S15 · Lesson 16 · Isolation amplifiers
Breaking the ohmic path between patient and mains-powered equipment, the isolation barrier's capacitance and the leakage it allows, isolation-mode rejection alongside common-mode rejection, transformer, optical and capacitive isolators, and the isolated power supplies that feed them
A monitor's electrodes connect the patient, through a few kilohms, to circuitry powered from the wall. If that circuitry's ground drifts, or a fault puts line voltage on it, current flows through the patient. An isolation amplifier puts a barrier in the signal path that passes the ECG but almost no current: kilovolts can stand across it while a few microamperes leak through its capacitance. Every patient-connected input in a hospital crosses one, and its capacitance, voltage rating and rejection set how safe and how quiet the measurement is.
- The curve is the current that leaks through an isolation barrier of capacitance C when the voltage across it is the line voltage, against frequency.
- The dashed line is 10 µA, the limit for patient leakage through a cardiac-connected input under normal conditions.
- At 60 Hz, a barrier must stay below about 220 pF at 120 V; fast transients and the higher harmonics of a switching supply push leakage up in proportion to frequency.
Predict
What isolation capacitance keeps the 60 Hz leakage from a 120 V line below 10 µA?
The barrier must present at least 120 V/10 µA = 12 MΩ at 60 Hz, so = 221 pF, Webster's Example 14.8. Practical isolators have a few picofarads, but the isolated power supply that feeds the patient side adds its own interwinding capacitance, and the total is what counts.
Name it
The isolation barrier
An isolation amplifier is an instrumentation amplifier followed by a unity-gain stage that carries the signal across a barrier with no ohmic connection: separate supplies and separate grounds on each side. The barrier is modeled by its isolation capacitance and resistance (typically a few picofarads and more than 10¹⁰ Ω). It must withstand an isolation voltage, between the input common and the output common, of 1 to 10 kV without breakdown, and pass leakage of only microamperes at line frequency: .
Three kinds of rejection
The output of an isolation amplifier with gain G is
The input stage rejects common-mode voltage between the electrodes and the floating input common, as any instrumentation amplifier does (above 100 dB). The barrier rejects the isolation-mode voltage between the two commons, often the full line voltage, with its isolation-mode rejection ratio, typically 120 to 160 dB at 60 Hz and falling with frequency.
How the signal crosses
- Transformer isolators modulate a carrier (FM or pulse width) through a small transformer, with signal bandwidths up to tens of kilohertz; a second winding at about 25 kHz, rectified and filtered, powers the input side.
- Optical isolators drive an LED whose light reaches a photodiode on the output side; a matched photodiode on the input side, in a feedback loop, cancels the LED's nonlinearity. They pass DC without a modulator.
- Capacitive isolators encode the signal digitally or as a frequency and pass it across a pair of small ceramic capacitors, with isolation up to 8 kV and bandwidths to 70 kHz.
Modern designs often digitize on the patient side and isolate the digital stream instead, so the analog signal never crosses the barrier.
Isolated power
The patient side needs power, from a battery or an isolated DC–DC converter. The converter's transformer has interwinding capacitance of tens of picofarads, often more than the signal isolator's, so it usually dominates the leakage budget. Standards such as IEC 60601-1 set patient leakage limits: 10 µA under normal conditions and 50 µA with a single fault for cardiac-connected (CF) inputs.
Worked example
A transformer-isolated ECG amplifier with gain 1000 has a CMRR of 100 dB and an IMRR of 120 dB. Its input sees 1 mV of ECG, 10 mV of common-mode voltage, and the full 120 V line between its commons. What's the output, and what fraction of it is error?
Show the solution
- Common-mode term. Referred to the input, with 100 dB = 10⁵:
- Isolation-mode term. With 120 dB = 10⁶:
- Output.
- Fraction that is error. The error terms add to 120.1 µV, nearly all of it line voltage leaking across the isolation barrier. That is 12% of the 1 mV signal.
- A better barrier. A 140 dB IMRR would cut it to 1.2%. The isolation-mode voltage is so large that the IMRR sets the floor, and the CMRR hardly matters.
Answer the output is 1.12 V. The error is 12% of the signal, nearly all of it line voltage leaking across the barrier.
Worked example
An isolated DC–DC converter with 20 pF of interwinding capacitance feeds the patient side of an amplifier whose signal isolator adds 2 pF. What leakage flows at 230 V, 50 Hz, and how much margin is left under a 10 µA limit?
Show the solution
- Barrier capacitance. The two capacitances are in parallel across the barrier, 22 pF in all.
- Leakage at 230 V, 50 Hz. a sixth of the 10 µA limit.
- The converter's share. It contributes 20 of the 22 pF, 91% of the leakage.
- At 60 Hz and 120 V. The same barrier leaks 1.0 µA.
- What the margin is for. It allows several such inputs to connect to one patient.
Answer 1.6 µA leaks at 230 V, 50 Hz, a sixth of the 10 µA limit, which leaves room for several such inputs on one patient.
Worked example
An isolation amplifier's IMRR is 140 dB at 60 Hz and falls by 20 dB per decade above. A switching supply puts a 10 V, 60 kHz ripple between the commons. What input-referred error does that cause?
Show the solution
- IMRR at 60 kHz. That's three decades above 60 Hz, so the IMRR has fallen to
- Input-referred error. 80 dB is a ratio of 10⁴: as large as the ECG itself, although at a frequency the ECG filter removes.
- Fast transients. Isolation-mode transients, such as an electrosurgical unit's RF or a defibrillator's pulse, couple the same way, which is why isolated inputs also need filtering and transient protection.
Answer the ripple appears as 1 mV at the input, as large as the ECG itself, though at a frequency the ECG filter removes.
Check
A barrier must limit leakage from a 120 V, 50 Hz line to 50 µA.
(a)
What impedance must it present?
Write units the way you normally would: 1.5 MΩ, 1500 kΩ and 0.0015 GΩ are the same answer.
(b)
What is the largest capacitance it can have?
Check
An isolation amplifier with gain 100 has CMRR 110 dB and IMRR 110 dB; its input sees 0.5 mV of signal, 50 mV of common-mode voltage and 120 V between its commons.
(a)
What input-referred error does the isolation-mode voltage cause?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What is the total output?
Check
An isolated converter with 10 pF of interwinding capacitance and a signal isolator with 1 pF feed one input; 1 such inputs connect to a patient, with 230 V at 60 Hz across the barriers.
(a)
How much leaks through one input's barrier?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
How much in total through all 1?
Check
An isolator's IMRR is 140 dB at 60 Hz and falls by 20 dB per decade above. A 10 V disturbance at 1000 Hz appears between its commons.
(a)
What is its IMRR at 1000 Hz?
(b)
What input-referred error results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An ECG input is protected from a 2 kV defibrillator pulse lasting 10 ms by a 10 kΩ series resistor ahead of clamp diodes that hold the amplifier's input near 0 V.
(a)
What current flows through the resistor?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
How much energy must the resistor absorb?
Check
Why do many modern monitors digitize the signal on the patient side and isolate the digital data instead of the analog signal?
Check
An isolation amplifier has gain G, CMRR C and IMRR M (as ratios); its input sees signal s, common-mode voltage c and isolation-mode voltage v.
Step 1
What is its output?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What fraction of the signal is error?
Explain why
Why does an isolation amplifier make a patient safer even though its barrier still has capacitance?
Without isolation, the electrodes connect the patient through the amplifier's input circuit to the equipment's ground and supplies, so a ground difference or a fault drives current straight through the patient, limited only by a few kilohms. The isolation barrier replaces that path with a tiny capacitance whose impedance at line frequency is tens of megohms, so even the full line voltage across it pushes only microamperes, below the level that can disturb the heart. The barrier still passes the signal by modulating a carrier, light or a digital code across it, so the measurement works while the dangerous low-frequency current path is broken.
Practice
Keep practicing
8 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.