C11 Measurement and lab practice · Practice
Problem set 1: uncertainty, calibration and meters
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
A meter specified as ±(0.03% + 3 counts), with a count of 0.01 V on its range, reads 9.87 V.
(a)
What is the uncertainty?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What is that as a percentage of the reading?
Check
Readings scatter with σ = 0.02 V.
(a)
What is the standard error of an average of 10?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
How many readings would bring it to 0.05 V?
Check
A 4½-digit meter is on its 2 V range.
(a)
How many counts can it show?
(b)
What is one count worth, in millivolts?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Independent measurements: absolute uncertainties 0.01 and 0.02; relative uncertainties 1% and 3%.
(a)
What is the uncertainty of the difference of the first two?
(b)
What is the relative uncertainty of the product of the last two?
Check
Power is computed as , and resistance as , from the same V (0.2%) and I (0.5%).
(a)
What is P's relative uncertainty?
(b)
What would 's be, with R known independently to 0.5%?
Check
A filter's corner uses R to 0.1% and C to 1%.
(a)
What is the corner's relative uncertainty?
(b)
What worst-case tolerance must a production design allow?
Check
A 1 MΩ meter measures a node with 200 kΩ of source resistance.
(a)
By how many percent does it read low?
(b)
Below what source resistance would the error stay under about 1%?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
An ammeter with a 0.5 Ω shunt measures 50 mA in a 3.3 V circuit.
(a)
What burden voltage does it drop?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What fraction of the supply is that?
Check
A 0.25 Ω resistor is measured through leads of 0.08 Ω each.
(a)
What does a two-wire measurement read?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
By what percentage is it too high?
Check
Waveforms of amplitude 2 V.
(a)
What is a triangle wave's true RMS?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What is the RMS of 1 V DC plus a sine of amplitude 2 V?
Check
An average-responding meter (calibrated for sines) reads waveforms of amplitude 2 V.
(a)
What does it show for a ±2 V square wave?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
By what percentage does it overstate the square wave's true RMS?
Check
Pulses of height 5 V have duty cycle 0.1.
(a)
What is their RMS value?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What crest factor must a true-RMS meter handle?
Check
A budget has a specification of ±0.02 (rectangular), a certificate with U = 0.01 (k = 2), and 20 readings with s = 0.02.
(a)
What is the combined standard uncertainty?
(b)
What is the expanded uncertainty at k = 2?
Check
A bridge balances with = 1 kΩ, = 3.3 kΩ and = 4.7 kΩ.
(a)
What is ?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
If each known resistor is uncertain by 0.1% (standard), how uncertain is ?
Check
A 4.7 µF capacitor has 0.02 Ω ESR and 15 nH ESL.
(a)
Where is its self-resonance?
Write units the way you normally would: 1.5 kHz, 1500 Hz and 0.0015 MHz are the same answer.
(b)
What is its impedance there?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in C11