S15 · Problem set 12

Problem set 12: design problems for front-end gain, common-mode rejection and resistor tolerance, anti-aliasing filters, catheter lengths, thermistor bridges, oximeter LED drive, isolation budgets, defibrillator capacitors, pacemaker cells, duty cycles, telemetry budgets, ultrasound depth, X-ray dose, stimulating electrodes and inductive links

The last set before the final turns the course around: each problem gives a requirement, from a resolution at the electrodes to a battery life or a link efficiency, and asks for the design value that meets it. Most combine two lessons, and several are the first calculation an engineer would make on a real instrument.

Check

An ECG front end must not saturate on ±300 mV of electrode offset with its first stage swinging ±2.5 V, and must resolve 2 µV at the electrodes with a 24-bit converter spanning ±2.5 V.

(a)

What is the largest gain the first, DC-coupled stage can have?

(b)

What total gain gives that resolution?

Check

Up to 10 mV of common mode must produce less than 1 µV of input-referred error, using a difference amplifier with gain 100.

(a)

What CMRR does that need?

(b)

What resistor tolerance guarantees it?

Check

An anti-aliasing Butterworth filter has its corner at 40 Hz and must attenuate by 60 dB at half the 500 Hz sampling rate.

(a)

What order does it need?

Check

A saline catheter of radius 0.5 mm feeds a transducer of 0.01 mm³ per 100 mmHg, and the system's natural frequency must be at least 50 Hz.

(a)

How long can the catheter be?

Write units the way you normally would: 1.5 m, 1500 mm and 0.0015 km are the same answer.

Check

A 10 kΩ (at 25 °C, B = 3500 K) thermistor in a half bridge with a fixed resistor equal to its resistance at 37 °C is excited at 3 V.

(a)

What fixed resistor does that call for?

Write units the way you normally would: 1.5 kΩ, 1500 Ω and 0.0015 MΩ are the same answer.

(b)

How many millivolts per °C does the bridge give at 37 °C?

Check

An oximeter's photodiode (responsivity 0.5 A/W) needs 0.1 µA; tissue passes 0.01 of the LED's light, and the LED converts electrical power to light at efficiency 0.2 from a 1.5 V drop.

(a)

What optical power must the LED emit?

Write units the way you normally would: 1.5 mW, 1500 µW and 0.0015 W are the same answer.

(b)

What LED current does that take?

Check

5 isolated inputs connect to one patient, whose total leakage must stay below 10 µA at 120 V, 60 Hz.

(a)

How much barrier capacitance can each input have?

Write units the way you normally would: 1.5 pF and 0.0015 nF are the same answer.

Check

An external defibrillator must store 150 J at no more than 2 kV.

(a)

What capacitance does it need?

Write units the way you normally would: 1.5 µF, 1500 nF and 0.0015 mF are the same answer.

Check

A pacemaker must last 8 years delivering 0.5 µC per pulse at 70 per minute with 6 µA of circuit current.

(a)

What cell capacity does it need?

Write units the way you normally would: 1.5 Ah, 1500 mAh and 0.0015 kAh are the same answer.

Check

A wearable draws 5 mA active and 2 µA asleep from a 300 mAh battery, which must last 30 days.

(a)

What fraction of the time can it be active?

Check

A telemeter's radio may average 50 µW at 2 nJ per bit, for 3 channels sampled at 500 Hz.

(a)

How many bits per sample can it send?

Check

An ultrasound scanner must make 30 frames a second of 128 lines each.

(a)

How deep can it image?

Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.