C11 · Lesson 17 · Grounding, ground loops and differential measurement

Where the hum comes from, and how to reject it

Connect a grounded scope to a grounded circuit across the lab and a 50 Hz hum appears on every trace. Its source isn't the circuit: the two "grounds" are at slightly different voltages, because mains currents flow in the building's wiring, and the scope's cable shield connects them, closing a ground loop. The cure is not to cut the safety ground but to measure differentially: look at the difference between two points and reject whatever voltage they share. How well an instrument does that is its common-mode rejection ratio, and it decides whether microvolt signals can be measured on a noisy bench.

  • The measured trace is a 10 mV, 1 kHz signal plus whatever fraction of a 50 Hz common-mode voltage Vcm leaks through, set by the measurement's CMRR.
  • A single-ended scope measurement (CMRR about 0 dB) shows the hum at full size, swamping the signal. A differential amplifier with 60 dB CMRR cuts 1 V of hum to 1 mV.
  • Raise the CMRR to 100 dB and the hum vanishes. The readout gives what's left of it.

Predict

A differential amplifier with a CMRR of 60 dB measures a small signal riding on 1 V of common-mode hum. How much hum reaches the output (referred to the input)?

Name it

Ground loops

When two pieces of equipment are each grounded through the mains and also joined by a cable shield, the shield and the building's ground wiring form a loop. Current from the ground potential difference, or induced by a changing magnetic field through the loop's area, flows in the shield, and its drop IRshield adds to the signal. A loop of area A in a field B oscillating at f picks up

V=2πfBA

Remedies: a single ground point, short cables, twisted pairs (small loop area), and above all differential measurement. Never remove the safety ground from an instrument to break a loop: its chassis can then rise to mains voltage.

Common-mode rejection

A voltage on both inputs at once is common mode; the difference is differential. A differential input rejects common mode by its CMRR:

Verror=Vcm×10−CMRR/20

Instrumentation amplifiers (C6) reach 100 to 120 dB at low frequencies, less at high.

Measuring across a floating component

To see the voltage across a component that isn't connected to ground (the high side of a motor driver, a current-sense resistor), use a differential probe, or two ordinary probes and the scope's A − B math. The second is limited by how well the two channels match: a 1% gain mismatch gives only about 40 dB of rejection.

Worked example

A thermocouple signal of 2 mV sits on 5 V of common mode from a heater supply. What CMRR keeps the error below 1 µV?

Show the solution
  1. Required rejection. The error must be at most 10−6 V out of 5 V: CMRR=20log10⁡510−6=134dB
  2. In practice. That's beyond most amplifiers at 50 Hz, so the design also needs isolation or filtering.

Answer at least 134 dB.

Worked example

A 1 cm² loop of probe ground lead sits near a transformer with a 1 mT field at 50 Hz. How much does it pick up?

Show the solution
  1. Induced voltage. V=2πfBA=2π×50×10−3×10−4V=31μV
  2. Near a transformer. Close to a power transformer the field can be ten times larger, which is why a probe's long ground lead picks up hum.

Answer about 31 µV.

Worked example

A sensor's return current of 1 A shares 10 mΩ of ground wiring with a measurement. What error does that cause single-ended, and with a differential amplifier of 80 dB CMRR?

Show the solution
  1. Single-ended. The drop appears directly in the reading: 1A×10mΩ=10mV
  2. Differential. It becomes common-mode, attenuated by 1080/20=104: 10mV104=1μV
  3. Where to measure. Measuring between the two ends of the sensor, rather than to a distant ground, removes the shared wiring from the signal path.

Answer a 10 mV error single-ended, and 1 µV with the differential amplifier.

Check

1 V of common mode reaches an input with 40 dB of CMRR. How much error does it cause?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

How much CMRR keeps 10 V of common mode to under 0.001 mV of error?

Check

A ground loop drives 100 mA through a cable shield of 0.005 Ω. How much voltage does it add to a single-ended measurement?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

A loop of 100 cm² sits in a 0.1 mT field at 50 Hz. How much voltage is induced?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

Check

Two scope channels' gains differ by 1%. About what CMRR does an A − B measurement achieve?

Check

A ground loop puts 50 Hz hum on a measurement. Which fix is unsafe?

Check

A differential amplifier with 60 dB CMRR sees 5 V of common mode.

(a)

What error does that add?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

(b)

And with 20 dB more CMRR?

Check

A 1 cm² loop sits in a 1 mT field at 60 Hz.

(a)

What peak voltage does it pick up?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

(b)

And with the loop area cut by ten (a twisted pair)?

Explain why

Why does a differential measurement reject a ground loop's hum while a single-ended one doesn't?