance: 0.5%

## Helium FRC: {V} L, {a}% → {b}%?

::: misses
- [any] $V_1(C_1 - C_2)/C_2$.
:::
:::

::: check
given: u = 0.15 | 0.3 | 0.5 | 0.8 | 1.2 | 2 | 3
answer: 1 - exp(-1/u)
tolerance: 0.5%

## FEV1/FVC for τ = {u} s?

::: misses
- [any] $1 - e^{-1/\tau}$.
:::
:::

::: check
given: V = 200 | 1000; D = 80 | 200; f = 6 | 40
answer: (V - D)*f/1000
unit: L/min
tolerance: 0.5%

## Alveolar ventilation: {V} mL, {D} mL dead space, {f}/min?

::: misses
- [any] $(V_T - V_D)f$.
:::
:::
:::