C11 · Lesson 14 · Network analyzers
Measuring S-parameters, and calibrating the errors away
A spectrum analyzer measures how much signal there is at each frequency. A vector network analyzer measures what a component does to a signal: it sends a known wave in and measures, in magnitude and phase, what reflects and what comes through, sweeping across frequency. Its raw readings are full of the analyzer's own imperfections, and the trick that makes it precise is calibration: measuring known standards and solving for those errors.
- The curves are the range a return-loss measurement can read for a load whose true return loss is on the bottom axis, when the analyzer's directivity is limited.
- Leakage of the incident wave straight into the reflection receiver adds a stray reflection of size D, in any phase. The measured |Γ| lies between |Γ| − D and |Γ| + D.
- Measuring a well-matched load whose return loss approaches the directivity gives meaningless readings. Calibration raises the effective directivity from about 20 dB to 40 dB or more.
Predict
An analyzer has 40 dB of directivity. A load's true return loss is 20 dB. What range can the measurement fall in?
About 19.2 to 20.9 dB. The true |Γ| is 0.1 and the stray reflection 0.01, so the reading is between 0.09 and 0.11. The closer the load's return loss to the directivity, the wider the spread: a 40 dB load could read anywhere from 34 dB to near infinity.
Name it
How a VNA works
A source sweeps frequency. Directional couplers sample the wave going toward the device and the wave coming back, and receivers measure each as a phasor. The ratios give the S-parameters: reflected over incident at port 1 is , transmitted to port 2 over incident is .
The errors
A real analyzer adds systematic errors, modeled for each port by:
- Directivity: incident wave leaking into the reflection receiver, a false reflection.
- Source match: the port's own imperfect 50 Ω, which re-reflects waves coming back from the device.
- Reflection tracking: the frequency response of the reflection path.
and for transmission, load match and transmission tracking, plus isolation. Twelve terms in all for two ports.
Calibration
Measuring known standards lets the analyzer solve for those terms at every frequency and remove them. SOLT uses a short, an open, a matched load at each port and a thru between them; TRL uses a thru, a reflection and a length of line, and suits fixtures and wafers. Calibration moves the reference plane to where the standards were connected; anything beyond it, such as an adapter, must be removed by port extension, which subtracts its electrical delay.
Dynamic range
The largest and smallest transmission an analyzer can measure is set by its source power and its receiver noise floor, which falls 10 dB for every tenfold narrowing of the IF bandwidth, at the cost of a slower sweep.
Worked example
A raw analyzer has 25 dB of directivity; after a one-port calibration its effective directivity is 46 dB. A load reads 26 dB return loss. How well is it known, before and after calibration?
Show the solution
- Stray term after calibration.
- The reading.
- Range of the true value. The stray term can add to the reading or subtract from it: That's 25.2 to 26.9 dB.
- Before calibration. The stray term, 0.056, is larger than the reading itself, so the true return loss could be anything from 19.5 dB to a perfect match.
- What calibration does. Calibration leaves the hardware as it is: it measures the error terms and subtracts them.
Answer after calibration the return loss is known to be 25.2 to 26.9 dB; before, it could be anything from 19.5 dB to a perfect match.
Worked example
A 5 cm air-filled adapter is left past the calibration plane at 3 GHz. What phase error does it add to ?
Show the solution
- Wavelength. so 5 cm is half a wavelength one way.
- Round trip. The reflection travels it twice: a full wavelength, 360°.
- Across a sweep. That happens to look like no error at exactly 3 GHz, but it rotates the phase by 120° per GHz across a sweep.
- Port extension. It removes the error by subtracting the adapter's electrical delay:
Answer 360°, which looks like no error at exactly 3 GHz but rotates the phase by 120° per GHz across a sweep.
Worked example
An analyzer outputs 0 dBm and its receiver floor is −110 dBm at 10 Hz IF bandwidth. What dynamic range does it have at 10 Hz and at 1 kHz?
Show the solution
- At 10 Hz.
- At 1 kHz. The noise bandwidth is 100 times wider, so the floor rises by 20 dB to −90 dBm: Each point is measured 100 times faster.
- The trade. Measuring a filter's deep stopband means choosing between speed and depth.
Answer 110 dB of dynamic range at 10 Hz, and 90 dB at 1 kHz.
Check
A load's true return loss is 20 dB, measured with 30 dB of directivity.
(a)
What is the lowest return loss it might read?
(b)
And the highest?
Check
An adapter 5 cm long with = 2.25 sits beyond the calibration plane at 1 GHz.
(a)
What round-trip delay does port extension subtract?
Write units the way you normally would: 1.5 ps and 0.0015 ns are the same answer.
(b)
How many degrees of phase does it add to at 1 GHz?
Check
An analyzer outputs 0 dBm and its floor is −120 dBm at 10 Hz IF bandwidth. It's set to 100000 Hz.
(a)
What is the floor at 100000 Hz?
(b)
What dynamic range does that leave?
Check
A passive two-port measures = −15 dB and = −0.5 dB.
(a)
What percentage of the incident power reflects?
(b)
What percentage is lost inside it?
Check
Which set of standards does an SOLT calibration of a two-port analyzer use?
Check
A load reflects |Γ| = g, and an analyzer's leakage adds a stray reflection D in an unknown phase.
Step 1
What is the largest |Γ| it might read?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
And the smallest?
Explain why
Why can a network analyzer measure far more accurately than its hardware is built?
Most of an analyzer's errors are systematic: leakage, mismatch and frequency response that stay the same from one measurement to the next. Measuring standards whose responses are known exactly, a short, an open, a load and a thru, gives enough equations at each frequency to solve for every error term. The analyzer then removes those errors from every later measurement by computation. What's left are the errors that don't repeat (noise, drift, connector variation) and the imperfection of the standards themselves, which are far smaller than the raw hardware errors, so a 20 dB directivity becomes 40 dB or more after calibration.