S645 of 55
0 XP
  1. Hook
  2. Play
  3. Predict
  4. Name it
  5. Check
  6. Explain

S6 · Lesson 45 · Network analyzer calibration

How a vector network analyzer measures, the one-port error model, short–open–load calibration, two-port models, TRL and residual uncertainty

A vector network analyzer measures S-parameters to a hundredth of a decibel, but only after it's been told what its own cables, connectors and couplers are doing. Raw, its readings are distorted by imperfect directivity (lesson 18), mismatches at its ports and frequency-dependent losses, and at microwave frequencies those errors can be larger than what's being measured. Calibration measures a few known standards, solves for the errors, and removes them mathematically from every later measurement. The error model is only a few lines of algebra, and it's the reason modern RF design can trust its numbers.

  • Reflection measured by an uncorrected one-port, against the phase of the device's reflection (as if the device sat at the end of a sliding line), for a device whose true return loss is R dB.
  • The analyzer's raw directivity and source match make the reading ripple around the truth (dashed) as the phases line up and cancel.
  • Improve the directivity or source match, or measure a worse load, and the ripple shrinks. Calibration removes it altogether, leaving only residual errors.

Predict

How many known standards does a one-port calibration need, at minimum?

Name it

The VNA

A source sweeps frequency; directional couplers or bridges at each port sample the incident and reflected waves (a1,b1,a2,b2); receivers down-convert them and measure their ratios in magnitude and phase. S11=b1/a1 with port 2 terminated, and so on.

The one-port error model

Everything between the analyzer's ideal reflectometer and the device is lumped into an error two-port with three terms:

  • e00, directivity (leakage straight to the reflected-wave receiver);
  • e11, source match (what the device sees looking back);
  • e10e01, reflection tracking (loss and phase of the path).

The measured reflection is

ΓM=e00+e10e01ΓA1−e11ΓA

which is lesson 4's terminated two-port.

Short–open–load calibration

Measure a load (Γ = 0), an open (Γ = +1) and a short (Γ = −1), giving ML, MO, MS. With a=MO−ML and b=MS−ML:

e00=ML,e11=a+ba−b,e10e01=a(1−e11)

and any later measurement is corrected by inverting the model:

ΓA=ΓM−e00e10e01+e11(ΓM−e00)

(Real standards aren't ideal: the open's fringing capacitance and the short's inductance are described by the calibration kit's model.)

Two ports, TRL and residual errors

  • SOLT adds a thru and gives the 12-term model (6 per direction: directivity, source match, reflection tracking, load match, transmission tracking, isolation).
  • TRL (thru, reflect, line) needs no well-known load: a thru, any high reflection (the same on both ports) and a line whose extra length differs from the thru by between about 20° and 160°, which limits one line to an 8 : 1 band. It's the standard for on-wafer and in-fixture work.
  • After calibration, residual directivity, source match and tracking, set by how well the standards are known, determine the uncertainty, and cable flexing and drift spoil a calibration over time.

Worked example

A one-port calibration measures ML = 0.05, MO = 0.80 and MS = −0.90 (real, for simplicity). Find the error terms and correct a reading of 0.30.

Show the solution
  1. Directivity. e00=ML=0.05
  2. Source match. With a = 0.75 and b = −0.95, e11=0.75−0.950.75+0.95=−0.118
  3. Reflection tracking. e10e01=0.75×1.118=0.838
  4. Correct the reading. ΓA=0.30−0.050.838−0.118×0.25=0.309
  5. Check. 0.05+0.838×0.3091+0.118×0.309=0.300

Answer e00 = 0.05, e11 = −0.118 and e10e01 = 0.838, and the device's corrected reflection is 0.309.

Worked example

A TRL line has 90° more delay than the thru at 5 GHz. Over what band is it usable?

Show the solution
  1. Phase limits. The line's extra phase scales with frequency, so 20° and 160° are reached at 5×20/90=1.11GHz5×160/90=8.89GHz
  2. Band. That's an 8 : 1 band.
  3. Wider bands. They use several lines, each covering part of the band.

Answer from 1.11 to 8.89 GHz, an 8 : 1 band.

Worked example

After calibration, an analyzer's residual directivity is −40 dB and its residual source match −35 dB. It measures a load whose true return loss is 20 dB. What could it read?

Show the solution
  1. True reflection. The true |Γ| is 0.1.
  2. Worst-case error. It is about e00+e11|Γ|2=0.010+0.0178×0.01=0.0102
  3. Range of readings. The reading lies between 0.0898 and 0.1102: a return loss between 19.2 and 20.9 dB.
  4. Which term matters. Directivity dominates when measuring good matches, and source match when measuring large reflections, which is why calibration kits specify both.

Answer a return loss anywhere from 19.2 to 20.9 dB.

Check

A one-port calibration measures 0.02 on the load, 0.75 on the open and −0.95 on the short (real). What's the source match e11?

Check

With the same measurements, what's the reflection tracking e10e01?

Check

A calibrated one-port has e00 = 0.02, e11 = −0.15 and e10e01 = 0.9 (real). It reads −0.4. What's the device's actual reflection?

Check

An uncorrected one-port has e00 = 0.02, e11 = −0.15 and e10e01 = 0.95 (real). What does it read for a device with Γ = −0.6?

Check

A TRL line has 90° more delay than the thru at 5 GHz, and the cal is trusted while the phase difference stays between 20° and 160°. What's the highest usable frequency?

Write units the way you normally would: 1.5 GHz, 1500 MHz and 0.0015 THz are the same answer.

Check

A TRL line must cover 4 to 4 × 3 GHz, so it's made a quarter wave long at the band's arithmetic center. How much longer than the thru is it, on a line with εeff = 6.5?

Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.

Check

After calibration an analyzer has residual directivity −40 dB and residual source match −20 dB. It measures a load whose true return loss is 14 dB.

(a)

What is the worst-case error in |Γ|?

(b)

What is the lowest return loss it could read?

Check

A one-port calibration measures m on the load, o on the open and s on the short (real).

Step 1

What is the directivity e00?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why can a network analyzer measure a nearly perfect load accurately after calibration, when its own directivity is only 25 dB?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.