S15 · Lesson 30 · Cardiovascular lab
A catheter system's natural frequency and damping from its dimensions with and without a bubble, systolic, mean and diastolic pressures from an oscillometric envelope, cardiac output from a dye curve with recirculation removed, and valve areas from gradients
Lessons 23 to 28 measured the circulation with catheters, cuffs, indicators and flowmeters, each through a formula. A monitor runs those formulas on raw data: a catheter's dimensions, a cuff's oscillation amplitudes, a dye concentration sampled every tenth of a second. In this lab you'll write the processing a bedside monitor does, and find where each method's assumptions bend the answer.
- A catheter and transducer form a second-order system. The curve is its amplitude ratio against frequency for the natural frequency and damping ratio on the sliders.
- Pressure waveforms carry harmonics to 20 Hz or so. Without a bubble a good system's is far above that; a tiny bubble drags it down near the harmonics and exaggerates them.
- Damping near 0.64 keeps the response flat over the widest fraction of .
Predict
A 1 m saline catheter of 0.5 mm radius feeds a transducer of compliance 0.01 mm³ per 100 mmHg. A 1 mm³ air bubble gets into the line. What happens to the natural frequency?
The bubble's compliance, = 1 mm³/101 kPa = 9.9 × 10⁻¹⁵ m³/Pa, is 13 times the transducer's 7.5 × 10⁻¹⁶ m³/Pa, so the total compliance rises fourteenfold, and , which goes as its inverse square root, falls 3.8 times: from 163 Hz to 43 Hz, as in lesson 23. Task 1 computes both from the dimensions.
Name it
What the lab computes
- Catheter dynamics: for a catheter of radius r and length L feeding a transducer with volume elasticity (a bubble's compliance, , adds to the transducer's), and .
- Oscillometry: the mean pressure is the cuff pressure at the largest oscillation; systolic and diastolic pressures are where the amplitude falls to 0.55 and 0.85 of the peak, above and below the mean, found by linear interpolation between samples.
- Indicator dilution: fit a line to ln c over the downslope between 70% and 30% of the peak, giving the exponential time constant τ; the area is the trapezoid sum up to the last fitted point plus the extrapolated tail , and the cardiac output is the dose over the area.
- Valves: the flow during ejection, F = (cardiac output)/(heart rate × ejection time), and the area .
Check
Write catheter(r, L, Ed, rho=1000, eta=0.001) returning (, ζ), and with_bubble(Ed, Vb, Patm=101325) returning the combined elasticity.
Python 3.14 with numpy and matplotlib, running in your browser. Tab indents; press Esc and then Tab to move on.
Check
Write oscillometric(P, A) for cuff pressures P falling with sample index and oscillation amplitudes A, returning (systolic, mean, diastolic).
Python 3.14 with numpy and matplotlib, running in your browser. Tab indents; press Esc and then Tab to move on.
Check
Write stewart_hamilton(t, c, dose): fit an exponential to the downslope between 70% and 30% of the peak, extrapolate the tail, and return (cardiac output in L/min, area, τ), with c in mg/L, t in seconds and the dose in mg.
Python 3.14 with numpy and matplotlib, running in your browser. Tab indents; press Esc and then Tab to move on.
Check
Write valve_area(co, hr, sep, dp_mmhg, cd=0.85, rho=1060) in cm², from cardiac output (L/min), heart rate (per minute), ejection time (s) and mean gradient (mmHg).
Python 3.14 with numpy and matplotlib, running in your browser. Tab indents; press Esc and then Tab to move on.
Explain why
Why is the mean arterial pressure the most reliable number an oscillometric cuff reports?
The cuff senses the artery's volume pulsations through the air in its bladder. The wall of the artery is most compliant when the pressure across it is near zero, that is, when the cuff pressure equals the pressure inside on average, so the oscillations are largest there: the peak marks the mean arterial pressure by physics. Systolic and diastolic pressures have no such marker in the envelope; the monitor estimates them where the amplitude falls to empirically chosen fractions of the peak, and those fractions shift with arterial stiffness and the envelope's shape, so they vary from person to person.