S15 · Lesson 28 · Plethysmography and impedance measurements
Venous-occlusion plethysmography for limb blood flow, impedance plethysmography from a cylinder model, four-electrode measurement at 100 kHz, impedance cardiography and apnea monitoring, and hot-film velocity sensors
The flowmeters of lesson 27 need a probe on or in the vessel. Often all a clinician can reach is the outside of a limb or the chest, yet they still want to know whether blood is getting into the calf, whether a vein is blocked by a clot, or whether a sleeping infant is still breathing. Each pulse of blood swells a limb by a fraction of a milliliter and each breath changes the chest's electrical impedance by a fraction of an ohm. How do instruments turn those small changes into volumes and flows?
- Venous-occlusion plethysmography of the calf. At t = 0 a thigh cuff is inflated to about 50 mmHg, which stops venous outflow but not arterial inflow, so the calf swells. At 20 s the cuff is released and the extra blood drains.
- The initial slope of the volume curve is the arterial inflow. As venous pressure rises toward the cuff pressure, the swelling levels off.
- After release a healthy calf empties with a time constant of a couple of seconds. Raise the outflow time constant to see the slow emptying that signals a deep-vein thrombosis.
Predict
An impedance plethysmograph drives a constant 1 mA at 100 kHz through a limb segment of 30 Ω. Each heartbeat lowers the impedance by 0.1 Ω. How big is the signal to be measured?
The segment's voltage is 1 mA × 30 Ω = 30 mV, and the pulsatile change is 1 mA × 0.1 Ω = 0.1 mV, a third of a percent. The instrument has to subtract the steady 30 mV and amplify the remainder, much as a strain-gauge bridge does in lesson 3.
Name it
Venous-occlusion plethysmography
A plethysmograph measures a change in volume. A rigid chamber around a limb, filled with water or air, measures the limb's swelling directly. A cuff upstream inflated to about 50 mmHg blocks venous return without stopping arterial inflow, so the initial rate of swelling is the arterial flow, , usually reported per 100 mL of tissue. When the cuff is released, the limb drains; slow drainage suggests a venous obstruction.
Impedance plethysmography
Model a limb segment of length L as a cylinder of resistance Z. Each pulse adds a volume ΔV of blood of resistivity in parallel, lowering the impedance by ΔZ. Following Swanson's derivation in Webster, with ΔZ ≪ Z,
It assumes the arteries expand uniformly, blood's resistivity stays constant (about 1.5 Ω·m) and the current runs parallel to the vessels, which is reasonable for most limb segments.
Measurements use four electrodes: two outer ones drive a constant current and two inner ones sense the voltage, so the large and variable skin–electrode impedances don't add to the reading. The current is about 1 mA at around 100 kHz, where electrode impedance is low and the current is far above any stimulation threshold, which makes a milliampere safe. A balancing circuit removes the steady voltage so the small pulsatile part can be amplified.
Applications
- Venous thrombosis: impedance replaces the chamber in venous-occlusion studies.
- Apnea monitoring: electrodes across the chest track breathing, since inhaled air raises the chest's impedance; infant monitors alarm when breaths stop, rejecting the heartbeat's smaller ripple by pattern recognition.
- Impedance cardiography: band electrodes around the neck and lower chest give a beat-by-beat estimate of stroke volume, by Kubicek's formula with T the ejection time. It works fairly well in healthy subjects and less well in very sick patients.
- Photoplethysmography, the optical version, tracks blood volume by absorbed light and underlies pulse oximetry (lesson 33).
Thermal velocity sensors
A heated film or bead in the bloodstream loses heat faster when the blood moves faster. Held at a constant temperature rise ΔT above the blood, its power follows King's law, , so measuring P gives the velocity v after calibration. These sensors are small and fast enough to follow velocity within a beat, but they need a separate temperature sensor and can't tell direction.
Worked example
A 10 cm limb segment measures 30 Ω, and each pulse lowers it by 0.1 Ω. Taking = 1.5 Ω·m, how much blood does each pulse add?
Show the solution
- Blood volume per pulse.
- Sanity check. That's a volume change of a few tenths of a percent of the segment, consistent with the 0.33% impedance change.
- The formula's assumptions. If the blood's resistivity drops with flow, as it does when red cells align in the stream, the computed volume is off by the same fraction.
Answer each pulse adds 1.67 mL of blood.
Worked example
Band electrodes 30 cm apart on a patient's chest read = 25 Ω, a peak dZ/dt of 1.2 Ω/s and an ejection time of 0.3 s. Estimate the stroke volume with = 1.35 Ω·m.
Show the solution
- Stroke volume. a typical resting stroke volume.
- Cardiac output. At 70 beats a minute that's 4.9 L/min, which can be compared against the indicator-dilution methods of lesson 26.
Answer the stroke volume is about 70 mL, a cardiac output of 4.9 L/min at 70 beats a minute.
Worked example
A mercury strain gauge around a calf of 35 cm circumference stretches by 0.35 mm in the first 10 s of venous occlusion. Taking a 10 cm segment as a cylinder, what's the arterial inflow per 100 mL of tissue per minute?
Show the solution
- Fractional volume change. For a cylinder, volume goes as circumference squared:
- Segment volume.
- Inflow. The segment gains 1.95 mL in 10 s, 11.7 mL/min.
- Per 100 mL of tissue. That's 1.2 mL per 100 mL per minute, a normal resting calf flow.
- Why the gauge lasted. It needs no electrodes and reads the whole limb's swelling, which is why it remained the standard for venous-occlusion studies.
Answer the arterial inflow is 1.2 mL per 100 mL of tissue per minute, a normal resting calf flow.
Check
A limb segment 5 cm long measures 50 Ω, and a pulse lowers it by 0.2 Ω. With blood resistivity 1.5 Ω·m, how much volume does the pulse add?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
After a venous-occlusion cuff inflates, an 800 mL calf segment swells at 0.5 mL/s. What's the blood flow in mL per minute per 100 mL of tissue?
Check
A four-electrode plethysmograph drives 1 mA through a 20 Ω segment. What's the pulsatile signal for a 0.2 Ω change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Impedance cardiography: bands 25 cm apart, = 30 Ω, = 0.8 Ω/s, ejection time 0.3 s, = 1.3 Ω·m. Estimate the stroke volume.
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
A hot-film sensor held 6 K above blood draws 0.1 W. Its calibration is with a = 2 mW/K and b = 4 mW/(K·(m/s)^½). What's the blood velocity?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
A strain gauge around a limb of 35 cm circumference stretches by 0.1 mm in 10 s of venous occlusion.
(a)
By what percent does the limb's volume grow?
(b)
What inflow is that, per 100 mL of tissue?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
A limb segment of length L (m) is modeled as a cylinder of blood-like conductor with resistivity ρ (Ω·m) and impedance Z.
Step 1
What volume does the impedance imply, in cubic meters?
Type it the way you would write it, like 3sin(2πft) or 3*sin(2*pi*f*t). Angles can be in radians or carry a degree sign, like 45°.
Step 2
What volume change does an impedance drop of ΔZ mean?
Explain why
Why does impedance plethysmography use four electrodes and a 100 kHz current?
The quantity of interest is a tiny change, a fraction of an ohm, in the impedance of the tissue between the electrodes. With only two electrodes, the measured impedance would include the skin–electrode impedances, which are hundreds to thousands of ohms, drift with sweat and pressure, and change with movement: they would swamp the signal. With four, the outer pair carries a known current and the inner pair senses voltage through a high-impedance amplifier that draws almost no current, so the inner electrodes' own impedances drop nothing and only the tissue's impedance appears. A high frequency helps twice: electrode impedance, which is largely capacitive, falls with frequency, and nerve and muscle can't respond to 100 kHz, so a current of a milliampere, large enough for a clean signal, is safe and unfelt.
Practice
Keep practicing
8 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.