S15 · Lesson 25 · Heart sounds and catheterization
Where the four heart sounds and murmurs come from, their frequencies and why the low ones go unheard, the stethoscope's quarter-wave resonances, phonocardiography, valve gradients and orifice areas from Bernoulli's equation, and the kinetic-energy error of a catheter facing the flow
Two of cardiology's oldest measurements are a stethoscope on the chest and a catheter in the heart. The stethoscope hears valves closing and blood turning turbulent, but it hears them through an acoustic system with resonances of its own and an ear deaf to their lowest notes. The catheter measures pressures on either side of a valve, and from the drop across it, with Bernoulli's equation, the valve's opening area. Both are instruments with errors worth understanding: one acoustic, one fluid-mechanical.
- The curve is the effective opening area of a heart valve against the mean pressure drop across it, from Bernoulli's equation with a discharge coefficient of 0.85, for the cardiac output, heart rate and ejection time on the sliders.
- At normal flow a healthy aortic valve, 3 to 4 cm², drops only a few mmHg; a drop of tens of mmHg means stenosis.
- The same gradient means a smaller valve at lower output, which is why gradients alone can understate stenosis in a failing heart.
Predict
A patient's cardiac output is 6.4 L/min at 78 beats per minute, with a 0.31 s ejection period, and the mean pressure drop across the aortic valve during ejection is 7.33 kPa (55 mmHg). Roughly how large is the valve's opening?
During ejection the flow is 6.4 L/min ÷ 78 beats ÷ 0.31 s = 2.65 × 10⁻⁴ m³/s. The jet's velocity is = 3.72 m/s, and with a discharge coefficient of 0.85 the area is = 0.84 cm², Webster's Example 7.6: severe aortic stenosis, a quarter of a normal valve.
Name it
Heart sounds and murmurs
Heart sounds come from the acceleration and deceleration of blood; murmurs from its turbulence. The first sound (S1) accompanies ventricular contraction and the closing of the atrioventricular valves; the second (S2) the deceleration and reversal of flow in the aorta and pulmonary artery as the semilunar valves close, at the end of the ECG's T wave. S2 often splits into aortic and pulmonary parts tens of milliseconds apart, more on inspiration. S3, at the end of rapid filling, is heard in children; S4, from atrial contraction, is recorded but seldom heard. Murmurs from stenosed or leaking valves are placed by their timing in the cycle and the chest site where they're loudest.
Heart sounds and murmurs have very small amplitudes and span about 0.1 to 2000 Hz, with most of the audible sound between about 20 and 200 Hz. Below about 20 Hz their amplitude is under the threshold of hearing, so that part can be recorded but not heard; the higher frequencies fall where the ear is most sensitive.
The stethoscope
An acoustic stethoscope's tubing, closed by the ear at one end, resonates where its length is an odd number of quarter wavelengths: . Its response is therefore uneven, with peaks and dips; a 3 dB loss at the wrong frequency can drop a faint sound below threshold. Pressing the chest piece firmly stretches the skin into a diaphragm that attenuates low frequencies, so clinicians switch between bell and diaphragm to emphasize low or high sounds.
Phonocardiography
A phonocardiogram records heart sounds with a microphone, filtered into bands and displayed with the ECG and pulse waveforms, so that sounds can be timed against electrical and mechanical events objectively. Pulse recordings use the 0.1 to 100 Hz band, sounds the higher ones.
Gradients and valve areas
A catheter laboratory measures pressures in each chamber and across each valve. For frictionless flow, Bernoulli's equation relates the drop across a narrowing to the jet's velocity, , so the effective orifice area is
with F the flow during the valve's open period and the discharge coefficient about 0.85 for semilunar valves and septal defects and 0.6 for mitral valves. Doppler echocardiography uses the same relation backward, as the "simplified Bernoulli equation" (mmHg, m/s), to estimate gradients from jet velocities.
Kinetic energy and the catheter's opening
A catheter whose end hole faces upstream reads the total pressure, ; a side hole reads the static pressure alone. The difference is small at rest in the aorta (about 4 mmHg at 1 m/s, 3% of systolic) but large in the pulmonary artery (13%) and in the venae cavae during exercise (over 50%), where pressures are low.
Worked example
Is the kinetic-energy term significant for a catheter facing the flow in the aorta, where the peak velocity is about 1.5 m/s and the systolic pressure 120 mmHg?
Show the solution
- Kinetic-energy term.
- Compare with systolic. That's 7.5% of 120 mmHg: noticeable, and larger during exercise when the velocity rises.
- Which catheter reads what. The end-hole catheter reads high at peak flow; a side-hole catheter, or one pointing downstream, reads the static pressure.
- Low-pressure vessels. In the low-pressure pulmonary artery and veins the same velocities make far larger relative errors.
Answer yes: the term is 8.9 mmHg, 7.5% of systolic pressure, so a catheter facing the flow reads noticeably high.
Worked example
A stethoscope's tubing is 0.5 m long. At what frequencies does it resonate, taking sound at 343 m/s?
Show the solution
- Resonance condition. Closed at the ear and open at the chest piece, the tube resonates at odd quarter wavelengths:
- First three resonances. For n = 1, 2 and 3 that's 172, 515 and 858 Hz.
- Effect on the sound. Sounds near those frequencies are boosted and those between them relatively weakened, so two stethoscopes with different tubing can make the same murmur sound different.
- Electronic stethoscopes. They replace the tube with a microphone and flat amplification, and can record the sound.
Answer the tubing resonates at 172, 515 and 858 Hz.
Worked example
Doppler echocardiography measures a 4 m/s jet through a stenosed aortic valve. What pressure gradient does the simplified Bernoulli equation give, and where does its factor of 4 come from?
Show the solution
- Simplified Bernoulli.
- Where the 4 comes from. It is in mmHg per (m/s)²:
- Clinical meaning. A gradient above 40 mmHg suggests severe stenosis, so this patient would be referred for valve replacement without needing a catheter at all.
Answer the gradient is 64 mmHg; the 4 is ρ/2, 530 Pa per (m/s)², converted to mmHg: 3.98.
Check
Cardiac output is 6.4 L/min at 60 beats per minute, the valve is open 0.33 s per beat, and the mean gradient across it is 80 mmHg; the discharge coefficient is 0.6.
(a)
What is the flow while the valve is open?
Write units the way you normally would: 1.5 mL/s, 1500 µL/s and 0.0015 L/s are the same answer.
(b)
What is the valve's effective area?
Check
A Doppler study finds a 3 m/s jet through a valve.
(a)
What gradient does Bernoulli's equation give (blood density 1060 kg/m³)?
Check
Blood moves at 0.9 m/s where the static pressure is 3 mmHg.
(a)
How much higher does a catheter facing upstream read?
(b)
What percent error is that?
Check
A stethoscope's tube is 0.5 m long, closed at the ear.
(a)
Where is its resonance number 1?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A phonocardiograph must capture murmurs up to 300 Hz, sampling at 5 times that frequency with 16-bit samples.
(a)
What sampling rate does it use?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
How much data does an hour of one channel take?
Check
Why can a phonocardiogram record parts of the heart sounds that a stethoscope doesn't let you hear?
Check
Flow F (m³/s) passes a valve with pressure drop ΔP (Pa), blood density ρ, and discharge coefficient .
Step 1
What is the jet's velocity?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What is the valve's effective area, in square meters?
Explain why
Why does a pressure gradient alone not tell how narrow a valve is?
The pressure drop across a narrowing is spent accelerating blood through it, so it depends on the jet's velocity, which is the flow divided by the opening's area. For a given valve, the gradient grows as the square of the flow: a stenotic valve in a heart that can only pump weakly produces a modest gradient, while a mildly narrowed valve can show a large one during exercise. Only by combining the gradient with the flow measured at the same time, as Bernoulli's equation does, can the opening's area be found, and the area is what describes the valve itself.
Practice
Keep practicing
8 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.