S15 · Lesson 14 · Biopotential electrodes

How electrodes turn ionic current into electronic current, half-cell potentials and the offsets they create, polarizable and nonpolarizable electrodes, the double-layer equivalent circuit and its impedance against frequency, skin, motion artifact, and microelectrodes

In the body, current is carried by ions; in the amplifier, by electrons. Every biopotential measurement depends on the junction where one becomes the other, and that junction is a small electrochemical cell with its own voltage, its own impedance and its own noise. A good electrode, like the silver/silver-chloride snap on an ECG pad, keeps its voltage steady and its impedance low. A poor one adds hundreds of millivolts of drifting offset and turns every movement of the patient into a signal larger than the heartbeat.

  • The magnitude of an electrode's impedance against frequency, from the model Rs+Rd‖Cd: a series resistance (electrolyte and gel) plus the charge-transfer resistance in parallel with the double-layer capacitance.
  • At low frequencies the impedance is Rs+Rd; above the corner 1/(2πRdCd) the capacitor shorts out Rd and only Rs remains.
  • Biopotentials sit at the low end, where the impedance is largest, so amplifier input impedance must be far larger still.

Predict

Two Ag/AgCl electrodes on the chest show a DC offset of 300 mV between them, the most a standard ECG input must tolerate. The amplifier's first stage runs from ±5 V supplies. What's the largest first-stage gain that avoids saturating on the offset alone?

Name it

The electrode–electrolyte interface

At a metal in contact with an electrolyte, oxidation and reduction reactions exchange charge, and ions arrange themselves into a charged double layer a few nanometers thick. At equilibrium the interface sits at its half-cell potential, measured against the standard hydrogen electrode: about +0.223 V for Ag/AgCl, +0.80 V for silver, −0.76 V for zinc. Two identical electrodes cancel, but differences in skin, gel and contact leave offsets of tens to hundreds of millivolts.

Polarizable and nonpolarizable electrodes

A perfectly polarizable electrode (platinum, roughly) passes no real charge across the interface; it behaves like a capacitor, and current shifts its voltage (the overpotential). A nonpolarizable electrode passes current freely without changing its potential. Ag/AgCl comes close: the silver chloride layer exchanges chloride ions with the gel, giving a stable, low-noise, low-drift contact, which makes it the standard for ECG.

The equivalent circuit

Z(f)=Rs+Rd1+j2πfRdCd

with the half-cell potential in series. Rs is the electrolyte and gel; Rd and Cd model charge transfer and the double layer. The stratum corneum of dry skin adds a similar, larger RC section, which abrasion and gel reduce.

Motion artifact

Moving the electrode disturbs the double layer and the skin's own potential, producing low-frequency artifacts of millivolts or more. Ag/AgCl with gel, firm attachment and good skin preparation reduce them; a high-pass filter removes what's slower than the ECG.

Microelectrodes

Glass micropipettes and fine metal tips for recording inside single cells have megohm resistances. Together with stray capacitance of picofarads they form a low-pass filter, fc=1/(2πRC), that can blur fast action potentials unless the amplifier cancels the capacitance.

Worked example

An electrode has Rs = 1 kΩ, Rd = 100 kΩ and Cd = 0.1 µF. Find its impedance at 0.1 Hz, 10 Hz and 1 kHz, and its corner frequency.

Show the solution
  1. Corner frequency. fc=12πRdCd=12π×105×10−7Hz=15.9Hz
  2. Impedance at each frequency. |Z(0.1Hz)|≈101kΩ|Z(10Hz)|=85.5kΩ|Z(1kHz)|=1.89kΩ approaching Rs at 1 kHz.
  3. Where the ECG sits. The ECG band sits mostly below the corner, at the high-impedance end.

Answer the corner is at 15.9 Hz, and |Z| is about 101 kΩ at 0.1 Hz, 85.5 kΩ at 10 Hz and 1.89 kΩ at 1 kHz.

Worked example

A 100 kΩ electrode drives an amplifier input of 10 MΩ. What fraction of the signal is lost? And what bandwidth does a 10 MΩ microelectrode with 10 pF of stray capacitance allow?

Show the solution
  1. Divider loss. loss=100k100k+10M=0.99% at low frequencies, and much more if the two electrodes differ, as lesson 15 shows.
  2. Microelectrode bandwidth. fc=12π×107×10−11Hz=1.59kHz low enough to round off a 1 ms action potential.

Answer 0.99% of the signal is lost, and the microelectrode's bandwidth is 1.59 kHz.

Worked example

A pair of electrodes presents 10 kΩ of resistance at body temperature. How much thermal noise does that add over the 0.05 to 150 Hz diagnostic band, compared with a 1 mV ECG?

Show the solution
  1. Noise power. For 10 kΩ at 310 K over the 150 Hz band: vn2=4kTRB=4×1.38×10−23×310×104×150V2
  2. Noise voltage. vn=0.16μV rms nearly 80 dB below the 1 mV ECG.
  3. What limits surface ECG. Thermal noise is rarely the limit; motion artifact and fluctuations of the half-cell potential, often tens of microvolts, are.
  4. EEG. At 10 to 100 µV, EEG is small enough that the same 0.16 µV starts to matter, and electrodes for small signals are prepared to bring their impedance down.

Answer the electrodes add 0.16 µV RMS of thermal noise, nearly 80 dB below a 1 mV ECG.

Check

An electrode has Rs = 0.5 kΩ, Rd = 50 kΩ and Cd = 0.5 µF. What's the magnitude of its impedance at 100 Hz?

Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.

Check

Where's the corner frequency of an electrode with Rd = 100 kΩ and Cd = 0.02 µF?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

Check

Electrodes present a 300 mV DC offset to an amplifier stage whose output can swing ±2.5 V. What's the largest DC-coupled gain it can use?

Check

Two electrodes with half-cell potentials 0.223 V and −0.76 V share an electrolyte. What voltage appears between them?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

Check

A microelectrode of 10 MΩ sees 2 pF of stray capacitance. Where's the resulting low-pass corner?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

Check

A 100 kΩ electrode drives an amplifier with 1 MΩ input resistance. What percent of the signal is lost?

Check

An electrode pair presents 20 kΩ at 37 °C, recorded over 40 Hz of bandwidth, for a 1000 µV signal.

(a)

How much thermal noise does it add?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

(b)

What signal-to-noise ratio does that leave, considering only this noise?

Check

An electrode's model has series resistance Rs, and a double layer of resistance Rd in parallel with capacitance Cd (kΩ and µF).

Step 1

What impedance does it present at DC, in kΩ?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why are Ag/AgCl electrodes the standard for recording the ECG?