S15 · Lesson 34 · Chemical biosensors
Potentiometric sensors with the Nernstian slope, pH glass electrodes and ion-selective electrodes with their selectivity, blood pH from bicarbonate and carbon dioxide, the amperometric Clark oxygen electrode and Faraday's law, enzyme glucose sensors and their saturation, and ISFETs
A blood-gas analyzer reports pH, oxygen and carbon dioxide within a minute from a drop of blood; a diabetic patient's glucose sensor reports every five minutes for two weeks from a filament under the skin. Underneath are two kinds of electrochemical measurement from lessons 10 and 14. Potentiometric sensors measure the voltage across a membrane that lets only one ion through, and read concentration through the Nernst equation's logarithm. Amperometric sensors hold an electrode at a fixed voltage and measure the current of a reaction, which counts molecules through Faraday's constant.
- An enzyme glucose sensor's current follows Michaelis–Menten kinetics, : linear at low concentration, saturating above .
- The dashed line is the linear calibration a meter would use; the gap shows the error at high glucose.
- Sensors put a diffusion-limiting membrane over the enzyme to raise the apparent and stretch the linear range over the clinical span.
Predict
A pH electrode at 37 °C sees the sample's pH change from 7.4 to 7.1. How much does its voltage change?
A glass electrode follows Nernst's equation for hydrogen ions: 2.303RT/F = 61.5 mV per pH unit at 37 °C. A change of 0.3 pH units gives 18.5 mV. The slope depends on temperature (59.2 mV/pH at 25 °C), which is why blood-gas analyzers hold samples at 37 °C.
Name it
Potentiometric sensors
An ion-selective membrane develops a Nernst potential for its ion against an internal reference solution, read against a reference electrode (Ag/AgCl, lesson 14) by a very high-impedance amplifier:
For pH, at 37 °C. No membrane is perfectly selective; the Nikolsky–Eisenman equation adds interfering ions j with selectivity coefficients:
so an interferent adds an apparent activity .
Blood pH and CO₂
Blood pH is set by the bicarbonate buffer (Henderson–Hasselbalch):
with bicarbonate in mM and in mmHg. The Severinghaus CO₂ electrode puts a pH electrode behind a gas-permeable membrane in a bicarbonate solution, so its pH tracks .
Amperometric sensors
The Clark electrode holds a platinum cathode at about −0.6 V behind an oxygen-permeable membrane; every O₂ molecule reaching it is reduced with four electrons, so the current is , proportional to .
Glucose sensors use glucose oxidase to turn glucose into gluconic acid and hydrogen peroxide, and measure the peroxide's oxidation current at about +0.6 V. Enzyme kinetics saturate:
so a linear calibration reads low by the fraction .
ISFETs
An ion-sensitive field-effect transistor replaces a MOSFET's metal gate with an ion-sensitive insulator in contact with the solution; the threshold voltage shifts with pH by up to the Nernstian 59 mV/pH at 25 °C. ISFETs are small and rugged and are the sensors in some sequencing chips.
Worked example
Blood has 24 mM bicarbonate and a of 40 mmHg. What's its pH? And what does a potassium electrode with = 2 × 10⁻⁴ read in 4 mM K⁺ and 140 mM Na⁺?
Show the solution
- pH.
- Apparent potassium from sodium. a 0.7% error.
- A poorer membrane. One with = 0.01 would read 35% high.
Answer the pH is 7.39, and the potassium electrode reads 0.028 mM high, a 0.7% error.
Worked example
A glucose sensor has = 100 nA and = 20 mM. What does it read at 5 and 10 mM, and how far off is a linear calibration at 5 mM? What current does one picomole of O₂ per second give in a Clark electrode?
Show the solution
- Current at 5 mM.
- Current at 10 mM.
- Linear calibration error. A linear calibration expects 25 nA at 5 mM, so it reads 20% low there.
- Clark electrode current. One picomole per second of O₂, at four electrons each, gives
Answer the sensor reads 20 nA at 5 mM and 33.3 nA at 10 mM, a linear calibration is 20% off at 5 mM, and the Clark electrode gives 386 nA.
Worked example
A Severinghaus CO₂ electrode reads the pH of a thin bicarbonate film behind a gas-permeable membrane. How much does its output change when doubles from 40 to 80 mmHg at 37 °C?
Show the solution
- pH change in the film. The bicarbonate stays fixed, so by Henderson–Hasselbalch the pH falls by
- Electrode output. At 61.5 mV per pH unit: so the sensor is logarithmic in , like any potentiometric sensor in its concentration.
- Response time. Because CO₂ must diffuse through the membrane and the film must equilibrate, it responds in a minute or two, much slower than an oxygen electrode.
Answer the output rises by 18.5 mV.
Check
A pH electrode at 37 °C sees the pH change by 0.1. How much does its voltage change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Blood has 24 mM bicarbonate and a of 30 mmHg. What's its pH?
Check
A potassium electrode with selectivity coefficient = 0.0002 measures 4 mM K⁺ in 140 mM Na⁺. By what percent does it read high?
Check
An amperometric sensor reacts 2 pmol/s of analyte, with 2 electrons per molecule. What current flows?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A glucose sensor has = 100 nA and = 10 mM. What current does it give at 15 mM?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A sensor with = 20 mM is calibrated with its initial linear slope. By what percent does it read low at 4 mM?
Check
A Severinghaus electrode at 37 °C sees rise from 20 to 60 mmHg.
(a)
By how much does the film's pH fall?
(b)
How much does its voltage change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Blood has b mM bicarbonate and a of p mmHg.
Step 1
What is its pH?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What would give a pH of t with the same bicarbonate?
Explain why
Why do potentiometric sensors read the logarithm of concentration while amperometric sensors read it linearly?
A potentiometric sensor draws essentially no current; its membrane comes to an equilibrium where the electrical work of moving an ion balances the change in its chemical potential, and chemical potential depends on the logarithm of activity, so the voltage follows the Nernst equation's logarithm: about 60 mV per tenfold change, whatever the absolute level. That gives a huge range but poor resolution of small changes. An amperometric sensor drives a reaction to completion at the electrode, so every molecule that diffuses to it is counted as a fixed number of electrons, and the current is proportional to the rate molecules arrive, which is proportional to their concentration, until enzyme kinetics or diffusion limits saturate it.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.