Proton spins precessing at the Larmor frequency, the tiny net magnetization, T1 and T2 relaxation and how echo and repetition times set image contrast, spatial encoding with gradients, slice selection and field of view, and imaging radioactive tracers with gamma cameras and PET

Magnetic resonance imaging listens to hydrogen nuclei. In a strong field, protons precess at a frequency proportional to the field, about 64 MHz at 1.5 T, and a radio pulse at exactly that frequency tips them over so they broadcast back. How quickly that signal fades depends on the tissue, which gives MRI its superb soft-tissue contrast, and making the field vary across the body makes each location broadcast at its own frequency, so a Fourier transform (C9) turns the signal into an image. Nuclear medicine takes the opposite approach: it images where an injected radioactive tracer goes, showing function rather than anatomy.

  • The spin-echo signal S∝(1−e−TR/T1)e−TE/T2 for two tissues with illustrative relaxation times: a solid tissue (T1 = 600 ms, T2 = 80 ms) and a fluid (T1 = 3000 ms, T2 = 1000 ms), against the echo time TE.
  • With a short repetition time TR the solid tissue is brighter (T1 weighting); with a long TR and a long TE the fluid is brighter (T2 weighting).
  • Choosing TR and TE is how the scanner decides which property the image shows.

Predict

At what frequency do protons precess in a 3 T scanner?

Name it

Spins in a field

A proton in a field B0 precesses at the Larmor frequency

f0=γ¯B0,γ¯=42.58 MHz/T

Slightly more spins align with the field than against it, by a fraction of about hf0/(2kT): only 5 parts per million at 1.5 T and body temperature (B11's Boltzmann factor). There are so many protons that this tiny excess still gives a measurable signal.

Excitation and relaxation

An RF pulse at f0 tips the net magnetization away from the field; it then precesses and induces a signal in a receive coil. It relaxes in two ways: the longitudinal part recovers as 1−e−t/T1 (energy given to the surroundings), and the transverse part, which carries the signal, decays as e−t/T2 (spins dephasing). In a spin-echo sequence repeated every TR and read at echo time TE,

S∝ρ(1−e−TR/T1)e−TE/T2

Short TR and short TE weight the image by T1; long TR and long TE weight it by T2; long TR and short TE leave mostly proton density ρ.

Spatial encoding

A gradient G makes the field, and the precession frequency, vary with position: Δf=γ¯GΔx. Then:

  • Slice selection: an RF pulse of bandwidth Δf excites a slice of thickness Δf/(γ¯G).
  • Frequency encoding: during readout, a receiver bandwidth BW covers a field of view FOV=BW/(γ¯G), and N samples give pixels FOV/N wide.
  • Phase encoding repeats the acquisition with steps of a second gradient; the 2D Fourier transform gives the image.

Nuclear medicine

A tracer decays as A=A02−t/T1/2. Gamma cameras (and SPECT) image single photons, such as technetium-99m's 140 keV gammas (half-life 6.0 h), through a collimator. PET images positron emitters like fluorine-18 (half-life 109.8 min): each positron annihilates into two 511 keV photons flying in opposite directions, and detecting both in coincidence locates the decay on a line, without a collimator.

Worked example

Compare the two tissues' signals for a T1-weighted (TR 500 ms, TE 15 ms) and a T2-weighted (TR 4000 ms, TE 100 ms) scan.

Show the solution
  1. T1-weighted, solid tissue. S=(1−e−500/600)e−15/80=0.469
  2. T1-weighted, fluid. S=(1−e−500/3000)e−15/1000=0.151 so tissue is three times brighter.
  3. T2-weighted. Tissue gives 0.286 and fluid 0.666, so fluid is now more than twice as bright.
  4. What it shows. The contrast reverses with the timing.

Answer on the T1-weighted scan tissue (0.469) is three times brighter than fluid (0.151); on the T2-weighted scan fluid (0.666) is more than twice as bright as tissue (0.286).

Worked example

A 10 mT/m readout gradient and a ±32 kHz receiver bandwidth give what field of view? What slice does a 1 kHz RF pulse select with the same gradient? And how much of a 370 MBq dose of ¹⁸F remains after an hour?

Show the solution
  1. Frequency spread of the gradient. G=42.58MHz/T×0.01T/m=4.26kHz per cm
  2. Field of view. The ±32 kHz receiver covers 64 kHz: FOV=644.26cm=15.0cm
  3. Slice thickness. Δz=1kHz4.26kHz/cm=2.35mm
  4. ¹⁸F after an hour. A=370×2−60/109.8=253MBq

Answer the field of view is 15.0 cm, the slice 2.35 mm, and 253 MBq of the ¹⁸F remains after an hour.

Worked example

An MRI scan with 1 mm cubic voxels has adequate signal-to-noise ratio. What happens to the ratio at 0.5 mm voxels, and how much longer must the scan run to recover it?

Show the solution
  1. Signal per voxel. The signal is proportional to the number of protons in a voxel, so halving each dimension divides it by 8, and the SNR with it: (12)3=18
  2. Recovering it by averaging. Averaging repeated acquisitions improves SNR only as the square root of time, so recovering a factor of 8 takes 82=64 times as long.
  3. In practice. High-resolution scans accept lower SNR, move to stronger magnets (SNR grows roughly in proportion to field strength) or use closer, smaller receive coils.

Answer the SNR falls by a factor of 8, and recovering it takes a scan 64 times as long.

Check

At what frequency do protons resonate in a 1.5 T field?

Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.

Check

What's the excess fraction of aligned protons, in parts per million, at 1.5 T and 300 K?

Check

A tissue has T1 = 1200 ms and T2 = 120 ms. What relative spin-echo signal does it give with TR = 1000 ms and TE = 15 ms?

Check

A readout gradient of 10 mT/m and a receiver bandwidth of 32 kHz cover what field of view?

Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.

Check

An RF pulse of 0.5 kHz bandwidth is applied with a 10 mT/m slice gradient. How thick a slice does it excite?

Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.

Check

A tracer with a 360.6 min half-life starts at 370 MBq. How much remains after 30 min?

Check

A scan's voxels are scaled by 0.7 in each dimension, with 2 times as many averages, and the field goes from 1.5 T to 3 T.

(a)

By what factor does the SNR change, taking it as proportional to field strength?

Check

An MRI scanner works at B tesla; separately, a tracer with half-life h minutes starts at activity A.

Step 1

At what frequency, in megahertz, do protons resonate?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

How does a gradient field turn MRI's radio signal into an image?

Practice

Keep practicing

10 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.