Problem set 11: problems that chain sensors to amplifiers, biopotentials to front ends, pressure and flow to their transducers, optics and chemistry to readings, imaging to its limits, and therapy and safety to their budgets

These problems draw on the whole course, and most chain several ideas: a transducer's output carried through an amplifier and converter, an ECG's common-mode problem traced from the power line to the input, a catheter's bubble turned into a pressure error, a Doppler shift turned into a flow, a pacemaker's pulse turned into a battery life. Work them as a review before the final.

Check

A 5 µV/V/mmHg transducer at 5 V feeds a gain of 500 and a 12-bit converter spanning 5 V.

(a)

Amplifier output at 150 mmHg?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

Resolution in mmHg?

Check

A full bridge of gauges (G = 2) at ±300 microstrain and 5 V, amplified 500 times into a 12-bit converter spanning ±5 V.

(a)

Amplified output?

Write units the way you normally would: 1.5 V, 1500 mV and 0.0015 kV are the same answer.

(b)

How many converter steps is that?

Check

An 8 kΩ (25 °C) thermistor with B = 3800 K near 39 °C, read to 0.1 °C.

(a)

Resistance change to resolve?

Write units the way you normally would: 1.5 Ω, 1500 mΩ and 0.0015 kΩ are the same answer.

(b)

As a percent of its resistance?

Check

A 120 V, 60 Hz line couples 2 pF to a patient grounded through 60 kΩ; electrodes differ by 30 kΩ; the amplifier has 50 MΩ common-mode input impedance and 90 dB CMRR.

(a)

Common-mode voltage?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

(b)

Total input-referred 60 Hz error?

Check

Lead I = 0.5 mV, lead II = 0.9 mV; R waves 25 mm apart at 25 mm/s; electrode offsets up to 300 mV into a ±4 V first stage; total gain 1000.

(a)

Axis?

(b)

Rate per minute?

(c)

Gain needed after the high-pass?

Check

A catheter (radius 0.4 mm, 1 m, saline) on a transducer of 0.03 mm³ per 100 mmHg compliance; a 2 mm³ bubble enters.

(a)

Natural frequency without the bubble?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

(b)

With the bubble?

Check

Oxygen uptake 230 mL/min, arteriovenous difference 40 mL/L, heart rate 85; a 5 MHz Doppler at 45° on blood at 0.4 m/s.

(a)

Stroke volume?

Write units the way you normally would: 1.5 mL, 1500 µL and 0.0015 L are the same answer.

(b)

Doppler shift?

Check

A 4 mg dye bolus: 40 mg·s/L before recirculation, then 1 mg/L falling with τ = 5 s; all of that flow leaves through an aorta 20 mm across, under an electromagnetic probe at 0.02 T.

(a)

Flow?

Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.

(b)

Mean velocity in the aorta?

(c)

Probe voltage?

Check

Red AC/DC = 0.02/1, infrared AC/DC = 0.045/1.4; 12 g/dL of hemoglobin and a PaO2 of 80 mmHg.

(a)

SpO₂ from the 110 − 25R calibration, in percent?

(b)

Oxygen content at that saturation?

Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.

Check

X-rays through 15 cm of tissue (μ = 0.18 cm⁻¹) at a target noise of 0.01; 5 MHz ultrasound to 8 cm at 0.5 dB/(cm·MHz).

(a)

Incident photons per measurement needed?

(b)

Ultrasound round-trip loss?

Check

A 1.5 T scanner with a 10 mT/m slice gradient and a 2 kHz RF pulse; a spin echo at TR = 2500 ms, TE = 20 ms for T1 = 900 ms, T2 = 90 ms.

(a)

RF frequency?

Write units the way you normally would: 1.5 MHz, 1500 kHz and 0.0015 GHz are the same answer.

(b)

Slice thickness?

(c)

Relative signal?

Check

A pacemaker: 2 V, 0.3 ms pulses into 700 Ω at 80 per minute, 5 µA of circuit current, a 1 Ah cell.

(a)

Pulse energy?

Write units the way you normally would: 1.5 µJ, 1500 nJ and 0.0015 mJ are the same answer.

(b)

Battery life in years?

Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.