Problem set 1: sensitivity, nonlinearity and resolution, first- and second-order sensor dynamics, strain gauges and bridges, capacitive and piezoelectric sensors, and phase-sensitive and optical detection

These problems cover lessons 1 to 5: transducer outputs and nonlinearity, input-referred resolution and combined errors, step, ramp and sinusoid responses of first-order probes, catheter amplitude ratios and overshoot, gauge resistance changes and bridge outputs, loaded potentiometers, capacitive and piezoelectric sensor signals, phase-sensitive demodulation, fibers and photodiodes.

Check

A transducer of 5 µV/V/mmHg at 3 V excitation; separately, an output sx+ax2 with a = 0.004 µV/mmHg² over 0 to 200 mmHg at the same s.

(a)

Output at 250 mmHg?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

(b)

Nonlinearity against the endpoint line, in percent?

Check

A gain of 2000 into a 12-bit converter spanning 3 V; independent errors 1.5, 0.4 and 1 mmHg.

(a)

Input-referred step?

Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.

(b)

Combined error in mmHg?

Check

A probe with τ = 8 s meets a 1.5 °C step.

(a)

Remaining error after 40 s?

(b)

Time to come within 0.2 °C?

Write units the way you normally would: 1.5 s and 1500 ms are the same answer.

Check

A first-order probe with τ = 10 s.

(a)

Lag behind a ramp of 0.5 °C/min?

(b)

Amplitude ratio at 0.5 Hz?

Check

A catheter system with fn = 20 Hz and ζ = 0.15.

(a)

Amplitude ratio at 15 Hz?

(b)

Step overshoot in percent?

Check

A 350 Ω gauge with G = 2 at 300 microstrain, in a bridge excited at 10 V.

(a)

Resistance change?

Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.

(b)

Approximate quarter-bridge output?

(c)

Output if four such gauges form a full bridge?

Check

A quarter bridge at 3 V with ΔR/R = 0.03 in its active arm.

(a)

Exact output?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

(b)

Percent error of the linear approximation?

Check

A foil gauge (G = 2) with a 15 ppm/°C temperature coefficient; a 2 kΩ pot at 0.5 of its travel with a 20 kΩ load.

(a)

Apparent strain for an 8 °C change, in microstrain?

(b)

Wiper output as a fraction of the excitation?

Check

A capacitive sensor: 0.5 cm² plates 0.1 mm apart, relative permittivity 5.

(a)

Capacitance?

Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.

(b)

Change for a 4 µm gap reduction?

Check

A piezoelectric sensor with d = 150 pC/N and 4 nF, read across 5 GΩ, under a 0.5 N step.

(a)

Initial voltage?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

Low-frequency corner?

(c)

Voltage after 1 s?

Check

A charge amplifier with Cf = 500 pF reads a sensor with d = 100 pC/N; a transducer whose sensitivity falls by a fraction 0.006 per °C.

(a)

Output size for 0.3 N?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

Percent gain error after 4 °C?

Check

A heart rate of 60 per minute whose first 8 harmonics must be reproduced by a catheter with fn = 80 Hz and ζ = 0.6.

(a)

Highest harmonic needed?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

(b)

Percent error in that harmonic's amplitude?

Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.

Practice

Keep practicing

15 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.