S15 · Lesson 44 · Neural stimulation

Stimulating nerves with charge-balanced current pulses, the electrode–tissue interface under stimulation with its double-layer capacitance and charge-injection limits, Shannon's safety criterion, cochlear implants and deep brain stimulators, and recording spikes with their noise and data rates

A pacemaker stimulates one heart through one or two electrodes. A cochlear implant stimulates the auditory nerve through 12 to 22, hundreds to thousands of times a second each, for decades; a deep brain stimulator does the same deep in the brain. The same electrodes must deliver enough charge to fire neurons without corroding or damaging tissue, and the recording versions must pick out microvolt spikes. The limits come from the electrode–tissue interface of lesson 14, now driven hard.

  • The axis across is the charge per phase of a stimulus pulse. The solid curve is Shannon's limit on charge density for safe stimulation, log⁡D=k−log⁡Q; above it, tissue damage has been observed.
  • The dashed line is the charge density an electrode of area A actually experiences, D = Q/A.
  • Where they cross is the largest safe charge per phase for that electrode: larger electrodes can deliver more, and the limit grows only as the square root of area.

Predict

A 1 mm² electrode (0.01 cm²) delivers charge-balanced pulses. With Shannon's k = 1.85, what is the largest safe charge per phase?

Name it

Stimulating with charge

A neuron fires when its membrane is depolarized past threshold, which takes a certain charge delivered quickly (the strength–duration curve of lesson 43). Stimulators therefore control current, not voltage, so the charge is fixed whatever the electrode's impedance, and they deliver biphasic, charge-balanced pulses: a stimulating phase followed by an equal and opposite one, so no net charge accumulates to drive electrochemical reactions over time. A series capacitor on each output guarantees zero DC current even under a fault.

The interface under stimulation

A stimulating electrode passes charge into tissue first by charging its double-layer capacitance (about 10 to 40 µF/cm² for smooth platinum), and beyond that by reversible surface reactions. The electrode's voltage during a constant-current pulse is the drop across the tissue's access resistance plus the double layer's charge:

V=IRa+QCdl

If the interface voltage leaves the electrode's safe "water window" (about −0.6 to +0.8 V for platinum), irreversible reactions such as electrolysis begin, generating gas and pH changes and dissolving the electrode. The charge a material can inject reversibly, its charge-injection limit, is about 50 to 150 µC/cm² for platinum and some 1 to 5 mC/cm² for activated iridium oxide or roughened coatings, which is why small electrodes use them. The stimulator's supply must provide the compliance voltage V for the largest pulse.

Tissue safety

Shannon's criterion, from animal data, marks damage above the line

log10⁡D=k−log10⁡Q

with D in µC/cm² per phase and Q in µC per phase; k = 1.85 is a commonly used safe value. For an electrode of area A, the largest safe charge per phase is 10kA.

Applications

Cochlear implants place 12 to 22 electrodes along the cochlea, each stimulating the nerve fibers tuned to one band of frequencies, at hundreds to a few thousand pulses a second per channel, with phases of tens of microseconds. Deep brain stimulators for Parkinson's disease run continuously at about 130 Hz with 60 to 90 µs phases. Spinal cord stimulators and functional electrical stimulation of muscles follow the same rules with larger electrodes.

Recording

Extracellular spikes are 50 to 500 µV and last about a millisecond, so recording amplifiers pass about 300 Hz to 6 kHz (below that lie local field potentials). The electrode's thermal noise, 4kTRB, sets a floor: a 1 MΩ electrode over 6 kHz contributes about 10 µV RMS. Digitized at 20 to 30 kHz per channel, a thousand-channel array produces hundreds of megabits a second, more than an implant can transmit, so implants detect spikes on the chip and send only their times or features.

Worked example

A cochlear implant drives 16 channels at 1000 biphasic pulses per second each, with 25 µs phases at 1 mA, from an 8 V compliance supply. What is the average stimulation current and power, and the charge per phase?

Show the solution
  1. Charge per phase. Q=1mA×25μs=25nC
  2. Average current per channel. Each channel draws current for 2 × 25 µs per pulse, 1000 times a second: I1=50μA
  3. All 16 channels. I=16×50μA=0.8mA
  4. Power. From 8 V: P=8V×0.8mA=6.4mW for stimulation alone, which is why the processor and radio link are designed around a budget of tens of milliwatts, delivered by an inductive link through the skin (lesson 49).

Answer each phase carries 25 nC, and stimulation averages 0.8 mA and 6.4 mW.

Worked example

A 1 mA, 100 µs constant-current phase passes through a 0.01 cm² platinum electrode with 20 µF/cm² of double-layer capacitance and 1 kΩ of access resistance. What voltage does the stimulator need, and how far is the interface pushed?

Show the solution
  1. Charge. Q=1mA×100μs=0.1μC
  2. Double-layer capacitance. Cdl=20×0.01μF=0.2μF
  3. Interface voltage. ΔV=0.1μC0.2μF=0.5V inside platinum's water window.
  4. Compliance needed. The access resistance adds 1 mA × 1 kΩ = 1 V, so the stimulator needs at least 1.5 V of compliance, plus headroom.
  5. Limits. Doubling the charge would push the interface past the window, which is when coatings with more capacitance are needed.

Answer the stimulator needs at least 1.5 V of compliance plus headroom, and the interface is pushed 0.5 V, inside platinum's water window.

Worked example

A neural recording array has 1024 channels sampled at 30 kHz with 12 bits. What raw data rate does it produce, and what does a 1 MΩ electrode contribute in thermal noise over 300 Hz to 6 kHz at 37 °C?

Show the solution
  1. Raw data rate. R=1024×30000×12=369Mb/s far beyond what an implant can transmit on a few milliwatts.
  2. Electrode noise. Over the 5700 Hz band: vn2=4kTRB=4×1.38×10−23×310×106×5700V2 so vn = 9.9 µV RMS.
  3. Spike against noise. A 50 µV spike stands only five times above it.
  4. Cutting the data. Detecting spikes on the chip and sending their times cuts the data by a factor of hundreds.

Answer the array produces 369 Mb/s of raw data, and the electrode contributes 9.9 µV RMS of thermal noise.

Check

A stimulating electrode has an area of 0.005 cm², and the safety criterion uses k = 1.7.

(a)

What is the largest safe charge per phase?

Write units the way you normally would: 1.5 µC, 1500 nC and 0.0015 mC are the same answer.

(b)

What charge density is that?

Check

A 1 mA constant-current phase lasts 50 µs through an electrode of 0.05 cm² with 40 µF/cm² of double-layer capacitance and 0.5 kΩ of access resistance.

(a)

How far does the double layer charge?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

What compliance voltage does the pulse need?

Check

A cochlear implant runs 16 channels at 500 biphasic pulses per second each, with 50 µs phases of 1.5 mA, from a 5 V supply.

(a)

What average current does stimulation draw?

Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.

(b)

What power is that?

Check

A recording array has 256 channels sampled at 20 kHz with 16 bits.

(a)

What raw data rate does it produce?

Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.

Check

A recording electrode of 1 MΩ is used over a 3 kHz band at 37 °C.

(a)

What RMS thermal noise does it contribute?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

(b)

What is a 200 µV spike's peak-to-noise ratio, considering only that noise?

Check

Why are stimulation pulses made biphasic and charge-balanced?

Check

A constant-current phase of I mA and t µs passes through an electrode of area A (cm²) with double-layer capacitance c (µF/cm²) and access resistance R (kΩ).

Step 1

How much charge does the phase carry, in µC?

Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.

Explain why

Why does a stimulating electrode have a charge limit that depends on its material?

Practice

Keep practicing

8 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.