Check
A load's true return loss is 20 dB, measured with 30 dB of directivity.
(a)
What is the lowest return loss it might read?
(b)
And the highest?
Check
An adapter 5 cm long with = 2.25 sits beyond the calibration plane at 1 GHz.
(a)
What round-trip delay does port extension subtract?
Write units the way you normally would: 1.5 ps and 0.0015 ns are the same answer.
(b)
How many degrees of phase does it add to at 1 GHz?
Check
An analyzer outputs 0 dBm and its floor is −120 dBm at 10 Hz IF bandwidth. It's set to 100000 Hz.
(a)
What is the floor at 100000 Hz?
(b)
What dynamic range does that leave?
Check
A passive two-port measures = −15 dB and = −0.5 dB.
(a)
What percentage of the incident power reflects?
(b)
What percentage is lost inside it?
Check
Lowest reading for an 18 dB load with 35 dB directivity?
Check
Phase from a 2 cm air adapter at 2 GHz, round trip?
Check
Floor at 100 Hz if it's −115 dBm at 10 Hz?