C116 of 30
0 XP
  1. Hook
  2. Problems

C11 · Problem set 1

Problem set 1: uncertainty, calibration and meters

These problems use lessons 1 to 5: accuracy, precision and uncertainty, propagating uncertainty, calibration and uncertainty budgets, multimeters in practice, and measuring impedance. Uncertainties are standard uncertainties, combined in quadrature when independent, unless a problem asks for the worst case.

Check

A meter specified as ±(0.03% + 3 counts), with a count of 0.01 V on its range, reads 9.87 V.

(a)

What is the uncertainty?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

What is that as a percentage of the reading?

Check

Readings scatter with σ = 0.02 V.

(a)

What is the standard error of an average of 10?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

How many readings would bring it to 0.05 V?

Check

A 4½-digit meter is on its 2 V range.

(a)

How many counts can it show?

(b)

What is one count worth, in millivolts?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

Independent measurements: absolute uncertainties 0.01 and 0.02; relative uncertainties 1% and 3%.

(a)

What is the uncertainty of the difference of the first two?

(b)

What is the relative uncertainty of the product of the last two?

Check

Power is computed as P=VI, and resistance as R=V/I, from the same V (0.2%) and I (0.5%).

(a)

What is P's relative uncertainty?

(b)

What would P=V2/R's be, with R known independently to 0.5%?

Check

A filter's corner 1/(2πRC) uses R to 0.1% and C to 1%.

(a)

What is the corner's relative uncertainty?

(b)

What worst-case tolerance must a production design allow?

Check

A 1 MΩ meter measures a node with 200 kΩ of source resistance.

(a)

By how many percent does it read low?

(b)

Below what source resistance would the error stay under about 1%?

Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.

Check

An ammeter with a 0.5 Ω shunt measures 50 mA in a 3.3 V circuit.

(a)

What burden voltage does it drop?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

(b)

What fraction of the supply is that?

Check

A 0.25 Ω resistor is measured through leads of 0.08 Ω each.

(a)

What does a two-wire measurement read?

Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.

(b)

By what percentage is it too high?

Check

Waveforms of amplitude 2 V.

(a)

What is a triangle wave's true RMS?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

What is the RMS of 1 V DC plus a sine of amplitude 2 V?

Check

An average-responding meter (calibrated for sines) reads waveforms of amplitude 2 V.

(a)

What does it show for a ±2 V square wave?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

By what percentage does it overstate the square wave's true RMS?

Check

Pulses of height 5 V have duty cycle 0.1.

(a)

What is their RMS value?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

What crest factor must a true-RMS meter handle?

Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.

Practice

Keep practicing

15 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.