Problem set 7: valve areas and Doppler gradients, kinetic-energy errors, catheters with bubbles, oscillometric envelopes, thermodilution, pulse-wave velocity, BTPS volumes, lung volumes, mechanics and work of breathing, oximetry and oxygen content, blood gases and enzyme sensors, and the laboratory's photometers and counters

These problems cover lessons 23 to 35, weighted toward what the earlier sets didn't reach: catheterization and the cardiovascular lab's processing, then the respiratory system, oxygen and carbon dioxide, and the chemical and laboratory measurements that read blood itself.

Check

Cardiac output is 4 L/min at 65 per minute, with 0.34 s of ejection per beat and a mean aortic-valve gradient of 50 mmHg; separately, Doppler measures a 2.5 m/s jet.

(a)

What is the valve's area?

(b)

What gradient does the simplified Bernoulli equation give for the jet?

Check

An end-hole catheter faces blood flowing at 1.5 m/s where the static pressure is 15 mmHg.

(a)

By how much does it read high?

(b)

What percent error is that?

Check

A saline catheter of radius 0.4 mm and length 0.9 m feeds a transducer of 0.05 mm³ per 100 mmHg, with a 2 mm³ air bubble trapped in the dome.

(a)

What is the natural frequency?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

(b)

What is the damping ratio?

Check

An oscillometric envelope follows e−((P−MAP)/w)2 with MAP = 85 mmHg and w = 25 mmHg; the monitor uses ratios of 0.55 and 0.85.

(a)

What systolic pressure does it report?

(b)

What diastolic pressure?

Check

5 mL of injectate at 0 °C enters blood at 37 °C; the temperature curve's area is 7 °C·s, with K = 1.08, at 90 beats per minute.

(a)

What is the cardiac output?

Write units the way you normally would: 1.5 L/min and 1500 mL/min are the same answer.

(b)

What is the stroke volume?

Check

An artery's wall has E = 0.4 MPa and thickness 0.4 mm with a 16 mm lumen; a pulse travels 0.9 m, and a pressure rise stiffens the wall by a factor of 1.1.

(a)

How long does the pulse take?

Write units the way you normally would: 1.5 ms, 1500 µs and 0.0015 s are the same answer.

(b)

How much shorter does the stiffening make it?

Check

A spirometer at 20 °C (water vapor 23.8 mmHg) and 760 mmHg measures a vital capacity of 3 L.

(a)

What is it at body conditions (BTPS)?

Write units the way you normally would: 1.5 L and 1500 mL are the same answer.

Check

A helium spirometer of 5 L starts at 8% helium and settles at 6.5% after the subject joins at FRC; separately, a body box at 1000 cmH₂O alveolar pressure shows a 20 mL compression for a 1 cmH₂O rise.

(a)

What FRC does helium dilution give?

Write units the way you normally would: 1.5 L and 1500 mL are the same answer.

(b)

What thoracic gas volume does the body box give?

Check

A respiratory system has R = 2 cmH₂O·s/L, C = 0.05 L/cmH₂O and inertance 0.02 cmH₂O·s²/L; it breathes 0.8 L at constant flow over 1 s, 12 times a minute.

(a)

At what frequency is it resonant?

Write units the way you normally would: 1.5 Hz, 1500 mHz and 0.0015 kHz are the same answer.

(b)

What power do the muscles deliver?

Check

An oximeter reads red AC/DC = 0.008/1 and infrared AC/DC = 0.02/1.3, calibrated as SpO₂ = 110 − 25R; the patient has 12 g/dL of hemoglobin and a PaO2 of 60 mmHg.

(a)

What saturation does it display?

(b)

What oxygen content does that imply?

Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.

Check

Blood has 24 mM bicarbonate and a PCO2 of 30 mmHg; a Severinghaus electrode at 37 °C then sees PCO2 rise by a factor of 3.

(a)

What is the blood's pH?

(b)

How much does the electrode's output change?

Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.

Check

A glucose sensor has imax⁡ = 50 nA and Km = 10 mM; separately, an amperometric sensor reacts 0.5 pmol/s with 2 electrons each.

(a)

What current does the glucose sensor give at 15 mM?

Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.

(b)

What current does the amperometric reaction carry?

Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.

Practice

Keep practicing

15 kinds of problem on this topic, dealt one at a time with new numbers every time, for as long as you like.