S15 Biomedical instrumentation · Practice
Every topic, mixed
Problems dealt one at a time, with new numbers every time. The kinds you miss come back more often, and you can keep going as long as you like.
Topics: all 52
Check
A pressure transducer's sensitivity is 10 µV per volt per mmHg, with 2.5 V excitation. What's its output at 250 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A sensor's output is with S = 20 and a = 0.001 (same units) over 0 to 300. What's its nonlinearity against the endpoint line, in percent of full scale?
Check
An amplifier with gain 5000 feeds a 10-bit converter spanning 5 V. What's one step, referred to the amplifier's input?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A transducer's zero drifts by 0.2 mmHg per °C. How large an error does a 5 °C temperature change cause?
Check
A transducer's sensitivity falls by 0.002 (as a fraction) per °C. By what percent is a reading wrong after a 5 °C rise, whatever the pressure?
Check
A measurement has independent errors of 1, 0.3 and 3 mmHg. What's the combined error?
Check
A sensor with output impedance 350 Ω drives an amplifier whose input impedance is 1 kΩ.
(a)
By what percent does loading lower the measured signal?
(b)
What total resistance does the signal source see?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
Check
Output: 2 µV/V/mmHg, 1 V, 300 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Input-referred step: 1 V, 8 bits, gain 10000?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Combined error: 0.1, 0.1, 5?
Check
A probe with a 5 s time constant touches a patient 2.5 °C warmer than the probe. How far below the patient's temperature does it read after 30 s?
Check
How long does a probe with time constant 5 s take to come within 0.1 °C of a 2.5 °C step?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A temperature rises steadily at 1 °C per minute. How far behind does a probe with a 5 s time constant read?
Check
A first-order sensor with τ = 0.1 s sees a 0.5 Hz sinusoid. What's its amplitude ratio?
Check
A catheter system has = 30 Hz and ζ = 0.1. What's its amplitude ratio at 20 Hz?
Check
A pressure system with damping ratio 0.4 sees a step. By what percent does it overshoot?
Check
A pressure system has natural frequency 30 Hz and damping ratio 0.1.
(a)
At what frequency is its resonant peak?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
How high is the peak?
Check
Remaining error: step 0.5 °C, τ = 2 s, after 120 s?
Check
Amplitude ratio: = 5 Hz, ζ = 0.05, f = 40 Hz?
Check
Overshoot for ζ = 0.2?
Check
A 1000 Ω strain gauge with gauge factor 2.1 is strained by 200 microstrain. By how much does its resistance change?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
Check
A single active gauge (G = 2) in a bridge excited at 10 V sees 200 microstrain. About what output does the bridge give?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A quarter bridge excited at 2.5 V has its active arm changed by ΔR/R = 0.05. What's the exact output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A bridge excited at 10 V has 4 active gauges (G = 2) at ±300 microstrain, arranged so their outputs add. What's its output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
An uncompensated foil gauge (G = 2) has a temperature coefficient of 20 ppm/°C. What apparent strain, in microstrain, does a 10 °C change produce?
Check
A 1 kΩ potentiometer at fraction 0.5 of its travel drives a 100 kΩ load. What fraction of the excitation appears at the wiper?
Check
A 120 Ω gauge (G = 2.1) is wired as a two-wire quarter bridge through leads totaling 0.2 Ω, of copper (α = 0.0039/°C).
(a)
By what percent do the leads reduce its sensitivity?
(b)
What apparent strain does a 20 °C rise in lead temperature cause?
Check
ΔR: G = 1.8, 50 microstrain, R = 5000 Ω?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
Check
Bridge output: 4 active arms, G = 2, 100 microstrain, 12 V?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Loaded pot: x = 0.1, = 0.5 kΩ, = 500 kΩ?
Check
A capacitive sensor has 1 cm² plates 0.05 mm apart with relative permittivity 4. What's its capacitance?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
Check
A 20 pF parallel-plate sensor has a 50 µm gap. How much does its capacitance change when the gap shrinks by 2 µm?
Write units the way you normally would: 1.5 fF and 0.0015 pF are the same answer.
Check
A piezoelectric sensor with d = 150 pC/N and 10 nF of capacitance is suddenly loaded with 0.5 N. What voltage appears at first?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A piezoelectric sensor of 0.5 nF is read across 1 GΩ. What's its low-frequency corner?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A piezoelectric sensor with d = 100 pC/N drives a charge amplifier with = 1000 pF. What's the size of the output for a 0.2 N force step?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A piezoelectric sensor jumps to 1 V under a steady load and is read across 2 GΩ with 2 nF of capacitance. What's left after 0.2 s?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A differential capacitor has 1 cm² plates, each 50 µm from a center plate in air, which moves 5 µm.
(a)
What is ?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
(b)
What is ?
Check
C: 0.1 cm², 0.01 mm, εr = 10?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
Check
Initial voltage: d = 2 pC/N, 0.1 N, 20 nF?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
Corner: 0.01 GΩ, 0.1 nF?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A 50 mV carrier arrives at 20° to the reference of a phase-sensitive demodulator. What's the average output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A demodulator's RC output filter must have its corner at 10 Hz with a 100 nF capacitor. What resistor?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
A fiber's core has index 1.48 and its cladding 1.4. What's its numerical aperture?
Check
For the same fiber (core 1.48, cladding 1.4), what's the acceptance half-angle in air?
Check
A photodiode with quantum efficiency 0.75 receives 50 µW at 530 nm. What current does it produce?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
At what wavelength does a tungsten filament at 2800 K radiate most strongly?
Write units the way you normally would: 1.5 µm, 1500 nm and 0.0015 mm are the same answer.
Check
A lock-in amplifier with an output time constant of 0.1 s sees noise of 1 µV/√Hz and an in-phase signal of 5 µV.
(a)
What RMS noise remains at its output?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What signal-to-noise ratio (amplitude) results?
Check
PSD average: 19 mV at 6.7°?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
NA for = 1.5, = 1.36?
Check
Photocurrent: η = 0.6, 460 nm, 18 µW?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A transducer of 5 µV/V/mmHg at 3 V excitation; separately, an output with a = 0.004 µV/mmHg² over 0 to 200 mmHg at the same s.
(a)
Output at 250 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Nonlinearity against the endpoint line, in percent?
Check
A gain of 2000 into a 12-bit converter spanning 3 V; independent errors 1.5, 0.4 and 1 mmHg.
(a)
Input-referred step?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Combined error in mmHg?
Check
A probe with τ = 8 s meets a 1.5 °C step.
(a)
Remaining error after 40 s?
(b)
Time to come within 0.2 °C?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A first-order probe with τ = 10 s.
(a)
Lag behind a ramp of 0.5 °C/min?
(b)
Amplitude ratio at 0.5 Hz?
Check
A catheter system with = 20 Hz and ζ = 0.15.
(a)
Amplitude ratio at 15 Hz?
(b)
Step overshoot in percent?
Check
A 350 Ω gauge with G = 2 at 300 microstrain, in a bridge excited at 10 V.
(a)
Resistance change?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
Approximate quarter-bridge output?
(c)
Output if four such gauges form a full bridge?
Check
A quarter bridge at 3 V with ΔR/R = 0.03 in its active arm.
(a)
Exact output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Percent error of the linear approximation?
Check
A foil gauge (G = 2) with a 15 ppm/°C temperature coefficient; a 2 kΩ pot at 0.5 of its travel with a 20 kΩ load.
(a)
Apparent strain for an 8 °C change, in microstrain?
(b)
Wiper output as a fraction of the excitation?
Check
A capacitive sensor: 0.5 cm² plates 0.1 mm apart, relative permittivity 5.
(a)
Capacitance?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
(b)
Change for a 4 µm gap reduction?
Check
A piezoelectric sensor with d = 150 pC/N and 4 nF, read across 5 GΩ, under a 0.5 N step.
(a)
Initial voltage?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
Low-frequency corner?
(c)
Voltage after 1 s?
Check
A charge amplifier with = 500 pF reads a sensor with d = 100 pC/N; a transducer whose sensitivity falls by a fraction 0.006 per °C.
(a)
Output size for 0.3 N?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
Percent gain error after 4 °C?
Check
A heart rate of 60 per minute whose first 8 harmonics must be reproduced by a catheter with = 80 Hz and ζ = 0.6.
(a)
Highest harmonic needed?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Percent error in that harmonic's amplitude?
Check
An LVDT's 4 kHz carrier reaches a phase-sensitive demodulator with an in-phase amplitude of 60 mV at 15° to the reference, plus 40 mV of quadrature pickup from cable capacitance.
(a)
Average output from the in-phase signal?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Average output with the quadrature pickup added?
(c)
Resistor for a 10 Hz RC output filter with 220 nF?
Check
A fiber with core index 1.49 and cladding 1.4 carries 5 µW at 880 nm to a photodiode with quantum efficiency 0.85.
(a)
The fiber's numerical aperture?
(b)
Acceptance half-angle in air?
(c)
Photocurrent?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A 120 Ω gauge (G = 2.1) measures 200 microstrain through two-wire leads totaling 0.5 Ω of copper (0.0039/°C), which warm by 25 °C during the test.
(a)
What strain does the bridge indicate, before the leads warm?
(b)
How much apparent strain does the warming add?
Check
A thermistor is 10 kΩ at 25 °C with B = 3400 K. What's its resistance at 40 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
What's the temperature coefficient, in %/°C, of a thermistor with B = 3950 K at 25 °C?
Check
What's the resistance of a platinum RTD with = 100 Ω (α = 0.00385/°C) at 20 °C?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
Check
A thermocouple with Seebeck coefficient 43 µV/°C has its junctions 5 °C apart. What voltage does it produce?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 3 kΩ thermistor carries 50 µA and has a dissipation constant of 5 mW/°C. How much does self-heating raise its temperature?
Check
At what wavelength does a body at 37 °C radiate most strongly?
Write units the way you normally would: 1.5 µm, 1500 nm and 0.0015 mm are the same answer.
Check
A thermistor is 10 kΩ at 25 °C with B = 3000 K.
(a)
What is its resistance at 40 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
What parallel resistor puts the combination's inflection at 40 °C?
Check
Thermistor: 1 kΩ at 25 °C, B = 3000 K, at 45 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
α: B = 2500 K at 0 °C?
Check
Self-heating: 1 µA, 1 kΩ, δ = 10 mW/°C?
Check
A transducer reads 0.3, 2.7 and 5.3 mV at 0, 50 and 100 mmHg.
(a)
What sensitivity, in mV per mmHg, does the least-squares line give?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What nonlinearity, as a percent of span, does the line leave?
Check
A sensor of 5 µV/unit and output impedance 300 Ω drives a 1 MΩ input; its sensitivity drifts by 0.005 per °C, and the room warms 25 °C.
(a)
What sensitivity per unit does the amplifier see?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
By what percent does the drift change every reading?
Check
A temperature probe has a 15 s time constant.
(a)
How far behind does it lag a temperature rising 0.2 °C per minute?
(b)
How long does it take to come within 0.05 °C after a 3 °C step?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A flush test's peaks shrink to 0.4 of the one before, 20 ms apart; the patient's heart rate is 120 per minute.
(a)
What is the damping ratio?
(b)
By what factor does the system scale the pressure wave's tenth harmonic?
Check
Four gauges (G = 2.1) at ±200 microstrain form a full bridge excited at 10 V; separately, one such gauge forms a quarter bridge.
(a)
What does the full bridge output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
By what percent does the quarter bridge's linear formula overstate its output?
Check
An active gauge (G = 2) at 100 microstrain and an identical unstrained dummy gauge, both with a temperature coefficient of 50 ppm/°C, form a half bridge excited at 5 V; both warm by 30 °C.
(a)
About what does the bridge output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What apparent strain would a single uncompensated gauge have shown?
Check
A capacitive pressure sensor's diaphragm, 1 cm² with relative permittivity 3 in the gap, sits 25 µm from a fixed plate and deflects 2 µm toward it.
(a)
What is its capacitance at rest?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
(b)
By how much does the capacitance rise?
Check
A piezoelectric sensor (d = 100 pC/N) under 1 N feeds a charge amplifier with = 1000 pF and = 100 GΩ.
(a)
What step output does it give?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What is its low-frequency corner?
Check
An LVDT's carrier reaches a phase-sensitive demodulator at 20 mV and 30° from the reference; the output filter's corner must be 10 Hz with a 100 nF capacitor.
(a)
What average output does the demodulator give?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What filter resistor does that corner need?
Check
A lock-in amplifier with a single-pole output filter of time constant 0.3 s sees noise of 5 nV/√Hz and an in-phase signal of 2 µV.
(a)
What noise remains at the output?
Write units the way you normally would: 1.5 nV, 1500 pV and 0.0015 µV are the same answer.
(b)
What signal-to-noise ratio (amplitude) does that give?
(c)
About how long must it wait after a change to settle within 5%?
Check
A photodiode with quantum efficiency 0.6 receives 1 µW at 940 nm, over a 10000 Hz bandwidth.
(a)
What photocurrent does it produce?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
How much shot noise does it carry?
Check
A thermistor is 10 kΩ at 25 °C with B = 3400 K.
(a)
What is it at 42 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
What is its temperature coefficient there, per °C?
Check
A platinum RTD with = 100 Ω (α = 0.00385/°C) at 37 °C carries 5 mA, and its dissipation constant is 10 mW/°C.
(a)
What is its resistance?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
How much does self-heating raise its reading?
Check
A thermocouple with a Seebeck coefficient of 40.6 µV/°C reads 0.5 mV, with its reference junction at 25 °C.
(a)
What temperature difference does that show?
(b)
What is the measuring junction's temperature?
Check
An infrared thermometer views 3 cm² of skin (emissivity 0.95) at 36 °C, in a 25 °C room.
(a)
What net power does the skin radiate?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
(b)
By what percent does that power change per °C of skin temperature?
Check
Potassium is 140 mM inside a cell and 3 mM outside. What's its Nernst potential at 37 °C?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Calcium is 1.2 mM outside a cell and 0.0001 mM inside. What's its Nernst potential at 37 °C?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A membrane is permeable only to K⁺ and Na⁺, with = 0.04. Potassium is 140 mM inside and 4 mM outside; sodium 12 mM inside and 145 mM outside. What's the membrane potential at 37 °C?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A membrane's conductances are = 1, = 0.02 and = 0.4 (relative units), with Nernst potentials −95, +66.6 and −90 mV. What's its potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
How many monovalent ions per square centimeter must cross a membrane of 1 µF/cm² to change its voltage by 50 mV?
Check
How many millivolts per tenfold concentration ratio does the Nernst equation give for an ion of charge 1 at 25 °C?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Potassium is 130 mM inside a cell; outside it rises from 3 to 9 mM, at 37 °C.
(a)
What is the Nernst potential before the rise?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
By how much does it rise?
Check
Nernst, z = 1: inside 5 mM, outside 2 mM?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
GHK with = 0.005, = 3 mM?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Parallel conductance: = 0.5, = 0.01, = 1?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
An axon is 2 µm across with = 5000 Ω·cm² and = 150 Ω·cm. What's its length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
Check
What's the membrane time constant for = 5000 Ω·cm² and = 0.8 µF/cm²?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
Check
A steady 20 mV depolarization is held at one point of an axon with λ = 1.5 mm. How much reaches 0.5 mm away?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A nerve stimulated at two points 25 cm apart gives response latencies of 11 ms and 3 ms. What's its conduction velocity?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
How long does an impulse take to travel 0.3 m along a myelinated fiber 8 µm across, at 6 m/s per µm?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
Check
A current dipole of 2.5 × 10⁻⁵ A·m sits in tissue of conductivity 0.4 S/m. What potential does it produce 20 cm away at 30° from its axis?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A nerve's myelinated fibers range from 3 to 10 µm across and conduct at 6 m/s per micrometer.
(a)
When do the first arrivals reach a site 0.8 m from the stimulus?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
(b)
Over how long are the arrivals spread?
Check
λ: d = 0.2 µm, = 1000 Ω·cm², = 300 Ω·cm?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
Check
Conduction velocity: 5 cm, latencies 20 and 1 ms?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
Dipole potential: 0.5 × 10⁻⁵ A·m, 0°, 40 cm, σ = 1?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Lead I reads 0.3 mV and lead II reads 0.5 mV. What does lead III read?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
With lead I = 0.3 mV and lead II = 0.5 mV, what does aVF read?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Lead I reads 0.4 mV and lead II reads 0.3 mV. What's the electrical axis?
Check
A heart vector of 0.8 mV points at 20°. What does a lead along 120° record?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
R waves are 20 mm apart on a trace recorded at 25 mm/s. What's the heart rate per minute?
Check
A QT interval of 0.38 s is measured at RR = 0.6 s. What's Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A recorder stores 3 ECG channels at 250 samples a second, 24 bits each, for 48 hours.
(a)
How much data does it collect?
(b)
What compression ratio fits it into 64 MB?
Check
Axis: I = 0.1 mV, II = 0.2 mV?
Check
Projection: M = 0.3 mV at −30°, lead at 120°?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Rate: 8 mm at 25 mm/s?
Check
A thermistor of 5 kΩ at 25 °C with B = 3500 K, at 39 °C.
(a)
Resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Coefficient in %/°C?
(c)
About how many ohms per °C?
Check
A platinum RTD with = 100 Ω; a thermocouple of 50 µV/°C.
(a)
RTD resistance at 30 °C?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
Thermocouple voltage for a 25 °C difference?
Check
A 4 kΩ thermistor carrying 20 µA with dissipation constant 4 mW/°C; a body at 40 °C.
(a)
Self-heating?
(b)
Peak emission wavelength?
Write units the way you normally would: 1.5 µm, 1500 nm and 0.0015 mm are the same answer.
Check
Potassium 130 mM inside and 4 mM outside; calcium 2.5 mM outside and 0.0003 mM inside; 37 °C.
(a)
Potassium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Calcium's?
Check
A K⁺/Na⁺ membrane with = 0.02 (140/3.5 mM K⁺, 12/145 mM Na⁺, inside/outside); separately, conductances 1.5, 0.3, 0.2 for K⁺, Na⁺, Cl⁻ with Nernst potentials −95, +66.6 and −90 mV.
(a)
GHK potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Parallel-conductance potential?
Check
A membrane of 1 µF/cm² and area 10 × 10⁻⁵ cm² swings by 80 mV.
(a)
Charge per area, in µC/cm²?
Write units the way you normally would: 1.5 µC/cm², 1500 nC/cm² and 0.0015 mC/cm² are the same answer.
(b)
Number of monovalent ions?
Check
An axon 1 µm across, = 4000 Ω·cm², = 200 Ω·cm, = 1.2 µF/cm².
(a)
Length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
Time constant?
(c)
What's left of a steady 15 mV at 0.5 mm?
Check
A conduction study with sites 20 cm apart and latencies 10 and 3 ms; a myelinated fiber 12 µm across.
(a)
Conduction velocity?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
(b)
Travel time over 0.5 m for the fiber?
Check
A 1.5 × 10⁻⁵ A·m dipole in tissue of 0.3 S/m.
(a)
Potential 25 cm away at 20°?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
At twice the distance?
Check
Lead I reads 0.4 mV and lead II 0.6 mV.
(a)
Lead III?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
aVL?
(c)
Electrical axis?
Check
A heart vector of 0.7 mV at 10°.
(a)
Lead I?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
The augmented lead along 90°?
Check
R waves 15 mm apart at 25 mm/s, with a QT of 0.42 s.
(a)
Heart rate per minute?
(b)
Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
Extracellular potassium rises from 4 to 7 mM; K⁺ is 140 mM inside, Na⁺ 12/145 mM and Cl⁻ 4/120 mM (inside/outside), with = 1 : 0.05 : 0.45, at 37 °C.
(a)
How far does the potassium Nernst potential rise?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
How far does the resting potential rise?
Check
A Holter recorder stores 3 channels at 250 samples a second and 16 bits for 24 hours, on 32 MB of memory.
(a)
How much raw data does a day produce?
(b)
What compression ratio fits it?
Check
A 5 kΩ thermistor (at 25 °C, B = 3400 K) is to be linearized around 37 °C with a parallel resistor.
(a)
What resistor puts the inflection at 37 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
An electrode has = 0.5 kΩ, = 50 kΩ and = 0.5 µF. What's the magnitude of its impedance at 100 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
Where's the corner frequency of an electrode with = 100 kΩ and = 0.02 µF?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Electrodes present a 300 mV DC offset to an amplifier stage whose output can swing ±2.5 V. What's the largest DC-coupled gain it can use?
Check
Two electrodes with half-cell potentials 0.223 V and −0.76 V share an electrolyte. What voltage appears between them?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A microelectrode of 10 MΩ sees 2 pF of stray capacitance. Where's the resulting low-pass corner?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A 100 kΩ electrode drives an amplifier with 1 MΩ input resistance. What percent of the signal is lost?
Check
An electrode pair presents 20 kΩ at 37 °C, recorded over 40 Hz of bandwidth, for a 1000 µV signal.
(a)
How much thermal noise does it add?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What signal-to-noise ratio does that leave, considering only this noise?
Check
Corner: = 5 kΩ, = 0.01 µF?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Max gain: 10 mV offset, ±1 V swing?
Check
Loss: 1 kΩ electrode, 0.5 MΩ input?
Check
An instrumentation amplifier has gain resistor = 5.6 kΩ, feedback resistors = 10 kΩ and a difference stage of gain 10. What's its gain?
Check
An amplifier has differential gain 100 and common-mode gain 0.001. What's its CMRR in decibels?
Check
A 20 mV common-mode voltage reaches an amplifier with 80 dB of CMRR. How large an input-referred error does that alone produce?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 20 mV common-mode voltage reaches inputs of 1000 MΩ common-mode impedance through electrodes that differ by 5 kΩ. What differential voltage does the mismatch create?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A difference amplifier with gain 5 uses resistors of tolerance 0.0001. What's its worst-case CMRR in dB?
Check
A 300 mV electrode offset must not push a first stage beyond ±3 V. If that stage uses all the gain it can, how much gain must follow the high-pass filter for a total of 2000?
Check
An amplifier's input noise is 4 nV/√Hz and 0.01 pA/√Hz; each input sees 1000 kΩ of source resistance at 37 °C, over a 150 Hz band.
(a)
What is the total noise density?
Write units the way you normally would: 1.5 nV/√Hz, 1500 pV/√Hz and 0.0015 µV/√Hz are the same answer.
(b)
What RMS noise is that over the band?
Check
INA gain: = 0.5 kΩ, = 5 kΩ, second stage 10?
Check
Mismatch error: 1 mV, ΔZ = 1 kΩ, = 5000 MΩ?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Difference-amp CMRR: G = 1, tolerance 5 × 10⁻⁵?
Check
A barrier must limit leakage from a 120 V, 50 Hz line to 50 µA.
(a)
What impedance must it present?
Write units the way you normally would: 1.5 MΩ, 1500 kΩ and 0.0015 GΩ are the same answer.
(b)
What is the largest capacitance it can have?
Check
An isolation amplifier with gain 100 has CMRR 110 dB and IMRR 110 dB; its input sees 0.5 mV of signal, 50 mV of common-mode voltage and 120 V between its commons.
(a)
What input-referred error does the isolation-mode voltage cause?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What is the total output?
Check
An isolated converter with 10 pF of interwinding capacitance and a signal isolator with 1 pF feed one input; 1 such inputs connect to a patient, with 230 V at 60 Hz across the barriers.
(a)
How much leaks through one input's barrier?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
How much in total through all 1?
Check
An isolator's IMRR is 140 dB at 60 Hz and falls by 20 dB per decade above. A 10 V disturbance at 1000 Hz appears between its commons.
(a)
What is its IMRR at 1000 Hz?
(b)
What input-referred error results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An ECG input is protected from a 2 kV defibrillator pulse lasting 10 ms by a 10 kΩ series resistor ahead of clamp diodes that hold the amplifier's input near 0 V.
(a)
What current flows through the resistor?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
How much energy must the resistor absorb?
Check
Largest barrier capacitance: 120 V, 50 Hz, 20 µA?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
Check
Input-referred error: 120 V at IMRR 120 dB?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Leakage: 5 pF at 120 V, 60 Hz?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A 120 V, 50 Hz line couples to a patient through 10 pF. What displacement current flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A displacement current of 0.2 µA leaves a patient through a 10 kΩ ground electrode. What's the body's common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
9 nA of line current couples into one ECG lead, and the electrode impedances differ by 5 kΩ. How large a differential pickup results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 60 Hz magnetic field of 1 µT threads a 0.01 m² loop formed by the leads and the body. What voltage is induced?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A driven-right-leg circuit has = 5 MΩ, = 5 MΩ and = 100 kΩ, with 0.1 µA of displacement current. What's the body's common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A 10 mV common-mode voltage must contribute less than 0.5 µV of input-referred error. What CMRR is needed?
Check
A floating monitor's common has 1000 pF to earth, and the patient couples 1 pF to a 120 V, 60 Hz line. The reference electrode is 100 kΩ, and the amplifier's CMRR is 60 dB.
(a)
What current flows through the patient?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
(b)
What common-mode voltage reaches the amplifier?
(c)
What input-referred error results?
Check
Displacement current: 100 V, 50 Hz, 20 pF?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
DRL: = 0.5 MΩ, = 0.05 µA, = 10 MΩ, = 200 kΩ?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Magnetic pickup: 0.1 µT, 0.005 m², 60 Hz?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 1 µV evoked response sits in 10 µV RMS of background. How many sweeps must be averaged for a signal-to-noise ratio of 10?
Check
A 16-bit converter spans 2 V. What's its step size?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A recorder has 64 channels sampled at 250 per second with 24-bit samples. What's its raw data rate?
Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.
Check
What's the time constant of a first-order high-pass filter with a 0.5 Hz corner?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A first-order high-pass has its corner at 0.05 Hz. How much phase lead does it give a 1 Hz component?
Check
About how wide a band does a 50 Hz notch with Q = 10 remove?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A spike detector's noise is Gaussian and flat from 0 to 5000 Hz; its threshold sits at 3 standard deviations.
(a)
How often does noise alone cross the threshold upward?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
How often does the noise cross zero?
Check
Sweeps for SNR 20: 0.1 µV in 5 µV?
Check
Data rate: 1 channels, 100 /s, 32 bits?
Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.
Check
High-pass lead: corner 0.01 Hz at 0.1 Hz?
Check
An electrode with = 0.5 kΩ, = 80 kΩ and = 0.2 µF.
(a)
Corner frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Impedance magnitude at 50 Hz?
(c)
Impedance at DC?
Check
Electrode offsets of 150 mV into a stage that swings ±3 V; half-cell potentials 0.8 and 0.34 V; a 5 MΩ microelectrode with 5 pF.
(a)
Largest DC-coupled gain?
(b)
Voltage between the two half-cells?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(c)
Microelectrode bandwidth?
Check
An instrumentation amplifier with = 2 kΩ, = 20 kΩ and a second stage of 5.
(a)
Gain?
(b)
Further gain needed for a total of 2000?
Check
An amplifier with differential gain 500 and common-mode gain 0.005, facing 40 mV of common-mode voltage.
(a)
CMRR?
(b)
Input-referred common-mode error?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
5 mV of common-mode voltage, electrodes differing by 10 kΩ, a common-mode input impedance of 500 MΩ; a difference stage of gain 10 with resistors of tolerance 0.0005.
(a)
Mismatch-induced differential voltage?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Worst-case CMRR of the difference stage?
Check
A 120 V, 50 Hz line couples to a patient through 5 pF; the patient is grounded through 80 kΩ.
(a)
Displacement current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Common-mode voltage?
Check
3 nA coupled into one lead with 10 kΩ of electrode mismatch; a 3 µT, 50 Hz field through a 0.1 m² loop.
(a)
Differential pickup from the lead current?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Magnetically induced voltage?
Check
A driven-right-leg circuit with = 2 MΩ, = 10 MΩ, = 50 kΩ and 0.1 µA of displacement current.
(a)
Common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Effective ground impedance?
Check
A common-mode voltage of 5 mV must leave less than 0.5 µV of error.
(a)
Required CMRR?
(b)
As a plain ratio?
Check
A 128-channel recorder sampling at 2000 per second with 18-bit converters spanning 2.5 V.
(a)
Step size?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Data rate?
Check
A first-order high-pass at 0.05 Hz; a 60 Hz notch with Q = 50.
(a)
Time constant?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
(b)
Phase lead at 1 Hz?
(c)
Notch width?
Check
An evoked response of 0.5 µV in 15 µV of background, with stimuli delivered 5 times per second.
(a)
Sweeps for an SNR of 6?
(b)
Recording time?
Check
An electrode pair presents 50 kΩ at 37 °C, recording a 1000 µV signal over 100 Hz.
(a)
How much thermal noise does it add?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What signal-to-noise ratio does that alone allow?
Check
Two electrodes present 20 and 10 kΩ at 60 Hz (treat them as resistive) to an amplifier with 100 MΩ common-mode input impedance; 20 mV of common mode is present.
(a)
What differential error does the mismatch create?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A notch filter at 50 Hz has Q = 10.
(a)
How wide a band does it remove?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What is the time constant of its ringing after a step?
Check
A membrane has = 1 : 0.05 : 0.6, with K⁺ 140/4 mM, Na⁺ 12/135 mM and Cl⁻ 4/120 mM (inside/outside), at 37 °C.
(a)
What is its resting potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
How far is it below the sodium Nernst potential (the driving force on Na⁺)?
Check
Calcium is 1.2 mM outside a cell and 0.0001 mM inside; the membrane sits at −60 mV, at 37 °C.
(a)
What is calcium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
What driving force pushes calcium inward?
Check
An action potential swings a membrane of 1 µF/cm² by 100 mV, in a cell 10 µm across (a sphere).
(a)
How many monovalent ions per square centimeter must cross?
(b)
How many ions in all, for the whole cell?
Check
An unmyelinated fiber 2 µm across has = 5000 Ω·cm² and = 150 Ω·cm; it's 4 mm long and sealed at its far end.
(a)
What is its length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What fraction of a steady voltage held at one end reaches the sealed end?
Check
A nerve's myelinated fibers range from 2 to 9 µm across, at 6 m/s per micrometer; a recording site is 0.6 m from the stimulus.
(a)
When does the response begin?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
(b)
Over how long is the volley dispersed?
Check
A current dipole of 1 × 10⁻⁵ A·m sits in tissue of 0.2 S/m.
(a)
What potential does it produce 20 cm away at 70° to its axis?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Lead I reads 0.4 mV and lead II 0.3 mV; R waves are 30 mm apart at 25 mm/s, and the QT interval is 0.4 s.
(a)
What is the electrical axis?
(b)
What is Bazett's corrected QT?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrode has = 0.5 kΩ, = 20 kΩ and = 2 µF.
(a)
What is its impedance's magnitude at 60 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
An instrumentation amplifier has = 5 kΩ, = 25 kΩ and a second stage of gain 50, with 90 dB of CMRR; 10 mV of common mode is present.
(a)
What is its gain?
(b)
What input-referred error does the common mode cause?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An amplifier's noise is 5 nV/√Hz and 0.05 pA/√Hz, and each input sees 200 kΩ at 37 °C, over 500 Hz.
(a)
What total input noise results?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
An isolated input's barrier has 20 pF and 140 dB of IMRR; 120 V at 60 Hz stands across it.
(a)
What leakage flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
What input-referred error does the isolation-mode voltage cause?
Check
A driven-right-leg circuit has = 1 MΩ, = 2 MΩ and = 50 kΩ, with 0.5 µA of displacement current.
(a)
What effective resistance ties the patient to the circuit?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
What common-mode voltage results?
Check
A battery-powered monitor's common has 100 pF to earth; the patient couples 1 pF to a 230 V, 60 Hz line, and the reference electrode is 100 kΩ.
(a)
What common-mode voltage reaches the amplifier?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
An evoked response of 1 µV sits in 10 µV RMS of EEG, recorded by a 24-bit converter spanning ±2 V after a gain of 1000.
(a)
How many sweeps reach a signal-to-noise ratio of 10?
(b)
What step does the converter resolve at the electrodes?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A spike detector's noise is flat from 0 to 3000 Hz, with its threshold at 3.5 standard deviations; the neuron fires 50 times a second.
(a)
How often does noise alone cross the threshold?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What fraction of detected events are false?
Check
A 4 µV/V/mmHg transducer at 4 V; a gain of 300 feeding a 14-bit converter spanning 10 V.
(a)
Output at 200 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Input-referred step?
(c)
That step in mmHg?
Check
A probe with τ = 6 s meets a 2 °C step; a catheter with = 50 Hz and ζ = 0.2.
(a)
Probe error after 30 s?
(b)
Catheter amplitude ratio at 10 Hz?
Check
Gauges with G = 2 at ±400 microstrain in a bridge at 10 V; separately, a quarter bridge with ΔR/R = 0.04.
(a)
Output with 4 active arms?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Exact quarter-bridge output?
Check
A capacitive sensor (0.5 cm², 0.05 mm, εr = 6); a piezoelectric sensor (d = 400 pC/N, 1 nF, read across 1 GΩ).
(a)
Capacitance?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
(b)
Initial piezo voltage for 0.5 N?
(c)
Piezo corner frequency?
Check
A thermistor of 8 kΩ at 25 °C with B = 3600 K, carrying 100 µA with dissipation constant 1 mW/°C.
(a)
Resistance at 40 °C?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Self-heating there?
Check
Potassium 135 mM inside and 3.5 mM outside; sodium 12 mM inside and 145 mM outside; = 0.06; 37 °C.
(a)
Potassium's Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
GHK resting potential (K⁺ and Na⁺ only)?
Check
An axon 1 µm across with = 5000 Ω·cm² and = 150 Ω·cm; a conduction study over 30 cm with latencies 8 and 3 ms.
(a)
Length constant?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
Conduction velocity?
Check
A 2 × 10⁻⁵ A·m dipole in tissue of 0.35 S/m.
(a)
Potential 20 cm away at 25°?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
On the axis at the same distance?
Check
Lead I reads 0.35 mV and lead II 0.8 mV; R waves are 24 mm apart at 25 mm/s; QT = 0.4 s.
(a)
Electrical axis?
(b)
Heart rate?
(c)
Bazett's QTc?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrode with = 0.5 kΩ, = 60 kΩ, = 0.3 µF; offsets up to 200 mV into a stage that swings ±4 V.
(a)
Impedance at 40 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Largest first-stage gain?
Check
An instrumentation amplifier ( = 3 kΩ, = 25 kΩ, unity second stage, CMRR 90 dB, common-mode input impedance 100 MΩ) with 25 mV of common mode and 50 kΩ of electrode mismatch.
(a)
Gain?
(b)
Total input-referred common-mode error, adding both causes?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
A 120 V, 50 Hz line couples through 4 pF to a patient on a driven-right-leg circuit ( = 8 MΩ, = 3 MΩ, = 25 kΩ).
(a)
Displacement current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Common-mode voltage?
Check
An evoked response of 0.5 µV in 10 µV of background; a 64-channel recorder at 500 per second with 16-bit samples.
(a)
Sweeps for an SNR of 8?
(b)
Data rate?
Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.
Check
A first-order high-pass at 0.05 Hz; a foil gauge (G = 2.1) with a 50 ppm/°C coefficient.
(a)
Phase lead at 0.5 Hz?
(b)
Apparent strain in microstrain for a 3 °C change?
Check
A pulse oximeter's 940 nm LED is switched at 1 kHz, and its photodiode (quantum efficiency 0.9) receives 2 µW of it. After amplification, the demodulator sees a 40 mV carrier at 20° to its reference.
(a)
Photocurrent from the LED's light?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Average demodulator output?
Check
An isolated front end has a 5 pF converter and a 2 pF signal isolator across its barrier, with 120 dB of IMRR and a gain of 500; 230 V at 60 Hz stands across the barrier.
(a)
What leakage flows through the patient?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
How much line interference appears at the output?
Check
A saline-filled catheter of radius 0.5 mm and length 0.5 m feeds a transducer whose compliance is 0.05 mm³ per 100 mmHg. What's the natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
For that catheter (radius 0.5 mm, length 0.5 m, transducer compliance 0.05 mm³ per 100 mmHg, η = 0.001 Pa·s), what's the damping ratio?
Check
A catheter system has = 150 Hz with a transducer compliance of 7.5 × 10⁻¹⁶ m³/Pa. A 3 mm³ air bubble enters. What's the new natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
In a fast-flush test, each oscillation peak is 0.25 of the one before. What's the damping ratio?
Check
A pressure transducer sits 10 cm below the heart, connected by a liquid column of density 1000 kg/m³. How much does it overread?
Check
What's 120 mmHg in kilopascals?
Write units the way you normally would: 1.5 kPa, 1500 Pa and 0.0015 MPa are the same answer.
Check
A pressure wave with heart rate 90 per minute is recorded through a catheter system with = 15 Hz and ζ = 0.6.
(a)
What is the frequency of its tenth harmonic?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
By what factor does the system scale that harmonic's amplitude?
Check
: r = 0.2 mm, L = 0.2 m, compliance 0.1 mm³/100 mmHg?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
ζ from a peak ratio of 0.2?
Check
Level error: 1 cm, ρ = 1000?
Check
Systolic pressure is 130 mmHg and diastolic 70 mmHg. What's the mean arterial pressure?
Check
An oscillometric monitor finds a mean pressure of 95 mmHg, and the diastolic is 65 mmHg. What systolic pressure is consistent with the one-third rule?
Check
An oscillometric envelope follows with MAP = 85 mmHg and w = 20 mmHg. Where on the high-pressure side does it fall to 0.6 of its peak?
Check
A cuff deflates at 3 mmHg/s with a heart rate of 50 per minute. How far does it fall between beats?
Check
What bladder width suits an arm with a 32 cm circumference?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
A cuff starts deflating from 180 mmHg at 3 mmHg/s. When does it reach a systolic pressure of 110 mmHg?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An artery's wall has Young's modulus 0.8 MPa and thickness 0.4 mm, with an inner diameter of 20 mm; blood density is 1060 kg/m³.
(a)
What pulse-wave velocity does the Moens–Korteweg equation give?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
(b)
How long does the pulse take to travel 1 m?
Check
MAP: 90/50 mmHg?
Check
Low-side ratio point: MAP 60, w = 10, k = 0.9?
Check
Drop per beat: 1 mmHg/s, 40 per minute?
Check
Cardiac output is 6.4 L/min at 60 beats per minute, the valve is open 0.33 s per beat, and the mean gradient across it is 80 mmHg; the discharge coefficient is 0.6.
(a)
What is the flow while the valve is open?
Write units the way you normally would: 1.5 mL/s, 1500 µL/s and 0.0015 L/s are the same answer.
(b)
What is the valve's effective area?
Check
A Doppler study finds a 3 m/s jet through a valve.
(a)
What gradient does Bernoulli's equation give (blood density 1060 kg/m³)?
Check
Blood moves at 0.9 m/s where the static pressure is 3 mmHg.
(a)
How much higher does a catheter facing upstream read?
(b)
What percent error is that?
Check
A stethoscope's tube is 0.5 m long, closed at the ear.
(a)
Where is its resonance number 1?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A phonocardiograph must capture murmurs up to 300 Hz, sampling at 5 times that frequency with 16-bit samples.
(a)
What sampling rate does it use?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
How much data does an hour of one channel take?
Check
Aortic valve area: 5 L/min, 70 per min, 0.32 s, 60 mmHg?
Check
Gradient for a 2.5 m/s jet?
Check
Stethoscope resonance 1 for a 0.4 m tube?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A patient takes up 250 mL of oxygen per minute. Arterial blood carries 180 mL/L and mixed venous blood 70 mL/L less. What's the cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
A 5 mg dye bolus gives a first-pass concentration curve whose area is 30 mg·s/L. What's the flow?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
A 5 mg bolus's curve accumulates 30 mg·s/L before recirculation, at which point the concentration is 0.5 mg/L and falling exponentially with a 6 s time constant. What's the flow?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
5 mL of injectate at 0 °C enters blood at 37 °C; the temperature curve's area is 8 °C·s and K = 1.08. What's the cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
A cardiac output of 5.5 L/min at 60 beats per minute means what stroke volume?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
What's the cardiac index, in L/min/m², for an output of 5.5 L/min and a body surface area of 1.6 m²?
Write units the way you normally would: 1.5 L/min/m² and 1500 mL/min/m² are the same answer.
Check
A 5 mg dye bolus gives a curve with an area of 30 mg·s/L and a mean transit time of 15 s.
(a)
What is the cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
What volume lies between injection and sampling?
Check
Fick: 150 mL/min uptake, 25 mL/L difference?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
Stewart–Hamilton: 1 mg, area 10 mg·s/L?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
Thermodilution: 3 mL at 0 °C, area 10 °C·s, K = 1.08?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
An electromagnetic flowmeter applies 0.03 T across a 3 mm vessel where blood moves at 1 m/s. What voltage appears?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Blood moves at a mean 0.1 m/s in a vessel 8 mm across. What's the volume flow?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
5 MHz ultrasound meets blood at 0.2 m/s, at 60° to the flow. What's the Doppler shift?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A 2 MHz Doppler at 60° measures a shift of 2000 Hz. What's the blood velocity?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
A transit-time probe's path is 10 mm long at 45° to blood moving at 0.2 m/s. What's the upstream–downstream time difference?
Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.
Check
A 3.5 MHz pulsed Doppler looks at a vessel 4 cm deep, at 60°. What's the fastest velocity it can measure without aliasing?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
A 5 MHz Doppler at 60° watches an 8 mm vessel with a parabolic profile peaking at 0.4 m/s.
(a)
What is the highest Doppler shift it measures?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What is the volume flow?
Check
Doppler shift: 1 MHz, 0.05 m/s, 70°?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
EM flowmeter: 0.005 T, 1 mm, 2 m/s?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
Check
Aliasing limit: 1 MHz, 1 cm, 60°?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
A limb segment 5 cm long measures 50 Ω, and a pulse lowers it by 0.2 Ω. With blood resistivity 1.5 Ω·m, how much volume does the pulse add?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
After a venous-occlusion cuff inflates, an 800 mL calf segment swells at 0.5 mL/s. What's the blood flow in mL per minute per 100 mL of tissue?
Check
A four-electrode plethysmograph drives 1 mA through a 20 Ω segment. What's the pulsatile signal for a 0.2 Ω change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Impedance cardiography: bands 25 cm apart, = 30 Ω, = 0.8 Ω/s, ejection time 0.3 s, = 1.3 Ω·m. Estimate the stroke volume.
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
A hot-film sensor held 6 K above blood draws 0.1 W. Its calibration is with a = 2 mW/K and b = 4 mW/(K·(m/s)^½). What's the blood velocity?
Write units the way you normally would: 1.5 m/s, 1500 mm/s and 0.0015 km/s are the same answer.
Check
A strain gauge around a limb of 35 cm circumference stretches by 0.1 mm in 10 s of venous occlusion.
(a)
By what percent does the limb's volume grow?
(b)
What inflow is that, per 100 mL of tissue?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
ΔV for L = 4.8 cm, Z = 44 Ω, ΔZ = 0.24 Ω, = 1.4?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
Flow per 100 mL per minute: 0.26 mL/s into 660 mL?
Check
A saline catheter of radius 0.4 mm and length 0.8 m on a transducer of compliance 0.03 mm³ per 100 mmHg.
(a)
Natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Damping ratio (η = 0.001 Pa·s)?
Check
A catheter system with = 100 Hz, ζ = 0.06 and transducer compliance 5 × 10⁻¹⁶ m³/Pa gains a 2 mm³ bubble.
(a)
New natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
New damping ratio?
Check
A flush test gives a peak ratio of 0.15 with peaks 30 ms apart; the transducer sits 15 cm above the heart (saline column).
(a)
Damping ratio?
(b)
Natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(c)
Size of the leveling error?
Check
A pressure of 115/70 mmHg measured with a cuff deflating at 4 mmHg/s, heart rate 90 per minute.
(a)
Mean arterial pressure?
(b)
Deflation between beats?
Check
An oscillometric envelope with M = 85 mmHg and w = 20 mmHg.
(a)
Systolic estimate at ratio 0.6?
(b)
Diastolic estimate at ratio 0.85?
Check
Oxygen uptake 220 mL/min with an arteriovenous difference of 35 mL/L; heart rate 95; body surface area 2.1 m².
(a)
Cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Stroke volume?
(c)
Cardiac index in L/min/m²?
Check
A 3 mg dye bolus; the first pass accumulates 35 mg·s/L before recirculation, where the concentration is 1 mg/L falling with τ = 5 s.
(a)
Flow with the tail included?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Percent error from ignoring the tail?
Check
5 mL of injectate at 0 °C into blood at 37 °C; curve area 6 °C·s; K = 1.08.
(a)
Cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Curve area, in °C·s, that the same injectate would give at 5 L/min?
Check
An electromagnetic probe applies 0.02 T across a 4 mm vessel with mean velocity 0.8 m/s.
(a)
Voltage?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Volume flow?
Check
A transit-time path of 8 mm at 45° to blood at 0.3 m/s.
(a)
Time difference?
Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.
(b)
Travel time with no flow?
Check
A 3 MHz Doppler at 60° to blood moving at 0.3 m/s.
(a)
Shift?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Percent velocity error if the true angle is 5° smaller than assumed?
Check
A 2 MHz pulsed Doppler at a depth of 5 cm and 60°.
(a)
Highest pulse repetition frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Highest unaliased velocity?
Check
A four-electrode impedance plethysmograph drives 1 mA at 100 kHz through an 8 cm limb segment of 40 Ω. Each pulse lowers the impedance by 0.12 Ω; blood resistivity is 1.4 Ω·m.
(a)
Steady voltage across the sensing electrodes?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Pulsatile signal?
(c)
Volume added by each pulse?
Check
In venous-occlusion plethysmography a 900 mL calf segment swells at 0.6 mL/s after the cuff inflates. After release, the extra volume drains with a time constant of 6 s.
(a)
Arterial inflow, in mL/min per 100 mL of tissue?
(b)
About how long does 95% of the extra volume take to drain?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A transducer sits 5 cm below the right atrium, connected by saline (1000 kg/m³).
(a)
What offset does that add?
(b)
What percent error is that on a central venous pressure of 5 mmHg?
Check
A spirometer at 23 °C (water vapor pressure 17.5 mmHg) and barometric pressure 760 mmHg measures a volume. By what factor must it be multiplied to get the BTPS volume?
Check
A 5 L spirometer starts with 8% helium; after the subject joins at FRC and the gas mixes, it reads 6.5%. What's the FRC?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
In a body box, panting against a closed shutter compresses the lung gas by 20 mL and raises its pressure by 15 cmH₂O, from an absolute dry-gas pressure of 950 cmH₂O. What's the lung volume?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
In the single-exponential model of a forced expiration with time constant 0.6 s, what's FEV1/FVC?
Check
A patient breathes 500 mL per breath at 20 breaths per minute with 120 mL of dead space. What's the alveolar ventilation?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
End-tidal gas contains 5% CO₂ (of the dry gas) at a barometric pressure of 700 mmHg. What's its CO₂ partial pressure in the saturated alveolar gas?
Check
Arterial is 40 mmHg and mixed expired 22 mmHg, with a tidal volume of 600 mL.
(a)
What fraction of each breath is dead space?
(b)
How many milliliters is that?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
Check
Helium FRC: 2 L, 5% → 4.5%?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
FEV1/FVC for τ = 0.5 s?
Check
Alveolar ventilation: 200 mL, 80 mL dead space, 40/min?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
Lung compliance is 0.15 L/cmH₂O and chest-wall compliance 0.1 L/cmH₂O. What's the total respiratory compliance, in L/cmH₂O?
Check
A Fleisch pneumotachometer has 400 tubes, each 0.8 mm in diameter and 5 cm long. What pressure drop does 2 L/s of air (η = 1.8 × 10⁻⁵ Pa·s) produce?
Write units the way you normally would: 1.5 Pa, 1500 mPa and 0.0015 kPa are the same answer.
Check
A patient's airway resistance is 2 cmH₂O·s/L and total compliance 0.05 L/cmH₂O. What's the time constant of passive exhalation?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A respiratory system has inertance 0.01 cmH₂O·s²/L and compliance 0.05 L/cmH₂O. At what frequency does its forced-oscillation reactance cross zero?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
After a nitrogen washout the collected 40 L of exhaled gas contains 3.5% nitrogen. Alveolar nitrogen started at 80% and ended near zero. What's the FRC?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
A single-breath test at sea level has = 5500 mL, an 8 s breath-hold, and alveolar CO falling from 0.3% to 0.2%. What's , in mL/(min·mmHg)?
Check
A breath of 0.4 L is drawn at constant flow over 2 s, with compliance 0.05 L/cmH₂O and resistance 8 cmH₂O·s/L, 12 times a minute.
(a)
How much work does each breath take?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
(b)
What power do the respiratory muscles deliver?
Check
Total compliance from = 0.14 and = 0.076 L/cmH₂O?
Check
Resonant frequency for I = 0.0069, C = 0.038?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Nitrogen washout: 42 L at 2.6%, from 79%. FRC?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
Red AC/DC is 0.02/0.8 and infrared AC/DC is 0.05/1.5. What's the ratio of ratios?
Check
An oximeter calibrated as SpO₂ = 110 − 25R measures R = 0.7. What does it display, in percent?
Check
Using extinction coefficients of 3227 and 320 (Hb, HbO₂) at 660 nm and 693 and 1214 at 940 nm, what R does the Beer–Lambert model predict at 85% saturation?
Check
Light passes 0.5 cm through hemoglobin at 0.002 M with an extinction coefficient of 1000 cm⁻¹/M. What fraction is transmitted?
Check
Blood has 12 g/dL of hemoglobin at saturation 0.8 and a of 100 mmHg. What's its oxygen content?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
A sample transmits a fraction 0.2 of the incident light. What's its absorbance?
Check
An oximeter's infrared perfusion index is 0.5%, with the DC level at 0.5 of a 16-bit converter's range.
(a)
How many counts does the pulse span?
Check
R: red 0.005/0.5, infrared 0.08/2?
Check
Beer–Lambert R at 80%?
Check
Oxygen content: Hb 6 g/dL, S = 0.6, = 500 mmHg?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
A pH electrode at 37 °C sees the pH change by 0.1. How much does its voltage change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
Blood has 24 mM bicarbonate and a of 30 mmHg. What's its pH?
Check
A potassium electrode with selectivity coefficient = 0.0002 measures 4 mM K⁺ in 140 mM Na⁺. By what percent does it read high?
Check
An amperometric sensor reacts 2 pmol/s of analyte, with 2 electrons per molecule. What current flows?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A glucose sensor has = 100 nA and = 10 mM. What current does it give at 15 mM?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A sensor with = 20 mM is calibrated with its initial linear slope. By what percent does it read low at 4 mM?
Check
A Severinghaus electrode at 37 °C sees rise from 20 to 60 mmHg.
(a)
By how much does the film's pH fall?
(b)
How much does its voltage change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
pH: 10 mM bicarbonate, = 20 mmHg?
Check
Michaelis–Menten: = 20 nA, = 2 mM, G = 30 mM?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
Faraday current: 0.1 pmol/s, n = 1?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A sample reads 25% transmittance. What's its absorbance?
Check
A standard of 8 mg/dL reads 20% transmittance; the unknown reads 60%. What's the unknown's concentration, in mg/dL?
Check
A photometer reads transmittance to within ±0.2% (of full scale). What's the relative concentration error for a sample with A = 0.434?
Check
A blood count gives hematocrit 42% and red cells 3.8 million per µL. What's the MCV, in fL?
Check
Hemoglobin is 13.5 g/dL and hematocrit 34%. What's the MCHC?
Check
An impedance-counting aperture is 50 µm in diameter and 100 µm long, filled with diluent of resistivity 1.2 Ω·m. What's its resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
Check
A protein with mobility 5 × 10⁻⁵ cm²/(V·s) migrates in a field of 5 V/cm. How far does it move in 60 minutes?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Blood with 5 million cells per µL is diluted 1:10000 and drawn through a 100 µm diameter aperture 100 µm long.
(a)
How many cells does the aperture hold on average?
(b)
What is the chance another cell is present when one passes?
Check
Absorbance at 8% transmittance?
Check
MCV in fL for HCT 39% and RBC 3.4 million/µL?
Check
Standard 4.1 mg/dL at 17%T, unknown at 57%T. Concentration in mg/dL?
Check
Cardiac output is 4 L/min at 65 per minute, with 0.34 s of ejection per beat and a mean aortic-valve gradient of 50 mmHg; separately, Doppler measures a 2.5 m/s jet.
(a)
What is the valve's area?
(b)
What gradient does the simplified Bernoulli equation give for the jet?
Check
An end-hole catheter faces blood flowing at 1.5 m/s where the static pressure is 15 mmHg.
(a)
By how much does it read high?
(b)
What percent error is that?
Check
A saline catheter of radius 0.4 mm and length 0.9 m feeds a transducer of 0.05 mm³ per 100 mmHg, with a 2 mm³ air bubble trapped in the dome.
(a)
What is the natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What is the damping ratio?
Check
An oscillometric envelope follows with MAP = 85 mmHg and w = 25 mmHg; the monitor uses ratios of 0.55 and 0.85.
(a)
What systolic pressure does it report?
(b)
What diastolic pressure?
Check
5 mL of injectate at 0 °C enters blood at 37 °C; the temperature curve's area is 7 °C·s, with K = 1.08, at 90 beats per minute.
(a)
What is the cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
What is the stroke volume?
Check
An artery's wall has E = 0.4 MPa and thickness 0.4 mm with a 16 mm lumen; a pulse travels 0.9 m, and a pressure rise stiffens the wall by a factor of 1.1.
(a)
How long does the pulse take?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
(b)
How much shorter does the stiffening make it?
Check
A spirometer at 20 °C (water vapor 23.8 mmHg) and 760 mmHg measures a vital capacity of 3 L.
(a)
What is it at body conditions (BTPS)?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
A helium spirometer of 5 L starts at 8% helium and settles at 6.5% after the subject joins at FRC; separately, a body box at 1000 cmH₂O alveolar pressure shows a 20 mL compression for a 1 cmH₂O rise.
(a)
What FRC does helium dilution give?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
What thoracic gas volume does the body box give?
Check
A respiratory system has R = 2 cmH₂O·s/L, C = 0.05 L/cmH₂O and inertance 0.02 cmH₂O·s²/L; it breathes 0.8 L at constant flow over 1 s, 12 times a minute.
(a)
At what frequency is it resonant?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
What power do the muscles deliver?
Check
An oximeter reads red AC/DC = 0.008/1 and infrared AC/DC = 0.02/1.3, calibrated as SpO₂ = 110 − 25R; the patient has 12 g/dL of hemoglobin and a of 60 mmHg.
(a)
What saturation does it display?
(b)
What oxygen content does that imply?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
Blood has 24 mM bicarbonate and a of 30 mmHg; a Severinghaus electrode at 37 °C then sees rise by a factor of 3.
(a)
What is the blood's pH?
(b)
How much does the electrode's output change?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A glucose sensor has = 50 nA and = 10 mM; separately, an amperometric sensor reacts 0.5 pmol/s with 2 electrons each.
(a)
What current does the glucose sensor give at 15 mM?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
(b)
What current does the amperometric reaction carry?
Check
A photometer's standard of 5 mg/dL reads 30% transmittance and a sample 70%; separately, blood with 5 million cells/µL is diluted 1:20000 into a counter whose 80 µm, 100 µm long aperture holds a volume of 5.03 × 10⁻⁴ µL.
(a)
What is the sample's concentration?
Write units the way you normally would: 1.5 mg/dL, 1500 µg/dL and 0.0015 g/dL are the same answer.
(b)
What is the chance of a second cell in the aperture when one passes?
Check
A forced expiration follows the single-exponential model with FVC = 4.5 L and time constant 0.3 s.
(a)
What is FEV₁/FVC?
(b)
What is the peak flow?
Write units the way you normally would: 1.5 L/s and 1500 mL/s are the same answer.
Check
Arterial is 38 mmHg and mixed expired 22 mmHg, in a patient breathing 600 mL at 24 breaths per minute.
(a)
What fraction of each breath is dead space?
(b)
What is the alveolar ventilation?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
Check
A detector 350 mm across must show 2 lp/mm. How many pixels does it need across?
Check
A pixel averages 2500 photons, and a feature changes the count by 0.01 (as a fraction). What's the feature's signal-to-noise ratio?
Check
How many photons per detail area are needed to see a contrast of 0.01 with a detection threshold of J = 4?
Check
An image has 1024 × 1024 pixels. About how many pixels exceed the mean by more than J = 3 standard deviations from noise alone?
Check
What's the MTF of a Gaussian blur with σ = 0.1 mm at 1 lp/mm?
Check
A Gaussian blur (σ = 0.05 mm) is followed by 0.1 mm pixels. What's the system MTF at 2 lp/mm?
Check
A detector absorbs a fraction 0.3 of 200 incident photons per pixel and adds 30 photons RMS of electronic noise.
(a)
What output signal-to-noise ratio does it achieve?
(b)
What is its detective quantum efficiency?
Check
SNR for 1200 photons and contrast 0.0074?
Check
Photons needed: contrast 0.01, J = 3.3?
Check
Gaussian MTF: σ = 0.046 mm at 0.87 lp/mm?
Check
What fraction of a narrow X-ray beam passes through 5 cm of tissue with μ = 0.19 cm⁻¹?
Check
What's the half-value layer of a material with μ = 0.19 cm⁻¹?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
A voxel has μ = 0.23 cm⁻¹ and water's is 0.19 cm⁻¹. What's its CT number, taking air as zero?
Check
A CT detector sees 0.02 of the unattenuated intensity. What's the projection value, ∫μ dl?
Check
How many detected photons give a relative noise of 0.01?
Check
Bone with μ = 0.45 cm⁻¹ replaces 5 cm of soft tissue with μ = 0.19 cm⁻¹ in a beam's path. What's the contrast, the fractional drop in transmitted intensity?
Check
A CT scan at 1 mSv is changed: pixel width scaled by 0.5, slice thickness by 0.4, and target noise (standard deviation) by a factor of 0.7.
(a)
What dose does the new protocol need?
Check
Transmission: μ = 0.1 cm⁻¹, x = 0.5 cm?
Check
HU: μ = 0, water 0.17?
Check
Photons for noise 0.01?
Check
A spirometer at 21 °C (water vapor 18.7 mmHg), barometric pressure 760 mmHg, records 5 L.
(a)
BTPS factor?
(b)
Volume in the lungs?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
Helium in a 4 L spirometer falls from 9% to 7%; a body box sees 25 mL of compression and 9 cmH₂O rise from 1000 cmH₂O.
(a)
FRC by helium dilution?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
Volume by plethysmography?
Check
A forced expiration with FVC = 3 L and τ = 0.35 s; quiet breathing at 650 mL, 170 mL dead space, 12 per minute.
(a)
FEV1?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
Alveolar ventilation?
(c)
Minute ventilation?
Check
End-tidal gas with 4.5% CO₂ (of the dry gas) at a barometric pressure of 520 mmHg.
(a)
End-tidal CO₂ partial pressure?
(b)
What CO₂ percentage would correspond to 40 mmHg at this pressure?
Check
Red AC/DC = 0.015/0.9; infrared AC/DC = 0.04/1.4.
(a)
Ratio of ratios?
(b)
Displayed SpO₂ with the 110 − 25R calibration, in percent?
Check
Saturation 75% in the Beer–Lambert model (coefficients 3227/320 at 660 nm, 693/1214 at 940 nm); blood with 10 g/dL Hb at saturation 0.97 and = 95 mmHg.
(a)
Theoretical R?
(b)
Oxygen content?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
A pH electrode at 25 °C; blood with 28 mM bicarbonate and = 55 mmHg.
(a)
Voltage change for 0.2 pH units?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Blood pH?
Check
A K⁺ electrode with = 0.0005 in 3 mM K⁺ and 145 mM Na⁺; an amperometric sensor reacting 5 pmol/s with 2 electrons each.
(a)
Percent overreading?
(b)
Sensor current?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A glucose sensor with = 60 nA and = 12 mM at 12 mM.
(a)
Current?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
(b)
What glucose does a meter using the initial linear slope report?
(c)
Percent low reading of a linear calibration?
Check
Soft tissue with μ = 0.16 cm⁻¹, 12 cm thick; a bone of μ = 0.7 cm⁻¹ replacing 4 cm of it.
(a)
Transmitted fraction through the tissue?
(b)
Half-value layer of the tissue?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
(c)
Bone contrast?
Check
A voxel with μ = 0.26 cm⁻¹ against water's 0.19; a target relative noise of 0.02.
(a)
CT number?
(b)
Detected photons needed?
Check
An 80 kV X-ray tube; a CT detector with 256 bins; a projection that transmits 0.05 of the beam.
(a)
Shortest wavelength?
Write units the way you normally would: 1.5 nm, 1500 pm and 0.0015 µm are the same answer.
(b)
Views needed?
(c)
Projection value ∫μ dl?
Check
A patient's lung compliance is 0.15 L/cmH₂O and chest-wall compliance 0.15 L/cmH₂O, with airway resistance 8 cmH₂O·s/L and inertance 0.02 cmH₂O·s²/L.
(a)
Total compliance, in L/cmH₂O?
(b)
Time constant of passive exhalation?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
(c)
Resonant frequency in a forced-oscillation test?
Check
A cholesterol standard of 100 mg/dL reads 30% transmittance and a patient's serum 65%. The same patient's blood count shows hematocrit 44% and 4.2 million red cells per µL.
(a)
The patient's cholesterol, in mg/dL?
(b)
MCV, in fL?
(c)
Relative concentration error from a 0.5% transmittance error at the patient's reading?
Check
An X-ray detector pixel averages 4000 photons. A detail has contrast 0.02, and detection needs an SNR of J = 5.
(a)
The detail's SNR at this exposure?
(b)
Photons per pixel needed to detect it?
(c)
By what factor must the dose rise?
Check
Ultrasound passes from a medium of 1.63 MRayl into one of 1.99 MRayl. What fraction of the intensity reflects?
Check
An echo arrives 65 µs after the pulse. How deep is the reflector?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue attenuates by 0.5 dB/(cm·MHz). How much is a 2 MHz echo from 15 cm deep attenuated?
Check
What's the axial resolution of a 2-cycle pulse at 2 MHz in soft tissue?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
Check
An image of 64 lines reaches 15 cm deep. What's the highest possible frame rate, in frames per second?
Check
What impedance, in MRayl, should a quarter-wave layer have to match a 30 MRayl transducer to 1.5 MRayl tissue?
Check
An ultrasound transducer 6 mm across works at 5 MHz in soft tissue (1540 m/s).
(a)
How long is its near field?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What lateral resolution does focusing at f-number 4 give?
Check
Intensity reflection: 0.0004 → 1 MRayl?
Check
Round-trip loss: α = 0.3, 1 MHz, 25 cm?
Check
Frame rate: 2 cm, 32 lines?
Check
At what frequency do protons resonate in a 1.5 T field?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
Check
What's the excess fraction of aligned protons, in parts per million, at 1.5 T and 300 K?
Check
A tissue has T1 = 1200 ms and T2 = 120 ms. What relative spin-echo signal does it give with TR = 1000 ms and TE = 15 ms?
Check
A readout gradient of 10 mT/m and a receiver bandwidth of 32 kHz cover what field of view?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
An RF pulse of 0.5 kHz bandwidth is applied with a 10 mT/m slice gradient. How thick a slice does it excite?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
Check
A tracer with a 360.6 min half-life starts at 370 MBq. How much remains after 30 min?
Check
A scan's voxels are scaled by 0.7 in each dimension, with 2 times as many averages, and the field goes from 1.5 T to 3 T.
(a)
By what factor does the SNR change, taking it as proportional to field strength?
Check
Larmor frequency at 1 T?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
Check
Spin echo: TR 100, TE 5, T1 4000, T2 2000 ms?
Check
Decay: 50 MBq, half-life 400 min, after 10 min?
Check
Tissue has a rheobase of 1 mA and a chronaxie of 0.3 ms. What threshold current does a 1 ms pulse need?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A pacing pulse of 2.5 V lasts 0.2 ms into an 800 Ω lead. What energy does it deliver?
Write units the way you normally would: 1.5 µJ, 1500 nJ and 0.0015 mJ are the same answer.
Check
A pacemaker draws 8 µA continuously and delivers 2 µC per pulse at 60 pulses per minute from a 2.5 Ah battery. How many years does the battery last?
Check
A defibrillator charges a 32 µF capacitor to 0.7 kV. How much energy does it store?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
Check
A 32 µF capacitor discharges into 50 Ω and is cut off after 6 ms. What fraction of its stored energy is delivered?
Check
A defibrillator at 3 kV discharges into a 40 Ω chest. What's the peak current?
Write units the way you normally would: 1.5 A, 1500 mA and 0.0015 kA are the same answer.
Check
An ICD stores 35 J in a 100 µF capacitor, charged through a converter drawing 12 W at efficiency 0.85.
(a)
To what voltage must it charge?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
How long does charging take?
Check
Threshold: rheobase 0.1 mA, chronaxie 0.1 ms, width 5 ms?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
Stored energy: 10 µF at 0.5 kV?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
Check
Delivered fraction: 25 Ω, 20 µF, cut at 10 ms?
Check
A stimulating electrode has an area of 0.005 cm², and the safety criterion uses k = 1.7.
(a)
What is the largest safe charge per phase?
Write units the way you normally would: 1.5 µC, 1500 nC and 0.0015 mC are the same answer.
(b)
What charge density is that?
Check
A 1 mA constant-current phase lasts 50 µs through an electrode of 0.05 cm² with 40 µF/cm² of double-layer capacitance and 0.5 kΩ of access resistance.
(a)
How far does the double layer charge?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
What compliance voltage does the pulse need?
Check
A cochlear implant runs 16 channels at 500 biphasic pulses per second each, with 50 µs phases of 1.5 mA, from a 5 V supply.
(a)
What average current does stimulation draw?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
What power is that?
Check
A recording array has 256 channels sampled at 20 kHz with 16 bits.
(a)
What raw data rate does it produce?
Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.
Check
A recording electrode of 1 MΩ is used over a 3 kHz band at 37 °C.
(a)
What RMS thermal noise does it contribute?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
What is a 200 µV spike's peak-to-noise ratio, considering only that noise?
Check
Safe charge per phase: 0.002 cm², k = 1.85?
Write units the way you normally would: 1.5 µC, 1500 nC and 0.0015 mC are the same answer.
Check
Double-layer voltage: 1 mA, 100 µs, 0.03 cm², 30 µF/cm²?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
Data rate: 128 channels, 20 kHz, 16 bits?
Write units the way you normally would: 1.5 Mb/s, 1500 kb/s and 0.0015 Gb/s are the same answer.
Check
A syringe pump pushes a 1.5 cm diameter plunger with a lead screw of 0.5 mm pitch, turned by a stepper of 400 steps per revolution. At what step rate does it deliver 20 mL/h?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
Blood enters a dialyzer at 300 mL/min with urea at 80 mg/dL and leaves at 45 mg/dL. What's the dialyzer's clearance, in mL/min?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
A dialyzer clears 220 mL/min for 3 hours on a 90 kg patient (body water 58% of weight). What's Kt/V?
Check
A session delivers Kt/V = 1.2. Ignoring urea generation, what's the urea reduction ratio?
Check
A ventilator delivers 0.5 L at 0.5 L/s with PEEP 5 cmH₂O to lungs with R = 20 cmH₂O·s/L and C = 0.07 L/cmH₂O. What's the peak airway pressure, in cmH₂O?
Check
Lungs with R = 15 cmH₂O·s/L and C = 0.03 L/cmH₂O exhale passively. About how long does it take to exhale 95% of the tidal volume?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
An electrosurgical unit sends 200 W through a 50 cm² return electrode. If only a fraction 0.5 of the pad stays in contact, by what factor does the heating per square centimeter under it rise?
Check
An electrosurgical unit returns 1 A RMS through a pad of 60 cm²; partial peeling leaves 30 cm² in contact.
(a)
What current density does the full pad carry?
Write units the way you normally would: 1.5 mA/cm^2, 1500 µA/cm^2 and 0.0015 A/cm^2 are the same answer.
(b)
By what factor does the heating rate rise after peeling?
Check
Dialyzer clearance in mL/min: 260 mL/min, urea 66 in, 30 out?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
Check
Peak pressure, cmH₂O: 0.48 L at 0.37 L/s, R = 22, C = 0.093, PEEP 1.5?
Check
Urea reduction ratio for Kt/V = 1.2?
Check
A person touches 120 V through a total path resistance of 1 kΩ. What current flows?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A device has 10000 pF between its 120 V, 50 Hz supply and its case. What leakage current flows through that capacitance?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
Two pieces of equipment connected to a patient differ in ground potential by 20 mV, and the path between them through a catheter is 300 Ω. What current flows?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
300 µA of leakage reaches a case connected to ground through 0.1 Ω. A person touching it offers a 10 kΩ path to ground. How much flows through the person?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
An isolated power system at 120 V, 50 Hz has 80 nF of capacitance to ground. What hazard current does its line isolation monitor report?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
A catheter path has 300 Ω, and the microshock limit is 10 µA. What's the largest ground-potential difference allowed between the two connected devices?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
Check
A hot-to-case fault occurs on a 120 V circuit with 0.1 Ω of wiring and 0.5 Ω in the ground conductor; a person of 5 kΩ touches the case.
(a)
What fault current flows?
Write units the way you normally would: 1.5 A, 1500 mA and 0.0015 kA are the same answer.
(b)
What current flows through the person until the breaker trips?
Check
Body current: 5 V, 0.5 kΩ?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
Check
Leakage: 100 V, 50 Hz, 20000 pF?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
Microshock: 1 mV over 100 Ω?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
Against a reference standard, a test is positive in 90 of the patients who have the disease and negative in 3 of them. What's its sensitivity?
Check
Among healthy subjects the test is negative in 900 and positive in 20. What's its specificity?
Check
A test with sensitivity 0.95 and specificity 0.9 is used where the prevalence is 0.2. What's its positive predictive value?
Check
For the same test and prevalence (sensitivity 0.95, specificity 0.9, prevalence 0.2), what's the negative predictive value?
Check
What's the positive likelihood ratio of a test with sensitivity 0.9 and specificity 0.9?
Check
An arrhythmia detector falsely flags 2 beats in 100,000. How many false alarms does it raise for a patient with a heart rate of 60 beats/min monitored for 72 hours?
Check
A study must show a sensitivity of about 0.95 with a 95% confidence half-width of 0.02, in a condition with prevalence 0.1.
(a)
How many patients with the condition must it include?
(b)
How many people would an unselected screening study need?
Check
PPV: sensitivity 0.74, specificity 0.64, prevalence 0.11?
Check
LR+ for sensitivity 0.68, specificity 0.55?
Check
A device draws 5 mA when active and 1 µA asleep, active a fraction 0.1 of the time. What's its average current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
A 220 mAh battery supplies an average of 20 µA. How many days does it last?
Check
A device streams 500 samples/s of 12-bit data all day by radio at 50 nJ/bit. How much energy does the radio use per day?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
Check
Tissue with conductivity 0.7 S/m and density 1050 kg/m³ sees an RMS field of 10 V/m. What's the specific absorption rate?
Write units the way you normally would: 1.5 W/kg, 1500 mW/kg and 0.0015 kW/kg are the same answer.
Check
A device has a constant failure rate with an MTBF of 50 years. What's the probability it survives 2 years?
Check
A device has three parts, any of which failing ends its life, with failure rates of 200, 100 and 80 per billion hours. What's its MTBF in years?
Check
A thermoelectric harvester gives 25 µW/cm² over 2 cm² through a converter of efficiency 0.7, powering a 60 µW average load.
(a)
What surplus (positive) or deficit (negative) power results?
Write units the way you normally would: 1.5 µW, 1500 nW and 0.0015 mW are the same answer.
(b)
What if the temperature difference falls to 0.5 of its value?
Check
Average current: 0.5 mA active, 0.2 µA asleep, duty 0.5?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
Check
Battery days: 20 mAh, 1 µA?
Check
SAR: σ = 0.1, E = 1 V/m, ρ = 1100?
Write units the way you normally would: 1.5 W/kg, 1500 mW/kg and 0.0015 kW/kg are the same answer.
Check
A telemeter sends 3 channels with content up to 40 Hz, sampled at five times that with 16 bits, at 50 nJ per bit.
(a)
What bit rate does it need?
Write units the way you normally would: 1.5 kb/s, 1500 b/s and 0.0015 Mb/s are the same answer.
(b)
What average radio power does that take?
Check
An implant's coil couples k = 0.03 to an external coil; their quality factors are 50 and 80, and the implant needs 50 mW.
(a)
What is the best efficiency the link can reach?
(b)
How much power must the external coil deliver?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
Tissue attenuates a radio wave's field by 34 Np/m.
(a)
How many decibels does a path of 1 cm cost?
(b)
At what depth has the field fallen to 1/e?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue has relative permittivity 57 and conductivity 0.5 S/m, at 2450 MHz.
(a)
What is its loss tangent, σ/ωε?
(b)
What is its field attenuation constant?
Check
An implant's link has a 100 kHz channel and needs 50 kb/s.
(a)
What spectral efficiency does it need?
Write units the way you normally would: 1.5 b/s/Hz and 0.0015 kb/s/Hz are the same answer.
Check
Bit rate: 2 channels to 100 Hz, 14 bits, 5× sampling?
Write units the way you normally would: 1.5 kb/s, 1500 b/s and 0.0015 Mb/s are the same answer.
Check
Link efficiency: k = 0.02, Q = 80 and 100?
Check
Attenuation: 20 Np/m over 2 cm?
Check
Ultrasound crosses from 1.38 MRayl into 1.75 MRayl; another echo returns after 150 µs.
(a)
Reflected intensity fraction?
(b)
Transmitted fraction?
(c)
Depth of the second reflector?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
Tissue at 0.5 dB/(cm·MHz), a 3 MHz 3-cycle pulse, 70 dB of dynamic range.
(a)
Deepest imageable depth?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
(b)
Axial resolution?
(c)
Frame rate for 128 lines to that depth?
Check
A 1.5 T scanner at 300 K with a 20 mT/m readout gradient and 100 kHz receiver bandwidth.
(a)
Larmor frequency?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
(b)
Polarization in ppm?
(c)
Field of view?
Check
A spin-echo sequence with TR = 450 ms and TE = 20 ms, for a tissue with T1 = 2500 ms and T2 = 900 ms.
(a)
Relative signal?
(b)
Fraction of the longitudinal magnetization recovered by each pulse?
Check
A 0.6 kHz RF pulse with a 6 mT/m slice gradient; 500 MBq of ¹⁸F (half-life 109.8 min).
(a)
Slice thickness?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
Activity after 120 min?
Check
Heart muscle with rheobase 0.8 mA and chronaxie 0.3 ms, stimulated with 0.5 ms pulses into 500 Ω.
(a)
Threshold current?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
Pulse energy at threshold?
(c)
Energy at the chronaxie instead?
Check
A pacemaker using 1 µC per pulse at 60 per minute, with 10 µA of circuit current, on a 2 Ah cell.
(a)
Average current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Battery life in years?
Check
A defibrillator: 30 µF at 1.5 kV into 75 Ω, truncated after 8 ms.
(a)
Stored energy?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
(b)
Peak current?
(c)
Energy delivered?
Check
A device with 4000 pF from its 120 V, 50 Hz supply to the case; ground resistance 1 Ω; a person offering 1 kΩ.
(a)
Leakage current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Current through the person with the ground intact?
Check
Grounds 8 mV apart across a 400 Ω catheter path; an isolated 230 V, 60 Hz system with 50 nF to ground.
(a)
Catheter current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Hazard current?
Check
A wearable with 3 mA active, 2 µA asleep, duty 0.02, on a 500 mAh cell.
(a)
Average current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Life in days?
Check
A 250 sample/s, 16-bit stream at 20 nJ/bit; tissue (σ = 1 S/m, ρ = 1000 kg/m³) in a 20 V/m field; an MTBF of 30 years.
(a)
Radio energy per day?
Write units the way you normally would: 1.5 J, 1500 mJ and 0.0015 kJ are the same answer.
(b)
SAR?
(c)
Probability of surviving 3 years?
Check
A dialyzer runs at 250 mL/min of blood flow, lowering urea from 90 to 40 mg/dL across it, for a 4-hour session on a 65 kg patient (body water 58% of weight).
(a)
Clearance, in mL/min?
Write units the way you normally would: 1.5 mL/min, 1500 µL/min and 0.0015 L/min are the same answer.
(b)
Kt/V?
(c)
Urea reduction ratio, ignoring generation?
Check
A screening test has sensitivity 0.9 and specificity 0.85, used where the prevalence is 0.1.
(a)
Positive predictive value?
(b)
Positive likelihood ratio?
(c)
Out of 1000 people screened, how many test positive?
Check
A study must estimate a sensitivity near 0.9 with a 95% confidence half-width of 0.03, in a condition of prevalence 0.1.
(a)
How many patients with the condition does it need?
(b)
How many people would an unselected study screen?
Check
A beam has two equal-weight energies with attenuation coefficients 0.3 and 0.15 cm⁻¹ in tissue.
(a)
What fraction passes 30 cm?
(b)
What effective attenuation coefficient does that imply?
Check
A CT protocol at 2 mSv changes its pixel width by a factor 0.5 and its slice thickness by 1, and allows the noise to change by a factor 1.5.
(a)
What dose does it need?
Check
A detector absorbs 0.5 of 100 photons per pixel and adds 20 photons RMS of electronic noise; a detail has contrast 0.05.
(a)
What is its detective quantum efficiency?
(b)
What contrast-to-noise ratio does the detail have per pixel?
Check
An ultrasound pulse at 7 MHz crosses 6 cm of tissue of 1.38 MRayl (0.5 dB/(cm·MHz)) to an interface with 1.7 MRayl.
(a)
How strong is the reflection, before attenuation?
(b)
How strong is the echo back at the transducer?
Check
A 3 MHz array focuses at f-number 3 and images 256 lines to 20 cm.
(a)
What lateral resolution does it reach at the focus?
Write units the way you normally would: 1.5 mm, 1500 µm and 0.0015 m are the same answer.
(b)
What is its highest frame rate?
Check
An MRI scan shrinks its voxels by 0.6 in each dimension, uses 2 averages, and runs at 3 T instead of 1.5 T; separately, two tissues have T2 of 300 and 50 ms.
(a)
By what factor does its SNR change?
(b)
What echo time best separates the two tissues?
Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.
Check
An ICD stores 25 J in 120 µF through a converter drawing 12 W at efficiency 0.9.
(a)
What voltage does the capacitor reach?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
How long does charging take?
Check
A pacemaker delivers 1 µC per pulse at 60 per minute and draws 8 µA continuously from a 2.5 Ah cell.
(a)
What average current does it draw?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
How many years does the cell last?
Check
A stimulating electrode of 0.002 cm² (double layer 20 µF/cm², access resistance 1 kΩ) delivers 2 mA phases of 200 µs; Shannon's k = 1.75.
(a)
What fraction of the safe charge per phase does each phase use?
(b)
What compliance voltage does the phase need?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
A cochlear implant drives 12 channels at 800 biphasic pulses per second with 40 µs, 1.2 mA phases from 6 V; its processor and link add 40 mW.
(a)
What total power does the implant draw?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
An electrosurgical unit returns 0.4 A RMS through an 80 cm² pad, of which only 40 cm² stays in contact.
(a)
What current density does the contact area carry?
Write units the way you normally would: 1.5 mA/cm^2, 1500 µA/cm^2 and 0.0015 A/cm^2 are the same answer.
(b)
How many times faster does the skin heat than with full contact?
Check
A device's case is grounded through 0.3 Ω; the supply is 120 V, 60 Hz, with 0.15 Ω of wiring; 2000 pF couples the supply to the case; a person of 1 kΩ touches it.
(a)
What flows through the person during a hot-to-case fault, before the breaker trips?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
Without a fault, what leakage would flow through the person if the ground wire were broken?
Check
A telemeter sends 1 channels with content to 100 Hz, sampled at five times that with 16 bits, at 20 nJ per bit.
(a)
What bit rate does it need?
Write units the way you normally would: 1.5 kb/s, 1500 b/s and 0.0015 Mb/s are the same answer.
(b)
How much radio energy does it use per day?
Check
An implant's coil couples k = 0.01 to an external coil with quality factors 60 and 80; the implant needs 30 mW.
(a)
What is the link's best efficiency?
(b)
How much must the external coil deliver?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
Check
Tissue between an implant and its reader costs 1.65 dB/cm, and the link has 15 dB of margin at zero depth.
(a)
How much does 4 cm of tissue cost?
(b)
How deep can the implant sit before the margin is gone?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
First-order high-pass coefficient a: corner 0.5 Hz, 250 samples/s?
Check
Approximate notch width: r = 0.95, 250 samples/s?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
60 beats are found in 50 s. What's the mean heart rate per minute?
Check
Beer–Lambert R at 80%?
Check
Beer–Lambert SpO₂ for R = 0.7, in percent?
Check
How far apart are the empirical line 110 − 25R and the Beer–Lambert theory at R = 0.7, in percentage points?
Check
A 5 µV/V/mmHg transducer at 5 V feeds a gain of 500 and a 12-bit converter spanning 5 V.
(a)
Amplifier output at 150 mmHg?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
Resolution in mmHg?
Check
A full bridge of gauges (G = 2) at ±300 microstrain and 5 V, amplified 500 times into a 12-bit converter spanning ±5 V.
(a)
Amplified output?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
(b)
How many converter steps is that?
Check
An 8 kΩ (25 °C) thermistor with B = 3800 K near 39 °C, read to 0.1 °C.
(a)
Resistance change to resolve?
Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.
(b)
As a percent of its resistance?
Check
A 120 V, 60 Hz line couples 2 pF to a patient grounded through 60 kΩ; electrodes differ by 30 kΩ; the amplifier has 50 MΩ common-mode input impedance and 90 dB CMRR.
(a)
Common-mode voltage?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Total input-referred 60 Hz error?
Check
Lead I = 0.5 mV, lead II = 0.9 mV; R waves 25 mm apart at 25 mm/s; electrode offsets up to 300 mV into a ±4 V first stage; total gain 1000.
(a)
Axis?
(b)
Rate per minute?
(c)
Gain needed after the high-pass?
Check
A catheter (radius 0.4 mm, 1 m, saline) on a transducer of 0.03 mm³ per 100 mmHg compliance; a 2 mm³ bubble enters.
(a)
Natural frequency without the bubble?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
With the bubble?
Check
Oxygen uptake 230 mL/min, arteriovenous difference 40 mL/L, heart rate 85; a 5 MHz Doppler at 45° on blood at 0.4 m/s.
(a)
Stroke volume?
Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.
(b)
Doppler shift?
Check
A 4 mg dye bolus: 40 mg·s/L before recirculation, then 1 mg/L falling with τ = 5 s; all of that flow leaves through an aorta 20 mm across, under an electromagnetic probe at 0.02 T.
(a)
Flow?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Mean velocity in the aorta?
(c)
Probe voltage?
Check
Red AC/DC = 0.02/1, infrared AC/DC = 0.045/1.4; 12 g/dL of hemoglobin and a of 80 mmHg.
(a)
SpO₂ from the 110 − 25R calibration, in percent?
(b)
Oxygen content at that saturation?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
X-rays through 15 cm of tissue (μ = 0.18 cm⁻¹) at a target noise of 0.01; 5 MHz ultrasound to 8 cm at 0.5 dB/(cm·MHz).
(a)
Incident photons per measurement needed?
(b)
Ultrasound round-trip loss?
Check
A 1.5 T scanner with a 10 mT/m slice gradient and a 2 kHz RF pulse; a spin echo at TR = 2500 ms, TE = 20 ms for T1 = 900 ms, T2 = 90 ms.
(a)
RF frequency?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
(b)
Slice thickness?
(c)
Relative signal?
Check
A pacemaker: 2 V, 0.3 ms pulses into 700 Ω at 80 per minute, 5 µA of circuit current, a 1 Ah cell.
(a)
Pulse energy?
Write units the way you normally would: 1.5 µJ, 1500 nJ and 0.0015 mJ are the same answer.
(b)
Battery life in years?
Check
Thermodilution: 5 mL of injectate at 0 °C, a curve area of 6 °C·s and K = 1.08, at 100 beats per minute, in a patient of 1.7 m² body surface area.
(a)
What is the cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
What is the cardiac index?
(c)
What is the stroke volume?
Check
An isolated ECG amplifier (gain 500, CMRR 120 dB, IMRR 120 dB) sees 1 mV of signal, 50 mV of common mode and 120 V across its barrier.
(a)
What fraction of the signal is error?
(b)
What is the output?
Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.
Check
An oximeter's infrared perfusion index is 0.5% with the DC at 0.6 of a 16-bit range; its red and infrared AC/DC ratios are 0.012 and 0.015, calibrated as 110 − 25R.
(a)
How many counts does the infrared pulse span?
(b)
What saturation does it display?
Check
An ECG front end must not saturate on ±300 mV of electrode offset with its first stage swinging ±2.5 V, and must resolve 2 µV at the electrodes with a 24-bit converter spanning ±2.5 V.
(a)
What is the largest gain the first, DC-coupled stage can have?
(b)
What total gain gives that resolution?
Check
Up to 10 mV of common mode must produce less than 1 µV of input-referred error, using a difference amplifier with gain 100.
(a)
What CMRR does that need?
(b)
What resistor tolerance guarantees it?
Check
An anti-aliasing Butterworth filter has its corner at 40 Hz and must attenuate by 60 dB at half the 500 Hz sampling rate.
(a)
What order does it need?
Check
A saline catheter of radius 0.5 mm feeds a transducer of 0.01 mm³ per 100 mmHg, and the system's natural frequency must be at least 50 Hz.
(a)
How long can the catheter be?
Write units the way you normally would: 1.5 m, 1500 mm and 0.0015 km are the same answer.
Check
A 10 kΩ (at 25 °C, B = 3500 K) thermistor in a half bridge with a fixed resistor equal to its resistance at 37 °C is excited at 3 V.
(a)
What fixed resistor does that call for?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
How many millivolts per °C does the bridge give at 37 °C?
Check
An oximeter's photodiode (responsivity 0.5 A/W) needs 0.1 µA; tissue passes 0.01 of the LED's light, and the LED converts electrical power to light at efficiency 0.2 from a 1.5 V drop.
(a)
What optical power must the LED emit?
Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.
(b)
What LED current does that take?
Check
5 isolated inputs connect to one patient, whose total leakage must stay below 10 µA at 120 V, 60 Hz.
(a)
How much barrier capacitance can each input have?
Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.
Check
An external defibrillator must store 150 J at no more than 2 kV.
(a)
What capacitance does it need?
Write units the way you normally would: 1.5 µF, 1500 nF and 0.0015 mF are the same answer.
Check
A pacemaker must last 8 years delivering 0.5 µC per pulse at 70 per minute with 6 µA of circuit current.
(a)
What cell capacity does it need?
Write units the way you normally would: 1.5 Ah, 1500 mAh and 0.0015 kAh are the same answer.
Check
A wearable draws 5 mA active and 2 µA asleep from a 300 mAh battery, which must last 30 days.
(a)
What fraction of the time can it be active?
Check
A telemeter's radio may average 50 µW at 2 nJ per bit, for 3 channels sampled at 500 Hz.
(a)
How many bits per sample can it send?
Check
An ultrasound scanner must make 30 frames a second of 128 lines each.
(a)
How deep can it image?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
Check
A radiograph must show a detail of contrast 0.01 at a Rose threshold of J = 3, with a detector of DQE 0.7.
(a)
How many detected-equivalent photons per detail area does that need?
(b)
How many must reach the detector?
Check
A stimulator must deliver 0.5 µC per phase within Shannon's limit with k = 1.7.
(a)
What is the smallest electrode area that allows it?
Check
An inductive link with coil quality factors 50 and 80 must reach an efficiency of 0.5.
(a)
What coupling coefficient does that need?
Check
A transducer of 5 µV/V/mmHg at 2.5 V; another with output µV (S = 30, a = 0.002) over 0 to 150 mmHg.
(a)
First transducer's output at 240 mmHg?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Second one's nonlinearity in percent of full scale?
Check
A probe with τ = 5 s meets a 2 °C step; a catheter has = 60 Hz and ζ = 0.1.
(a)
Probe error after 25 s?
(b)
Catheter amplitude ratio at 8 Hz?
(c)
Catheter step overshoot, percent?
Check
Gauges (G = 2) at ±250 microstrain in a bridge at 10 V with 4 active arms; a piezoelectric sensor (d = 100 pC/N) into a charge amplifier with = 100 pF.
(a)
Bridge output?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Charge amplifier output for 0.5 N?
Check
A thermistor of 3 kΩ at 25 °C, B = 3300 K, at 41 °C carrying 150 µA with δ = 1 mW/°C.
(a)
Resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Coefficient in %/°C?
(c)
Self-heating?
Check
K⁺ at 130 mM inside and 4 mM outside, Na⁺ at 12 and 145 mM, = 0.1, 37 °C; an axon 10 µm across with = 8000 Ω·cm² and = 70 Ω·cm.
(a)
Potassium Nernst potential?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
GHK potential?
(c)
Length constant?
Check
Lead I = 0.3 mV, lead II = 0.7 mV; R waves 30 mm apart at 25 mm/s; QT = 0.42 s.
(a)
Lead III?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Axis?
(c)
Bazett QTc?
Check
An electrode with = 0.5 kΩ, = 50 kΩ, = 0.3 µF; an instrumentation amplifier with = 3 kΩ and = 25 kΩ.
(a)
Electrode impedance at 40 Hz?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Amplifier first-stage gain?
Check
5 mV of common mode, 10 kΩ electrode mismatch, 500 MΩ common-mode input impedance; a driven right leg with = 6 MΩ, = 4 MΩ, = 25 kΩ and 0.1 µA of displacement current.
(a)
Mismatch error?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Driven-right-leg common-mode voltage?
Check
An evoked response of 0.5 µV in 10 µV of background; a high-pass at 0.5 Hz; a 60 Hz notch with Q = 20.
(a)
Sweeps for an SNR of 6?
(b)
High-pass phase lead at 1 Hz?
(c)
Notch width?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
Check
A saline catheter (radius 0.4 mm, 0.8 m) on a transducer of 0.04 mm³ per 100 mmHg; a pressure of 150/70 mmHg; a transducer 5 cm below the heart.
(a)
Natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Mean arterial pressure?
(c)
Leveling error?
Check
Fick: 220 mL/min uptake and 40 mL/L difference; thermodilution: 10 mL at 20 °C, area 3 °C·s, K = 1.08; Doppler: 2 MHz at 30° on 0.3 m/s.
(a)
Fick output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Thermodilution output?
(c)
Doppler shift?
Check
Helium dilution from 9% to 7% in a 3 L spirometer; a forced expiration with τ = 1 s; an oximeter reading R = 0.5; blood with 10 g/dL Hb at saturation 0.9, 90 mmHg.
(a)
FRC?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
FEV1/FVC?
(c)
Calibrated SpO₂ in percent?
(d)
Oxygen content?
Check
A pH electrode at 25 °C; blood with 28 mM bicarbonate at = 50 mmHg; a glucose sensor with = 80 nA and = 15 mM at 5 mM.
(a)
Voltage change for 0.2 pH units?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Blood pH?
(c)
Glucose current?
Check
Tissue with μ = 0.16 cm⁻¹, 10 cm thick; a voxel of μ = 0.4 against water's 0.2; an ultrasound boundary from 1.38 to 1.9 MRayl; a 3 MHz echo from 5 cm.
(a)
X-ray transmission?
(b)
CT number?
(c)
Reflected intensity fraction?
(d)
Round-trip loss at 0.5 dB/(cm·MHz)?
Check
A 1.5 T scanner with a 10 mT/m readout gradient and 100 kHz bandwidth; 400 MBq of ¹⁸F; a spin echo (TR 500 ms, TE 15 ms) for T1 = 1000 ms, T2 = 100 ms.
(a)
Larmor frequency?
Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.
(b)
Field of view?
(c)
Activity after 60 min?
(d)
Relative signal?
Check
Tissue with rheobase 0.5 mA and chronaxie 0.3 ms; a defibrillator: 100 µF at 2 kV into 50 Ω, truncated after 4 ms.
(a)
Threshold for 0.5 ms pulses?
Write units the way you normally would: 1.5 mA, 1500 µA and 0.0015 A are the same answer.
(b)
Stored energy?
(c)
Fraction delivered?
Check
A 120 V, 60 Hz device with 1000 pF to its case; 30 mV between grounds across an 800 Ω catheter path; a wearable (3 mA active, 3 µA asleep, duty 0.005) on 200 mAh.
(a)
Leakage current?
Write units the way you normally would: 1.5 µA, 1500 nA and 0.0015 mA are the same answer.
(b)
Catheter current?
(c)
Wearable life in days?
Check
A volume-controlled breath delivers 0.45 L at 0.5 L/s with PEEP 5 cmH₂O to lungs with R = 12 cmH₂O·s/L and C = 0.06 L/cmH₂O.
(a)
Peak airway pressure, in cmH₂O?
(b)
Plateau pressure, in cmH₂O?
(c)
Time for 95% of passive exhalation?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A rhythm test with sensitivity 0.85 and specificity 0.9 is used where the prevalence is 0.15.
(a)
Positive predictive value?
(b)
Negative predictive value?
Check
A hematology aperture is 60 µm in diameter and 120 µm long, filled with diluent of resistivity 1.2 Ω·m and driven by 1 mA. Red cells of 95 fL pass through it.
(a)
Aperture resistance?
Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.
(b)
Resistance change for one cell, from ?
(c)
Pulse height at 1 mA?
A kind of problem counts as solid once you've got it right on the first try twice in a row. Back to S15