S15 · Final exam
Final exam: biomedical instrumentation
Twelve problems from the whole course: sensors and their dynamics, temperature measurement, membrane potentials and conduction, the ECG, electrodes and biopotential amplifiers, interference, signal processing, blood pressure and flow, respiration, oximetry and chemical sensors, the clinical laboratory, image quality and medical imaging, therapeutic devices, electrical safety, diagnostic accuracy and device power budgets. Give yourself three hours and work without the lessons open. "Another set" deals a different exam.
Check
A transducer of 5 µV/V/mmHg at 2.5 V; another with output µV (S = 30, a = 0.002) over 0 to 150 mmHg.
(a)
First transducer's output at 240 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Second one's nonlinearity in percent of full scale?
Check
A probe with τ = 5 s meets a 2 °C step; a catheter has = 60 Hz and ζ = 0.1.
(a)
Probe error after 25 s?
(b)
Catheter amplitude ratio at 8 Hz?
(c)
Catheter step overshoot, percent?
Check
Gauges (G = 2) at ±250 microstrain in a bridge at 10 V with 4 active arms; a piezoelectric sensor (d = 100 pC/N) into a charge amplifier with = 100 pF.
(a)
Bridge output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Charge amplifier output for 0.5 N?
Check
A thermistor of 3 kΩ at 25 °C, B = 3300 K, at 41 °C carrying 150 µA with δ = 1 mW/°C.
(a)
Resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Coefficient in %/°C?
(c)
Self-heating?
Check
K⁺ at 130 mM inside and 4 mM outside, Na⁺ at 12 and 145 mM, = 0.1, 37 °C; an axon 10 µm across with = 8000 Ω·cm² and = 70 Ω·cm.
(a)
Potassium Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
GHK potential?
(c)
Length constant?
Check
Lead I = 0.3 mV, lead II = 0.7 mV; R waves 30 mm apart at 25 mm/s; QT = 0.42 s.
(a)
Lead III?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Axis?
(c)
Bazett QTc?
Check
An electrode with = 0.5 kΩ, = 50 kΩ, = 0.3 µF; an instrumentation amplifier with = 3 kΩ and = 25 kΩ.
(a)
Electrode impedance at 40 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Amplifier first-stage gain?
Check
5 mV of common mode, 10 kΩ electrode mismatch, 500 MΩ common-mode input impedance; a driven right leg with = 6 MΩ, = 4 MΩ, = 25 kΩ and 0.1 µA of displacement current.
(a)
Mismatch error?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Driven-right-leg common-mode voltage?
Check
An evoked response of 0.5 µV in 10 µV of background; a high-pass at 0.5 Hz; a 60 Hz notch with Q = 20.
(a)
Sweeps for an SNR of 6?
(b)
High-pass phase lead at 1 Hz?
(c)
Notch width?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A saline catheter (radius 0.4 mm, 0.8 m) on a transducer of 0.04 mm³ per 100 mmHg; a pressure of 150/70 mmHg; a transducer 5 cm below the heart.
(a)
Natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Mean arterial pressure?
(c)
Leveling error?
Check
Fick: 220 mL/min uptake and 40 mL/L difference; thermodilution: 10 mL at 20 °C, area 3 °C·s, K = 1.08; Doppler: 2 MHz at 30° on 0.3 m/s.
(a)
Fick output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Thermodilution output?
(c)
Doppler shift?
Check
Helium dilution from 9% to 7% in a 3 L spirometer; a forced expiration with τ = 1 s; an oximeter reading R = 0.5; blood with 10 g/dL Hb at saturation 0.9, 90 mmHg.
(a)
FRC?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
FEV1/FVC?
(c)
Calibrated SpO₂ in percent?
(d)
Oxygen content?
Check
A pH electrode at 25 °C; blood with 28 mM bicarbonate at = 50 mmHg; a glucose sensor with = 80 nA and = 15 mM at 5 mM.
(a)
Voltage change for 0.2 pH units?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Blood pH?
(c)
Glucose current?
Check
Tissue with μ = 0.16 cm⁻¹, 10 cm thick; a voxel of μ = 0.4 against water's 0.2; an ultrasound boundary from 1.38 to 1.9 MRayl; a 3 MHz echo from 5 cm.
(a)
X-ray transmission?
(b)
CT number?
(c)
Reflected intensity fraction?
(d)
Round-trip loss at 0.5 dB/(cm·MHz)?
Check
A 1.5 T scanner with a 10 mT/m readout gradient and 100 kHz bandwidth; 400 MBq of ¹⁸F; a spin echo (TR 500 ms, TE 15 ms) for T1 = 1000 ms, T2 = 100 ms.
(a)
Larmor frequency?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
(b)
Field of view?
(c)
Activity after 60 min?
(d)
Relative signal?
Check
Tissue with rheobase 0.5 mA and chronaxie 0.3 ms; a defibrillator: 100 µF at 2 kV into 50 Ω, truncated after 4 ms.
(a)
Threshold for 0.5 ms pulses?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
Stored energy?
(c)
Fraction delivered?
Check
A 120 V, 60 Hz device with 1000 pF to its case; 30 mV between grounds across an 800 Ω catheter path; a wearable (3 mA active, 3 µA asleep, duty 0.005) on 200 mAh.
(a)
Leakage current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Catheter current?
(c)
Wearable life in days?
Check
A volume-controlled breath delivers 0.45 L at 0.5 L/s with PEEP 5 cmH₂O to lungs with R = 12 cmH₂O·s/L and C = 0.06 L/cmH₂O.
(a)
Peak airway pressure, in cmH₂O?
(b)
Plateau pressure, in cmH₂O?
(c)
Time for 95% of passive exhalation?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A rhythm test with sensitivity 0.85 and specificity 0.9 is used where the prevalence is 0.15.
(a)
Positive predictive value?
(b)
Negative predictive value?
Check
A hematology aperture is 60 µm in diameter and 120 µm long, filled with diluent of resistivity 1.2 Ω·m and driven by 1 mA. Red cells of 95 fL pass through it.
(a)
Aperture resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Resistance change for one cell, from ?
(c)
Pulse height at 1 mA?
Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.
Practice
Keep practicing
20 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.