C11 · Lesson 1 · Accuracy, precision and uncertainty
How much to trust a reading
A meter reads 12.34 V. Is the voltage really 12.34 V? Probably not exactly: the meter has a calibration error, its last digit flickers with noise, and its display can't show anything finer than 10 mV. Every measurement is a number with an uncertainty, and a reading without one says less than it seems. Engineers separate the errors that repeat the same way every time (bias, which sets accuracy) from the ones that scatter (noise, which sets precision), estimate each, and report the result as a value plus or minus a stated uncertainty.
- Each draw is one reading of a true 5.000 V by a meter with a bias b and random noise of standard deviation s. The histogram piles up the readings.
- Raise the noise: the pile spreads, but its center stays put. Add bias: the whole pile shifts, and no amount of averaging moves it back.
- Increase the number of readings: the average (readout) settles, with a standard error of s/√N. Averaging fixes noise, never bias.
Predict
A meter specified as ±(0.5% of reading + 2 counts) reads 12.34 V on its 20.00 V range. What is the uncertainty?
±0.08 V: 0.5% of 12.34 V is 0.062 V, and 2 counts of the last digit (10 mV each on this range) add 0.02 V. Report it as 12.34 ± 0.08 V, so the true value is between 12.26 and 12.42 V.
Name it
Kinds of error
| Term | Meaning |
|---|---|
| accuracy | how close readings are to the true value; limited by bias |
| precision | how close repeated readings are to each other; limited by noise |
| resolution | the smallest change the instrument can show (one count) |
| bias (systematic error) | an offset that repeats every time: calibration, loading |
| random error | scatter that differs each time: noise, interference |
Instrument specifications
Meters are specified as ±(% of reading + counts), where a count is one step of the last digit on the range in use. A "3½-digit" meter shows up to 1999 counts; a "6½-digit" one up to 1,999,999. Resolution is not accuracy: a meter can show six digits and be wrong in the third.
Random error and averaging
N readings with standard deviation σ have a mean whose standard deviation, the standard error, is
So quadrupling the readings halves the random error, and reaching an error e needs readings. Bias is untouched by averaging; it's removed by calibration against a better reference.
Reporting
Give the value with its uncertainty, rounded to match: 12.34 ± 0.08 V, not 12.3417 ± 0.08 V. Relative error is the error divided by the value, often in percent.
Worked example
A 3½-digit meter on its 2 V range reads 1.234 V. Its spec is ±(0.8% + 3 counts). What is the uncertainty?
Show the solution
- Size of one count. On the 2 V range one count is 1 mV.
- Percentage part.
- Count part.
- Add them.
Answer the uncertainty is ±12.9 mV, so the reading is 1.234 ± 0.013 V.
Worked example
Readings scatter with σ = 0.2 V. How many must be averaged to bring the random error down to 0.05 V?
Show the solution
- Readings needed.
- Check ten readings. Ten would not quite be enough:
Answer 16 readings must be averaged.
Worked example
A meter's specification allows ±0.05 V, treated as a rectangular distribution. Ten readings scatter with s = 0.03 V. What are the combined standard uncertainty and the expanded uncertainty (k = 2)?
Show the solution
- Type B, from the specification. A rectangular half-width is divided by :
- Type A, from the scatter.
- Combine in quadrature.
- Expand with k = 2.
- What dominates. The specification dominates, so more readings would hardly help; a better meter would.
Answer the combined standard uncertainty is 0.030 V, and the expanded uncertainty is 0.061 V (k = 2).
Check
A meter specified as ±(0.05% of reading + 5 counts) reads 4.987 V on a range where one count is 0.01 V. What is the uncertainty?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
Readings scatter with a standard deviation of 0.2 V. What is the standard error of the average of 4 readings?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
Readings have σ = 0.5 mV. How many must be averaged for a standard error of 0.01 mV?
Check
How many counts can a 4½-digit display show at most?
Check
A reference of exactly 5 V reads as 10.2 V. What is the relative error?
Check
A meter's readings of a stable source are always within 1 mV of each other, but all are 50 mV above the true value. How would you describe it?
Check
A meter reads 12.34 V with resolution 0.01 V. Its spec is ±(0.05% of reading + 5 counts).
(a)
What is the uncertainty?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What is that as a percentage of the reading?
Check
Readings scatter with s = 0.2 and the random error of their mean must be at most 0.01.
(a)
How many readings must be averaged?
(b)
What standard error does that give?
Explain why
Why does averaging many readings reduce random error but not bias?
Random errors push each reading up or down by different, unrelated amounts, so in a sum they partly cancel: the spread of a sum of N readings grows only as , and dividing by N to get the average leaves a spread of . A bias pushes every reading the same way by the same amount, so it survives averaging intact: the average of a thousand readings that are all 50 mV high is still 50 mV high. Removing a bias needs information from outside the measurement, such as a comparison with a calibrated reference.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.