title: "Therapeutic devices: electrosurgery, dialysis, ventilators and infusion pumps"
heading: Cutting and coagulating with radio-frequency current and keeping the return electrode cool, the artificial kidney's clearance and Kt/V, a ventilator's airway pressure from resistance and compliance, stepper-driven infusion pumps, and a short tour of lithotripters, incubators, lasers and assist devices
eyebrow: "S15 · Lesson 45 · Therapeutic devices: electrosurgery, dialysis, ventilators and infusion pumps"
xp: 50
---

::: hook
Most of this course has been about measuring the body. The devices in this lesson act on it. A surgeon cuts with a current of half an ampere that the patient can't feel. A dialysis machine does a kidney's filtering three times a week. A ventilator breathes for a patient who can't, and an infusion pump delivers a drug a fraction of a milliliter at a time for days. Each is an engineered system with a physical model at its core, and each can hurt the patient when the model's assumptions fail.
:::

::: play
- A ventilator delivers a volume-controlled breath: a constant flow Q = 0.5 L/s for 1 s (a 500 mL tidal volume), then lets the patient exhale passively against the set end-expiratory pressure (PEEP).
- The lungs are modeled as an airway resistance R in series with a compliance C. The solid curve is the pressure the ventilator sees at the airway; the dashed one is the pressure in the alveoli.
- At the start of the breath the airway pressure jumps by RQ, the drop across the airways, then climbs as the lungs fill. Stiff lungs (small C) raise the peak; narrowed airways (large R) raise the gap between airway and alveolar pressure and slow the exhalation, whose time constant is RC.
:::

```sim plot
view: graph
x: t = 0..4 s
y: P = 0..45
slider: R = 5..30 = 10
slider: C = 0.02..0.1 = 0.05
slider: E = 0..15 = 5
plot: step(1 - t)*(E + R*0.5 + 0.5*t/C) + step(t - 1)*E | samples: 800 | label: airway pressure, cmH₂O
plot: step(1 - t)*(E + 0.5*t/C) + step(t - 1)*(E + 0.5/C*exp(-(t - 1)/(R*C))) | samples: 800 | dashed | color: ch3 | label: alveolar pressure
readout: peak airway pressure, cmH₂O = E + R*0.5 + 0.5/C | digits: 3
readout: plateau pressure, cmH₂O = E + 0.5/C | digits: 3
readout: exhalation time constant RC, s = R*C | digits: 3
```

::: predict
## An electrosurgical unit passes about 0.6 A of radio-frequency current through the patient to cut tissue. Why doesn't that current stop the heart, when 0.1 A at 60 Hz can?

- [ ] The current flows only on the skin's surface
- [x] At 500 kHz each half-cycle is far too short to depolarize nerve and muscle membranes
- [ ] The return electrode absorbs the current before it reaches the heart
- [ ] The current is too small to matter at any frequency

::: reveal
Excitable membranes integrate current: lesson 43's strength–duration curve shows that short pulses need much more current to stimulate. A half-cycle at 500 kHz lasts 1 µs, so the threshold rises far above any current used, and the only effect left is heating. Electrosurgery runs from about 250 kHz to a few megahertz for exactly this reason.
:::
:::

::: nameit
## Electrosurgery

An electrosurgical unit drives radio-frequency current from a small active electrode, where the current density is high enough to heat tissue rapidly, into a large **dispersive (return) electrode** on the patient's skin, where it spreads out and heats little. Webster gives the ranges: **cutting** uses continuous waves at 500 kHz to 2.5 MHz, open-circuit voltages up to 9 kV and 100 to 750 W, vaporizing cells in a spark; **coagulation** uses bursts of damped sinusoids repeated about 120 times a second, at 300 to 2000 V and 80 to 200 W into 500 Ω, drying tissue and sealing vessels; a **blend** combines them. **Bipolar** forceps keep the current between the two tips.

Heating per unit area at the return electrode goes as the current density squared. If a pad partly peels off, the current crowds into what's left, and halving the contact area quadruples the heating there. Return-electrode monitors split the pad in two and check the impedance between the halves continuously. Stray capacitance from cables and the patient to ground gives the current other paths at radio frequencies, which is why burns can appear at ECG electrodes or wherever skin touches a grounded surface.

## Hemodialysis

An artificial kidney passes blood on one side of a semipermeable membrane, today usually 10,000 to 15,000 hollow fibers about 0.2 mm across, and a dialysate on the other. Small waste molecules such as urea diffuse across. A dialyzer's performance is its **clearance**, the blood flow it cleans completely:

$$K = Q_B\,\frac{C_\text{in} - C_\text{out}}{C_\text{in}}$$

Treating urea as spread through a volume V of body water (about 58% of body weight), its concentration falls as $e^{-Kt/V}$ over a session. The dose is stated as **Kt/V**, with about 1.2 per session the usual minimum for three sessions a week, and the **urea reduction ratio** is $1 - e^{-Kt/V}$. The machine monitors the dialysate's conductivity (its electrolyte concentration) and temperature, blood-leak and air detectors, and the pressures that drive ultrafiltration, the removal of excess water.

## Ventilators

A ventilator pushes gas into the lungs, which behave to first order like a resistance R (the airways) in series with a compliance C (the lungs and chest wall). The airway pressure during a breath is

$$P_\text{aw} = \text{PEEP} + RQ + \frac{V}{C}$$

where Q is the flow and V the volume delivered so far. In **volume control** the ventilator sets the flow and tidal volume and the pressure follows; in **pressure control** it sets the pressure and the volume follows from the lungs' mechanics. **PEEP**, a few cmH₂O kept at the end of exhalation, holds small airways open. Exhalation is passive and decays with time constant RC, so about 3RC is needed to empty; too short an exhalation traps gas. Tidal volumes are now set by body size, about 6 to 8 mL per kilogram of ideal body weight, to avoid overstretching injured lungs. High-frequency ventilators move small volumes hundreds of times a minute.

## Infusion pumps

Peristaltic pumps squeeze a tube with rollers; syringe pumps push a plunger with a lead screw. Both are driven by stepper motors, so each step delivers a fixed volume and the rate is set by the step frequency. They detect **occlusion** (rising line pressure) and **air in line** (by ultrasound) and stop with an alarm. Closed-loop systems adjust the rate to keep a measured variable, such as blood pressure or glucose, at a target.

## Other therapeutic devices

- **Lithotripters** break kidney stones with focused shock waves: a source at one focus of an ellipsoidal reflector concentrates the pulse at the other focus, positioned on the stone. Water couples the shock into the body because soft tissue's acoustic impedance is nearly the same as water's (S16 covers why that matters).
- **Infant incubators** servo-control air or skin temperature, usually around 36 to 37 °C at the skin, and humidity.
- **Therapeutic lasers** cut, coagulate or ablate by choosing a wavelength the target absorbs strongly, such as CO₂ at 10.6 µm in water.
- **Mechanical heart valves** and **ventricular assist devices** replace or support the heart's pumping; their engineering problems are flow, wear and clotting.
:::

::: example
## A volume-controlled breath delivers 500 mL at 0.5 L/s with PEEP = 5 cmH₂O to lungs with R = 10 cmH₂O·s/L and C = 50 mL/cmH₂O. Find the peak and plateau pressures and how long exhalation takes.

::: solution
1. **Pressures the breath must supply.** Pushing 0.5 L/s through the airways and filling the compliance with 0.5 L cost
   $$\begin{aligned} RQ &= 10 \times 0.5 \\ &= 5\ \text{cmH}_2\text{O} \\ \frac{V}{C} &= \frac{0.5}{0.05} \\ &= 10\ \text{cmH}_2\text{O} \end{aligned}$$
2. **Peak pressure.** PEEP plus both:
   $$\begin{aligned} P_\text{peak} &= 5 + 5 + 10 \\ &= 20\ \text{cmH}_2\text{O} \end{aligned}$$
3. **Plateau pressure.** During an inspiratory pause the flow stops and the resistive drop disappears:
   $$\begin{aligned} P_\text{plat} &= 5 + 10 \\ &= 15\ \text{cmH}_2\text{O} \end{aligned}$$
   the alveolar pressure.
4. **What the two measure.** The difference between peak and plateau measures resistance; the plateau minus PEEP measures compliance.
5. **Exhalation.**
   $$\begin{aligned} \tau &= RC \\ &= 10 \times 0.05 \\ &= 0.5\ \text{s} \end{aligned}$$
   so about 1.5 s is needed to exhale 95%.
6. **Time available.** At 20 breaths a minute with 1 s of inspiration, the patient has 2 s: enough here, but not for a patient with asthma whose resistance is three times higher.

**Answer:** the peak pressure is 20 cmH₂O and the plateau 15 cmH₂O, and exhaling 95% takes about 1.5 s.
:::
:::

::: example
## A dialyzer with 300 mL/min of blood flow lowers urea from 100 to 30 mg/dL across it. What dose does a 4-hour session give a 70 kg patient?

::: solution
1. **Clearance.**
   $$\begin{aligned} K &= 300 \times \frac{100 - 30}{100} \\ &= 210\ \text{mL/min} \end{aligned}$$
2. **Volume cleared in the session.** In 240 minutes:
   $$Kt = 50.4\ \text{L}$$
3. **Kt/V.** Urea spreads through about 0.58 × 70 = 40.6 L of body water:
   $$\begin{aligned} \frac{Kt}{V} &= \frac{50.4}{40.6} \\ &= 1.24 \end{aligned}$$
   just above the usual minimum of 1.2.
4. **Urea reduction.** The urea concentration at the end is $e^{-1.24}$ = 29% of the start, a urea reduction ratio of 71%.
5. **Limits of the estimate.** This single-pool estimate ignores urea made during the session and the rebound from tissues afterward, which clinical formulas correct for.

**Answer:** the session gives Kt/V = 1.24, just above the usual minimum of 1.2, with a urea reduction ratio of 71%.
:::
:::

::: example
## An electrosurgical unit returns 1 A RMS through a 100 cm² dispersive pad. What current density does the skin see, and how much faster does it heat if the pad peels back to 20 cm² of contact?

::: solution
1. **Current density on the full pad.**
   $$\begin{aligned} J &= \frac{1\ \text{A}}{100\ \text{cm}^2} \\ &= 10\ \text{mA/cm}^2 \end{aligned}$$
   low enough that blood flow carries the heat away.
2. **On 20 cm².**
   $$\begin{aligned} J &= \frac{1\ \text{A}}{20\ \text{cm}^2} \\ &= 50\ \text{mA/cm}^2 \end{aligned}$$
3. **Heating rate.** Heating goes as the square of current density:
   $$\left(\frac{50}{10}\right)^2 = 25$$
   so the skin under the remaining contact heats 25 times faster, enough to burn.
4. **Protection.** That's why modern units monitor the pad's contact impedance with split electrodes and stop if it rises.

**Answer:** the skin sees 10 mA/cm² on the full pad; peeled back to 20 cm², it sees 50 mA/cm² and heats 25 times faster.
:::
:::

::: check
given: d = 1 | 1.5 | 2; p = 0.5 | 1; n = 200 | 400; q = 2 | 5 | 20
answer: q/(pi*(d/2)^2*(p/10)/n)/3600
unit: Hz
tolerance: 0.5%

## A syringe pump pushes a {d} cm diameter plunger with a lead screw of {p} mm pitch, turned by a stepper of {n} steps per revolution. At what step rate does it deliver {q} mL/h?

::: right
Right. Each step moves the plunger {p}/{n} mm and delivers $\pi d^2/4$ times that: {answer} steps per second.
:::

::: misses
- [q/(pi*(d/2)^2*(p/10)/n)] That's steps per hour. Divide by 3600.
- [q/(pi*d^2*(p/10)/n)/3600] Use the radius, d/2, for the plunger's area.
- [any] Not yet. Find the volume per step, then divide the flow by it.
:::
:::

::: check
given: Q = 200 | 300 | 400; a = 80 | 100 | 150; b = 20 | 35 | 45
answer: Q*(a - b)/a
unit: mL/min
tolerance: 0.5%

## Blood enters a dialyzer at {Q} mL/min with urea at {a} mg/dL and leaves at {b} mg/dL. What's the dialyzer's clearance, in mL/min?

::: right
Right. $K = Q_B(C_\text{in} - C_\text{out})/C_\text{in}$ = {answer}.
:::

::: misses
- [Q*(a - b)/b] Divide by the inlet concentration.
- [Q*b/a] That's the fraction left, not the fraction removed.
- [any] Not yet. Clearance is the blood flow times the fraction removed.
:::
:::

::: check
given: K = 180 | 220 | 260; t = 3 | 3.5 | 4; W = 60 | 75 | 90
answer: K*t*60/1000/(0.58*W)
tolerance: 0.5%

## A dialyzer clears {K} mL/min for {t} hours on a {W} kg patient (body water 58% of weight). What's Kt/V?

::: right
Right. Kt = {K} × {t} × 60 mL, V = 0.58 × {W} L: Kt/V = {answer}.
:::

::: misses
- [K*t/1000/(0.58*W)] Convert hours to minutes.
- [K*t*60/1000/W] V is the body water, 58% of the weight.
- [any] Not yet. Multiply clearance by time and divide by the water volume.
:::
:::

::: check
given: k = 0.9 | 1.2 | 1.5 | 1.8
answer: 100*(1 - exp(-k))
unit: %
tolerance: 0.5%

## A session delivers Kt/V = {k}. Ignoring urea generation, what's the urea reduction ratio?

::: right
Right. $1 - e^{-Kt/V}$ = {answer}%.
:::

::: misses
- [100*k] The decay is exponential, not linear.
- [100*exp(-k)] That's the fraction left. The reduction ratio is what's removed.
- [any] Not yet. $C_\text{end}/C_\text{start} = e^{-Kt/V}$.
:::
:::

::: check
given: V = 0.4 | 0.5 | 0.6; Q = 0.5 | 0.8; R = 8 | 12 | 20; C = 0.03 | 0.05 | 0.07; E = 5 | 8
answer: E + R*Q + V/C
tolerance: 0.5%

## A ventilator delivers {V} L at {Q} L/s with PEEP {E} cmH₂O to lungs with R = {R} cmH₂O·s/L and C = {C} L/cmH₂O. What's the peak airway pressure, in cmH₂O?

::: right
Right. PEEP + RQ + V/C = {answer} cmH₂O.
:::

::: misses
- [E + V/C] That's the plateau. The peak adds the resistive drop RQ.
- [R*Q + V/C] Add the PEEP.
- [any] Not yet. Sum PEEP, the resistive drop and the elastic pressure.
:::
:::

::: check
given: R = 8 | 15 | 30; C = 0.03 | 0.05 | 0.08
answer: 3*R*C
unit: s
tolerance: 0.5%

## Lungs with R = {R} cmH₂O·s/L and C = {C} L/cmH₂O exhale passively. About how long does it take to exhale 95% of the tidal volume?

::: right
Right. Three time constants, 3RC = {answer} s.
:::

::: misses
- [R*C] One time constant empties 63%. 95% takes three.
- [any] Not yet. The decay has time constant RC.
:::
:::

::: check
given: P = 100 | 200 | 400; A = 50 | 100 | 150; f = 0.25 | 0.5
answer: 1/f^2
tolerance: 0.5%

## An electrosurgical unit sends {P} W through a {A} cm² return electrode. If only a fraction {f} of the pad stays in contact, by what factor does the heating per square centimeter under it rise?

::: right
Right. The current density rises by 1/{f}, and heating goes as its square: {answer} times.
:::

::: misses
- [1/f] Heating goes as the square of the current density.
- [any] Not yet. Heating per area is proportional to $J^2$.
:::
:::

::: check
kind: parts
given: I = 0.5 | 1 | 1.5; A = 60 | 150; a = 10 | 30

## An electrosurgical unit returns {I} A RMS through a pad of {A} cm²; partial peeling leaves {a} cm² in contact.

::: part
answer: I*1000/A
unit: mA/cm^2
tolerance: 0.5%

### What current density does the full pad carry?

::: misses
- [any] Current over area.
:::
:::

::: part
answer: (A/a)^2
tolerance: 0.5%

### By what factor does the heating rate rise after peeling?

::: misses
- [A/a] Heating goes as the square of current density.
- [any] $(A/a)^2$.
:::
:::
:::

::: check
kind: derivation
vars: d = 0.5..3, p = 0.5..2, s = 48..400, r = 0.1..100

## A syringe pump drives a plunger of diameter d (cm) with a lead screw of pitch p (mm), turned by a stepper with s steps per revolution at r steps per second.

::: step
answer: pi*d^2/4*p/10/s

### What volume does one step deliver, in milliliters?
:::

::: step
answer: pi*d^2/4*p/10/s*r*3600

### What flow results, in mL per hour?
:::

::: right
Right. Smaller syringes give finer steps, so low-rate infusions use small syringes to avoid pulsatile flow.
:::
:::

::: explain
## Why do electrosurgical units run at hundreds of kilohertz, and why does that create burn risks that a 60 Hz device wouldn't?

::: model
Nerve and muscle membranes respond to the charge delivered in a pulse, and a half-cycle at 500 kHz lasts only a microsecond, so the stimulation threshold at that frequency is far above the currents used. The current can therefore heat and cut without causing shocks, muscle contractions or fibrillation. The cost of the high frequency is that small capacitances become low impedances: a 100 pF path from the patient to a grounded table or monitor lead has about 3 kΩ of reactance at 500 kHz, against 27 MΩ at 60 Hz. Current can return through those paths instead of the dispersive pad, and wherever it leaves the body through a small area, such as an ECG electrode or a fingertip touching metal, the current density is high and burns result. Isolated generator outputs, return-electrode monitoring and careful placement of the pad reduce these risks.
:::
:::

::: practice

::: check
given: Q = 150..450; a = 50..200; b = 10..40
answer: Q*(a - b)/a
unit: mL/min
tolerance: 0.5%

## Dialyzer clearance in mL/min: {Q} mL/min, urea {a} in, {b} out?

::: misses
- [any] $Q_B(C_\text{in} - C_\text{out})/C_\text{in}$.
:::
:::

::: check
given: V = 0.3..0.8; Q = 0.3..1; R = 5..30; C = 0.02..0.1; E = 0..15
answer: E + R*Q + V/C
tolerance: 0.5%

## Peak pressure, cmH₂O: {V} L at {Q} L/s, R = {R}, C = {C}, PEEP {E}?

::: misses
- [any] PEEP + RQ + V/C.
:::
:::

::: check
given: k = 0.5..2.5
answer: 100*(1 - exp(-k))
unit: %
tolerance: 0.5%

## Urea reduction ratio for Kt/V = {k}?

::: misses
- [any] $1 - e^{-Kt/V}$.
:::
:::
:::