S15 · Problem set 6
Problem set 6: catheter natural frequency and damping, bubbles and flush tests, leveling, cuff methods and mean pressure, Fick and indicator dilution, thermodilution, electromagnetic, transit-time and Doppler flowmeters, and plethysmography
These problems cover lessons 23, 24 and 26 to 28: catheter dynamics from dimensions, the effect of a bubble, damping from a flush test, hydrostatic errors, mean pressure and oscillometric ratios, deflation errors, Fick and Stewart–Hamilton outputs with their tails, thermodilution, stroke volume, flow from magnetic induction, transit times and Doppler shifts, and volumes from venous occlusion and impedance plethysmography.
Check
A saline catheter of radius 0.4 mm and length 0.8 m on a transducer of compliance 0.03 mm³ per 100 mmHg.
(a)
Natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Damping ratio (η = 0.001 Pa·s)?
Check
A catheter system with = 100 Hz, ζ = 0.06 and transducer compliance 5 × 10⁻¹⁶ m³/Pa gains a 2 mm³ bubble.
(a)
New natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
New damping ratio?
Check
A flush test gives a peak ratio of 0.15 with peaks 30 ms apart; the transducer sits 15 cm above the heart (saline column).
(a)
Damping ratio?
(b)
Natural frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(c)
Size of the leveling error?
Check
A pressure of 115/70 mmHg measured with a cuff deflating at 4 mmHg/s, heart rate 90 per minute.
(a)
Mean arterial pressure?
(b)
Deflation between beats?
Check
An oscillometric envelope with M = 85 mmHg and w = 20 mmHg.
(a)
Systolic estimate at ratio 0.6?
(b)
Diastolic estimate at ratio 0.85?
Check
Oxygen uptake 220 mL/min with an arteriovenous difference of 35 mL/L; heart rate 95; body surface area 2.1 m².
(a)
Cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Stroke volume?
(c)
Cardiac index in L/min/m²?
Check
A 3 mg dye bolus; the first pass accumulates 35 mg·s/L before recirculation, where the concentration is 1 mg/L falling with τ = 5 s.
(a)
Flow with the tail included?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Percent error from ignoring the tail?
Check
5 mL of injectate at 0 °C into blood at 37 °C; curve area 6 °C·s; K = 1.08.
(a)
Cardiac output?
Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.
(b)
Curve area, in °C·s, that the same injectate would give at 5 L/min?
Check
An electromagnetic probe applies 0.02 T across a 4 mm vessel with mean velocity 0.8 m/s.
(a)
Voltage?
Write units the way you normally would: 1.5 µV, 1500 nV and 0.0015 mV are the same answer.
(b)
Volume flow?
Check
A transit-time path of 8 mm at 45° to blood at 0.3 m/s.
(a)
Time difference?
Write units the way you normally would: 1.5 ns, 1500 ps and 0.0015 µs are the same answer.
(b)
Travel time with no flow?
Check
A 3 MHz Doppler at 60° to blood moving at 0.3 m/s.
(a)
Shift?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Percent velocity error if the true angle is 5° smaller than assumed?
Check
A 2 MHz pulsed Doppler at a depth of 5 cm and 60°.
(a)
Highest pulse repetition frequency?
Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.
(b)
Highest unaliased velocity?
Check
A four-electrode impedance plethysmograph drives 1 mA at 100 kHz through an 8 cm limb segment of 40 Ω. Each pulse lowers the impedance by 0.12 Ω; blood resistivity is 1.4 Ω·m.
(a)
Steady voltage across the sensing electrodes?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Pulsatile signal?
(c)
Volume added by each pulse?
Check
In venous-occlusion plethysmography a 900 mL calf segment swells at 0.6 mL/s after the cuff inflates. After release, the extra volume drains with a time constant of 6 s.
(a)
Arterial inflow, in mL/min per 100 mL of tissue?
(b)
About how long does 95% of the extra volume take to drain?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
Check
A transducer sits 5 cm below the right atrium, connected by saline (1000 kg/m³).
(a)
What offset does that add?
(b)
What percent error is that on a central venous pressure of 5 mmHg?
Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.
Practice
Keep practicing
15 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.