S15 · Problem set 8
Problem set 8: BTPS corrections, helium dilution and body plethysmography, FEV1 and alveolar ventilation, ratio of ratios and oxygen content, respiratory mechanics, Nernstian and amperometric sensors, spectrophotometry and blood counts, image quality, attenuation, Hounsfield units and photon noise
These problems cover lessons 31 to 38: gas-volume corrections, lung volumes by dilution and Boyle's law, forced expiration and ventilation, end-tidal CO₂, compliance, resistance and forced oscillation, oximeter ratios and calibrations, oxygen content, pH and ion-selective electrodes, Henderson–Hasselbalch, Faraday currents and enzyme saturation, absorbance and red-cell indices, photon-limited detection and MTF, and X-ray attenuation, contrast, CT numbers and dose.
Check
A spirometer at 21 °C (water vapor 18.7 mmHg), barometric pressure 760 mmHg, records 5 L.
(a)
BTPS factor?
(b)
Volume in the lungs?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
Check
Helium in a 4 L spirometer falls from 9% to 7%; a body box sees 25 mL of compression and 9 cmH₂O rise from 1000 cmH₂O.
(a)
FRC by helium dilution?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
Volume by plethysmography?
Check
A forced expiration with FVC = 3 L and τ = 0.35 s; quiet breathing at 650 mL, 170 mL dead space, 12 per minute.
(a)
FEV1?
Write units the way you normally would: 1.5 L and 1500 mL are the same answer.
(b)
Alveolar ventilation?
(c)
Minute ventilation?
Check
End-tidal gas with 4.5% CO₂ (of the dry gas) at a barometric pressure of 520 mmHg.
(a)
End-tidal CO₂ partial pressure?
(b)
What CO₂ percentage would correspond to 40 mmHg at this pressure?
Check
Red AC/DC = 0.015/0.9; infrared AC/DC = 0.04/1.4.
(a)
Ratio of ratios?
(b)
Displayed SpO₂ with the 110 − 25R calibration, in percent?
Check
Saturation 75% in the Beer–Lambert model (coefficients 3227/320 at 660 nm, 693/1214 at 940 nm); blood with 10 g/dL Hb at saturation 0.97 and = 95 mmHg.
(a)
Theoretical R?
(b)
Oxygen content?
Write units the way you normally would: 1.5 mL/dL, 1500 µL/dL and 0.0015 L/dL are the same answer.
Check
A pH electrode at 25 °C; blood with 28 mM bicarbonate and = 55 mmHg.
(a)
Voltage change for 0.2 pH units?
Write units the way you normally would: 1.5 mV, 1500 µV and 0.0015 V are the same answer.
(b)
Blood pH?
Check
A K⁺ electrode with = 0.0005 in 3 mM K⁺ and 145 mM Na⁺; an amperometric sensor reacting 5 pmol/s with 2 electrons each.
(a)
Percent overreading?
(b)
Sensor current?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
Check
A glucose sensor with = 60 nA and = 12 mM at 12 mM.
(a)
Current?
Write units the way you normally would: 1.5 nA, 1500 pA and 0.0015 µA are the same answer.
(b)
What glucose does a meter using the initial linear slope report?
(c)
Percent low reading of a linear calibration?
Check
Soft tissue with μ = 0.16 cm⁻¹, 12 cm thick; a bone of μ = 0.7 cm⁻¹ replacing 4 cm of it.
(a)
Transmitted fraction through the tissue?
(b)
Half-value layer of the tissue?
Write units the way you normally would: 1.5 cm, 15 mm and 0.015 m are the same answer.
(c)
Bone contrast?
Check
A voxel with μ = 0.26 cm⁻¹ against water's 0.19; a target relative noise of 0.02.
(a)
CT number?
(b)
Detected photons needed?
Check
An 80 kV X-ray tube; a CT detector with 256 bins; a projection that transmits 0.05 of the beam.
(a)
Shortest wavelength?
Write units the way you normally would: 1.5 nm, 1500 pm and 0.0015 µm are the same answer.
(b)
Views needed?
(c)
Projection value ∫μ dl?
Check
A patient's lung compliance is 0.15 L/cmH₂O and chest-wall compliance 0.15 L/cmH₂O, with airway resistance 8 cmH₂O·s/L and inertance 0.02 cmH₂O·s²/L.
(a)
Total compliance, in L/cmH₂O?
(b)
Time constant of passive exhalation?
Write units the way you normally would: 1.5 s and 1500 ms are the same answer.
(c)
Resonant frequency in a forced-oscillation test?
Check
A cholesterol standard of 100 mg/dL reads 30% transmittance and a patient's serum 65%. The same patient's blood count shows hematocrit 44% and 4.2 million red cells per µL.
(a)
The patient's cholesterol, in mg/dL?
(b)
MCV, in fL?
(c)
Relative concentration error from a 0.5% transmittance error at the patient's reading?
Check
An X-ray detector pixel averages 4000 photons. A detail has contrast 0.02, and detection needs an SNR of J = 5.
(a)
The detail's SNR at this exposure?
(b)
Photons per pixel needed to detect it?
(c)
By what factor must the dose rise?
Have paper and a calculator ready. Work each problem out by hand, then enter the answer. Your first try on each one counts toward your score.