S15 · Lesson 14 · Biopotential electrodes
How electrodes turn ionic current into electronic current, half-cell potentials and the offsets they create, polarizable and nonpolarizable electrodes, the double-layer equivalent circuit and its impedance against frequency, skin, motion artifact, and microelectrodes
In the body, current is carried by ions; in the amplifier, by electrons. Every biopotential measurement depends on the junction where one becomes the other, and that junction is a small electrochemical cell with its own voltage, its own impedance and its own noise. A good electrode, like the silver/silver-chloride snap on an ECG pad, keeps its voltage steady and its impedance low. A poor one adds hundreds of millivolts of drifting offset and turns every movement of the patient into a signal larger than the heartbeat.
- The magnitude of an electrode's impedance against frequency, from the model : a series resistance (electrolyte and gel) plus the charge-transfer resistance in parallel with the double-layer capacitance.
- At low frequencies the impedance is ; above the corner the capacitor shorts out and only remains.
- Biopotentials sit at the low end, where the impedance is largest, so amplifier input impedance must be far larger still.
Predict
Two Ag/AgCl electrodes on the chest show a DC offset of 300 mV between them, the most a standard ECG input must tolerate. The amplifier's first stage runs from ±5 V supplies. What's the largest first-stage gain that avoids saturating on the offset alone?
300 mV times a gain of 16.7 reaches 5 V, so the first stage can only have a gain of about 16 (less with real output swing). The rest of the gain must come after a high-pass filter removes the DC offset, which is why biopotential amplifiers are built in stages.
Name it
The electrode–electrolyte interface
At a metal in contact with an electrolyte, oxidation and reduction reactions exchange charge, and ions arrange themselves into a charged double layer a few nanometers thick. At equilibrium the interface sits at its half-cell potential, measured against the standard hydrogen electrode: about +0.223 V for Ag/AgCl, +0.80 V for silver, −0.76 V for zinc. Two identical electrodes cancel, but differences in skin, gel and contact leave offsets of tens to hundreds of millivolts.
Polarizable and nonpolarizable electrodes
A perfectly polarizable electrode (platinum, roughly) passes no real charge across the interface; it behaves like a capacitor, and current shifts its voltage (the overpotential). A nonpolarizable electrode passes current freely without changing its potential. Ag/AgCl comes close: the silver chloride layer exchanges chloride ions with the gel, giving a stable, low-noise, low-drift contact, which makes it the standard for ECG.
The equivalent circuit
with the half-cell potential in series. is the electrolyte and gel; and model charge transfer and the double layer. The stratum corneum of dry skin adds a similar, larger RC section, which abrasion and gel reduce.
Motion artifact
Moving the electrode disturbs the double layer and the skin's own potential, producing low-frequency artifacts of millivolts or more. Ag/AgCl with gel, firm attachment and good skin preparation reduce them; a high-pass filter removes what's slower than the ECG.
Microelectrodes
Glass micropipettes and fine metal tips for recording inside single cells have megohm resistances. Together with stray capacitance of picofarads they form a low-pass filter, , that can blur fast action potentials unless the amplifier cancels the capacitance.
Worked example
An electrode has = 1 kΩ, = 100 kΩ and = 0.1 µF. Find its impedance at 0.1 Hz, 10 Hz and 1 kHz, and its corner frequency.
Show the solution
- Corner frequency.
- Impedance at each frequency. approaching at 1 kHz.
- Where the ECG sits. The ECG band sits mostly below the corner, at the high-impedance end.
Answer the corner is at 15.9 Hz, and |Z| is about 101 kΩ at 0.1 Hz, 85.5 kΩ at 10 Hz and 1.89 kΩ at 1 kHz.
Worked example
A 100 kΩ electrode drives an amplifier input of 10 MΩ. What fraction of the signal is lost? And what bandwidth does a 10 MΩ microelectrode with 10 pF of stray capacitance allow?
Show the solution
- Divider loss. at low frequencies, and much more if the two electrodes differ, as lesson 15 shows.
- Microelectrode bandwidth. low enough to round off a 1 ms action potential.
Answer 0.99% of the signal is lost, and the microelectrode's bandwidth is 1.59 kHz.
Worked example
A pair of electrodes presents 10 kΩ of resistance at body temperature. How much thermal noise does that add over the 0.05 to 150 Hz diagnostic band, compared with a 1 mV ECG?
Show the solution
- Noise power. For 10 kΩ at 310 K over the 150 Hz band:
- Noise voltage. nearly 80 dB below the 1 mV ECG.
- What limits surface ECG. Thermal noise is rarely the limit; motion artifact and fluctuations of the half-cell potential, often tens of microvolts, are.
- EEG. At 10 to 100 µV, EEG is small enough that the same 0.16 µV starts to matter, and electrodes for small signals are prepared to bring their impedance down.
Answer the electrodes add 0.16 µV RMS of thermal noise, nearly 80 dB below a 1 mV ECG.
Check
An electrode has = 0.5 kΩ, = 50 kΩ and = 0.5 µF. What's the magnitude of its impedance at 100 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
Where's the corner frequency of an electrode with = 100 kΩ and = 0.02 µF?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Electrodes present a 300 mV DC offset to an amplifier stage whose output can swing ±2.5 V. What's the largest DC-coupled gain it can use?
Check
Two electrodes with half-cell potentials 0.223 V and −0.76 V share an electrolyte. What voltage appears between them?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A microelectrode of 10 MΩ sees 2 pF of stray capacitance. Where's the resulting low-pass corner?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A 100 kΩ electrode drives an amplifier with 1 MΩ input resistance. What percent of the signal is lost?
Check
An electrode pair presents 20 kΩ at 37 °C, recorded over 40 Hz of bandwidth, for a 1000 µV signal.
(a)
How much thermal noise does it add?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What signal-to-noise ratio does that leave, considering only this noise?
Check
An electrode's model has series resistance , and a double layer of resistance in parallel with capacitance (kΩ and µF).
Step 1
What impedance does it present at DC, in kΩ?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
And at very high frequencies?
Step 3
Where is the corner between them, in hertz?
Explain why
Why are Ag/AgCl electrodes the standard for recording the ECG?
An electrode must pass small currents between ionic tissue and electronic wires without its own potential changing, because any change in its half-cell potential looks like signal. Silver coated with silver chloride in a chloride gel is nearly nonpolarizable: the reaction Ag + Cl⁻ ⇌ AgCl + e⁻ runs freely in both directions, so charge crosses the interface easily and the potential stays fixed by the chloride concentration. That gives a stable, reproducible half-cell potential, low impedance and low noise, and much smaller motion artifacts than polarizable metals like stainless steel or platinum, whose double layer behaves like a capacitor whose charge shifts whenever the electrode moves.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.