S15 · Lesson 24 · Indirect blood pressure and heart sounds

Cuff measurements by auscultation and Korotkoff sounds, the oscillometric method with its peak at mean arterial pressure and fixed-ratio estimates of systolic and diastolic pressure, mean pressure from systolic and diastolic, deflation rate and cuff size errors, and the phonocardiogram

Most blood pressures are measured with a cuff, without breaking the skin. A clinician listening with a stethoscope hears tapping sounds start as the cuff pressure falls below systolic and fade near diastolic. An automatic monitor hears nothing at all: it watches the small pressure oscillations that each heartbeat makes in the cuff, finds where they're largest, and estimates systolic and diastolic pressures from ratios calibrated against direct measurements. Both methods are simple to use and subtle in their errors.

  • The envelope of cuff oscillation amplitude against cuff pressure, modeled as a bell centered at the mean arterial pressure with width w.
  • An oscillometric monitor finds the peak (MAP), then reports systolic where the amplitude has fallen to 55% of the peak on the high-pressure side, and diastolic where it's 85% on the low side.
  • The estimates depend on the envelope's shape, which varies between people: the fixed ratios are empirical.

Predict

A cuff deflates at 3 mmHg per second and the heart rate is 75 per minute. How precisely can the auscultatory method pin down systolic pressure?

Name it

Auscultation

The cuff is inflated above systolic pressure, closing the brachial artery, then deflated steadily. When cuff pressure falls just below systolic, blood spurts through during each peak, and the turbulence makes the tapping Korotkoff sounds heard below the cuff. They muffle and disappear as cuff pressure reaches diastolic, when the artery stays open throughout the cycle. Each reading is uncertain by the pressure drop between beats.

Oscillometry

The arterial pulse makes the cuff pressure oscillate slightly. The oscillation amplitude is largest when the cuff pressure equals the mean arterial pressure, where the arterial wall is unloaded and most compliant. Monitors filter the oscillations out of the cuff-pressure record (a band-pass of roughly 0.5 to 10 Hz), find the peak, and estimate systolic and diastolic pressures where the envelope falls to fixed fractions of the peak, typically around 0.5 to 0.6 on the high side and 0.7 to 0.85 on the low side, chosen empirically.

Mean arterial pressure

The arterial waveform spends longer near diastolic than near systolic, so the mean is closer to diastolic:

MAP≈DBP+SBP−DBP3

The difference SBP − DBP is the pulse pressure.

Sources of error

A cuff that's too narrow for the arm reads high, because cuff pressure isn't fully transmitted to the artery; the inflatable bladder should be about 40% of the arm's circumference wide. Deflating too fast, a cuff above heart level, movement and irregular rhythms all add error.

Heart sounds

The closing of the heart's valves makes the first and second heart sounds, S1 and S2, mostly between about 20 and 200 Hz; murmurs from turbulent flow reach higher. A phonocardiogram records them with a chest microphone, an accelerometer or a piezoelectric sensor (lesson 4), filtered into bands to separate sounds from murmurs.

Worked example

A pressure is 120/80 mmHg. What's its mean? And for an oscillometric envelope centered at 93 mmHg with w = 30 mmHg, what do the 0.55 and 0.85 ratios report?

Show the solution
  1. Mean pressure. MAP=80+403=93.3mmHg
  2. Systolic estimate. At the 0.55 ratio: Ps=93+30−ln⁡0.55=116.2mmHg
  3. Diastolic estimate. At the 0.85 ratio: Pd=93−30−ln⁡0.85=80.9mmHg
  4. A limitation. A wider envelope from a stiffer artery would move both estimates, an error the fixed ratios can't see.

Answer the mean is 93.3 mmHg, and the envelope reports a systolic pressure of 116.2 mmHg and a diastolic of 80.9 mmHg.

Worked example

What bladder width suits an arm of 32 cm circumference? And how long after deflation starts at 180 mmHg (3 mmHg/s) should the first sound come for a systolic of 120?

Show the solution
  1. Bladder width. w=0.4×32=12.8cm
  2. Time to reach 120 mmHg. t=180−1203=20s

Answer a 12.8 cm bladder suits the arm, and the first sound should come 20 s after deflation starts.

Worked example

An artery has a wall with Young's modulus 0.5 MPa, 0.5 mm thick, and an inner diameter of 10 mm, filled with blood of density 1060 kg/m³. What pulse-wave velocity does the Moens–Korteweg equation give, how long does the pulse take to travel 0.8 m, and how much shorter is that if a pressure rise stiffens the wall by 20%?

Show the solution
  1. Pulse-wave velocity. c=Ehρd=0.5×106×0.5×10−31060×0.01=4.86m/s
  2. Transit time over 0.8 m. t=0.84.86s=165ms
  3. Stiffer wall. A 20% stiffer wall speeds the wave by 1.2: c′=4.86×1.2=5.32m/s cutting the transit time to 150 ms, 14 ms shorter.
  4. Why it matters. Because arterial stiffness rises with pressure, pulse transit time, measured from the ECG's R wave to the pulse at a finger, tracks blood pressure without a cuff, once it's calibrated for each person.

Answer the pulse travels at 4.86 m/s and takes 165 ms to cover 0.8 m; a 20% stiffer wall makes that 14 ms shorter, 150 ms.

Check

Systolic pressure is 130 mmHg and diastolic 70 mmHg. What's the mean arterial pressure?

Check

An oscillometric monitor finds a mean pressure of 95 mmHg, and the diastolic is 65 mmHg. What systolic pressure is consistent with the one-third rule?

Check

An oscillometric envelope follows e−((P−MAP)/w)2 with MAP = 85 mmHg and w = 20 mmHg. Where on the high-pressure side does it fall to 0.6 of its peak?

Check

A cuff deflates at 3 mmHg/s with a heart rate of 50 per minute. How far does it fall between beats?

Check

What bladder width suits an arm with a 32 cm circumference?

Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.

Check

A cuff starts deflating from 180 mmHg at 3 mmHg/s. When does it reach a systolic pressure of 110 mmHg?

Write units the way you normally would: 1.5 s and 1500 ms are the same answer.

Check

An artery's wall has Young's modulus 0.8 MPa and thickness 0.4 mm, with an inner diameter of 20 mm; blood density is 1060 kg/m³.

(a)

What pulse-wave velocity does the Moens–Korteweg equation give?

Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.

(b)

How long does the pulse take to travel 1 m?

Check

A pressure reads s over d mmHg.

Step 1

What does the usual formula give for the mean arterial pressure?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why are oscillations in cuff pressure largest at the mean arterial pressure?