C11 Measurement and lab practice · Practice
Problem set 3: spectra, networks, power, noise and grounding
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Check
An analyzer with a 12 dB noise figure is set to 100 Hz RBW.
(a)
What noise floor does it show?
(b)
What would it show at a tenth of that RBW?
Check
A swept analyzer covers 10 MHz at 0.1 kHz RBW (k = 2).
(a)
About how long is a sweep?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
(b)
How long at three times the RBW?
Check
A carrier is at 3 dBm, its second harmonic at −40 dBm and its third at −60 dBm.
(a)
What is the second harmonic in dBc?
(b)
What is the carrier's power in milliwatts?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
A repeating signal with 5 mV of random noise is averaged over 100 sweeps.
(a)
What noise is left?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
By how many dB has the SNR improved?
Check
A lock-in with time constant 1 s measures a source with 5 nV/√Hz of noise.
(a)
What is its noise bandwidth?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What noise does it see?
Check
A resistor of 10000 Ω at 300 K is measured over 1 Hz.
(a)
What is its thermal noise?
Write units the way you normally would: 1.5 nV, 1500 pV and 0.0015 µV are the same answer.
(b)
What is its noise density, in nV/√Hz?
Check
A differential amplifier with 50 dB CMRR sees 5 V of common mode.
(a)
What input-referred error results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What CMRR would cut that error tenfold?
Check
A loop of 50 cm² sits in a 0.5 mT field at 50 Hz.
(a)
What does it pick up?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What would twisting the wires to cut the effective area tenfold leave?
Check
Two probes and A − B math measure a floating node with 1 V of common mode; the channels' gains differ by 0.2%.
(a)
What CMRR does that give?
(b)
What error appears on the difference?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A −60 dBm tone is measured on an analyzer with 4 dB noise figure at 1000 Hz RBW.
(a)
What SNR does the display show?
(b)
What is the tone's power in picowatts?
Write units the way you normally would: 1.5 pW and 0.0015 nW are the same answer.
Check
A measurement needs 20 dB more SNR.
(a)
How many sweeps must be averaged?
(b)
By what factor must a lock-in's time constant grow instead (noise bandwidth shrinking in proportion)?
Check
A ground loop drives 20 mA through a 0.01 Ω shield; the signal is 10 mV.
(a)
How much hum does a single-ended measurement show?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What is the signal-to-hum ratio, in dB?
Check
A VNA with 30 dB directivity measures |Γ| = 0.1.
(a)
What stray reflection does the directivity add?
(b)
What is that as a percentage of the reading?
Check
A meter reads apparent power of 20 kVA on a load with power factor 0.5.
(a)
What is the real power?
Write units the way you normally would: 1.5 kW, 1500 W and 0.0015 MW are the same answer.
(b)
By how much does the apparent power overstate it?
Check
A signal averages 17 dBm with 6 dB PAPR.
(a)
What is its peak power in dBm?
(b)
In milliwatts?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
A low-level setup reads 230 µV forward and −88 µV reversed; elsewhere, a 50 V node is insulated by 100 GΩ.
(a)
What's the true signal?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
How much current leaks through the insulation, unguarded?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to the lesson · Practice every topic in C11