S15 · Lesson 12 · The electrocardiogram
The heart's conduction system and the waves of the ECG, the heart vector, Einthoven's limb leads and his law, the augmented leads and the hexaxial reference system, finding the electrical axis, and reading rate and intervals from the trace
Each heartbeat starts at a small cluster of pacemaker cells and spreads through the atria, pauses at the atrioventricular node, then races down specialized fibers into the ventricles. At every instant the wavefront of depolarizing muscle acts like one large current dipole, the heart vector, and lesson 11's volume conduction carries its potential to the skin. Einthoven's insight, in 1903, was that potential differences between the arms and leg project that vector onto three fixed directions. From those projections a clinician reads the heart's rate, rhythm, conduction times and electrical axis.
- The heart vector of size M at angle θ in the frontal plane (0° points to the patient's left, +90° straight down), and the three limb-lead axes: I at 0°, II at 60°, III at 120°.
- Each lead records the vector's projection onto its axis: .
- The readouts obey Einthoven's law, I + III = II, at every angle.
Predict
At the peak of the QRS complex, lead I reads 0.6 mV and lead II reads 1.2 mV. What does lead III read?
Lead I is , lead II is and lead III is , so I + III = II identically: III = 1.2 − 0.6 = 0.6 mV. Only two of the three limb leads carry independent information.
Name it
The heartbeat and its waves
The sinoatrial node fires; atrial depolarization gives the P wave. Conduction pauses in the atrioventricular node (the flat PR segment), then the His bundle and Purkinje fibers spread excitation through the ventricles, giving the large, fast QRS complex. Ventricular repolarization gives the T wave. Typical adult values: PR interval 0.12 to 0.20 s, QRS under 0.12 s, peak amplitudes around 1 mV. Diagnostic ECGs record 0.05 to 150 Hz; monitors use a narrower band.
Limb leads and Einthoven's law
With electrodes on the right arm (RA), left arm (LA) and left leg (LL):
Treating the electrodes as corners of an equilateral triangle around the heart, lead I lies at 0°, II at 60° and III at 120°, and each records .
Augmented leads
Each augmented lead compares one limb with the average of the other two: , and , along −150°, −30° and +90°. Projected, they read . With the limb leads they make the six-axis hexaxial system. The precordial leads V1 to V6 measure chest electrodes against the average of the three limbs (Wilson's central terminal) and see the horizontal plane.
The electrical axis
From I = M cos θ and II = M cos(θ − 60°):
Normal adult axes fall between about −30° and +90°.
Rate and intervals
At the standard 25 mm/s paper speed, an R-to-R spacing of D millimeters means a heart rate of 1500/D per minute. The QT interval shortens as the rate rises, so it's corrected, for instance by Bazett's formula with RR in seconds.
Worked example
Finish the predict question: the augmented leads, the axis and the size of the heart vector.
Show the solution
- Lead III.
- Augmented leads.
- Axis. so θ = 60°.
- Size of the heart vector. pointing straight along lead II.
- Check against the leads. That's why II is the largest and aVL, perpendicular to it, is zero.
Answer III = 0.6 mV, aVF = 0.9 mV, aVL = 0 and aVR = −0.9 mV; the axis is 60°, and the heart vector is 1.2 mV, along lead II.
Worked example
R waves are 20 mm apart at 25 mm/s and the QT interval is 0.36 s. Find the rate and the corrected QT.
Show the solution
- RR interval.
- Rate.
- Corrected QT. within the normal range (roughly up to 0.44 to 0.46 s).
Answer the rate is 75 per minute, and the corrected QT is 0.402 s, within the normal range.
Worked example
A 24-hour Holter recorder stores three ECG channels at 250 samples a second and 12 bits each. How much data does it collect, and what compression ratio would fit it in 32 MB?
Show the solution
- Data per day.
- Compression needed. so fitting 32 MB needs about 3:1 compression.
- Why lossless methods get there. Successive ECG samples differ little, so storing differences instead of values cuts most samples to a few bits.
- Sample rate. The 250 Hz rate covers monitoring bandwidth; diagnostic recordings sample at 500 Hz or more to keep the 150 Hz band and the fine detail of pacemaker spikes.
Answer the recorder collects 7.78 × 10⁸ bits, 97.2 MB, a day, so fitting it in 32 MB needs about 3:1 compression.
Check
Lead I reads 0.3 mV and lead II reads 0.5 mV. What does lead III read?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
With lead I = 0.3 mV and lead II = 0.5 mV, what does aVF read?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Lead I reads 0.4 mV and lead II reads 0.3 mV. What's the electrical axis?
Check
A heart vector of 0.8 mV points at 20°. What does a lead along 120° record?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
R waves are 20 mm apart on a trace recorded at 25 mm/s. What's the heart rate per minute?
Check
A QT interval of 0.38 s is measured at RR = 0.6 s. What's Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A recorder stores 3 ECG channels at 250 samples a second, 24 bits each, for 48 hours.
(a)
How much data does it collect?
(b)
What compression ratio fits it into 64 MB?
Check
Lead I reads a and lead II reads b.
Step 1
What does lead III read?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What does aVR read?
Step 3
What does aVL read?
Explain why
Why do the three limb leads carry only two independent signals, and what does that say about the heart vector?
The three limb leads are differences between the same three electrode potentials, so they're linearly dependent: lead I plus lead III equals lead II at every instant. In Einthoven's model the frontal-plane heart vector has two components, so two leads at different angles are enough to determine it completely, and every other frontal lead, including the augmented ones, is a combination of those two. The redundancy is still useful: it lets a clinician see the projection that best shows a particular wave, and it gives a check on electrode placement. To capture the third, front-to-back component, the precordial chest leads are needed.
Practice
Keep practicing
10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.